KPS 60 SS AND KPS 150 SS OPERATIONS MANUAL E MAN05312022
PAGE 5 OF 8
CALIBRATION
1. You need a test weight to perform this calibration.
2.
3.
Press and hold down the ZERO key and Tare key and then switch the power to ON which is on the left side of
the indicator, the display will show CAL then the A/D value.
Press TARE key. The display will show the calibration weight which you set.
4.Place the calibration weight on the scale’s platter.
5.Press the TARE key.
6.The display will show several dashes (- - - - -) and then the A/D value.
7.Remove the weight.
8.The calibration is now complete.
9.Troubleshoong (if calibraon is ineecve):
a. Try a more stable surface
b. Move to an interference free locaon
c. Replace baeries
If you are encountering problems calibrating your scale, please contact your dealer for further assistance. This is
necessary because the weight of a mass in one location is not necessarily the same in another location. Also, with time
and use, mechanical deviations may occur.
NOTE: Calibration has been done in factory already. Do not recalibrate it unless it is extremely inaccurate. If you have
any problems about calibration, please contact with the dealer who will provide you professional guidance.
FUNCTION SETTING
1.A-OFF:Time for auto-off (120s / 180s / 300s / OFF)
2.bL: Backlight (On/Auto/OFF/STB)
3. Units : Weighing units ( kg / lb/oz)
4.Por.u: Power on units ( kg / lb/oz/kg.lb.oz)
5.G.sw : Gravity (On/OFF)
6.Bmi.sw: Body mass index mode(On/OFF)
7.bAud : Baut rate (1200 / 2400/4800 /9600)
8.PAr: Parity bit(none/odd/EwEn)
9.Mod : Transmission mode(kEy/Stb/coN/OFF)
10. Ani,sw:Animal mode(On/OFF)
11. FiLd.L: Filter level(STr/WEEK/nor)
12. LocK.r:Locking range (10/30/50/OFF )
13. Pzr: Zero range when power on(3/4/10/20 )
14. Kzr:Key zero range(3/4/5/10 )
This scale requires 50% of the capacity in Test weights. Please make sure you have them PRIOR to performing the calibration.