Kingfisher KI7343C-InGaAs User manual

kingfisherfiber.com
KI 7340C Series
PREMIUM TWO WAY OPTICAL LOSS TEST SET
WITH OPTICAL RETURN LOSS (ORL)
Optical Communications
Test Applications
•Attenuation testing
•System power testing
•ORL testing
•Continuity testing
Revision 24
The KI 7340C series is the industry's fastest
and easiest bi-directional loss tester. Average
fiber optic link loss and ORL each end is
automatically displayed in real time on both
instruments, at multiple wavelengths, via a
single fiber.
Featuring zero warm up, high speed and
high accuracy, results can either be stored in
internal memory, or inserted directly into a
customized acceptance report on a PC, with
one mouse click.
Detector & calibration options cover a wide
range of connector types, fiber types and
CWDM wavelengths from +27 to -70 dBm
with 1% Traceable Accuracy.
It is a robust, reliable and easy to use
instrument for high performance single
mode or multimode fiber optic cable testing.
Features
• Very high productivity
Reliable, rugged & field proven
Zero warm up & high accuracy
Full feature ORL testing
Autotest compatibility with other instruments
Mode controlled multimode sources
Multimode sources come with 50 & 62.5 µm fiber
mandrel wraps
3 ~ 7 year warranty
3 year calibration cycle
Interchangeable connectors
Long battery life
Large memory
Flexible real-time PC software
Instant Pass / Fail indication
Up to 4 LED or laser sources
Compact & light weight
Made in Australia

kingfisherfiber.com
KI 7340C Series - PREMIUM TWO WAY OPTICAL LOSS TEST SET WITH OPTICAL RETURN LOSS (ORL)
This is the industry fastest and easiest bi-directional loss test set. From start
of test to acceptance report takes one mouse click and 4 seconds per
wavelength.
The real-time loss and ORL display on both instruments means that cable
certification and rectification use the same procedure, which simplifies
training and operating procedures.
Autotest is available on both Test and Meter ports and is compatible with all
other Autotest instruments. When a pair of identical models are used, the
HOLD functions provide handy communication method between two users,
so they can synchronize while working through successive fibers.
High availability is the result of zero warm up, >190 hour battery life, patented
interchangeable optical connectors for both ports, 3 year calibration cycle
and superior reliability.
The instrument is also a standalone traceable power meter, multi- light source
and Optical Return Loss Tester.
The ORL Zero function compensates for residual reflections, and provides
extended measurement range with improved linearity.
The ORL User Calibration Mode compensates for stray losses in a test set-up,
which improves overall accuracy.
Multimode LED sources feature standards compliant beam geometry &
modal distribution across the fiber core, results in greatly improved
measurement accuracy.
The new InGaAs detectors have wider wavelength response range from 600
~ 1700 nm. It provides good response for all common wavelengths.
Flexible KITS™ PC software is a real-time measurement, Pass/Fail assessment
and reporting solution. Easily customized for any language and reporting
format, it also supports memory download, data logging, label printing,
legacy instruments and enterprise level data management.
POWER METER SPECIFICATIONS
Response
(Nm)
Damage
level
(dBm)
Calibration (nm)
Power range
(dBm)
Tone &
Autotest
Min
(dBm)
Midrange
linearity1
(d)
Calibration
Accuracy2
(%)
Polarization
Sensitivity
(dB)
Total
Uncertainty3
(dB)
Sensitivity ± 30
nm5(dB)
InGaAs detector
600 ~ 1700
+15
780, 820, 850, 980
1270, 1290, 1300, 1310,
1330, 1350, 1370, 1390,
1410, 1430, 1450, 1470,
1490, 1510, 1530, 1550,
1570, 1590, 1610, 1625,
1650
+5 ~ -60
+5 ~ -70
-45
-50
0.02
1 %
(0.06 dB)
< 0.005
0.3
0.03
H5 (InGaAs) detector
800 ~ 1700
+25
820, 850, 980
1270, 1290, 1300, 1310,
1330, 1350, 1370, 1390,
1410, 1430, 1450, 1470,
1490, 1510, 1530, 1550,
1570, 1590, 1610, 1625, 1650
+15 ~ -50
+15 ~ -60
-35
-40
0.02
1 %
(0.06 dB)
< 0.005
0.3
0.03
H3B (InGaAs) detector
800 ~ 1700
+304
820, 850, 980
1270, 1290, 1300, 1310,
1330, 1350, 1370, 1390,
1410, 1430, 1450, 1470,
1490, 1510, 1530, 1550,
1570, 1590, 1610, 1625, 1650
+27 ~ -40
+27 ~ -50
-25
-30
0.02
1 %
(0.06 dB)
< 0.005
0.35
0.03
Ge detector
600 ~ 1650
+25
780, 820, 850, 980
1270, 1290, 1300, 1310,
1330, 1350, 1370, 1390,
1410, 1430, 1450, 1470,
1490, 1510, 1530, 1550,
1570, 1590, 1610, 1625, 1650
+15 ~ -60
+15 ~ -60
-45
-50
0.04
1 %
(0.06 dB)
< 0.005
0.5
0.03
Si detector
350 ~ 1100
+5
635, 650, 660, 780, 850,
980
+0 ~ -70
-50
0.02
1 %
(0.06 dB)
< 0.005
0.3
0.03
typical
typical
max
typical
Note 1: Midrange linearity excludes top 5 dB and bottom 10 dB of range.
Note 2: Calibration condition: non coherent light, -35±5 dBm, 23±1°C, ±1 nm, 10±3 nm FWHM, PC ceramic connector, 100 µm fiber.
Note 3: Includes contributions of: varying optical connector types, calibration uncertainty, linearity over temperature & range, and fiber core diameter up to 200 µm.
Note 4: H3B can sustain the damage level for 2 minutes.
Note 5: At calibration wavelengths in bold type.

kingfisherfiber.com
LIGHT SOURCE SPECIFICATIONS
Parameters
1310/1550 nm laser
Other lasers
LED
Comments
2 source power
3 or 4
source power
-7 dBm
-10 dBm
-7 dBm
-10 dBm
-26 dBm to 62.5 µm,
-29 dBm to 50 µm,
-41 dBm to 10 µm
± 1 dB for Laser
± 3 dB for LED
Short term stability, dB
0.035
0.055
0.01
15 min, max, no warm up, 3C
Stability over temperature, dB
0.2
0.2
0.35
Max, over temperature
tolerance, nm
20
6.5
N/A
At 25 C
width, nm
3
< 1
N/A
FWHM, typical
Mode Controlled Source
N/A
N/A
Yes
Mode controlled6
nm/C
0.4
0.1
0.4
typical
Reconnection repeatability, dB
0.1
0.05
N/A
95 % confidence
Laser output adjustment
Adjustable over 6 dB in 0.01 dB steps
N/A
Modulation
270 Hz, 1, 2 KHz, 2 %
Note 5: For ORL < -25 dB.
Note 6: Multimode source mode distribution @ 50/125 is compliant with the following standards: IEC 61280-4-1 {Ed.1.0}, TIA/EIA 526-14A and TIA TSB-178.
ORL SPECIFICATIONS
Parameters
Laser
LED
No of wavelength
1 or 2
3 or 4
Measurement range7
0 ~ 65 dB
0 ~ 60 dB
0 ~ 40 dB (for 62.5 µm fiber)
0 ~ 37 dB (for 50 µm fiber)
Port isolation8
Standard > 30 dB; Optional > 50 dB
> 22 dB
ORL accuracy
0 ~ 50 dB: 0.5 dB
50 ~ 65 dB: 1 dB after zero
offset
0 ~ 45 dB: 0.5 dB
45 ~ 60 dB: 1 dB after zero
offset
0 ~ 30 dB: 0.5 dB
30 ~ 45 dB: 1 dB after zero offset
Resolution
0 ~ 50 dB: 0.01 dB
50 ~ 65 dB: 0.1 dB
0 ~ 45 dB: 0.01 dB
45 ~ 60 dB: 0.1 dB
0 ~ 30 dB: 0.01 dB
30 ~ 45 dB: 0.1 dB
available
See source options
Note 7: After a zero offset, measurement range will be 10 dB better than the connector or other residual ORL. So PC connectors have reduced range.
Note 8: Port Isolation is the passive return loss isolation of an ORL test port (with the instrument turned off). This is not relevant for manual ORL testing, when
only one active ORL meter is connected. However, in Autotest, the ORL range is limited to the port isolation value when the link attenuation is small. If the
standard 30 dB port isolation is inadequate, order the 50 dB port isolation built-in option -ISO50, e.g. KI 73416C-InGaAs-ISO50. In any test situation, the
Autotest ORL range limit (in dB) will be: port isolation (dB) + 2 x link loss (dB), up to the instrument specification.
Australian and international patents, technical data is subject to change without notice as part of our program of continuous improvements. Class 1
Laser/LED product, complies IEC60825-1 and 21CFR1040.10

kingfisherfiber.com
GENERAL SPECIFICATIONS
Parameters
Values
Battery life
360 hours Power Meter / 190 hours laser in Autotest
Size
190 x 130 x 70 mm, 7.5 x 5.1 x 2.8”
Weight
500 gm, 1.1 lb. Shipping 1.5 Kg, 3.3 lb.
Temperature / Relative Humidity
-15 to 55 °C (Operating), -25 to 70 °C (Storage) / 0 ~95%
LCD size
68 x 26 mm / 2.7 x 1.0 “
Hidden keypad
For setting advanced functions
Case
Polycarbonate, 1-meter drop tested on concrete
PC interface
USB Type B
Memory
1269/874/667 bi-directional 2 /3 /4 loss and ORL test results
Power
2 alkaline C cells (7.6 A/Hour); External DC with 2.5mm +ve pin or via USB
port. Selectable auto-off, low battery indicator, backlit display
Tone detection
150 ~ 9999 Hz 1 %
Pass/Fail
Insertion & Return loss pass/fail criteria can be set for all
Max / min
Recording feature for stability testing
Power meter resolution
0.01 dB
ORDERING INFORMATION
Description
Part number
P/N with Optical Return Loss (ORL)
Instrument, LTS-2W ORL 1310-1550 nm u/s9, InGaAs
Instrument, LTS-2W ORL 1310-1550 nm u/s, APC, InGaAs
KI7343C-InGaAs
KI7343C-InGaAs-APC
Instrument, LTS-2W ORL 850-1300 nm LED, 62.5 um, APC, Ge
Instrument, LTS-2W ORL 850/1300 nm LED, 50 µm, APC, Ge
KI7344C-Ge-APC
KI7344C-GE-APC-50u
Instrument, LTS-2W ORL 850 nm LED, APC, Si
Instrument, LTS-2W ORL, 850 nm LED, 50 µm APC, Si
KI73411C-Si-APC
KI73411C-Si-APC-50u
Instrument, LTS-2W ORL 1310-1490-1550 nm u/s, InGaAs
KI7347C-InGaAs
Instrument, LTS-2W ORL 1310-1490-1550 nm u/s, APC, InGaAs
KI7347C-InGaAs-APC
Instrument, LTS-2W ORL 1310-1550-1625 nm u/s, APC, InGaAs
KI73410C-InGaAs-APC
Instrument, LTS-2W ORL 1310-1490-1550-1625 nm u/s, APC, InGaAs
KI73416C-InGaAs-APC
P/N without Optical Return Loss (ORL)
Instrument, LTS-2W 1310-1625nm u/s, APC, InGaAs
KI7345nC-InGaAs-APC
Instrument, LTS -2W 1310-1550-1625 nm u/s, APC, InGaAs
KI73410nC-InGaAs-APC
Note 9: Ultra Stable laser source
STANDARD ACCESSORIES
Description
Quantity
Description
Quantity
SC connector adaptor (OPT046)
2
Carry Pouch
1
LC connector adaptor (OPT076)
2
Leather protective holster
1
ST connector adaptor (OPT040)
2
C cell batteries
1
SC PC Terminator to check ORL reading (OPT703)
1
& AA-to-C size battery converter (OPT101)
1
SC APC Terminator to check ORL reading (OPT704)
1
Operation manual
1
PC-to-APC Test Lead to check ORL reading (OPT705)
1
Quick reference guide
1
USB cable: A/B type
1
QA certificates
1 set
KITS™ recording/reporting software
Downloadable
free from website
Calibration certificates including: Power Meter, Light
Source, Two-way detector & ORL
1 set

AUTHORIZED DEALER
Kingfisher International Pty Ltd
720 Springvale Road, Mulgrave
VIC 3170 Australia
T+61 3 8544 1700
F +61 3 8544 1793
E sales@kingfisher.com.au
kingfisherfiber.com
OPTIONAL ACCESSORIES
Description
Part number
Option, Carry Case, KI2x/KI7x/KI3x, small
OPT153
Option, Carry Case, Cletop, Cleaning Sticks, KI2x/KI3x/KI7x, large
OPT154A
Option, Power Pack, KI7x IEC 100-240 V 2.5 mm Plug
OPT103B
Please visit kingfisherfiber.com for a wide range of FiberTester kits
OPTIONAL INTERCHANGEABLE CONNECTOR ADAPTORS
This instrument is supplied with metal-free sleeve optical interchangeable connector adaptors. The source ferrule type is fixed and customer specified as
either PC or APC. The power meter is for both PC & APC. Green is associated with APC. Green is associated with APC. You can order any number of
connector adaptors. Order quantity two of each type.
Description
Part number
Option, Hybrid Adaptor, Ceramic Sleeve, SC/ST
OPT040
Option, Hybrid Adaptor, Ceramic Sleeve, SC/FC
OPT051
Option, Hybrid Adaptor, Ceramic Sleeve, SC/D4
OPT055
Option, Hybrid Adaptor, Ceramic Sleeve, SC/LC, metal body
OPT076
Option, Hybrid Adaptor, Ceramic Sleeve, SC/E2000 Green
OPT060G
Option, Hybrid Adaptor, Ceramic Sleeve, SC/E2000
OPT060
Option, Hybrid Adaptor, Ceramic Sleeve, SC/LSA-DIN47256
OPT071
Option, Hybrid Adaptor, Ceramic Sleeve, SC/MU
OPT080
Option, Hybrid Adaptor, Ceramic Sleeve, SC/Universal 2.5 mm
OPT081
Option, Hybrid Adaptor, Metal Sleeve, SC/SMA 905/906
OPT082
Other manuals for KI7343C-InGaAs
1
This manual suits for next models
12
Table of contents
Other Kingfisher Test Equipment manuals

Kingfisher
Kingfisher KI 2400 Series Installation guide

Kingfisher
Kingfisher KI7343C-InGaAs User manual

Kingfisher
Kingfisher KI 2300 Series User manual

Kingfisher
Kingfisher KI7400 Series Installation guide

Kingfisher
Kingfisher KI021-1 User manual

Kingfisher
Kingfisher KI9800 User manual

Kingfisher
Kingfisher KI7347C Series User manual

Kingfisher
Kingfisher KI2000 Series Operating manual

Kingfisher
Kingfisher KI 2400 Series User manual

Kingfisher
Kingfisher KI7300 Series Instruction Manual