Dimensions (Units : mm)
340
216
127
11017
70
204200047.4
30.2
10
φ44
10
30.2
φ44
φ30
φ49
232200047.4
88
98
108
116
126
136
Main Body Universal Measuring Probe
LCD Flicker Measuring Probe
Small Universal Measuring Probe
Small LCD Flicker Measuring Probe
ISO screw
ISO screw
Tripod socket
ISO screw
ISO screw
Tripod socket
φ26
φ49
CA-210 (
Universal Measuring Probe
)
φ27 mm
±2.5˚
30±10 mm
0.01 to 1000 cd/m2
0.10 to 1000 cd/m2
0.10 to 0.99 cd/m20.2%+1 digit
1.00 to 1000 cd/m20.1%+1 digit
0.10 to 1000 cd/m2
0.10 to 4.99 cd/m2±0.008
for white
5.00 to 39.99 cd/m2±0.005
for white
40.00 to 1000 cd/m2±0.003
for white
160 cd/m2±0.002
for white
(±0.004 for monochrome)∗2
0.10 to 0.19 cd/m20.015 (2 σ)
0.20 to 0.49 cd/m20.008 (2 σ)
0.50 to 1.99 cd/m20.003 (2 σ)
2.00 to 1000 cd/m20.001 (2 σ)
5 (4.5) 0.10 to 3.99 cd/m2
20 (17 ) 4.00 cd/m2or higher
Main body: 340 (W) ×127 (H) ×216 (D)
mm
,
Probe: φ49 ×204
mm
Main body: 3.58 kg, Probe: 520 g
Detector: Silicon photo cell
By LED
Displayed in 4 or 3-digit value (Can be chosen)
±2%±1 digit of reading
(temperature : 23°C±2°C,relative humidity : (40±10)%)
-------
-------
-------
-------
-------
-------
-------
-------
-------
xyLv, XYZ, T∆uvLv, u'v'Lv, RGB analyze
Chromaticity is displayed up to fourth decimal place.
(Three-digit indication can be chosen.)
∆x∆y∆Lv, R/G B/G ∆G, ∆R B/R G/R
16 characters by 2 lines (with backlight)
NTSC, PAL, EXT, UNIV, INT
Vertical synchronizing frequency: 40 to 200 Hz
100 channels
Standard function
RS-232C (38,400 bps or below), USB (Rev.1.1)
Max. 5 points(Use 4-Probe Expansion Board CA-B14)
SDK software (supplied as standard accessory)
100 – 240 V~, 50–60 Hz, 50 VA
CA-210 (
Small Universal Measuring Probe
)
φ10 mm
±5˚
30±5 mm
0.01 to 3000 cd/m2
0.30 to 3000 cd/m2
0.30 to 2.99 cd/m20.2%+1 digit
3.00 to 3000 cd/m20.1%+1 digit
0.30 to 3000 cd/m2
0.30 to 14.99 cd/m2±0.008
for white
15.00 to 119.99 cd/m2
±0.005
for white
120.00 to 3000 cd/m2
±0.003
for white
160 cd/m2±0.002
for white
(±0.004 for monochrome)∗2
0.30 to 0.59 cd/m20.015 (2 σ)
0.60 to 1.49 cd/m20.008 (2 σ)
1.50 to 5.99 cd/m20.003 (2 σ)
6.00 to 3000 cd/m20.001 (2 σ)
5 (4.5) 0.30 to 11.99 cd/m2
20 (17 ) 12.00 cd/m2or higher
Main body: 340 (W) ×127 (H) ×216 (D)
mm
,
Probe: φ49 ×232
mm
Main body: 3.58 kg, Probe: 540 g
CA-210 (
LCD Flicker Measuring Probe
)
φ27 mm
±2.5˚
30±10 mm
0.01 to 1000 cd/m2
0.10 to 1000 cd/m2
0.10 to 0.99 cd/m20.2%+1 digit
1.00 to 1000 cd/m20.1%+1 digit
0.10 to 1000 cd/m2
0.10 to 4.99 cd/m2±0.005
for white
5.00 to 19.99 cd/m2±0.004
for white
20.00 to 1000 cd/m2±0.003
for white
160 cd/m2±0.002
for white
(±0.004 for monochrome)∗2
0.10 to 0.19 cd/m20.010 (2 σ)
0.20 to 0.49 cd/m20.005 (2 σ)
0.50 to 0.99 cd/m20.002 (2 σ)
1.00 to 1000 cd/m20.001 (2 σ)
5 (4.5 ) 0.10 to 1.99 cd/m2
20 (17 ) 2.00 cd/m2or higher
Main body: 340 (W) ×127 (H) ×216 (D)
mm,
Probe: φ 49 ×204
mm
Main body: 3.58 kg, Probe: 520 g
5 cd/m2or higher
0.0 to 100 %
±1 % (Flicker frequency: 30 Hz AC/DC 10 % sine wave)
±2 % (Flicker frequency: 60 Hz AC/DC 10 % sine wave)
1 % (2 σ) (Flicker frequency: 20 to 65 Hz AC/DC 10 % sine wave)
5 cd/m2or higher
±0.5 dB (Flicker frequency: 30 Hz AC/DC 10 % sine wave)
0.3 dB (2 σ) (Flicker frequency: 30 Hz AC/DC 10 % sine wave)
16 measurements/sec. (16 measurements/sec.)
0.5measurements/sec. (0.3 measurements/sec.) ∗5
xyLv, XYZ, T∆uvLv, u'v'Lv, RGB analyze
Chromaticity is displayed up to fourth decimal place.
(Three-digit indication can be chosen.)
Flicker (Contrast method) ∗3
∆x∆y∆Lv, R/G B/G ∆G, ∆R B/R G/R, Flicker (Contrast method) ∗3
Vertical synchronization frequency: 40 to 200 Hz (Luminance or chromaticity measurement), 40 to 130 Hz (Flicker measurement)
CA-210 (
Small LCD Flicker Measuring Probe
)
φ10 mm
±5˚
30±5 mm
0.01 to 3000 cd/m2
0.30 to 3000 cd/m2
0.30 to 2.99 cd/m20.2%+1 digit
3.00 to 3000 cd/m20.1%+1 digit
0.30 to 3000 cd/m2
0.30 to 14.99 cd/m2±0.005 for white
15.00 to 59.99 cd/m2±0.004 for white
60.00 to 3000 cd/m2±0.003 for white
160 cd/m2±0.002 for white
(±0.004 for monochrome)∗2
0.30 to 0.59 cd/m20.010 (2 σ)
0.60 to 1.49 cd/m20.005 (2 σ)
1.50 to 2.99 cd/m20.002 (2 σ)
3.00 to 3000 cd/m20.001 (2 σ)
15 cd/m2or higher
15 cd/m2or higher
5 (4.5) 0.30 to 5.99 cd/m2
20 (17) 6.00 cd/m2or higher
Main body: 340 (W) ×127 (H) ×216 (D)
mm
,
Probe: φ49 ×232
mm
Main body: 3.58 kg, Probe: 540 g
Item
Receptor
Measurement area
Acceptance angle
Pointing function
Measurement distance
Display range Luminance
Chromaticity
Luminance
Measurement range
Accuracy (for white)
∗1
Repeatability(2 σ)
∗1
Chromaticity
Measurement range
Accuracy∗1
(temperature : 23°C±2°C,
relative humidity : (40±10)%)
Repeatability∗1
Flicker Contrast
Measurement range
method Display range
Accuracy
Repeatability
Flicker JEITA
Measurement range
method ∗3Accuracy
Repeatability
Measurement xyLv∗4
speed
(measurements/sec. )
Flicker Contrast
Flicker JEITA∗3
Display Digital
Analog
LCD
SYNC mode
Object under measurement
Memory channel
Analyzer function
Interface
Multi-point Measurement
Software
Operating temperature/humidity range
Storage temperature/humidity range
Input voltage range
Size
Weight
Specifications
∗1: The chromaticity and luminance are measured under Konica Minolta’s condition (standard LCD(6500 K, 9300 K) is used).
∗2: The luminance for monochrome is measured when the reading of luminance for white is 160 cd/m2.
∗3: Measurement of flicker (JEITA method) is supported by SDK software.
∗4:
Measuring probe connected to probe connector P1 only,used USB (used RS-232C
Baud rate: 38400 bps
)
∗5: Measured by Konica Minolta’s PC (P3-600 MHz)
• Specifications are subject to change without notice.
0 to 28°C : relative humidity 70 % or less with no condensation 28 to 40°C : relative humidity 40 % or less with no condensation
Temperature : 10 to 28°C; relative humidity 70 % or less with no condensation Luminance change : ±2 % ±1 digit of reading for white
Chromaticity change ±0.002 for white, ±0.006 for monochrome from reading of Konica Minolta's standard LCD∗1, 160.0 cd/m2, with 23°C 40 %