KILOTECH INC.
www.kilotech.com 12
10. Percent Weighing Mode
10.1 In this mode, scale will weigh goods weight on it, calculate and display its percentage after the
unit-percentage-weight of goods is obtained.
(NOTE: If % display format is set to 100%, 100.0% or 100.00% in CONFIG-FUNC-PERCEN menu
item, then, the unit-percentage-weight is the weight of 1%, 0.1% or 0.01%)
10.2 To make percent weighing function be available, CONFIG-FUNC-PERCEN menu item shouldn’t be
set to NONE.
10.3 To enter percent weighing mode, in normal weighing or counting mode, long press FUNC key,
WEIGH/COUNT will be shown, use ACC or PRINT key to select PERCEN, then press TARE to
confirm go to percent weighing mode. The last set unit-percentage-weight will be used.
10.4 In percent weighing mode, the function of ZERO, TARE, PRINT, SETUP, PRESET TARE, ACC,
SETUP, ON/OFF are available.
10.5 There’re two ways to obtain the unit-percentage-weight: (1) input weight and its percentage, then
scale calculates the unit-percentage-weight, refer operation of step 10.5.1 (2) weigh samples weight
which percentage is known, refer operation of step 10.5.2
10.5.1 Input weight and its percentage from keypad, and calculate unit-percentage-weight: in percent
weighing mode, press UNIT key, When InP.Pct is shown, press TARE key to enter this mode: Before
input weight, use PRINT or ACC key to select the percentage from 1%, 2%, 5%, 10%, 20%, 50% and
100%, this percentage is corresponding to the weight you will input in following steps
10.5.1.2 When UNIT.KG is shown, use PRINT or ACC key to select the unit of input weight, use TARE key
to confirm and go to next. Press ZERO key to exit and back to percent weighing mode.
10.5.1.3 When last stored unit-percentage-weight data is shown, use PRINT, ACC, UNIT key to input new
unit-percentage-weight, press SETUP key more than 3s to input decimal point. Press TARE key to
confirm and save it, then go back to percent weighing mode. If the calculated
unit-percentage-weight is less than 0.5d, the indicator will display Pct.Er and go back to percent
weighing mode.
10.5.2 Obtain unit-percentage-weight by weighing samples weight which percentage is known: in percent
weighing mode, press UNIT key, When InP.Pct is shown, use PRINT or ACC key to select SPL.Pct,
press TARE key to weigh samples (which percentage is known) weight, calculate piece weight. Press
ZERO key to exit and back to percent weighing mode.
10.5.2.1 When SPL.Lo is shown, remove everything from the platter and press TARE key to confirm, before
scale is stable, SPL.Lo will be flashed. After it is stable, it will go to next step. Press ZERO key to
exit and back to percent weighing mode.
10.5.2.2 When SPL.Hi is shown, put samples (its percentage is known) onto the scale, Press TARE key to
confirm reading weight. Before scale is stable, SPL.Hi will be flashed. After it is stable, it will go to
next step. Press ZERO key to exit and back to percent weighing mode.
10.5.2.3 After INP.PCT being shown, 000000(position of decimal point is determined by