Konica Minolta CM-2600d User manual

SPECTROPHOTOMETER
CM-2600d
Portable, Compact, Easy to Use
Performs Like a Desktop Spectrophotometer
World's first portable spectrophotometer equipped with automatic
UV adjustment function.
Advanced Numerical UV Control dramatically reduces evaluation time.
Mac OS 9.2J illustrator 10J
CM-2600d_cover.ai 2006.11.21
MAV ø8mm
SAV ø3mm
For plastics, paints, resins and consumer products
For textile, paper, fluorescent and consumerproducts
For pharmaceuticals, cosmetics, printing, building materials, textiles etc.
Relativity Gloss Value
Numerical UV Control
Numerical Gloss Control
Expanded LCD display (64 x 240 dots)
Displays a large quantity of information
(simultaneous display of SCI and SCE data etc.)
High-accuracy sensor
Measures at 10nm intervals
for the full wavelength range.
Excellent repeatability
Easy-to-carry, compact
and lightweight
670g (without batteries)
In various applications
Powerful partnership between
CM-2600d and
Simultaneous measurement of SCI and SCE displays data on the LCD in 1.5 seconds. Unlike conventional spectrophotometers, there is no need
to switch between SCI and SCE mode. This improves working efficiency, and provides stable measured data since the measurement area does
not shift when the mode is switched. And also Relativity Gloss Value can be displayed by using Numerical Gloss Control.
• SCI is a method in which measurements are taken with
the specular reflection included. For this reason, it
minimizes influences of the surface condition of a sample,
and is especially suitable for color quality control and
Computer Color Matching.
• SCE is a method in which measurements are taken
excluding the specular reflection. For this reason, it
provides measurement results similar to those observed
by the human eye.
Light sources including and excluding UV component flash sequentially to provide
sample data taken with UV-included energy as well as UV-excluded energy
(UV400nm cutoff filter)
The battery-powered small, compact body allows the instrument to be placed in any position at the
sample surface.
The large LCD reverse video display provides easy reading, irrespective of which hand it is held in.
Using your finger, the
navigation wheel allows
simple and user-friendly
operation.
( Turn) ( Push)
The optics use an integrating sphere to provide diffuse
illumination/8-degree viewing system.
The CM-2600d conforms to all widely accepted industry standards
including ISO, JIS, DIN, CIE and ASTM, and generates
measurements in color spaces such as L*a*b*, Yxy, Munsell and CMC.
Example of data taken with UV-excluded energy Example of data taken after UV adjustment
Promotes accurate color communication.
Conforms to all widely accepted industry
standards and allows measurements in
all commonly used color spaces.
The instrument is portable and it allows
measurements to be taken using two
different areas of view (ø8mm and
ø3mm). The user can choose the most
suitable measurement area for the
target. The lightweight, easy-to-carry body
with the illuminated viewfinder enables
the user to position the instrument on the
target quickly and accurately.
Measures the target with
high accuracy.
Easy-to-carry stylish body with
an illuminated viewfinder.
Allows measurement in any position.
Compact, lightweight with an easy-to-operate navigation
wheel and large LCD display.
World's first portable spectrophotometer equipped
with instantaneous UV adjustment function.
UV evaluation time has been reduced
revolutionarily due to the introduction of
advanced Numerical UV Control.
Simultaneous measurement of SCI (specular component included) and SCE
(specular component excluded).
Advanced Numerical Gloss Control.
The number of moving parts used in the instrument is minimized through the
introduction of numerical control technology. The CM-2600d can be used with
confidence, since it has been developed, manufactured and calibrated to meet
ISO 9001 requirements.
High reliability and long life.
Maintenance-free design.
Light source for SCI
For paints, plastics, automobiles, ceramics, architectural interiors, textile, paper, food etc.
UV included
UV excluded
Finder
Mac OS 9.2J illustrator 10J
CM-2600d_inside.ai 2006.11.21
The CM-2600d is a portable integrating sphere spectrophotometer designed for
versatility in various applications.
UV adjustment function, data can be easily taken using the desired light source (e.g. D65, D50, C).
UV calibration can be performed simply by measuring a standard with known fluorescent values
under the desired light source.
Once the UV calibration is completed,
sample data can be taken by measuring
the sample under the desired light
source. Since the conventional UV
adjustment that requires the UV cutoff
filter to be moved is not necessary,
measurement time can be reduced
drastically.
(Note on UV-Adjustment : Numerical UV-
calibration requires for the adjustment
calculation
SpectraMagicTM
NX
.)
Since the instrument has a
built-in instantaneous
The LCD specifications are subject to
change without prior notice.
Color Data Software
Light source for evaluation
of specular component
Illuminated viewfinder
d/8 integrating sphere optics that
conform to industry standards
SpectraMagicTM NX enables you to perform comprehensive color
inspection and analysis of incoming raw materials, in process
production, and outbound color critical goods and materials in
virtually any industry. With SpectraMagicTM NX you can insert digital
images with measured data. Measure samples in any of 8
universally accepted color spaces. Select from 15 illuminants, and
up to 40 indices to determine specific color and appearance
properties, such as strength, brightness, haze, yellowness, opacity
and strength. You can even configure up to 3 customized color
equations. Reports range from simple Pass/Fail to trend charts,
histograms, color plots, and spectral graphs. SpectraMagicTM NX
comes with predefined templates using skin technology, or you can
create your own templates. For illustrations and explanations to
understanding color and color measurement technology, there is a
link to Konica Minolta's well known and respected “Precise Color
Communication”. Step by step navigation help. SpectraMagicTM NX
conforms to FDA 21 CFR Part 11 assuring integrity and reliability of
data records.
•Windows®is a trademark of Microsoft Corporation in the USA and other countries.
(Optional)
Supports Windows®2000/XP
SpectraMagicTM
NX
SpectraMagicTM NX
AGMGPK
Printed in Japan9242-4879-11
2001 KONICA MINOLTA SENSING, INC.
SAFETY PRECAUTIONS
For correct use and for your safety, be sure to read
the instruction manual before using the instrument.
Always connect the instrument to the
specified power supply voltage. Improper
connection may cause a fire or electric shock.
Be sure to use the specified batteries. Using
improper batteries may cause a fire or electric
shock.
9
Certificate No : JQA-E-80027
Registration Date : March 12, 1997
Certificate No : YKA 0937154
Registration Date : March 3, 1995
System Diagram
Standard Accessories
*1 Operating temperature/humidity range of products for North America : 5 to 40˚C, relative humidity 80% or less (at 31˚C) with no condensation
Specifications
Optional Accessories
193
69
96
AA-size
battery (x4)
AC Adapter AC-A17
AC Adapter AC-A23
(for Korea)
Vinyl cover
Dust cover CM-A152
Target mask ø3
mm
CM-A147
Target mask ø8
mm
CM-A146
CM-2600d
RS232C cable
IF-A16
CM-S100w
Hard case CM-A148
Dust cover set
CM-A149
Zero calibration box
CM-A32
White calibration plate
CM-A145
d/8 (diffuse illumination, 8-degree viewing ), equipped with simultaneous
measurement of SCI (specular component included)/SCE (specular
component excluded) Conforms to CIE No.15,ISO 7724/1,ASTM E1164,
DIN 5033 Teil7 and JIS Z8722 Condition C standard.
ø52mm
Silicon photodiode array (dual 40 elements)
Diffraction grating
360nm to 740nm
10nm
Approx. 10nm
0 to 175%, resolution: 0.01%
3 pulsed xenon lamps
Approx. 1.5 seconds (approx. 2 seconds for fluorescent measurement)
3 seconds for SCI/SCE (4 seconds for fluorescent measurement)
Alkaline manganese:approx. 1000 measurements
MAV: ø8mm/ø11mm
SAV: ø3mm/6mm (Selectable between MAV and SAV)
Spectral Reflectance:Standard deviation within 0.1% (360 to 380nm within 0.2%)
Colorimetric Value : Standard deviation within ∆E*ab 0.04(Measurement
conditions:White calibration plate measured 30 times
at 10-second intervals after white calibration was performed)
within ∆E*ab 0.2 (MAV/SCI) Average for 12BCRA Series
II
color
tiles compared to values measured with master body.
Instantaneous numerical adjustment (no mechanical adjustment required)
With UV400nm cut filter
Single measurement/automatic averaging of multiple measurements
(auto mode: 3, 5, 8 times/manual mode)
RS-232C standard
2/10 degrees (CIE 1931/2°,CIE 1964/10°)
A, C, D50, D65, F2, F6, F7, F8, F10, F11, F12 (simultaneous evaluation
is possible using two light sources)
Spectral value/graph, colorimetric value, color difference value/graph, PASS/FAIL result
L*a*b*, L*C*h, CMC (1:1), CMC (2:1), CIE94, Hunter Lab, Yxy, Munsell, XYZ,
MI, WI (ASTM E313), YI (ASTM E313/ASTM D1925), ISO Brightness (ISO 2470),
Density status A/T, WI/Tint (CIE/Ganz), CIE0
0
1700 pieces of data (as SCI/SCE 1 data) * 700 pieces of data in the “ defined in COND.” mode.
* Total of the sample data for the COND and TASK modes and color difference target data
Tolerance for color difference (both box and eliptical tolerances can be set)
4 AA-size battery or AC adapter
69 x 96 x 193mm
Approx. 670g (without batteries)
5 to 40˚C, relative humidity 80% or less (at 35˚C) with no
condensation
0 to 45˚C, relative humidity 80% or less (at 35˚C) with no
condensation
White calibration plate, Target mask ø8mm, Target mask ø3mm ,
RS-232C cable, AC adapter, AA-size battery (x4)
Hard case, Dust cover set, Dust cover,
SpectraMagic
TM
NX
(software), Zero calibration box
Mac OS 10.3.7J illustrator 10J
CM-2600d(E)_back.ai 2006.11.21
Personal Computer
(commercially available)
Color control network by spectrophotometer
High inter-instrument agreement between the
portable CM-2600d spectrophotometer and the
desk top CM-3000 series make it easy to build a
total color control network.
R&D
Quality control
Production
CM-3700d
CM-3600d
Company
B
In-house color control Color control among companies
R&D
Quality
Production
Company
A
R&D
Quality
Production
CM-2600d CM-2600d
Dimensions (Units:mm) CM-2600d
Illumination/
observation system
Sphere Size
Light-receiving element
Spectral separation device
Wavelength range
Wavelength pitch
Half bandwidth
Reflectance range
Light source
Measurement time
Minimum interval
between measurements
Battery perfomance
Measurement/
illumination area
Repeatability
Inter instrument
agreement
UV adjustment
Measurement
mode
Interface
Observer
Illuminant
Display data
Color space/
colorimetric data
Data memory
Tolerance judgment
Power source
Size (WxHxD)
Weight
Operating temperature/
humidity range
(*1)
Storage temperature/
humidity range
Standard
accessories
Optional
Accessories
SpectraMagicTM NX
* The specifications and
drawings given here are
subject to change without
prior notice.

Other manuals for CM-2600d

1

Other Konica Minolta Measuring Instrument manuals

Konica Minolta CM-3700D User manual

Konica Minolta

Konica Minolta CM-3700D User manual

Konica Minolta CA-210 User manual

Konica Minolta

Konica Minolta CA-210 User manual

Konica Minolta CM-23d User manual

Konica Minolta

Konica Minolta CM-23d User manual

Konica Minolta CM-36dGV User manual

Konica Minolta

Konica Minolta CM-36dGV User manual

Konica Minolta CM-26dG User manual

Konica Minolta

Konica Minolta CM-26dG User manual

Konica Minolta VIVID 9i User manual

Konica Minolta

Konica Minolta VIVID 9i User manual

Konica Minolta CM3600D User manual

Konica Minolta

Konica Minolta CM3600D User manual

Konica Minolta CR-10 Plus User manual

Konica Minolta

Konica Minolta CR-10 Plus User manual

Konica Minolta CL-70F User manual

Konica Minolta

Konica Minolta CL-70F User manual

Konica Minolta CM-700d User manual

Konica Minolta

Konica Minolta CM-700d User manual

Konica Minolta REGIUS 110 User manual

Konica Minolta

Konica Minolta REGIUS 110 User manual

Konica Minolta CHROMA METER CS-200 User manual

Konica Minolta

Konica Minolta CHROMA METER CS-200 User manual

Konica Minolta CECF-9 User manual

Konica Minolta

Konica Minolta CECF-9 User manual

Konica Minolta CS-3000HDR User manual

Konica Minolta

Konica Minolta CS-3000HDR User manual

Konica Minolta CM2500C User manual

Konica Minolta

Konica Minolta CM2500C User manual

Konica Minolta JM-103 User manual

Konica Minolta

Konica Minolta JM-103 User manual

Konica Minolta CA-310 User manual

Konica Minolta

Konica Minolta CA-310 User manual

Konica Minolta CM-3700A User manual

Konica Minolta

Konica Minolta CM-3700A User manual

Konica Minolta CIVI-36clG User manual

Konica Minolta

Konica Minolta CIVI-36clG User manual

Konica Minolta SPECTRAMAGIC NX - User manual

Konica Minolta

Konica Minolta SPECTRAMAGIC NX - User manual

Konica Minolta CR-10 Plus User manual

Konica Minolta

Konica Minolta CR-10 Plus User manual

Konica Minolta CM-3630 User manual

Konica Minolta

Konica Minolta CM-3630 User manual

Konica Minolta CS-150 User manual

Konica Minolta

Konica Minolta CS-150 User manual

Konica Minolta CM-36d User manual

Konica Minolta

Konica Minolta CM-36d User manual

Popular Measuring Instrument manuals by other brands

Electro Industries/GaugeTech Shark 100T Installation & operation manual

Electro Industries/GaugeTech

Electro Industries/GaugeTech Shark 100T Installation & operation manual

TRUGLO TG7630G manual

TRUGLO

TRUGLO TG7630G manual

Novatech 1638 manual

Novatech

Novatech 1638 manual

Dremel Micro 8050 Operating/safety instructions

Dremel

Dremel Micro 8050 Operating/safety instructions

PCB Piezotronics PCB-(M)356A70 Installation and operating manual

PCB Piezotronics

PCB Piezotronics PCB-(M)356A70 Installation and operating manual

HEIDENHAIN ND 231B user manual

HEIDENHAIN

HEIDENHAIN ND 231B user manual

Spectra Precision HD150 user manual

Spectra

Spectra Precision HD150 user manual

Endress+Hauser MODBUS RS485 operating instructions

Endress+Hauser

Endress+Hauser MODBUS RS485 operating instructions

Avtech AVO-6HF-B instructions

Avtech

Avtech AVO-6HF-B instructions

KPS DCM3000 instruction manual

KPS

KPS DCM3000 instruction manual

PLS HR 1000 operating manual

PLS

PLS HR 1000 operating manual

A-Plus Water DuraLine Guide

A-Plus Water

A-Plus Water DuraLine Guide

VibrAlign Fixturlaser Laser Kit Quick tips

VibrAlign

VibrAlign Fixturlaser Laser Kit Quick tips

Micronics portaflow 220 user manual

Micronics

Micronics portaflow 220 user manual

Advanced Instruments GPR-IN190 owner's manual

Advanced Instruments

Advanced Instruments GPR-IN190 owner's manual

Davis Instruments Electronic Rain Counter instruction manual

Davis Instruments

Davis Instruments Electronic Rain Counter instruction manual

Raycus XZ Series user guide

Raycus

Raycus XZ Series user guide

INTEGRA Merering Calec energy master Installation and operating instructions

INTEGRA Merering

INTEGRA Merering Calec energy master Installation and operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.