JUMO ecoTRANS Lf 03 User manual

ecoTRANS Lf 03
Transmitter /
Switching Device
for Conductivity
Type 202732
B 20.2732.0
Operating Manual
02.05/00429233


1 Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2 Identifying the instrument version . . . . . . . . . . . . . . . . . . 6
2.1 Type designation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
5 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.1 Fundamentals of conductivity measurement . . . . . . . . . . . . . . . . . . 12
5.2 Applying the supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
6 Setting / altering the instrument functions . . . . . . . . . . 13
6.1 Actual-value display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.3 Function of keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.4 Level selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.5 The operator level (USER) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.6 The administrator level (ADMIN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
6.7 The enabling level (RIGHT) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6.8 The calibration level (CALIB) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
7 Setting ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
8 Configurable parameters . . . . . . . . . . . . . . . . . . . . . . . . . 23
8.1 Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8.1.1 Measurement input for conductivity (submenu CON) . . . . . . . . . . . . 24
8.1.2 Measurement input for temperature (submenu TEMP) . . . . . . . . . . . 28
8.2 Relay / open-collector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
8.2.1 Logic output 1 (submenu BIN.1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
8.2.2 Logic output 2 - option (submenu BIN.2) . . . . . . . . . . . . . . . . . . . . . 33
8.3 Analog outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
8.3.1 Conductivity (submenu CO.OUT) . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
8.3.2 Temperature (submenu TE.OUT) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
8 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
8.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
8.2 Activating the calibration mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8.3 Selecting the calibration procedure . . . . . . . . . . . . . . . . . . . . . . . . . 43
8.4 Calibrating the temperature coefficient using automatic
temperature measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

8.5 Calibrating the temperature coefficient using manual
temperature entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
8.6 Calibrating the relative cell constant . . . . . . . . . . . . . . . . . . . . . . . . 49
9 Analog output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
9.1 Response during calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
9.2 Response of the output signal in fault condition . . . . . . . . . . . . . . . 52
9.3 Response of the output signal on leaving the
scaling range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
9.4 Manual operation of the analog output . . . . . . . . . . . . . . . . . . . . . . 53
10 Relay output / open-collector . . . . . . . . . . . . . . . . . . . . . 54
10.1 Response of the relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
10.2 Manual operation of the relay outputs . . . . . . . . . . . . . . . . . . . . . . . 54
10.3 Response of the relay during calibration . . . . . . . . . . . . . . . . . . . . . 56
10.4 Pulse function of the relay output . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
10.5 Fault output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
11 The USP contact (for high-purity water) . . . . . . . . . . . . . 57
11.1 USP pre-alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
12 Display and LED messages . . . . . . . . . . . . . . . . . . . . . . . 59
12.1 Operating states of the JUMO ecoTrans Lf 03 . . . . . . . . . . . . . . . . . 59
12.2 Underrange . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
12.3 Overrange . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
12.4 Probe break . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
12.5 Short circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
12.6 Initialization of dependent parameters . . . . . . . . . . . . . . . . . . . . . . . 60
12.7 Calibration timer has run down . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
13 Operation via setup interface . . . . . . . . . . . . . . . . . . . . . 61
13.1 Operation through Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
14 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
15 Environment / waste disposal . . . . . . . . . . . . . . . . . . . . . 66

5
1 Notes
All necessary settings are described in this Operating
Manual. However, if any difficulties should still arise
during start-up, please do not carry out any
manipulations on the unit. You could endanger your
rights under the instrument warranty! Please contact
the nearest subsidiary or the head office in such a
case.
Please read this operating manual before starting up
the instrument. Keep the manual in a place which is
accessible to all users at all times. Please assist us to
improve this operating manual, where necessary.

6
2 Identifying the instrument version
The nameplate with the order code is glued to the side of the
housing. The supply voltage must correspond to the voltage given
on the nameplate.
2.1 Type designation
(1) Basic type
202732 JUMO ecoTRANS Lf 03,
Microprocessor transmitter / switching device for
conductivity (freely programmable ranges)
(2) Output I (conductivity / resistivity)
888 analog signal output, freely programmable
(3) Output II (temperature)
888 analog signal output, freely programmable
(4) Output III (switching)
101 1 x relay, changeover contact
177 2 x open-collector
(5) Extra codes
000 none
024 delivery includes PC setup software
999 customer-specific programming
(1) (2) (3) (4) (5)
Order code 202732 / 888 - 888 - /
Order example 202732 / 888 - 888 - 177 / 024

7
3 Installation
4 Electrical connection
The choice of cable, the installation, the fusing and the electrical
connection must conform to the requirements of VDE 0100
“Regulations on the Installation of Power Circuits with Nominal
Voltages below 1000 V” or the appropriate local regulations.
❏The electrical connection must only be carried out by qualified
personnel.
❏Electromagnetic compatibility conforms to the standards and
regulations cited in the technical data.
❏The instrument is not suitable for use in areas with an explosion
hazard (Ex areas).
Apart from faulty installation, incorrect settings on the instrument
may also affect the proper functioning of the subsequent process or
lead to damage. Safety devices should always be provided that are
independent of the instrument (such as overpressure valves or
temperature monitors/limiters) and only capable of adjustment by
specialist personnel. Please observe the relevant safety regulations
for such matters.
£ 75
-10
+60
°C

8
❏The load circuit must be fused for the maximum relay current, in
order to prevent the output relay contacts becoming welded in
the event of a short circuit.
❏Do not connect any additional loads to the screw terminals for
the supply of the instrument.
❏Any electrical connection other than that specified in the
connection diagram may result in the destruction of the
instrument.
❏Run input, output and supply cables separately and not parallel
to one another.
❏Probe leads must be implemented as twisted and shielded,
uninterrupted cables (do not run them via terminal blocks or
similar).
❏Supply fluctuations are only permissible within the specified
tolerances (see Data Sheet 20.2732).
❏The instrument must only be assembled or disassembled in the
de-energized state or when the cables are not connected.

9
Terminal assignment
Outputs Terminals Symbol
I
Analog
signal output for
conductivity
(electrically
isolated)
5
6
+
-
II
Analog
signal output for
temperature
(electrically
isolated)
7
8
+
-
P
K1
EXIT
4321
5876
13 L-L+14
9121110
56
+-
78
+-

10
III
Relay
1
3
4
common
n.c. (break)
n.o. (make)
Open-collector
output 1
(electrically
isolated)
1
3
GND
+
Open-collector
output 2
(electrically
isolated)
1
4
GND
A+
Measurement
inputs Terminals Symbol
Conductivity
cell
14
13
outer electrode,
on coaxial cells
inner electrode,
on coaxial cells
Resistance
thermometer
in 2-wire circuit
9
12
temperature sensor
Resistance
thermometer
in 3-wire circuit
9
11
12
temperature sensor
Resistance
thermometer
in 4-wire circuit
9
10
11
12
temperature sensor
413
PS
Ö
13
GND +
14
GND +
14 13
J
91
2
J
911
12
J
11 12
9 10

11
Connection of the conductivity cell
Supply Terminals Symbol
Supply
voltage
(with reverse-
polarity
protection)
L-
L +
Conductivity cell
(JUMO types)
JUMO
ecoTRANS Lf 03
Plug-in head Attached cable
Outer electrode white 14
Inner electrode 2 brown 13
Temperature
compensation
1
3
yellow
green
9
12
L- L+
-+

12
5 Commissioning
5.1 Fundamentals of conductivity measurement
Principle of measurement
Electrolytic conductivity measuring cells basically consist of two
metal plates arranged opposite each other which are immersed in the
solution to be measured. The conductivity of the solution is
determined through the measuring voltage and the resulting
measuring current.
The current between the metal plates depends on their geometry
(distance and area). The
cell constant
describes this dependence.
Subject to manufacturing tolerances, or because of dirt or wear, the
real cell constant of a measuring cell often deviates from its nominal
value. This deviation is reflected by the
relative cell constant
that is
adjustable on the conductivity transmitter.
The conductivity of a solution depends on temperature (i.e. the
conductivity of a solution increases with rising temperature). The
temperature coefficient
of the solution describes the dependence
of conductivity and temperature. Since conductivity is not always
measured at the reference temperature, automatic temperature
compensation has been integrated. The transmitter uses the
temperature coefficient to calculate the conductivity that would be
present at the reference temperature from the current conductivity
and the current temperature, and then display this value.
5.2 Applying the supply voltage
When the instrument has been correctly connected, all LCD
segments will be shown briefly, immediately after applying the supply
voltage.
Note
After initializing the controller, the output signal is 0 V or 0 mA. The
logic outputs or relays are in the quiescent state (inactive). After
approx. 2 sec, the JUMO ecoTRANS Lf 03 operates according to its
configuration.

13
6 Setting / altering the instrument functions
Alterations can be carried out in the setup program or from the keys
of the JUMO ecoTRANS Lf 03.
6.1 Actual-value display
The actual value is displayed either in the
- static mode or in the
- alternating mode
Static display (default setting)
Compensated conductance with current unit
Alternating display (fixed rhythm: approx. 3 seconds)
Compensated conductance with Current temperature with
current unit current unit
Change from static display (compensated conductivity only) to
alternating display
✱Press (for less than 2 sec)
Change from alternating display to static display
✱Press (for less than 2 sec)
approx. 3 sec
P
P

14
6.2 Operation
The operation of the instrument is arranged on levels.
The access to all levels (exception: operator level) is protected by
different codes2.
At the operator level (USER), all parameters can be viewed or
altered in accordance with the user rights1(see enabling level).
At the calibration level (CALIB), the cell constant and/or
temperature coefficient can be calibrated.
At the enabling level (RIGHT), the user rights can be defined.
At the administrator level (ADMIN), all parameters can be set
(configured).
The different codes and the settings at the enabling level make it
possible to assign different rights to the user.
1In the factory default setting, all parameters are set to READ,
which means that all parameters can only be read at the operator
level but not altered.
2The codes for the administrator and enabling levels can only be
altered through the setup program,
see Chapter 13 “Operation via setup interface”, page 61.
6.3 Function of keys
- Use the UP and DOWN keys to select a submenu in the main
menu or to scroll forwards or backwards.
- Use the P key to change to the corresponding submenu.
- If you want to alter (edit) a parameter, press the P key.
- If the parameter has been enabled for editing, the value flashes, if
the parameter is inhibited, LOCK is displayed.
After 60 seconds without operator action (key stroke)
the instrument jumps back to displaying the actual
value.
This time-out function is inactive during calibration!

15
- In order to alter a parameter, it must be enabled at the enabling
level (set from “rEAd” to “Edit”).
- Press the UP or DOWN key to increase or decrease the value.
- Accept the value by pressing the P key.
- Use the EXIT key to cancel the entry and change to the next-
higher level.

16
6.4 Level selection
1or time-out (automatic return after 60 sec without operator action)
see Chapter 6.5 “The
operator level (USER)”,
page 17
see Chapter 6.6 “The
administrator level
(ADMIN)”, page 18
see Chapter 6.7 “The
enabling level (RIGHT)”,
page 19
see Chapter 6.8 “The
calibration level
(CALIB)”, page 20

17
6.5 The operator level (USER)

18
6.6 The administrator level (ADMIN)

19
6.7 The enabling level (RIGHT)
Parameters at the operator level (USER)
Value is shown can be
altered
EDIT X X
READ X -

20
6.8 The calibration level (CALIB)
1see Chapter 8 “Calibration”, page 41.
2see Chapter 8.4 “Calibrating the temperature coefficient using
automatic temperature measurement”, page 44 or Chapter 8.5
“Calibrating the temperature coefficient using manual
temperature entry”, page 47.
3see Chapter 8.6 “Calibrating the relative cell constant”, page 49.
The time-out function is not active during calibration!
Table of contents
Other JUMO Transmitter manuals

JUMO
JUMO B 90.7027.1 User manual

JUMO
JUMO dTRANS T01 Junio User manual

JUMO
JUMO dTRANS T07 User manual

JUMO
JUMO CTI-750 User manual

JUMO
JUMO dTRANS p02 DELTA User manual

JUMO
JUMO dTRANS T09 DS User manual

JUMO
JUMO dTRANS T02 Ex User manual

JUMO
JUMO CTI-750 User manual

JUMO
JUMO dTRANS T02 PCP User manual

JUMO
JUMO JUMO dTRANS p20 DELTA User manual

JUMO
JUMO HART dTRANS p20 Ex d User manual

JUMO
JUMO B 707025.0 User manual

JUMO
JUMO AQUIS 500 CR User manual

JUMO
JUMO dTRANS O2 01 User manual

JUMO
JUMO AQUIS 500 pH User manual

JUMO
JUMO AQUIS 500 Ci User manual

JUMO
JUMO 404304 Series User manual

JUMO
JUMO dTRANS T06 Ex User manual

JUMO
JUMO dTRANS p02 DELTA User manual

JUMO
JUMO 40.2005 User manual