MSA M1 Installation guide

Basic Manual
M1 LGDV
Positive Pressure
Lung Governed Demand Valve
Order No.: D2018702-EN/00
MSA Europe GmbH
Schlüsselstrasse 12
8645 Rapperswil-Jona
Switzerland
www.MSAsafety.com
1 Safety Regulations
1.1 Correct Use
The M1 LGDV, in this manual also referred to as device, is a lung governed
demand valve for a Compressed Air Breathing Apparatus or Compressed
Airline Breathing Apparatus. It is connected between compressed air supply
and full face mask and controls the inhaled air.
The M1 lung governed demand valve series is only permitted for specific
system components.
WARNING!
The device is not an independent breathing apparatus. It must be used
exclusively mounted on a full face mask and an SCBA / CABA.
The device is not suitable for underwater diving.
WARNING!
This product is supporting life and health. Inappropriate use, maintenance
or servicing may affect the function of the device and thereby seriously
compromise the user‘s life.
Before use, the product operability must be veried. The product must not
be used if the function test is unsuccessful, it is damaged, a competent
servicing/maintenance has not been made, genuine MSA spare parts
have not been used..
1.2 Liability Information
MSA accepts no liability in cases where the device has been used
inappropriately or not as intended. The selection and use of the device are
the exclusive responsibility of the individual operator.
Product liability claims, warranties and guarantees made by MSA with
respect to the device are voided, if it is not used, serviced or maintained in
accordance with the instructions in this manual.
1.3 Series
This manual applies to the following lung governed demand valves from the
M1 LGDV series:
AS (push to connect / quick connect)
• M1-AS LGDV, short
• M1-AS LGDV, long
• M1-AS LGDV, fix
M45x3 (threaded version)
• M1-AE LGDV M45x3, short
• M1-AE LGDV M45x3, long
• M1-AE LGDV M45x3, fix
ESA (push to connect / quick connect according to DIN 58600)
• M1-ESA LGDV, short
• M1-ESA LGDV, long
AS-B (push to connect / quick connect with Bypass)
• M1-AS-B LGDV, short
• M1-AS-B LGDV, long
Explanations on model designation
AS: Positive pressure, plug-in connection, suitable for full face
masks PS-MaXX
AE: Positive pressure, thread connection M45x3 according to
DIN EN 148-3, suitable for full face masks -PF
ESA: Positive pressure, plug-in connection according to DIN 58600,
suitable for full face masks ESA
AS-B: Positive pressure, bypass for constant air flow
2 Description
1
2
3
4
5
6
8a 8b
7
9
M1-AS LGDV M1-AE LGDV M1-ESA LGDV M1-AS-B LGDV
1 Flush button 6
2 Cover 7
3 Positive pressure spring 8a
4 Diaphragm 8b
5 Valve assembly (inclusive housing) 9
Dust cap
Medium pressure hose
Short / long version for coupling
Fixed version
Bypass
3 Use
3.1 Preparing the Device
WARNING!
The lung governed demand valve may only be put into use in a fully main-
tained and tested condition. If malfunctions or defects are noticed prior to
use, do not use the lung governed demand valve under any circumstan-
ces. Failure to follow this warning can result in serious personal injury or
death.
Get the apparatus checked and repaired by an authorised service centre.
Switching off the Device
Before connecting the device to medium pressure, in order to prevent
accidental air delivery, push the red operating button to switch off the device.
3.2 Connecting the Device
1. Insert plug-in nipple of the device into the coupling of the medium
pressure hose of an approved SCBA until it distinctly snaps into place (not
for fixed LGDV).
2. Don full face mask and perform tightness check according to the mask
operating manual.
3. Open cylinder valve(s) fully.
4. Screw the device into the mask (M1-AE) or insert it into the adapter of the
mask until it distinctly snaps into place (M1-AS, M1-AS-B, M1-ESA).
Support during Donning
MSA recommends that the device must be attached to the mask and the
connection inspected by a second person.
3.3 Putting the Device into Operation
To activate the device, inhale once forcefully or push the flush button.
WARNING!
When using the ush button, the compressed air consumption increases.
In case of an emergency, push the flush button to flush out smoke
and gases from the mask.
3.4 Switching off the Device
To switch off the device, push red operating button until stopper.
GB
3.5 Removing the Device from Mask
For M1-AE:
1. Switch off the device by pushing the red operating button.
2. Unscrew handwheel and separate device from the mask.
For M1-AS, M1-AS-B and M1-ESA:
1. Press both operating buttons simultaneously.
2. Separate the device from the mask.
3.6 Removing the Device from SCBA
1. If necessary, uncouple medium pressure hose (not for fixed LGDV):
Push the plug-in nipple into the coupling of the medium pressure hose
forcefully and pull back the coupling sleeve at the same time.
3.7 Bypass System (only for M1-AS-B LGDV)
A bypass system is available for the M1-AS-B LGDV.
It is a safety system that supplies the user with continuously flowing air
independent of the lung governed demand valve. By turning the red
handwheel, the bypass is opened and can be set individually.
WARNING!
When using the bypass, the compressed air consumption increases.
4 Maintenance
4.1 Maintenance Intervals
Component Work to be
performed
After use Annually Every 3
years
Every 10
years1)
M1 LGDV
Cleaning X X
Disinfection X
Sight, function
and tightness
test
X X2)
Diaphragm
replacement X
Overhaul X
1) For compressed air breathing apparatus that are frequently used, MSA
recommends a complete overhaul after approx. 540 hours. (This corresponds to
1080 uses with a usage duration of 30 minutes.)
2) Visual test of diaphragm only after use in aggressive media or under extreme
conditions like i. e. excessive heat; function and tightness test always.
4.2 Cleaning and Disinfection
For cleaning and disinfection, see washing instructions.
4.3 Storage
Always store positive pressure lung governed demand valves when switched
off. To do this, push the red operating button.
Store the device with protective cap attached. The storage should be in a
clean and dry condition at normal climate free of hazards, protected against
direct sunlight and heat.
The storage room must be cool, dry, dust-free and moderately ventilated.
The optimal storage temperature is between +15 °C and 25 °C. The optimal
relative humidity is below 65 %.
4.4 Malfunction
In case of malfunctions, the device must be inspected and if necessary
repaired by MSA or trained service personnel.
5 Technical Data
Operating pressure : 4 - 9 bar dynamic
Operating temperature : -40 °C to 60 °C
Operating gas : Breathing air according to EN12021 or
better
Max. flow bypass (AS-B) : 650 l/min
Transponder : Amplitude modulation: 13.56 MHz, 1024 bit,
IP 68
Temperature range: -40 °C up to +70 °C
6 Certication
Approvals The device complies with the Regulation (EU) 2016/425.
0158
EN 137 Lung governed demand valves of the M1 LGDV series
are approved as components of a compressed air
breathing apparatus in compliance with EN 137.
DEKRA Dinnendahlstr. 9,44809 Bochum, Germany
The Declaration of Conformity can be found under the following link:
https://MSAsafety.com/DoC
6.1 Marking
For ATEX marking for use in potentially explosive atmospheres see M1
LGDV operating manual.
7 Approved System Components
Compressed air breathing apparatus - Series
M1 ATO
Full face mask - Series
3S
Ultra Elite
G1
ATO = Assemble To Order
Manual Part Number
Operating Manual M1 LDGV 10190261
Operating Manual G1 Facepiece 10161150
Operating Manual 3S and Ultra Elite full face masks 10126346
See MSAsafety.com.

Other manuals for M1

9

Other MSA Control Unit manuals

MSA SM5000 User manual

MSA

MSA SM5000 User manual

MSA Cylinder Valve Setup guide

MSA

MSA Cylinder Valve Setup guide

MSA FireHawk M7 Operating instructions

MSA

MSA FireHawk M7 Operating instructions

MSA SMC 5100-28-IT User manual

MSA

MSA SMC 5100-28-IT User manual

MSA Chillgard LC User manual

MSA

MSA Chillgard LC User manual

MSA AutoMaXX-AE User manual

MSA

MSA AutoMaXX-AE User manual

MSA GasGard XL User manual

MSA

MSA GasGard XL User manual

MSA 9010 SIL User manual

MSA

MSA 9010 SIL User manual

MSA ICU User manual

MSA

MSA ICU User manual

MSA 9010 LCD User manual

MSA

MSA 9010 LCD User manual

MSA M1 User manual

MSA

MSA M1 User manual

MSA M1 User manual

MSA

MSA M1 User manual

MSA 610A User manual

MSA

MSA 610A User manual

MSA SM5000 User manual

MSA

MSA SM5000 User manual

MSA 9010 User manual

MSA

MSA 9010 User manual

MSA M1 User manual

MSA

MSA M1 User manual

MSA ASV User manual

MSA

MSA ASV User manual

MSA AutoMaXX-ASE User manual

MSA

MSA AutoMaXX-ASE User manual

Popular Control Unit manuals by other brands

Bardiani BBZS1 Instruction, use and maintenance manual

Bardiani

Bardiani BBZS1 Instruction, use and maintenance manual

Viessmann VITOTRONIC 300 Installation and service instructions

Viessmann

Viessmann VITOTRONIC 300 Installation and service instructions

GITER G5040-48 installation manual

GITER

GITER G5040-48 installation manual

Vimar by-me Eikon 20546 manual

Vimar

Vimar by-me Eikon 20546 manual

Granville-Phillips Micro-Ion 354 E Series instruction manual

Granville-Phillips

Granville-Phillips Micro-Ion 354 E Series instruction manual

Texas Instruments TPS2HCS10-Q1 user guide

Texas Instruments

Texas Instruments TPS2HCS10-Q1 user guide

Azbil Actival VY5199J Series Specifications and Instructions

Azbil

Azbil Actival VY5199J Series Specifications and Instructions

GEM 639 eSyStep operating instructions

GEM

GEM 639 eSyStep operating instructions

MICRO-EPSILON IF1032/ETH instruction manual

MICRO-EPSILON

MICRO-EPSILON IF1032/ETH instruction manual

Digi XBee-PRO 900HP user guide

Digi

Digi XBee-PRO 900HP user guide

GE 22158 user manual

GE

GE 22158 user manual

Jung 2224 TSM operating instructions

Jung

Jung 2224 TSM operating instructions

Richardson RFPD RadioCarbon RFPD-RC-4450-50 quick start guide

Richardson RFPD

Richardson RFPD RadioCarbon RFPD-RC-4450-50 quick start guide

Nexa ZRT-106 quick guide

Nexa

Nexa ZRT-106 quick guide

RSP P0010A product manual

RSP

RSP P0010A product manual

handtmann 12503 operating manual

handtmann

handtmann 12503 operating manual

Schweitzer Engineering Laboratories SEL-2245-4 manual

Schweitzer Engineering Laboratories

Schweitzer Engineering Laboratories SEL-2245-4 manual

Emona Instruments DxIQ-45G manual

Emona Instruments

Emona Instruments DxIQ-45G manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.