Memmert hpp 1400 User manual

www.memmert.com |www.atmosafe .net
100% ATMOSAFE. MADE IN GERMANY.
HPP
OPERATING
INSTRUCTIONS
CONSTANT CLIMATE CHAMBER HPP 1400/2200

Manufacturer and customer service
Memmert GmbH + Co. KG
Willi Memmert Straße 90-96
D-91186 Büchenbach
Deutschland
Phone: +49 (0)9122 925-0
Fax: +49 (0)9122 14585
E-mail: [email protected]
Internet: www.memmert.com
Customer service:
Service hotline: +49 (0)9171 9792 911
Service fax: +49 (0)9171 9792 979
E-mail: [email protected]
When contacting customer service, always quote the product serial number on the nameplate
(see page 12).
Shipping address for repairs:
Memmert GmbH + Co. KG
Kundenservice
Willi-Memmert-Str. 90-96
DE-91186 Büchenbach
Germany
Please contact our customer service before sending appliances for repair or before returning
equipment, otherwise, we have to refuse acceptance of the shipment.
© 2018 MEMMERT GmbH + Co. KG
D33464 | Date 09/2018
We reserve the right to make changes

D33464 | Date 09/2018 3
About this manual
About this manual
Purpose and target group
This manual describes the assembly, function, transport and operation of constant climate
chambers HPP 1400 and 2200. It is intended for use by trained personnel of the owner, who
have the task of operating and/or maintaining the respective appliance.
If you are asked to work on the appliance, read this manual carefully before starting. Familiar-
ise yourself with the safety regulations. Only perform work that is described in this manual. If
there is something you do not understand, or certain information is missing, ask your superior
or contact the manufacturer. Do not do anything without authorisation.
Versions
The appliances are available in different configurations and sizes. If specific equipment fea-
tures or functions are available only for certain configurations, this is indicated at the relevant
points in this manual.
The functions described in this manual refer to the latest firmware version.
Due to individual configurations and sizes, illustrations in this manual may be slightly different
from the actual appearance. Function and operation are identical.
Other documents that have to be observed:
►For operation of the appliance with MEMMERT AtmoCONTROL, observe the separate
software manual. To open the AtmoCONTROL software manual, click on “Help” in the
AtmoCONTROL menu bar.
►for service and repair work (see page 57) – a separate service manual
Storage and forwarding
This instruction manual belongs with the appliance and should always be stored where
persons working on the appliance have access to it. It is the responsibility of the owner to
ensure that persons who are working or will work on the appliance are informed as to the
whereabouts of this instruction manual. We recommend that it is always stored in a protected
location close to the appliance. Make sure that the instruction manual is not damaged by heat
or humidity. If the appliance is sold on or transported and then set up again at a different
location, the operating instructions must go with it.
You will find the current version of our operating manual as pdf file if you go to
www.memmert.com/de/service/downloads/bedienungsanleitung/.

4 D33464 | Date 09/2018
Contents
Contents
1. For your Safety 6
1.1 Terms and signs used........................................................................................................... 6
1.2 Product safety and dangers ................................................................................................ 7
1.3 Requirements of the operating personnel.......................................................................... 7
1.4 Responsibility of the owner................................................................................................. 7
1.5 Intended use........................................................................................................................ 8
1.6 Changes and alterations...................................................................................................... 8
1.7 Behaviour in case of malfunctions and irregularities.......................................................... 8
1.8 Switching off the appliance in an emergency .................................................................... 8
2. Construction and description 9
2.1 Construction ........................................................................................................................ 9
2.2 Description......................................................................................................................... 10
2.3 Material.............................................................................................................................. 10
2.4 Electrical equipment.......................................................................................................... 10
2.5 Connections and interfaces............................................................................................... 10
2.6 Designation (nameplate)................................................................................................... 12
2.7 Technical data .................................................................................................................... 13
2.8 Applied directives and standards ...................................................................................... 14
2.9 Declaration of conformity ................................................................................................. 14
2.10 Ambient conditions........................................................................................................... 15
2.11 Scope of delivery ............................................................................................................... 15
2.12 Optional accessories.......................................................................................................... 15
3. Delivery, transport and setting up 16
3.1 Safety regulations.............................................................................................................. 16
3.2 Delivery .............................................................................................................................. 16
3.3 Transport............................................................................................................................ 16
3.4 Unpacking ......................................................................................................................... 16
3.5 Storage after delivery ........................................................................................................ 17
3.6 Setting up .......................................................................................................................... 17
4. Putting into operation 21
4.1 Connecting the appliance .................................................................................................21
4.2 Filling and connecting the water tank .............................................................................. 21
4.3 Switching on...................................................................................................................... 22
5. Operation and control 23
5.1 Operating personnel.......................................................................................................... 23
5.2 Opening and closing the doors......................................................................................... 23
5.3 Loading the appliance....................................................................................................... 24
5.4 Operating the appliance.................................................................................................... 25
5.5 Monitoring function.......................................................................................................... 31
5.6 Graph................................................................................................................................. 35
5.7 Ending operation............................................................................................................... 36

D33464 | Date 09/2018 5
Contents
6. Malfunctions, warning and error messages 37
6.1 Warning messages of the monitoring function................................................................ 37
6.2 Malfunctions, operating problems and appliance errors ................................................ 38
6.3 Power failure...................................................................................................................... 40
7. Menu mode 41
7.1 Overview............................................................................................................................ 41
7.2 Basic operation in menu mode using the example of language selection....................... 42
7.3 Setup.................................................................................................................................. 43
7.4 Date and Time ................................................................................................................... 47
7.5 Calibration ......................................................................................................................... 48
7.6 Program............................................................................................................................. 53
7.7 Sound ................................................................................................................................ 54
7.8 Protocol ............................................................................................................................. 55
7.9 User-ID ............................................................................................................................... 56
8. Maintenance and service 57
8.1 Cleaning............................................................................................................................. 57
8.2 Regular maintenance.........................................................................................................58
8.3 Repairs and service ............................................................................................................ 58
9. Storage and disposal 59
9.1 Storage .............................................................................................................................. 59
9.2 Disposal ............................................................................................................................. 59
Index 60

6 D33464 | Date 09/2018
For your Safety
1. For your Safety
1.1 Terms and signs used
In this manual and on the appliance itself, certain common terms and signs are used to warn
you of possible dangers or to give you hints that are important in avoiding injury or damage.
Observe and follow these hints and regulations to avoid accidents and damage. These terms
and signs are explained below.
1.1.1 Terms used
"Warning" is used whenever you or somebody else could be injured if you do not
observe the accompanying safety regulation.
"Caution" is used for information that is important for avoiding damage.
1.1.2 Signs used
Warning signs (warning of a danger)
Danger of
electrocution
Danger of
explosion
Dangerous
gases / vapours
Danger of top-
pling over
Hazard area!
Observe the opera-
ting instructions
Prohibition signs (forbidding an action)
Do not lift Do not tilt Do not enter
Regulation signs (stipulating an action)
Disconnect
the mains
plug
Wear gloves Wear safety
boots
Observe information in separate manual
Other icons
Important or useful additional information

D33464 | Date 09/2018 7
For your Safety
1.2 Product safety and dangers
The appliances described in this manual are technically sophisticated, manufactured using
high-quality materials and subject to many hours of testing in the factory. They contain the
latest technology and comply with recognised technical safety regulations. However, there are
still risks involved, even when the appliances are used as intended. These are described below.
Warning!
After removing covers, live parts may be exposed. You may receive
an electric shock if you touch these parts. Disconnect the mains plug
before removing any covers. Only electrical technicians may work on
the electrical equipment of the appliances.
Warning!
When loading the appliance with an unsuitable load, poisonous or
explosive vapours or gases may be produced. This could cause the
appliance to explode, and people could be severely injured or poi-
soned. The appliance may only be loaded with materials/test objects
which do not form any toxic or explosive vapours when heated up
(see also chapter Intended use on page 8).
Warning!
If the door is open while the appliance is in operation, the appliance
may overheat and pose a fire hazard. Do not leave the door open
during operation.
Warning!
In case of appliances of a certain size, you can get accidentally
locked in, which is life-threatening. Do not climb into the appliance!
1.3 Requirements of the operating personnel
The appliance may only be operated and maintained by persons who are of legal age and
have been instructed accordingly. Personnel who are to be trained, instructed or who are un-
dergoing general training may only work with the appliance under the continuous supervision
of an experienced person.
Repairs may only be performed by qualified electricians. The regulations in the separate service
manual must be observed.
1.4 Responsibility of the owner
The owner of the appliance
►is responsible for the flawless condition of the appliance and for it being operated in ac-
cordance with its intended use (see page 8);
►is responsible for ensuring that persons who are to operate or service the appliance are
qualified to do this, have been instructed accordingly and are familiar with the operating
instructions at hand;
►must know about the applicable guidelines, requirements and operational safety regula-
tions, and train staff accordingly;

8 D33464 | Date 09/2018
For your Safety
►is responsible for ensuring that unauthorised persons have no access to the appliance;
►is responsible for ensuring that the maintenance plan is adhered to and that maintenance
work is carried out properly (see page 57);
►has to ensure that the appliance and its surroundings are kept clean and tidy, for example
through corresponding instructions and inspections;
►is responsible for ensuring that personal protective clothing is worn by operating person-
nel, e.g. work clothes, safety shoes and protective gloves.
1.5 Intended use
Constant climate chambers HPP 1400 and 2200 may be used exclusively for temperature and
climate testing of materials and substances in the context of the procedures and specifications
described in this manual. Any other use is improper, and may result in hazards and damage.
The appliance is not explosion-proof (does not comply with the German workplace health &
safety regulation VBG 24). The appliance may only be loaded with materials and substances
which cannot form any toxic or explosive vapours at the set temperature and which cannot
explode, burst or ignite.
The appliance may not be used for drying, vaporising and branding paints or similar materi-
als the solvents of which could form an explosive mixture when combined with air. If there
is any doubt as to the composition of materials, they must not be loaded into the appliance.
Potentially explosive gas-air mixtures must not form, neither in the working chamber nor in
the direct vicinity of the appliance.
1.6 Changes and alterations
No unauthorised changes or alterations may be made to the appliance. No parts may be
added or inserted which have not been approved by the manufacturer.
Unauthorised changes or alterations result in the EC declaration of conformity losing its valid-
ity, and the appliance may no longer be operated.
The manufacturer is not liable for any damage, danger or injuries that result from unauthor-
ised changes or alterations, or from non-observance of the regulations in this manual.
1.7 Behaviour in case of malfunctions and irregularities
The appliance may only be used in a flawless condition. If you as the operator notice irregu-
larities, malfunctions or damage, immediately take the appliance out of service and inform
your superior.
You can find information on eliminating malfunctions from page 37.
1.8 Switching off the appliance in an emergency
Push the On/Off switch on the control panel
(Fig. 1) and disconnect power plug. This
disconnects the appliance from the power
supply at all poles. ON
O
O
ON
N
N
N
O
O
N
O
Fig. 1
Switch off the appliance by pressing the On/
Off switch

D33464 | Date 09/2018 9
Construction and description
2. Construction and description
2.1 Construction
1
2
3
4
5
6
8
9
7
Fig. 2 Construction
1 ControlCOCKPIT with capacitive function
keys and LCD displays (see page 25)
2 On/Off switch (see page 22)
3 Door handle (see page 23)
4 Slide-in units
5 Heated full-sight glass door
6 Lockable castors with extendable feet
(see page18)
7 Nameplate (see page 12)
8 Door handle (see page 23)
9 USB interface (see page 11)

10 D33464 | Date 09/2018
Construction and description
2.2 Description
The appliance can heat the interior up to 60 °C and cool it down to 15 °C. Low-noise, long-life
and energy-saving Peltier cooling and heating technology is used for this. In heating opera-
tion, a part of the required energy is extracted from the surroundings (heat pump principle).
Additionally, the humidity in the interior can be regulated. The humidity is increased by the
evaporation of water from a tank which is then fed into the interior and reduced by conden-
sation on a Peltier module.
2.3 Material
For the outer housing, MEMMERT uses stainless steel (Mat.No. 1.4016 – ASTM 430) and for
the interior, stainless steel (Mat.No. 1.4301 – ASTM 304) is used, which stands out through
its high stability, optimal hygienic properties and corrosion resistance to many (but not all!)
chemical compounds (caution for example with chlorine compounds).
The chamber load for the appliance must be carefully checked for chemical compatibility with
the materials mentioned. A material resistance table can be requested from the manufacturer.
2.4 Electrical equipment
►Operating voltage and current consumption: See nameplate
►Protection class I, i.e. operating insulation with PE conductor
in accordance with EN 61010
►Protection type IP 20 acc. to EN 60 529
►Interference suppression acc. to EN 55011 class B
►Appliance fuse: Safety fuse 250 V/15 A, quick-blow
►The temperature sensor is equipped with a 100 mA miniature fuse.
2.5 Connections and interfaces
2.5.1 Electrical connection
This appliance is intended for operation on an electrical power system with a system imped-
ance Zmax of a maximum of 0.292 ohm at the point of transfer (service line).
The operator must ensure that the appliance is operated only on an electrical power system
that meets these requirements. If necessary, you can ask your local energy supply company
what the system impedance is.
Observe the country-specific regulations when making connections (e.g. in Germany
DIN VDE 0100 with earth leakage circuit breaker).

D33464 | Date 09/2018 11
Construction and description
2.5.2 Communication interfaces
The interfaces are intended for appliances which meet the requirements of IEC 60950-1.
USB interface
The appliance is fitted by default with a USB
interface in accordance with the USB specifica-
tion. This way, you can
►transfer software stored on a USB storage
medium to the appliance (see page 53).
►export protocol logs from the appliance to
a USB storage medium (see page 55).
►transfer user ID data stored on a USB stor-
age medium to the appliance (see page
56).
The USB port is located on the right side of the ControlCOCKPIT (Fig. 3).
Ethernet interface
Via Ethernet interface, the appliance can be
connected to a network, so that you can trans-
fer programmes created with the
AtmoCONTROL software to the appliance and
read out protocols. The Ethernet interface is
located on the rear of the appliance (Fig. 4).
For identification purposes, each appliance con-
nected must have its own unique IP address.
Setting the IP address is described on page
43.
You will find a description of how to transfer programmes via Ethernet
in the enclosed AtmoCONTROL manual.
With an optional USB to Ethernet converter, the appliance can be directly connected
to a computer / laptop (see Optional accessories on page 15).
Fig. 3 USB interface
Fig. 4 Ethernet interface

12 D33464 | Date 09/2018
Construction and description
2.6 Designation (nameplate)
The nameplate (Fig. 5) provides information about the appliance model, manufacturer and
technical data. It is attached to the front of the appliance on the right, behind the right door
(see page 9).
Typ: HPP 1400 F.-Nr.: 0109.0088
230 V~ 13,5 A 50/60 Hz 3100 W
DIN12880-Kl.3.1 Nenntemp.: 60 °C
1
2
3
4
5
6
10
8
9
7
Fig. 5 Nameplate (example)
1 Type designation
2 Operating voltage
3 Applicable standard
4 Protection type
5 CE conformity
6 Address of manufacturer
7 Disposal note
8 Temperature range
9 Connection / power ratings
10 Appliance number

D33464 | Date 09/2018 13
Construction and description
2.7 Technical data
Appliance size 1400 2200
Appliance width D1[mm] 1435 2157
Appliance height E1(incl. castors) [mm] 1905 1905
Appliance depth F1[mm] 1005 1005
Depth of door lock [mm] 56
Chamber width A1[mm] 1250 1972
Chamber height B1[mm] 1450 1450
Chamber depth C1[mm] 750 750
Chamber volume [litres] 1360 2140
Net weight [kg] 450 493
Current consumption [A] 230 V, 50/60 Hz 13.5 15,2
115 V, 50/60 Hz 27 30,4
Power [W] 3100 3500
max. number of sliding grids 28 (2 x 14) 42 (3 x 14)
max. load per sliding grid [kg] 30
max. load per appliance [kg] 250
Temperature Adjustment range 15 °C to 60 °C2
Adjustment precision 0.1 K
Humidity Adjustment range [% rh] 15 % to 80 %
1see Fig. 6.
2The minimum temperature depends on the outdoor temperature (see ambient conditions
on page 15). If the interior is illuminated, the temperature range is further restricted.
3 In the range of 15 to 45 °C

14 D33464 | Date 09/2018
Construction and description
D
AFC
(93) (93) 56
35
256
E
B
82
295 147
180
Fig. 6 Dimensions
2.8 Applied directives and standards
►Directive 2004/108/EC amended (Directive of the council for harmonisation
of the laws of the member states on electromagnetic compatibility).
Standards complied with:
DIN EN 61326:2004-05, EN 61326:1997, EN 61326/A1:1998, EN 61326/A2:2001
EN 61326/A2:2003
►Directive 2006/95/EC amended (Directive of the council for harmonisation of the laws of
member states relating to electrical equipment designed for use within certain voltage
limits). Standards complied with:
DIN EN 61 010-1 (VDE 0411 part 1):2002-08
DIN EN 61 010-2-010 (VDE 0411 part 2-010):2004-06
EN 61 010-1:2001, EN 61 010-2-010:2003
2.9 Declaration of conformity
You can download the EC declaration of conformity of the appliance online:
English: http://www.memmert.com/en/service/downloads/ce-statement/
German: http://www.memmert.com/de/service/downloads/eg-konformitaetserklaerung/

D33464 | Date 09/2018 15
Construction and description
2.10 Ambient conditions
►The appliance may only be used in enclosed areas and under the following ambient condi-
tions:
Ambient temperature 16 ºC to 40 ºC
Humidity rh max. 70 % non-condensing
Overvoltage category II
Pollution degree 2
Altitude of installation max. 2,000 m above sea level
►The appliance may not be used in areas where there is a risk of explosion. The ambient air
must not contain any explosive dusts, gases, vapours or gas-air mixtures. The appliance is
not explosion-proof.
►Heavy dust production or aggressive vapours in the vicinity of the appliance could lead to
sedimentation in the interior and, as a consequence, could result in short circuits or dam-
age to electrical parts. For this reason, sufficient measures to prevent large clouds of dust
or aggressive vapours from developing should be taken.
2.11 Scope of delivery
►Power cable
►Tilt protection (see page 19)
►Sliding grids (load capacity 30 kg each)
►USB storage medium with software and AtmoCONTROL manual
►the operating instructions at hand
►Calibration certificate
►Water tank with connection hose
2.12 Optional accessories
►USB to Ethernet converter (Fig. 7). Makes it
possible to connect the Ethernet connection
interface (see page 11) to the USB port of
a computer / laptop.
Fig. 7 USB to Ethernet converter

16 D33464 | Date 09/2018
Delivery, transport and setting up
3. Delivery, transport and setting up
3.1 Safety regulations
Warning!
Because of the heavy weight of the appliance, you could
injure yourself if you try to lift it. The appliance may only
be transported by a manual pallet jack or forklift truck.
Warning!
You may get your hands or feet squashed when transport-
ing and installing the appliance. Wear protective gloves and
safety boots.
Warning!
The appliance could fall over and seriously injure you. Never
tilt the appliance. Only transport in upright position and
without load (except standard accessories such as steel grids
or shelves).
Appliances with castors always have to be moved by two
people.
3.2 Delivery
The appliance is packaged in a wooden crate and delivered on a wooden pallet.
3.3 Transport
The appliance can be transported in three ways:
►With a forklift truck; move the forks of the truck entirely under the pallet.
►On a manual pallet jack
►on its own castors, for which the catch on the (front) castors must be released
3.4 Unpacking
To avoid damage, do not unpack the appliance until you reach the installation site.
Unscrew wooden crate and remove. Remove the protective film.
3.4.1 Checking for completeness and transport damage
►Check the delivery note to ensure that the delivery is complete.
►Check the appliance for damage.
If you notice deviations from the delivery note, damage or irregularities, do not put the appli-
ance into operation but inform the haulage company and the manufacturer.
3.4.2 Remove the transport protection
Remove the transport protection. It is located between the door hinge, the door and the
frame and has to be removed after opening the doors.

D33464 | Date 09/2018 17
Delivery, transport and setting up
3.4.3 Disposing of packaging material
Dispose of the packaging material (cardboard, wood, foil) in accordance with the
applicable disposal regulations for the respective material in your country.
3.5 Storage after delivery
If the appliance is first to be stored after delivery: Read the storage conditions from page
59.
3.6 Setting up
Warning!
Due to its centre of gravity, the appliance can fall over to the front
and injure you or other people. Always attach the appliance to a wall
with the tilt protection (see page 19). If there is not enough space,
do not begin operating the appliance and do not open the doors.
Contact the Memmert service (see page 2).
3.6.1 Preconditions
The appliance may only be installed on the bottom. The installation site must be flat and
horizontal and must be able to reliably bear the weight of the appliance (see Technical data on
page 13). Do not place the appliance on a flammable surface.
A suitable power source (see type plate and technical data on page 13) must be available at
the installation site.
The distance between the wall and the rear of the appliance must be at least 15 cm.
The clearance from the ceiling must not be less than 20 cm and the side clearance from walls
or nearby appliances must not be less than 5 cm (Fig. 8). Sufficient air circulation in the vicinity
of the appliance must be guaranteed at all times.
For appliances with castors, always position and lock the castors facing forwards.
FP FP FP
FP
Fig. 8 Minimum clearance from walls and ceiling

18 D33464 | Date 09/2018
Delivery, transport and setting up
3.6.2 Level and secure the device against rolling away
The height of the appliance can be adjusted using the heavy-duty castors attached to the bot-
tom of the appliance. It can also be secured against rolling away or being shifted. To do this,
the feet must be extended.
1. To unlock the adjustment mechanism, pull out the ratchet lever on the ring (Fig. 9).
2. To retract and extend the stand, actuate the ratchet lever (Fig. 10).
Fig. 9 To unlock the mechanism, pull out
the ratchet lever
Fig. 10 To retract and extend the stand,
actuate the ratchet lever
The direction of movement (up/down) can be adjusted using the rocker above the ratchet
lever:
►Press in the rocker on the right side to extend the stand with the ratchet lever (Fig. 11).
►Press in the rocker on the left side to retract the stand with the ratchet lever (Fig. 12).
Fig. 11 To extend the stand, press in the
right-hand rocker
Fig. 12 To retract the stand, press in the
left-hand rocker
In the end position, push in the ratchet lever again to fix the stand.
Use a spirit level to ensure that all four feet are adjusted to the same height.
To move the cabinet, all four feet must be retracted.

D33464 | Date 09/2018 19
Delivery, transport and setting up
3.6.3 Tilt protection
Attach the appliance to a
wall with the tilt protection.
The tilt protection is included
in the delivery.
3. Screw the tilt protection
onto the back of the ap-
pliance as illustrated.
4. Bend the tilt protec-
tion upwards by 90° in
the desired distance to
the wall (observe the
minimum distance to the
wall, see Fig. 8).
5. Drill a hole, insert
a dowel and screw
the tilt protection
to a suitable wall.

20 D33464 | Date 09/2018
Delivery, transport and setting up
3.6.4 Adjusting the doors
You can adjust the doors if necessary, for example if they are warped due to uneven flooring.
There are two adjusting screws each at the top and the bottom of each door for this purpose
(Fig. 13).
First, adjust the setting at the top of the door and, if this is not sufficient, adjust the screws
at the bottom of the door.
1. Open the door.
2. Loosen the screws.
3. Adjust the position of the door.
4. Tighten the screws again.
5. Check the position of the door.
6. Readjust if required.
Fig. 13 Adjusting the doors
This manual suits for next models
1
Table of contents
Other Memmert Climate Chamber manuals

Memmert
Memmert HCP105 User manual

Memmert
Memmert ICO Series User manual

Memmert
Memmert ICO User manual

Memmert
Memmert ICO User manual

Memmert
Memmert HPPeco User manual

Memmert
Memmert WTB 6 User manual

Memmert
Memmert HPPlife Operating instructions

Memmert
Memmert HPPeco User manual

Memmert
Memmert HPPeco User manual

Memmert
Memmert ICO User manual