Page 2/13Data sheet 202569
Display modes
Three display modes are available:
Large digits
In this mode, the measured values are dis-
played as digits as usual.
Trend display
In this mode, the numerical value is supple-
mented by a symbol indicating the direction
and speed at which the measured value is
changing.
This can be useful for control optimization, for
example.
Left to right:
Rising quickly, at moderate speed, and slowly,
stable,
falling slowly, at moderate speed, and quickly.
Bar graph
In this display, a cursory glance shows the
range in which the current measured value is
found.
The bar graph is freely scalable.
Measurement of dissolved oxy-
gen
We recommend the use of the JUMO ecoLine
O-DO sensor (type 202613) for measuring
dissolved oxygen.
This sensor operates optically according to
the principle of luminescence quenching. For
displaying the measured value, the user can
choose between % Sat, mg/l, and ppm. To
improve the operating life of the sensor's lu-
minescent layer, the sampling rate of the
JUMO AQUIS 500 RS can be adjusted in the
range 1 to 60 s, depending on the applica-
tion.
To compensate the salinity and ambient pres-
sure, these can be connected to the JUMO
AQUIS 500 RS.
The temperature can be measured either in-
ternally, with the sensor, or using an external
temperature sensor connected to the analog
temperature input of the JUMO
AQUIS 500 RS.
Functional description
The device is designed for on-site use. A ro-
bust case protects the electronics and electri-
cal connections against aggressive
environmental conditions (IP67). Alternatively,
the device can also be installed in a panel, al-
lowing frontal protection type IP65 to be
achieved. The electrical connections are
made with installation-friendly screw connec-
tors. A ventilation screw with PTFE membrane
prevents the formation of condensate.
Display and control elements
(1) Switching output 1 or 2 is active
(2) Binary output 1 is controlled
(3) Keypad is locked
(4) Alarm was activated
(5) Device is in manual mode
(6) Device status
(7) Medium temperature
(8) Main measured value
(9) Unit of main measured value
The user can specify what should be dis-
played at positions (7) and (8) on the display:
•No display
•Compensated or uncompensated mea-
sured value
•Temperature
•Output level 1 or 2
•Setpoint value 1 or 2
Operation
All parameters are displayed in clear levels
and in plain text to ensure ease of program-
ming and operation. Operation is protected
with a code word. Individual modification is
possible if parameters are generally approved
and/or are assigned to the protected area.
The setup program (optionally available) for
PC enables easy configuration of the device.
(1) (2) (4) (5)
(6) (7) (8) (9)
(3)
Measurement of free chlorine,
total chlorine, ozone, hydrogen
peroxide, peracetic acid
For the measurement of the most important
disinfectants, the following membrane-cover-
ed, amperometric sensors with digital inter-
face are available:
•JUMO tecLine Cl2 (free chlorine),
type 202630/50 and 202630/51
•JUMO tecLine TC (total chlorine),
type 202631/52
•JUMO tecLine O3 (ozone),
type 202634/65 and 202634/66
•JUMO tecLine H2O2 (hydrogen peroxide),
Typ 202636/80
•JUMO tecLine PAA (Peressigsäure),
Typ 202636/75
The sensors are designed for continuous me-
asurement operation. The sampling rate of the
AQUIS 500 RS is fixed at 500 ms when using
these sensors.
The temperature of the measuring medium
can be measured either internally, with the
sensor, or using an external temperature sen-
sor connected to the analog temperature in-
put of the JUMO AQUIS 500 RS.
Turbidity measurement
We recommend using the JUMO ecoLine
NTU sensor (type 202670) for measuring the
turbidity to monitor water with low to medium
turbidity.
This sensor operates based on the optical
(nephelometric) measurement with 90° scat-
tered light according to DIN EN ISO 7027 and
covers a total range of 0 to 4000 NTU with 4
measuring range sections. The measured val-
ues are called up by the JUMO AQUIS 500 RS
with a fixed sampling rate of 1 s with the sen-
sor. For displaying the measured value, the
user can choose between NTU and FNU.
The temperature of the measuring medium
can be measured either internally, with the
sensor, or using an external temperature sen-
sor connected to the analog temperature in-
put of the JUMO AQUIS 500 RS.
Measurement with freely configu-
rable sensor
To enable measurement with Modbus sensors
not included in the internal list of predefined
sensors, the sensor type can be selected as
"freely configurable sensor".
Using this sensor type allows the temperature
and the compensated and uncompensated
measured value to be read out. The sensor
must be capable of continuous independent
measurement.
To allow the sensor to be connected, the
Modbus addresses for the temperature value
in °C, the compensated and uncompensated
measured value, the byte sequence for trans-