MRC CENVCC-3 User manual

CENVCC-3
Table Vacuum Centrifugal concentrator
Instruction & Using Manual
Please read this manual carefully before use
3, Hagavish st. Israel 58817 Tel: 972 35595252, Fax: 972 3 5594529
MRC.8.22

Content
Ⅰ、FEATURES AND USES ..............................................................................1
Ⅱ、WORKING PRINCIPLE ............................................................................2
Ⅲ、VACUUM CONCENTRATOR SPECIFICATION ..................................2
Ⅳ、ROTOR SPECIFICATION ........................................................................ 4
Ⅴ、PRECAUTION .............................................................................................5
Ⅵ、APPEARANCE AND STRUCTURE ........................................................ 6
Ⅶ、WORKING PROCEDURE ........................................................................ 7
VIII、TROUBLES & SOLUTIONS ................................................................. 8
IX、INSTALLATION......................................................................................9
X、MAINTENANCE....................................................................................... 10
XI、WARRANTY...........................................................................................11
XII、CERTIFICATE OF CONFORMITY..................................................12
XIII、PACKING LIST...................................................................................13

1/17
一、Features and Uses
CENVCC-3 is a microcomputer-controlled desktop vacuum concentrator. made by full steel body. the
surface is spray-coated, with good rigidity and high strength. It has novel shape, beautiful appearance, low
noise, rising temperature slowly, safety and reliability etc
This machine use frequency conversion motor. stable speed and maintenance free.Adpot high-quality
solenoid valve and fully closed microporous filter ensure smooth and safe return air. LCD display
temperature, PID algorithm temperature control, highly temperature accuracy. adopts 304 stainless steel
inner cavity, transparent high-strength lid and aluminum rotor. with strong corrosion resistance, not easy to
scratch,has a long service life
Ⅱ、Working principle
The vacuum concentrator is a method in which the sample is continuously pumped by a
vacuum pump while being centrifuged, or even close to vacuum, so that the solvent in the
sample is quickly evaporated, the sample system is rapidly concentrated or dried, and the cold
trap and the filtration method can be effectively used,which collected evaporated solvent. This
model is suitable for RNA/DNA in the fields of molecular biology, biochemistry, genetics,
analytical chemistry, quality control, etc., as well as protein concentration or drying. The
sample after centrifugation and concentration can be conveniently used for various qualitative
and analysis of chemistry, biochemistry, bioanalysis, immunoscreening and instrumental
analysis.
Ⅲ、Vacuum Concentrator Specification
Max RPM:1800r/min
Max capacity:62*1.5mm
Max RCF: 409×g
Speed accuracy:±20r/min
Time control:1~99min59s
Time control accuracy:±1S
Temperature stability range:Room temp~+60℃
Temperature Accuracy:±2℃
Ultimate vacuum: -0.1Mpa

2/17
Noise:≤50dB
Power:AC220V±10% 50Hz 10A
Machine size:430 mm×400 mm×330 mm (L×W×H)
Net Weight:35Kg
Ⅳ、Rotor Specification
Rotor
Number
Capacity
Max Speed
Max RCF
Angle rotor
No.1
62×1.5ml
1800r/min
409×g
No.2
135×0.5ml
1800r/min
273×g
Ⅴ、PRECAUTION
1. In order to safety working and correct centrifuging, please take the machine in a solid,
shockproof and horizontal table, please ensure the four feet equal force.
2. Before using, please move out the foreign body; check the rotor is correctly installed in the
rotor seat, please ensure the ro
3. tor and motor shaft with very good connection.
4. Set the speed strictly under the speed range allows, It is strictly prohibit to work
over-speed.
5. No unbalance working, centrifuge tube must be placed symmetrically, no single tube
working, the tube solution must be uniform to ensure balanced operation.
6. No high-speed working without rotor.
7. Remove the rotor must use the special handle which together with the machine or use
non-metallic tools gently tap the rotor first, repeated until the shock between the rotor and the
motor shaft connection. It is strictly prohibited to release the screw directly after the hard pull
up, so as not to damage the motor's flexible support.
8. Before using the rotor and centrifuge tube to check whether there are cracks, corrosion
traces and aging phenomena, if necessary to immediately replace, it is strictly prohibited to
use the rotor with crack or corrosion.

3/17
9. Check whether the external power supply meets the requirements, the instrument is
reliable grounding.
10. No moving and No lid-open during the centrifuge running, Do not open the lid when the
motor and rotor are not completely stopped.
11. The instrument or rotor disabled for three month, please running at low speed 10min, then
allow to work on high speed.
12. To prevent damage to the instrument and personal safety, No mixed using other company
Rotors.
13. To prevent the machine and spare parts corrode, Do Not place a container with liquid on
the instrument.
14. Please clean the machine and close the power switch and power plug after using.
Ⅵ、Appearance and structure
1、Concentrator appearance
2、Display Panel

4/17
1.Information window 2、Speed window 3、RCF window 4、Time
window 5、Rotor no. 6、Acceleration gear 7、Deceleration gear 8、
Program group number 9 、Setting key 10 、increasing key 11 、
instantaneous centrifugal key 12、Running indicator 13、Start key 14、
Confirmation key 15、Reduction key 16、Speed/Centrifugal Force Mode
Switching Key 17、Stop indicator 18、Stop key/open key

Ⅶ、Working procedure
1) the centrifuge placed on the platform, visual to balance, shake and check whether the
centrifuge placed smoothly.
2) confirm the placement is stable, turn on the power, press the power switch, press the "stop
/ open" button, the lid can be opened.
3) Open the lid, the selected rotor gently placed on the motor shaft, screw on the screw so
that the rotor and the motor shaft tightly connected together (Note: it should be careful
and gentle, to avoid shaking and pull up and down, so as not to Damage the motor
support), and then put the tube set or adapter on the rotor, complete gently turn the rotor,
the rotor should be flexible rotation, Tube set (adapter) on the rotor should be tilted freely,
no block and rub Touch the phenomenon.
4) Put the tube filled with the same amount of test solution symmetrical into the tube set to
ensure that the rotor balance operation.
5) close the lid and check whether it has been locked (closed gently lift the lid, the lid is not
open. If the lid is not closed, press the "Start", the "time display" window will display "
E3 ").
6) set the rotor number, speed, time and other operating parameters.
(1)Set the speed: press the "Set" key, move the cursor to the "speed setting" position,
when the speed digital flicker, that is, enter the speed setting, then press the "increase
key" or "reduce key" to set the speed of this operation, after setting, press the
confirmation key to confirm the parameters.
Special reminder: The horizontal rotors and different adapters of this machine have their
own rotor numbers and maximum speed limits. Users must strictly observe the speed
limits of the rotor parameters when setting the speed parameters. Overspeed operation is
strictly prohibited to ensure safe operation.
(2)RCF set:Press“Set”to move the cusrsor to“Speed indicator”, when flashing,it is speed &
RCF switching state,Press“+”or“-” Switch the speed to RCF, “RCF”indicator,Press“Set”
move to,“Speed/RCF set”,when flashing,Press“+”or“-” to set and press “Enter” confirm
(3)Time:press“Set” to move the cursor to“Time Set”, when flashing,Press "+" or "-" to time
set,then press“Enter” confirm.
(4)Acc/Dec Set:press“Set” to move the cursor to“Acc” or “Dec”, when flashing,Press "+"
or "-" to acc/Dec set,then press“Enter”to confirm.
The system supports 0- 9 a total of 10 files, 0 is the fastest, 9 is the slowest, when the

use of larger volume of the angle rotor or swing-out rotor, you can choose a little bit of the
slows, the other can choose a little faster file Bit. Generally recommended to use 5 files;
Finally press "Enter" to confirm.
7) When the "Short" button is pressed, the machine centrifuges, and when the mobile phone
is loosened, centrifugation stops.
Remind: each step is completed, should immediately press the "Enter" to be confirmed.
8) Finally, check the parameters on the screen confirm correct, press the "Start" " to run.
9) The running time countdown to "0", the centrifuge will automatically stop, the buzzer
ring, when the speed is 0, press the "Stop / Open" button, you can open the lid. During the
event of an emergency, press the "Stop / Open" button, the buzzer does not ring when the
centrifuge stops running.
Note: Do not open frequently, the proposed interval of 10 seconds to avoid
damage to components
10) Open, take out the tube carefully, complete the separation process.
11) turn off the power switch, cut off the centrifuge power.
VIII、Troubles & Solutions
Troubles
Cause
Solution
No display when power
plugged in
No power
Power check
Fuse blown
Check & replace
Cable connection loose
Check & reconnect
Large vibration
Tube No Symmetrically placed
Check & adjust
Tube broken
Check & replace
rotor not tightened
Check & adjust
Partial damage to shock
Check & replace
Display 0000 ,Start no
running
Circuit board or transformer
damaged
replace
control system connector is
loose
Re-tighten

Key damaged
replace
Motor damage or leakage
replace
Running but not speed, the
machine has strange sound or
smell
Control system or motor failure
manufacturers
maintenance
The machine with fault prompt, when the machine fails, the "display window" will
display the following items:
E1
Unbalance
E2
Over-speed
E3
Lid-open
E4
Lid exception
E5
Temperature sensor error
E6
Over temperature
E7
Motor stall
E8
Communication error
E9
System error
EA
Low voltage
EB
Overcurrent protection
EC
Motor fault (Hall state error)
ED
Brake resistance
Note:
The above alarm is displayed in the "display window", when appears, the buzzer
alarm for a long time short and alarm shutdown. In the event of an alarm, the machine
will be correct display only when the fault is released. If the fault is not released, press
the “Enter”to clear the buzzer alarm. At this time, the fault display will not be cleared.
When the hardware fault is solved, press the Enter key again to clear the fault display
IX、INSTALLATION

1、Out of the box check
a) Before receiving the instrument, you should carefully check for transport damage. If
so, please reject the goods and inform the Company in writing and take pictures of
the damaged machine.
b) check whether the goods match the order list, if not, please contact the company.
c) 1.3, remove the packaging, and remove the contents of the centrifugal chamber.
2、Installation
a) Installation site requirements
the installation of the table should be strong, dry room, clean, to avoid direct
sunlight.
the room should have a separate ground, to ensure safe use of electricity.
b) the installation of the centrifuge
A), the centrifuge installed to a solid table to ensure that the four foot contact with
the table.
B), centrifuge around the gap with the wall should be greater than 16cm, to ensure
good ventilation.
C), power exchange 220V 50Hz 10A or 110v 60Hz, the wire cross-sectional area of
not less than 2 square millimeters, the socket capacity of not less than 10A, the
protective grounding wire must be connected firmly.
X、MAINTENANCE
1. After using, such as the day is no longer used, please use a clean soft cloth to clean the
centrifuge chamber, rotor and centrifuge tube.
2. Regularly check the rotor and centrifuge tube is cracked.
3. If not use the machine more than a month time, please remove the rotor from the
centrifuge chamber, the rotor center hole and centrifuge spindle coated with grease, keep
the rotor dry.
4. regular inspection of moving parts, such as gas spring seat, lid hinges and other screws
fastening.
XI、WARRANTY

1. 1.Vacuum centrifugal concentrator warranty: 1 year
Free warranty is one year since the date of purchase upon non-human damage.
Technical guide for the whole life.
2. the company supporting a variety of rotor quality assurance period of one year.
the company does not assume any responsibility the following:
More than the max speed of the rotor used;
The rotor has scratches and bumps;
The rotor is seriously corroded by acid and alkali;
Unauthorized professional maintenance.
3. without the permission of the Company, non-authorized professionals shall not open the
machine; all maintenance and adjustment must be carried out by professionals in
accordance with the maintenance manual.
4. the Company will not be responsible for the maintenance and adjustment of unauthorized
personnel!

XII、CERTIFICATE OF CONFORMITY
Table Vacuum Concentrator
CERTIFICATE OF CONFORMITY
Max RPM:1800r/min
Max RCF: 409×g
Serial number:
This product has been tested and approved.
Inspectors:

XIII、PACKING LIST
Packing List
No.
Name
Specification
Qty.
Mark
一
Table Vacuum
concentrator
1Set
二
Accessories & tools
1
Power plug
1Pcs
2
Angle rotor
62×1.5ml
1Set
3
Vacuum Pump
1set
4
Cold Trap
1set
5
6
7
8
9
三
Document
1
Instruction manual
1pc
This manual
2
Certificate of
conformity
1pc
This manual
3
Packing list
1pc
This manual
Packing Inspector:
July 19,2022

Cold Trap Operating Instructions
1. The overview of the cold trap is as follows::
Basic parameters of cold trap:
Power supply: 220V ±22V power: 500W
Overall dimension: 665 ×four hundred and ten ×420 (long) ×wide ×Height in mm
Pipe joint size: φ10mm
Temperature setting range: - 40 ℃~ normal temperature
Temperature control accuracy: ±1℃
Volume of freezing chamber: 4L
2. The control interface of the cold trap is as follows:
Control panel, easy to operate and use

1. Temperature display window 2. Defrosting key 3. Compressor operation indicator
4. Compressor start / stop button
3、The operating procedures of the instrument are as follows:
1. Place the cold trap on the table stably, and the table shall be flat without strong vibration
and strong heat source.
2. After confirming that it is placed stably, connect the air inlet with the pipe for
concentration instrument, and the air outlet with the pipe for vacuum pump, and lock it
with clamps.
3. Plug the power cord into the power socket.
4. Turn on the power switch.
5. Press the start button of the compressor to cool the compressor. At this time, the
operation indicator lights up. If you need to stop cooling, press the stop button directly (the
compressor will also stop when the temperature reaches the set temperature).
The temperature control settings are as follows:
Code
Explain
Set range
Factory set
P0
Starting
temperature
-40 ℃~room
temperature
-40℃
P1
Stop temperature
-40℃~room
temperature
-40℃
P0 set the starting temperature: press the setting key to enter the internal menu. The default

display is P0. Press the setting key again to set the required starting temperature through
the up and down keys;
P1 set stop temperature: press the set key to enter the internal menu, the digital display is
P0, switch to P1 through the up and down keys, and then press the set key to set the
required stop temperature through the up and down keys;
6. Defrosting: the cold trap is equipped with a one button defrosting button. When the
frost in the freezing chamber is serious after shutdown, press the defrosting button to
defrost (at this time, ensure that the compressor is in the stopped refrigeration state). When
defrosting is completed, press the defrosting button again to reset the button. If there is a
large amount of effusion in the freezing cavity, use a tool to unscrew the water drain valve
on the left side of the instrument to discharge the effusion (confirm that the seal is
complete when tightening).
7. Please turn off the power after use.
8. The cold trap host provides one year warranty. In case of non-human damage to the
product, the company will provide free maintenance service within 1 year from the date of
purchase. In case of human damage within 1 year or failure beyond the 1 year warranty
period, the company will charge a certain cost for maintenance. Repair by unauthorized
professionals. Unauthorized professionals shall not open the machine without the
permission of the company; All repairs and adjustments must be carried out by
professionals in accordance with the maintenance manual. The company will not be
responsible for the maintenance and adjustment of unauthorized personnel!
Table of contents
Other MRC Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Thermo Scientific
Thermo Scientific Omni FTIR Multi Gas CEMS instruction manual

COMINOX
COMINOX STERILCLAVE 18 Series Service manual

Egger
Egger EL 1 plus N2 Operating instruction

Germfree
Germfree LAF user manual

Thermo Scientific
Thermo Scientific MEGA-PURE 11A Water Still 677 Series Operating manual and parts list

Thermo Scientific
Thermo Scientific MaxQ 2000 CO2 Plus Operation manual