MSA S5000 User manual

MSAsafety.com
Operating Manual
General Monitors S5000
Gas Monitor
Order No.: MANS5000/03
CR 800000032562

26776 Simpatica Circle
Lake Forest, CA 92630 USA
For countries of Russian Federation, Republic of Kazakhstan and Republic of Belarus, the gas detector
will be delivered with a passport document that includes valid approval information. On the CD with
manual instruction attached to the gas detector the user will find the documents "Type Description" and
"Test Method" - appendixes to Pattern Approval Certificate of Measuring instrument, valid in the coun-
tries of use.

3
US
General Monitors S5000
Contents
1 Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 Correct Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Product Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2 Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2 No Tool Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3 Bluetooth®Wireless Technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.4 Dual Sensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.5 Retrofit Installations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.6 XCell Sensors Optimized for Fixed Gas Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.7 TruCal Sensing Technology for CO and H2S Electrochemical Sensors . . . . . . . . . . . . . . . . . . . . . 9
2.8 SafeSwap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.9 Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.10 Component Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.11 Label Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1 Installation Warnings - Read before Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2 Reviewing Shipment and Identifying Product Model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.3 Product Installation Check List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.4 Mounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.4.1 Sensor Mounting Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.4.2 Transmitter Mounting Location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.4.3 Sensor Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.4.4 Connecting Sensor to Transmitter Housing or Remote Junction Box . . . . . . . . . . . . . . . . . . . . . . 18
3.4.5 Integrated Mounting Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.4.6 Adjustable Pipe Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.4.7 Duct Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.4.8 Mounting with a Sunshield . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.5 Installing a Remote Sensor Junction Box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.6 Electrical Power Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.6.1 Electrical Warnings - Read before Connecting Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.6.2 Retrofit Applications with S4000CH, S4000TH, or TS4000H. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.6.3 Electrical Hardware Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.6.4 Power Load Requirements and Maximum Mounting Distances. . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.6.5 Instructions for Power and Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.6.6 Relay Electrical and Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.1 Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.1.1 Initial Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.1.2 Sensor Warm Up Times . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.1.3 Startup after Power Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.2 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.2.1 Instrument Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.2.2 Sensor Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.2.3 Relay Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
4.2.4 HART Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
4.2.5 HART AO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
4.2.6 Bluetooth®. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

4
US
General Monitors S5000
4.2.7 Modbus Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
4.2.8 EZ Touch Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
4.2.9 Swap Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ... 51
4.2.10 UI Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
4.2.11 Transmitter Setting Reset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
4.3 Info Menu - Viewing Device Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4.3.1 View or Change System Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4.3.2 Last Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4.3.3 Non-Critical Fault Log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4.3.4 Device Tag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.3.5 Bluetooth®FCC/IC ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
4.3.6 Sensor Life and Health Status (only displayed if XCell Sensor is connected) . . . . . . . . . . . . . . . . 54
4.4 Setting only configurable via Bluetooth®, Modbus, or HART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4.4.1 AO Custom Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4.4.2 Relay Zone - Horn Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4.4.3 Unit Device ID/Tag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4.4.4 Alarm Direction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
4.4.5 Alarm Enable/Disable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
5.1 Calibration Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
5.2 Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
5.3 Calibration Frequency for XCell Sensors with TruCal (H2S & CO only). . . . . . . . . . . . . . . . . . . . . 58
5.4 Calibration Types: Zero vs. Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
5.5 How to Zero Calibrate XCell and IR400 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.6 How to Calibrate XCell Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.7 How to Calibrate an Oxygen XCell Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
5.8 How to Calibrate an IR400 Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
5.9 How to Calibrate a Passive Sensor (Catalytic Bead or MOS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
5.10 XCell Catalytic Bead Failsafe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
5.11 Calibration Confirmation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
6 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
6.1 IR400 Cleaning Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
6.2 Replacing an XCell Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.3 Replacing a Passive Sensor (Cat Bead or MOS). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
6.4 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
7 Ordering Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7.1 Replacement Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
8 Appendix: Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
9 Appendix: General Certification Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
10 Appendix: HART Specific Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80

5
Safety Regulations
US
General Monitors S5000
1 Safety Regulations
1.1 Correct Use
The S5000 Gas Monitor, hereafter also called device, is a gas monitor for measuring toxic and
combustible gases as well as oxygen. Using sensors, the device tests the ambient air and triggers
the alarm as soon as the gas exceeds a specific concentration level.
WARNING!
Read this manual carefully. The device will perform as designed only if it is used, installed and
maintained in accordance with the manufacturer's instructions. Otherwise, it could fail to perform
as designed and persons who rely on this device for their safety could sustain serious personal
injury or loss of life.
WARNING!
Do not use silicone-type lubricants in assembling the device and do not allow silicone vapors to
be drawn into the flow system while in operation. Silicone can desensitize the combustible gas
sensor, thereby giving erroneously low readings.
Use only genuine MSA replacement parts when performing any maintenance procedures
provided in this manual. Failure to do so may seriously impair sensor and gas monitoring perfor-
mance, alter flameproof/explosionproof characteristics or void agency approvals.
Failure to follow the above warnings can result in serious personal injury or loss of life.
WARNING!
The Digital Sensor Xcell sensor module utilizes thread locker suitable to an ambient temperature
of -55°C to +74°C. If the Digital Sensor is exposed to temperatures outside of its listed ratings,
reapplication of thread locker when changing out the Xcell sensor module may be required.
Failure to follow the above warning can result in serious personal injury or loss of life.
NOTICE
This is a Class A product in accordance with CISPR 22. In a domestic environment, this product
may cause radio interference, in which case the user may be required to take adequate measures.
NOTICE
The XCell sensor refers to the sensor portion of the Digital Sensor throughout this manual.
FCC Warning Statements
Changes or modifications not expressly approved by the manufacturer could void the user's
authority to operate the equipment.
This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions:
• this device may not cause harmful interference, and
• this device must accept any interference received, including interference that may
cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in
a commercial environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.

6
Safety Regulations
US
General Monitors S5000
Industry Canada (IC) Warning Statements
The installer of this radio equipment must ensure that the antenna is located or pointed such that
it does not emit RF field in excess of Health Canada limits for the general population; consult
Safety Code 6, obtainable from Health Canada's website www.hc-sc.gc.ca.
1.2 Product Warranty
The warranties made by GM with respect to the product are voided if the product is not installed,
used and serviced in accordance with the instructions in this manual. Please protect yourself and
your employees by following the instructions.
This warranty does not cover filters, fuses, etc. Certain other accessories not specifically listed
here may have different warranty periods. This warranty is valid only if the product is maintained
and used in accordance with Seller's instructions and/or recommendations. The Seller shall be
released from all obligations under this warranty in the event repairs or modifications are made by
persons other than its own or authorized service personnel or if the warranty claim results from
physical abuse or misuse of the product. No agent, employee or representative of the Seller has
any authority to bind the Seller to any affirmation, representation or warranty concerning this
product. Seller makes no warranty concerning components or accessories not manufactured by
the Seller, but will pass on to the Purchaser all warranties of manufacturers of such components.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED, IMPLIED OR
STATUTORY, AND IS STRICTLY LIMITED TO THE TERMS HEREOF. SELLER SPECIFI-
CALLY DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR OF FITNESS FOR A
PARTICULAR PURPOSE.
Exclusive Remedy
It is expressly agreed that Purchaser's sole and exclusive remedy for breach of the above
warranty, for any tortious conduct of Seller, or for any other cause of action, shall be the replace-
ment at Seller's option, of any equipment or parts thereof, which after examination by Seller is
proven to be defective. Replacement equipment and/or parts will be provided at no cost to
Purchaser, F.O.B. Seller's Plant. Failure of Seller to successfully replace any nonconforming
equipment or parts shall not cause the remedy established hereby to fail of its essential purpose.
Exclusion of Consequential Damage
Purchaser specifically understands and agrees that under no circumstances will seller be liable to
purchaser for economic, special, incidental or consequential damages or losses of any kind what-
soever, including but not limited to, loss of anticipated profits and any other loss caused by reason
of non-operation of the goods. This exclusion is applicable to claims for breach of warranty,
tortious conduct or any other cause of action against seller.
ITEM WARRANTY PERIOD
S5000 Gas Monitor
MSA warrants that this product will be free from mechanical
defects and faulty workmanship for the period specified in this
table for each component, provided it is maintained and used in
accordance with MSA’s instructions and/or recommendations.
Warranty shall not exceed.
Main Transmitter Housing
and PCBA
2 years from date of shipment.
Shall not exceed 2 years and 6 months from date of manufacture.
XCell Sensors 3 years from date of shipment.
Shall not exceed 3 years and 6 months from date of manufacture.
IR400 2 years from date of shipment.
Shall not exceed 2 years and 6 months from date of manufacture.
Passive Catalytic Bead 2 years from date of shipment.
Shall not exceed 2 years and 6 months from date of manufacture.
Passive MOS 2 years from date of shipment.
Shall not exceed 2 years and 6 months from date of manufacture.

7
Description
US
General Monitors S5000
2 Description
2.1 Display
The S5000 utilizes a dot matrix LED display, capable of displaying four alphanumeric characters
at a time. The display will scroll words that exceed four letters. Most of the messages scroll twice
across the screen before moving onto next selection.
Fig. 1 S5000 Main Display
In addition to the red LED display, the S5000 uses six icons to indicate status. Green LED indi-
cates power supply status. A yellow triangle and red bell indicate fault and warning or alarm condi-
tions respectively. The Bluetooth®icon indicates that the Bluetooth®wireless technology is
enabled on the device. Yellow "1" and "2" icon indicate which sensor gas reading is being
displayed, or during configuration which sensor's options are being accessed.

8
Description
US
General Monitors S5000
2.2 No Tool Interface
The S5000 does not require any tools or third party devices to change settings, reset alarms or
perform any maintenance operation. The EZ touch button works through the glass and does not
require opening the explosion proof enclosure. The EZ touch button works with bare fingers or
with gloved hands, so long as the gloves are not black. See section 4 for more information on navi-
gating the menu with the EZ touch button.
Fig. 2 Interface
2.3 Bluetooth®Wireless Technology
The S5000 comes by default with Bluetooth®communication. Using the X/S Connect App on an
appropriate smart phone or tablet, users are able to interface with the S5000 menu options in a
larger and more user friendly setting. The Bluetooth® communication can interact with the device
within a maximum transmission distance of 70 feet (21 m).
WARNING!
Bluetooth®Operation is dependent upon signal availability of the wireless service(s) necessary to
maintain the communication link. Loss of wireless signal will prevent communication of alarms and
other information to linked devices. Take appropriate precautions in the event a loss of wireless
signal occurs.
Failure to follow the above warning can result in serious personal injury or loss of life.
The S5000 and user provided communication device will need to be paired. This requires both
devices to be in range and for a pairing sequence inputting a 6 digit pairing code. The instructions
will be displayed on both the S5000 and communication device.
The user menu can also be accessed using the round GM magnet on the General Moni-
tors logo.
If ordered with Bluetooth®, the device will be shipped with Bluetooth®enabled. See
section 4 for instructions on disabling Bluetooth®.
If the device was not ordered with Bluetooth®it cannot be upgraded.
There are communication devices capable of being used in classified areas. Please
contact your MSA representative for additional information.

9
Description
US
General Monitors S5000
2.4 Dual Sensing
The S5000 supports two Digital Sensors, or one IR400 point IR detector and one Digital Sensor
simultaneously with four wire connections. However, the device will only support one passive
sensor, either combustible catalytic bead or metal oxide semiconductor (MOS) sensors, based on
the ATO configuration. Passive catalytic bead uses three wires, passive MOS uses four wires.
The S5000 Gas Monitor generates two independent analog outputs; one for each sensor
connected to the transmitter. The analog output associated with Sensor 1 also has the digital
HART (Highway Addressable Remote Transducer) communication superimposed on the analog
signal. If two sensors are connected, the digital HART communication carries information for both
sensors.
2.5 Retrofit Installations
The S5000 has the conduit entries located in the exact same orientation and distance from the
wall and the mounting holes for attaching to a wall are identical to the S4000 Series detectors.
Users will be able to re-use the existing wiring for the remotely mounted sensors.
2.6 XCell Sensors Optimized for Fixed Gas Applications
XCell toxic and combustible catalytic bead sensors are developed and manufactured by MSA.
Now optimized for fixed gas applications, the XCell sensor platform is available in the S5000 and
provides multiple benefits, including a standard 3-year warranty on all XCell sensors.
One important optimization for fixed gas was incorporating the GM catalytic bead into the XCell
sensor.
The XCell Oxygen sensor does not use lead, but rather a non-consuming reaction chemistry. The
XCell Oxygen sensor is expected to last well over 3 years and can be safely stored on the shelf
for at least 1 year without sensor performance degradation.
2.7 TruCal Sensing Technology for CO and H2S Electrochemical Sensors
Using patented pulse check technology and proprietary Adaptive Environmental Compensation
(AEC) algorithms, XCell sensors wtih TruCal verify operation.
Automated pulse checks are the muscle behind the TruCal technology. Every six hours, an elec-
trical pulse stimulates the XCell sensor similar to having actual calibration gas applied, providing
a snapshot of the sensor's sensitivity at the time of the pulse. Using this sensitivity snapshot, the
sensor can diagnose sensor failures like electrode poisoning, electrolyte leaking, or electrical
connectivity issues. The sensitivity snapshots are mapped over time to provide an overall Life &
Health status as either "Good" or "Fair" if the sensor is approaching end of life.
But that's not all. The brains of the TruCal technology lie within the proprietary Adaptive Environ-
mental Compensation (AEC) algorithms. AEC uses the sensitivity snapshots provided by the
pulse check to adjust sensor output, compensating for environmental impacts on sensor accu-
racy. If the AEC adjustment is greater than expected based on typical environmental impact vari-
ations, the transmitter LED’s will slowly flash GREEN, indicating that the sensor should be
calibrated to reset the AEC cycle. The result is a sensor that actively self-monitors for operation
and accuracy, with far fewer manual calibrations.
Actual TruCal sensor performance will depend on the application, background gas exposure, and
environment. To validate XCell sensors with TruCal, it is recommended that users follow their
regular calibration cycle and record the "as found" and "as left" values. This data can be used to
extend the time between calibrations depending on the required specification of the application.

10
Description
US
General Monitors S5000
2.8 SafeSwap
The S5000 comes with patented SafeSwap technology, which allows users to change or replace
XCell sensors without needing to power down the instrument. Swap delay is enabled on the
S5000 by default; a feature that gives users a 2 minute window to change sensors without trig-
gering a fault condition. SafeSwap and Swap Delay are only applicable for XCell sensors. For
more information on SafeSwap and Swap Delay, see section 6.
WARNING!
As part of the product certification, it was verified that optional communication functions of this
gas detection instrument while operating at the maximum transaction rate do not adversely
affect the gas detection operation and functions of the instrument. The product certification,
however, does not include or imply approval of the SafeSwap feature, communications
protocol or functions provided by the software of this instrument or of the communications
apparatus and software connected to this instrument.
Follow the warnings below when removing or replacing sensors. Reference Fig. 3 for compo-
nent overview.
- Never remove or replace a Sensor Body Assembly or an IR400 while under power or when
explosive hazards are present.
- Confirm that the area is free of explosive hazards before removing or replacing an XCell
Sensor under power.
- To remove an XCell Sensor, unscrew XCell Sensor three full turns, wait 10 seconds, and then
remove the XCell Sensor completely.
Failure to follow the above warnings can result in serious personal injury or loss of life.
2.9 Housing
The S5000 comes in 316 Stainless Steel for the highest corrosion resistance. All housings have
¾” NPT conduit entries. Custom tags are available and easily attach to an integral ring.

11
Description
US
General Monitors S5000
2.10 Component Overview
Fig. 3 Exploded View

12
Description
US
General Monitors S5000
2.11 Label Overview
Fig. 4 Transmitter - Position of Labels
Fig. 5 Board stack - Position of Labels
Fig. 6 Digital Sensor - Position of Labels

13
Description
US
General Monitors S5000
Fig. 7 IR400 - Position of Labels
Fig. 8 Passive Sensor - Position of Labels

14
Installation
US
General Monitors S5000
3 Installation
3.1 Installation Warnings - Read before Installation
WARNING!
Refer to section 9 "Appendix: General Certification Information" before installation and opera-
tion.
Some Digital Sensors are provided in a fritless sensor housing. The fritless sensor housing is
labeled as Div 2 or Zone 2 and is approved for Div 2 or Zone 2 installations only. The protection
method is Nonincendive or Type n respectively. Ensure all components are approved for the
wiring method being used and in accordance with the National Electric Code of the country of
use, any applicable local regulations, and this manual.
As part of the product certification, it was verified that optional communication functions of this
gas detection instrument while operating at the maximum transaction rate do not adversely
affect the gas detection operation and functions of the instrument. The product certification,
however, does not include or imply approval of the SafeSwap feature, communications
protocol or functions provided by the software of this instrument or of the communications
apparatus and software connected to this instrument.
Follow the warnings below when removing or replacing sensors. Reference Fig. 3 for compo-
nent overview.
- Never remove or replace a Sensor Body Assembly or an IR400 while under power or when
explosive hazards are present.
- Confirm that the area is free of explosive hazards before removing or replacing an XCell
Sensor under power.
- To remove an XCell Sensor, unscrew XCell Sensor three full turns, wait 10 seconds, and then
remove the XCell Sensor completely.
Do not paint the device. Avoid painting in areas where the S5000 and remote sensor junction
box are located. If painting is required in an area where an S5000 or remote sensor has been
installed, exercise caution to ensure paint is not deposited on the sensor inlet fitting. Paint
solvents can also cause an alarm condition to occur or potentially poison electrochemical
sensors.
Protect the device from extreme vibration.
Do not mount the sensing head in direct sunlight without a sunshield (Part Number 10180254).
The IR400 sensor contains no user- or field-serviceable parts and must be returned to the
factory for repair. Any attempt to open the monitor will damage the unit and void the warranty.
Failure to follow the above warnings can result in serious personal injury or loss of life.
NOTICE
When installing the IR400 sensor, under no circumstances should a pry-bar be applied to the two
legs that support the unit's reflectors during installation or removal of the sensor. Applying force
to the legs can permanently damage the IR400 sensor.

15
Installation
US
General Monitors S5000
3.2 Reviewing Shipment and Identifying Product Model
To determine the sensor type and options, check the shipping carton.
Fig. 9 Shipping Label
Digital and IR400 sensors are not shipped attached to the S5000. The IR400 is a one piece
sensor, but all other Digital Sensors are comprised of two parts; the Sensor Body Assembly and
the XCell Sensor. Sensor Body Assemblies must be installed and tightened using a strap wrench.
Check the sensor details before attaching to the S5000 housing. The sensor details are listed on
the inside of the XCell Sensor. Unscrew the XCell Sensor from the Sensor Body Assembly and
check the label on the inside for gas type, range, configuration ordered, serial number, and firm-
ware revision number.
Passive sensors are shipped attached to the housing.
3.3 Product Installation Check List
Before Installation
• Review national electrical codes
• Review local procedural and building codes
• Determine optimum transmitter placement
• Determine wire requirements
• Determine mounting hardware requirements
• Review approvals and ensure suitability for installation
Mounting
• Attach appropriate sensor to housing or junction box (see section 3.4.3 for proper sensor orien-
tation)
• Mount transmitter or junction box using appropriate mounting hardware
• Confirm free air flow around the sensor

16
Installation
US
General Monitors S5000
3.4 Mounting
WARNING!
Refer to section 9 "Appendix: General Certification Information" before installation.
Some toxic gases are provided in a fritless sensor housing. The fritless sensor housing is labeled
as Div 2 or Zone 2 and is approved for Div 2 or Zone 2 installations only. The protection method
is Nonincendive or Type n respectively. Ensure all components are approved for the wiring
method being used and in accordance with the National Electrical Code of the country of use, any
applicable local regulations, and this manual.
Failure to follow the above warning can result in serious personal injury or loss of life.
3.4.1 Sensor Mounting Location
The best location for the transmitter and the sensor may not be the same location. Sensors should
be placed in a location where a gas leak is most likely to be detected. When the best sensor place-
ment would not allow the transmitter display to be easily viewed or accessed, a remote junction
box can be used to mount the sensor remotely from the transmitter, allowing both to be installed
in the optimum location.
Two main factors should be considered when choosing a sensor location. The first is the density
of the target gas relative to the air. Gases, such as propane, that are heavier than air should be
placed near ground level while gases that are lighter than air should be placed above potential
leak sources.
Optimum sensor placement will depend on the surrounding processing equipment, such as pipes,
valves, or turbines. MSA offers a gas and flame mapping service that systematically evaluates
potential sources of leaks and recommends detector quantity and placement to create the most
effective detection system.
3.4.2 Transmitter Mounting Location
The transmitter display should be mounted so that the screen is visible and easily accessed after
installation. The electronics assembly inside the enclosure have one orientation inside the cylin-
drical housing. Take care to position the conduit entries and display so that the display is read in
the correct orientation.

17
Installation
US
General Monitors S5000
3.4.3 Sensor Orientation
WARNING!
Mount the IR400 with the sensor inlet fitting extended horizontally from the main enclosure (Fig.
10) to prevent the build-up of particulate or liquid matter on the monitor's optical surfaces.
Mount the digital sensor with the sensor inlet fitting (Fig. 11) pointed downward; otherwise, the
inlet may become clogged with particulate matter or liquids.
Failure to follow the above warnings may result in serious personal injury or loss of life.
Sensor orientation will depend on the sensor type. If mounting an IR400 point IR sensor, whether
locally on the transmitter or via remote junction box, the sensor should be mounted horizontally.
If the IR400 sensor is not mounted horizontally, the sensor will be prone to more frequent beam
blocking issues due to accumulated dust and condensation on the surface of the IR400 sensor.
Fig. 10 shows the correct and incorrect mounting orientations for the IR400.
Fig. 10 Correct and Incorrect Mounting Orientations for IR400 Point IR Sensor
All other sensors, including electrochemical, catalytic bead, oxygen, passive catalytic bead, and
passive MOS should be mounted vertically with the gas inlet pointed downward. If the sensor is
not mounted with the gas inlet facing down, it is more likely to become clogged with particulate
matter or liquids. Fig. 11 shows the correct and incorrect mounting orientation for digital sensors.
Passive catalytic bead and MOS sensors come already installed on the transmitter housing.
Fig. 11 Correct and Incorrect Mounting Orientation for Digital Sensors

18
Installation
US
General Monitors S5000
3.4.4 Connecting Sensor to Transmitter Housing or Remote Junction Box
Digital and IR400 sensors are not shipped attached to the main enclosure. The main sensor input
is provided via a four-terminal connection that provides a digital interface for all sensor modules.
Up to two sensors (excluding passive sensors) can be connected to a single transmitter with two
analog outputs capable of representing the readings of the individual sensors.
If planning to install the sensor attached to the transmitter or remote junction box, connect the
sensor before selecting the final mounting location so that the dimensions of the sensor can be
considered.
Passive sensors are shipped already attached and electrically wired to the device. Only one
passive sensor can be used on a single S5000, and they are not interchangeable with other
passive sensors or Digital Sensors. Consider the sensor dimensions when choosing a mounting
location for the transmitter or junction box.
To connect the sensor:
(1) Loosen the set screw located on the lid using a 1.5 mm Allen wrench.
(2) Turn the lid counter-clockwise to remove.
(3) Pull out display module to expose terminal connections.
Fig. 12 Terminal Connections
(4) Route the cable from the sensor through a conduit entry hole in the enclosure so that the
sensor is oriented in the correct position (see section 3.6 for details).
(Repeat to attach a second sensor to the S5000 transmitter).
(5) Connect the sensor to the "Sensor 1" position on the electronics assembly.
a) If using a second sensor, connect it to the second sensor position.
NOTICE
If only using one sensor and it is connected to “Sensor 2” position, the S5000 will enter Sensor
Missing fault. See Disable Sensor in section 4.2.2 for details on how to clear this fault.

19
Installation
US
General Monitors S5000
Fig. 13 Connecting Sensor to the Stack and Grounding Terminal
(6) Verify the sensor connector is firmly seated on the terminal board.
(7) Attach the sensor's ground to either of the grounding screws inside the S5000 housing.
(8) Replace the display module. Push firmly on the board stack.
NOTICE
Ensure that the electronics assembly is fully engaged in the mounting holes. If not fully seated,
the touch interface performance can be negatively affected.
(9) Replace the cover by turning clockwise.
(10) Tighten the set screw located on the lid using a 1.5 mm Allen wrench.

20
Installation
US
General Monitors S5000
3.4.5 Integrated Mounting Points
The S5000 transmitter can be surface mounted without any additional brackets using the inte-
grated mounting tabs.
Fig. 14 Integral Mounting Tabs (not compatible with S4000 IR400 sensors)
3.4.6 Adjustable Pipe Mount
A Universal Pipe Mount Kit (Part Number 10176946) can be used to mount the S5000 on pipes
ranging from 20-150 mm in diameter. Two brackets are mounted over the top of the integrated
mounting tabs and fitted with an adjustable pipe band (not included).
Fig. 15 Adjustable Pipe Mount
Other manuals for S5000
1
Table of contents
Other MSA Gas Detector manuals
Popular Gas Detector manuals by other brands

Detcon
Detcon SmartWireless CXT Operator's installation and instruction manual

Sygonix
Sygonix 2483186 operating instructions

Riken Keiki
Riken Keiki GD-K88Ai operating manual

Medem
Medem SEC-L v4 installation instructions

RKI Instruments
RKI Instruments RIKEN RX-516 Operation manual

Riken Keiki
Riken Keiki SD-1 operating manual

Xtralis VESDA
Xtralis VESDA Sensepoint XCL Frequently asked questions

RKI Instruments
RKI Instruments SC-04 Operator's manual

Det-Tronics
Det-Tronics Eagle Quantum 95-8470-05 instructions

Honeywell
Honeywell PHD6 Quick reference card

RKI
RKI GX-6000 Operator's manual

RKI Instruments
RKI Instruments GX-3R Pro Operator's manual