2-8-611HAN/2-8-611CAS HOOKUP CONNECTION GUIDE
Using MEDIVATORS®ADVANTAGE®Automated Endoscope Reprocessors
Port ID Port Name Connector Assembly Description Order Number
1LEAK TEST Leak Test Connector Assembly H28611-1
2JET/AUX Air/Water Connector Assembly H28611-26
3BIOPSY Biopsy Connector Assembly H28611-3
4AIR Air Probe Connector Assembly H28611-4
5ELEVATOR Elevator Wire/Auxiliary Water Connector Assembly H28611-5
6WATER Air/Water Connector Assembly H28611-26
7SUCTION Suction Connector Assembly H28611-7
ANA Air/Water & Suction Port Separator 1-5-112
If a Connector Assembly needs replacement, locate the port number/name and use
the corresponding order number found in the Connector Assembly Identication Chart.
Connector Assemblies are specic to individual hookups. Check each Hookup Connection
Guide for relevant information.
ADVANTAGE HOOKUP and CHANNEL SEPARATOR (Figure 1)
* Representative endoscope for illustrative purposes only. For exact
port locations, refer to endoscope manufacturer’s instruction guide.
Each Advantage Hookup is comprised of multiple ports and connector assemblies. Each
port is labeled with a Port ID number (circled). This same number also identies the
corresponding Connector Assembly. Each Connector Assembly is tted with a specic
endoscope connector. (see Fig. 2).
EXAMPLE CONNECTOR ASSEMBLY (Figure 2)
READ ALL INSTRUCTIONS PRIOR TO CONNECTION
Locate the endoscope port* and follow the instructions for attaching the
corresponding Endoscope Connector. All connectors must be attached.
CONNECTOR ASSEMBLY IDENTIFICATION CHART
Ferrule Endoscope
Connector
example
Tubing
Connector
Coupling
Label
Ensure proper alignment of all connectors with the
endoscope mating parts prior to attachment.


A



Leak Test Connector
Align the slot on connector with the pin on the
scope. Push on and turn clockwise to engage.
To remove leak test connector, turn connector
counter-clockwise until slot is aligned with pin and
pull off. Do not pull until slot is aligned with pin.

Suction Connector
Align silicone tubing with
suction barb and push tubing
onto barb until fully engaged.

Elevator Wire Channel Connector
(endoscope model dependent)
Align the connector with the endoscope tting
and screw clockwise until fully engaged. (Check
alternate location ve)

Air/Water and Suction
Port Separator
Align the separator with
piston ports and push to
compress springs before
sliding the frame forward to
fully engage.
AAir Probe Connector
Align with Air Probe, push and
slide until fully engaged.

Biopsy Connector
Align the biopsy
connector disc over the
biopsy inlet and push
down to compress spring
and slide the frame
forward to fully engage.

Air/Water Connector
Align the connector
openings with the air/water
ports of the endoscope as
shown. Ensure side latch
is in. Push connector onto
ports until latch moves
out and clicks. To remove,
push latch to release.

Auxiliary Water Connector
(endoscope model dependent)
Align the connector with the endoscope
tting and screw clockwise until fully
engaged. (Check alternate location ve)

If you have any questions regarding this hookup, in the U.S., please contact MEDIVATORS Technical Support
at 1-800-444-4729. Outside the U.S., please contact your MEDIVATORS Distributor.
DESIGNATED HOOKUP FOR THE FOLLOWING OLYMPUS®
ENDOSCOPES*:
CF-1T140I CF-HQ290I CF-Q180AL JF-140 PJF-240
CF-1T140L CF-HQ290L GIF-1T140 JF-140F TJF-140
CF-FH260AZI CF-Q145I GIF-1T140L JF-140R TJF-140F
CF-FH260AZL CF-Q145L GIF-1T160 JF-1T40 TJF-140R
CF-H180AI CF-Q160 GIF-1T240 JF-240 TJF-145
CF-H180AL CF-Q160AI GIF-1TQ160 PCF-H180AI TJF-160F
CF-H80DI CF-Q160AL GIF-H180J PCF-H180AL TJF-160R
CF-H185I CF-Q160I GIF-H190 PCF-H190I TJF-160VF
CF-H185L CF-Q160L GIF-H190J PCF-H190L TJF-160VR
CF-H190I CF-Q160S GIF-HQ190 PCF-Q180AI TJF-240
CF-H190L CF-Q165I GIF-HQ290 PCF-Q180AL TJF-260V
CF-HQ190I CF-Q165L GIF-XK240 PJF-16 TJF-M20
CF-HQ190L CF-Q180AI GIF-XTQ160 PJF-160
* This list of corresponding endoscope models is subject to change without notice. For a full
list, please go to:
www.medivators.com/hookuplookup
© 2014 MEDIVATORS Inc. 50097-585/H
HOOKUP CONNECTOR BLOCK LABEL
The Hookup Connector Block Label contains useful information for quick reference.
Handle Locking Instructions - insert block onto basin channel manifold and move handle
to the left to lock. Move to the right to unlock.
Lot Number - identies date of manufacture.
Hookup Part Number - identies hookup.
Hookup Barcode - barcode of part number.
Port Name - identier of port location in ADVANTAGE system. Error logs identify issues using
these names. Also identier of Connector Assembly replacements for older model hookups.
Port ID Number - identier for Connector Assembly replacements. Not found on older
model hookups.
Alternative Channel Connector Names - connector names vary depending on specic
endoscope channel conguration.
WARNINGS & CAUTIONS
This hookup is intended for connection only with the specic endoscope models
identied in the appropriate MEDIVATORS online hookup guide. Use of this hookup
with endoscopes other than those specied by the hookup application guide may
result in inadequate HLD.
Users operating MEDIVATORS AERs and hookups must be trained and competent in
the understanding of endoscope channel systems.
Prior to reprocessing, users must verify that the hookup used contains connections
for all appropriate channels that require a separate reprocessing connection
according to the hookup instructions
All connectors and adapters must remain rmly attached and unrestricted for the
entire disinfection cycle to ensure adequate HLD. Users must inspect the endoscope
and hookup combination to verify proper connection and ow.
Modications or repairs to hookups which do not correspond to the manufacturer’s
specications may result in inadequate HLD and/or damage to endoscopes.
Failure to properly clean and prepare endoscope for immersion prior to hookup
connection may result in inadequate HLD and/or damage to the endoscope.
Endoscopes must be inspected for damage and veried to be in proper working
order prior to hookup connection. Connection of hookups to damaged endoscopes
or endoscopes with obstructed/restricted channels could result in inadequate HLD.
Manual leak testing must be carried out in accordance with manufacturer’s
recommendations prior to reprocessing.
Install all necessary waterproof caps, plugs and cleaning adapters prior to
immersion.
Do NOT autoclave MEDIVATORS hookups.
MEDIVATORS makes no claim on high-level disinfection efcacy when these
reprocessing instructions are not followed, or when this hookup kit is applied to
endoscopes other than those specied in the online interactive hookup guide.
Avoid biological contamination. Always wear personal protective equipment when
handling endoscopes.
CASSETTE-STYLE HOOKUPS (not available in US)
An ADVANTAGE Hookup Cassette combines an endoscope tray with the hookup connector
block. Specic endoscope connections for this hookup are identical to non-cassette
versions. Cassette-style hookup part numbers are designated with a CAS sufx.
ADVANTAGE Hookup Cassettes are available outside the United States only. Connector
Assemblies are listed in the Connector Assembly Identication Chart.
LOT: XXXXXX ID: xxxX
P/N: X-X-XXXHAN
Hookup Part Number
Hookup
Barcode
Port ID
Number
Port Name
Alternative
Channel
Connector
Names
Handle Locking
Instructions
Lot Number

This manual suits for next models

1

Other Medivators Medical Equipment manuals

Medivators ADVANTAGE PLUS Operating instructions

Medivators

Medivators ADVANTAGE PLUS Operating instructions

Medivators 2-8-633HAN Instruction Manual

Medivators

Medivators 2-8-633HAN Instruction Manual

Medivators SCOPE BUDDY User manual

Medivators

Medivators SCOPE BUDDY User manual

Medivators DSD-201 User manual

Medivators

Medivators DSD-201 User manual

Medivators DSD EDGE User manual

Medivators

Medivators DSD EDGE User manual

Medivators CER-1 User manual

Medivators

Medivators CER-1 User manual

Medivators DSD-201 Operating instructions

Medivators

Medivators DSD-201 Operating instructions

Medivators SCOPE BUDDY PLUS User manual

Medivators

Medivators SCOPE BUDDY PLUS User manual

Medivators CER-1 User manual

Medivators

Medivators CER-1 User manual

Medivators DSD-201 User manual

Medivators

Medivators DSD-201 User manual

Medivators ADVANTAGE PLUS Technical specifications

Medivators

Medivators ADVANTAGE PLUS Technical specifications

Medivators SCOPE BUDDY Series User manual

Medivators

Medivators SCOPE BUDDY Series User manual

Medivators ADVANTAGE PLUS User manual

Medivators

Medivators ADVANTAGE PLUS User manual

Medivators Advantage Instruction Manual

Medivators

Medivators Advantage Instruction Manual

Medivators DSD EDGE User manual

Medivators

Medivators DSD EDGE User manual

Medivators DSD-201 User manual

Medivators

Medivators DSD-201 User manual

Popular Medical Equipment manuals by other brands

BIONIME Rightest GD500 Instructions for use

BIONIME

BIONIME Rightest GD500 Instructions for use

Stryker Medical Trio 1033 Operation manual

Stryker Medical

Stryker Medical Trio 1033 Operation manual

KaWe SwiSto3 user manual

KaWe

KaWe SwiSto3 user manual

Smiths Medical acapella duet Vibratory PEP Therapy System reference guide

Smiths Medical

Smiths Medical acapella duet Vibratory PEP Therapy System reference guide

Stryker SV2 Maintenance manual

Stryker

Stryker SV2 Maintenance manual

Physio Control LIFEPAK 9 Service manual

Physio Control

Physio Control LIFEPAK 9 Service manual

Vetario S40 User instructions

Vetario

Vetario S40 User instructions

Atmos Variotherm plus operating instructions

Atmos

Atmos Variotherm plus operating instructions

Barco CORONIS 1MP Installation & user manual

Barco

Barco CORONIS 1MP Installation & user manual

BARDOMED CarePump Lite 6 user manual

BARDOMED

BARDOMED CarePump Lite 6 user manual

Boston Scientific precision spectra Information for Prescribers

Boston Scientific

Boston Scientific precision spectra Information for Prescribers

Holtek BMH08004 Operation guide

Holtek

Holtek BMH08004 Operation guide

MPS PHARM/PUMP III operating instructions

MPS

MPS PHARM/PUMP III operating instructions

Orliman E-250 INSTRUCTIONS FOR USE AND PRESERVATION

Orliman

Orliman E-250 INSTRUCTIONS FOR USE AND PRESERVATION

Trulife PRESSURECARE Oasis Extremity Belt Instructions for use

Trulife

Trulife PRESSURECARE Oasis Extremity Belt Instructions for use

Arjo Nimbus 4 instructions

Arjo

Arjo Nimbus 4 instructions

Invacare Octave user manual

Invacare

Invacare Octave user manual

SAALMANN Saalio Textile Electrodes Instructions for use

SAALMANN

SAALMANN Saalio Textile Electrodes Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.