NARDA PMM PL-01 User manual

Other NARDA Measuring Instrument manuals

NARDA L3HARRIS SMARTS II User manual

NARDA

NARDA L3HARRIS SMARTS II User manual

NARDA ESM-10 RadMan Mobile User manual

NARDA

NARDA ESM-10 RadMan Mobile User manual

NARDA AMS-8063 Series User manual

NARDA

NARDA AMS-8063 Series User manual

NARDA RadMan 2LT User manual

NARDA

NARDA RadMan 2LT User manual

NARDA IDA User manual

NARDA

NARDA IDA User manual

NARDA nbm-520 User manual

NARDA

NARDA nbm-520 User manual

NARDA nbm-520 User manual

NARDA

NARDA nbm-520 User manual

NARDA SRM-3006 User manual

NARDA

NARDA SRM-3006 User manual

NARDA EFA-200 User manual

NARDA

NARDA EFA-200 User manual

NARDA NBM-550 User manual

NARDA

NARDA NBM-550 User manual

NARDA SignalShark 3310 User manual

NARDA

NARDA SignalShark 3310 User manual

NARDA FieldMan User manual

NARDA

NARDA FieldMan User manual

NARDA EHP-50-G User manual

NARDA

NARDA EHP-50-G User manual

NARDA nbm-520 User manual

NARDA

NARDA nbm-520 User manual

NARDA AMS-8061 User manual

NARDA

NARDA AMS-8061 User manual

NARDA HP-01 User manual

NARDA

NARDA HP-01 User manual

NARDA NBM-550 User manual

NARDA

NARDA NBM-550 User manual

NARDA NIM-511 User manual

NARDA

NARDA NIM-511 User manual

NARDA FieldMan User manual

NARDA

NARDA FieldMan User manual

NARDA 8718B User manual

NARDA

NARDA 8718B User manual

NARDA PMM CA0010 User manual

NARDA

NARDA PMM CA0010 User manual

NARDA PMM 6600 User manual

NARDA

NARDA PMM 6600 User manual

NARDA SRM-3006 User manual

NARDA

NARDA SRM-3006 User manual

NARDA SRM-3000 User manual

NARDA

NARDA SRM-3000 User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

NARDA
Safety
Test
Solutions
S.r.l. Socio Unico
Sales & Support:
Via Rimini, 22
20142 - Milano (MI)
Tel.: +39 02 581881
Fax: +39 02 58188273
Manufacturing Plant:
Via Benessea, 29/B
17035 - Cisano sul Neva (SV)
Tel.: +39 0182 58641
Fax: +39 0182 586400
http://www.narda-sts.it
SERIAL NUMBER OF THE INSTRUMENT
You can find the Serial Number in a side of the box.
Serial Number is in the form : 0000X00000.
The first four digits and the letter are the Serial Number prefix, the last five digits are
the Serial Number suffix. The prefix is the same for identical instruments, it changes
only when a configuration change is made to the instrument.
The suffix is different for each instrument.
Document PL01EN-00708 – Copyright © NARDA 2020
User’s Manual
PMM PL-01
PULSE LIMITER
II Note and symbols
NOTE:
If the instrument is used in any other way than as described in this Users Manual, it may become unsafe
Before using this product, the related documentation must be read with great care and fully understood to
familiarize with all the safety prescriptions.
To ensure the correct use and the maximum safety level, the User shall know all the instructions and
recommendations contained in this document.
This product has a Pollution Degree II normally only non-conductive pollution occurs. Occasionally,
however, a temporary conductivity caused by condensation must be expected.
The information contained in this document is subject to change without notice.
KEY TO THE ELECTRIC AND SAFETY SYMBOLS:
You now own a high-quality instrument that will give you many years of reliable service.
Nevertheless, even this product will eventually become obsolete. When that time
comes, please remember that electronic equipment must be disposed of in accordance
with local regulations. This product conforms to the WEEE Directive of the European
Union (2002/96/EC) and belongs to Category 9 (Monitoring and Control Instruments).
You can return the instrument to us free of charge for proper environment friendly
disposal. You can obtain further information from your local NARDA Sales Partner or by
visiting our website at www.narda-sts.it .
Warning, danger of electric shock
Earth
Read carefully the Operating Manual and its
instructions, pay attention to the safety symbols.
Unit Earth Connection
Earth Protection
Equipotential
KEY TO THE SYMBOLS USED IN THIS DOCUMENT:
DANGER
The DANGER sign draws attention to a potential risk to a person’s
safety. All the precautions must be fully understood and applied before
proceeding.
WARNING
The WARNING sign draws attention to a potential risk of damage to the
apparatus or loss of data. All the precautions must be fully understood
and applied before proceeding.
CAUTION
The CAUTION sign draws attention against unsafe practices for the
apparatus functionality.
NOTE:
The NOTE draw attention to important information.
Contents III
Contents
Safety recommendations and instructions.……….......
EC Declaration of Conformity ……………………………
Page
V
VI
1 General Information
1.1 Documentation…….……………………………………
1.2 Introduction to PL01 Pulse Limiter……………...........
1.3 Insertion loss………………..............................……...
1.4 Environment………….................................…............
1.5 Return for service………….......................…...…........
2 Main Specification
Page
1-1
1-1
1-2
1-2
1-2
Page
2.1 Main Specification ………………………………………. 2-1
3 Preparation for use Page
3.1 Introduction….…………………………………………… 3-1
3.2 Packing unpacking……………...………………………. 3-1
3.3 Initial inspection…………….…………………………… 3-1
3.4 Preparation for use……………………………………... 3-2
3.5 Equipment cleaning …………………………………….. 3-2
IV Contents
Figures
Figure
1-1
PMM PL01 Typical frequency response……………….
Page
1-1
Tables
Table
2-1
Main Specification…………………….….………………
Page
2-1
Safety consideration V
SAFETY RECOMMENDATIONS AND INSTRUCTIONS
This product has been designed, produced and tested in Italy, and it left the factory in conditions fully
complying with the current safety standards. To maintain it in safe conditions and ensure correct use,
these general instructions must be fully understood and applied before the product is used.
•When the device must be connected permanently, first provide effective grounding;
•If the device must be connected to other equipment or accessories, make sure they are all safely
grounded;
•In case of devices permanently connected to the power supply, and lacking any fuses or other
devices of mains protection, the power line must be equipped with adequate protection
commensurate to the consumption of all the devices connected to it;
•In case of connection of the device to the power mains, make sure before connection that the voltage
selected on the voltage switch and the fuses are adequate for the voltage of the actual mains;
•Devices in Safety Class I, equipped with connection to the power mains by means of cord and plug,
can only be plugged into a socket equipped with a ground wire;
•Any interruption or loosening of the ground wire or of a connecting power cable, inside or outside the
device, will cause a potential risk for the safety of the personnel;
•Ground connections must not be interrupted intentionally;
•To prevent the possible danger of electrocution, do not remove any covers, panels or guards installed
on the device, and refer only to NARDA Service Centers if maintenance should be necessary;
•To maintain adequate protection from fire hazards, replace fuses only with others of the same type
and rating;
•Follow the safety regulations and any additional instructions in this manual to prevent accidents and
damages.
VI
EC Decl ara tio n of Conf or m i ty
In accordance with the Decision 768/2008/EC
Compliant to the Directives: EMC 2014/30/
EU, Low Voltage 2014/35/EU, RoHS 2011/65/EU
Also compliant to the ISO/IEC standards 17050-1 and 17050-2
The manufacturer,
NARDA Safety Test Solutions s.r.l.
via Benessea 29/B
17035 Cisano sul Neva (SV) – ITALY
based on the following harmonized European Standards, successfully applied:
Safety: EN 61010-1
(undated reference, applies to all editions)
EMC: EN 61326-1
(undated reference, applies to all editions)
declares, under its sole responsibility, that the product: PMM PL-01 Pulse Limiter
conforms with the essential requirements of the Low Voltage Directive 2014/35/EU, of the EMC Directive
EMC 2014/30/EU and of the RoHS directive 2011/65/EU.
Cisano sul Neva, 24/08/2016 Egon Stocca, General Manager
General Information
1-1
1 - General Information
1.1 Documentation
1.2 Introduction to
PL01 Pulse Limiter
Enclosed with this manual are a service questionnaire to send back to
NARDA in case that equipment service is needed.
The Pulse Limiter PL01 is an accessory designed to protect the input circuit
of a Receiver against high energy interfering voltage pulses.
The Pulse Limiter does not cause any significant change of the measured
signal and loss in sensitivity on the measuring receiver.
Its main function is to suppress the transients that can occur when the test
item, connected to an artificial mains network LISN, is switched on or off.
When operating the receiver from artificial mains network LISN, the Pulse
Limiter PL01 must be used particularly when very strong interfering signal is
suspected. The precision input attenuator and the input mixer of the
receiver are therefore protected from being damaged.
On the other hand if the disturbance signal and spikes do not exceed the
maximum input voltage permissible for the receiver, the use of the Pulse
Limiter can be omitted.
Designed according to criteria of cheaply and compactness it can be used
together with PMM 9010F for conducted interferences measurement, or
any other RF receiver.
The following figure shows the Pulse Limiter typical frequency response.
CH1
S
21
log MAG
2 dB/
REF 0 dB
START .010 000 MHz
STOP 100.000 000 MHz
Cor
PRm
MARKER 1
30 MHz
14 Jul 2004 11:22:24
1
1_:-.3274 dB
30.000 000 MHz
Fig. 1-1 Pulse Limiter typical frequency response
Document PL01EN-00708 - © NARDA 2020
1-2 General Information
1.3 Insertion loss
The insertion loss of the Pulse Limiter is calibrated in a 50 ohm system over
the frequency range of 9 kHz to 30 MHz, and is 0.4 dB maximum.
1.4 Environment The operating environment is specified to be within the following limitations:
•Temperature
•Humidity
0° to +45° C
< 90% relative
The instrument should be stored in a clean, dry environment
The storage and shipping environment is specified to be within the following
limitations :
•Temperature
•Humidity
-25° to + 70° C
< 95% relative
1.5 Return for service If the instrument should be returned to NARDA for service, please complete
the service questionnaire enclosed with the Operating Manual and attach it
to the instrument.
To minimize the repair time, be as specific as possible when describing the
failure. If the failure only occurs under certain conditions, explain how to
duplicate the failure.
If possible, reuse of the original packaging to ship the equipment is
preferable.
In case other package should be used ensure to wrap the instrument in
heavy paper or plastic.
Use a strong shipping container and use enough shock absorbing material
around all sides of the equipment to provide a firm cushion and prevent
movement in the container.
Seal the shipping container securely with shipment tape.
Mark the shipping container FRAGILE to encourage careful handling.
Main specifications
2-1
2 - Main specifications
2.1 Main specifications
Table 2-1 lists the PMM PL01 Pulse Limiter performance specifications.
The following conditions apply to all specifications :
•The ambient temperature must be 0° to 45°.
TABLE 2-1 Main specifications
Electrical characteristics Performance Limits
Insertion loss max:
Input impedance:
VSWR with 50 Ohm termination:
Frequency response:
Max permissible input power:
RF input:
RF output:
Rated temperature:
Storage temperature:
0.4 dB
50 Ohm
< 1.25
9 kHz ÷30 MHz
0,5 W
BNC female connector
BNC male connector
0 to + 45 C
- 25 to + 70 C
Document PL01EN-00708 - © NARDA 2020
2-2 Main specifications
This page has been intentionally left blank