Nissei DSK-1031 User manual

DSK-1031 INSTRUCTOIN MANUAL
DIGITAL BLOOD PRESSURE MONITOR DSK-1031
INSTRUKCJA OBSŁUGI
CIŚNIENIOMIERZA CYFROWEGO DSK-1031
РЪКОВОДСТВО ЗА ЕКСПЛОАТАЦИЯ
НА ЦИФРОВ ТОНОМЕТЪР DSK-1031
UŽIVATELSKÝ NÁVOD
K DIGITÁLNÍMU TONOMETRU DSK-1031
A DSK-1031 TÍPUSÚ DIGITÁLIS TONOMÉTER
HASZNÁLATI UTASÍTÁSA
MANUAL DE UTILIZARE
A TENSIOMETRULUI DIGITAL DSK-1031
ROU HUN CZE BGR POL ENG

2
ENG
GENERAL INFORMATION
INDICATIONS FOR USE
This Instruction Manual is designed to assist the user with safe and eective operation
of the digital blood pressure monitor DSK-1031 (hereinafter – the “Device”). The device
is designed to measure systolic and diastolic blood pressure and heart rate reading in
patients aged 12 years and older. This device should not be used for newborn children.
The incorrect measurement result is also possible when the device is operated in a baby
hand. Consult your doctor for blood pressure measurement in children. The device is rec-
ommended for use in patients with unstable (nonpermanent) blood pressure or hyper-
tension at home as a supplement to medical surveillance. The cu is suitable for a upper
arm with a circumference of about 22 - 42 cm. Blood pressure is measured in the range
from 40 to 250 mm Hg and heart rate, in the range from 40 to 160 heartbeats per minute.
OPERATION PRINCIPLES
The device uses the oscillometric method of measurement. The cu connected to
the electronic module is wrapped around the upper arm.
When you press the START/STOP button, the device starts the automatic ination,
which is followed by the blood pressure measurement. The sensor element of the
device detects slight pressure oscillations in the cu produced by the expansion
and contraction of the brachial artery in response to each heartbeat. Ination is
stopped when the cu is pumped enough to determine diastolic and systolic pres-
sure, whereupon the air is discharged from the cu. Rhythm and amplitude of the
pressure waves are measured and displayed on the LC display as a numerical value
in millimeters of mercury. The device has an arrhythmia indicator, as well as two
memories with 60 cells in each calculating the average value.
NISSEI New Technologies
Fuzzy Ination Algorithm – is an algorithm for automatic selection
of the cu ination pressure. Using this algorithm, the device by itself
determines the pressure level to which it is necessary to inate the cu
based on the patient’s systolic pressure. Owing to the Fuzzy Ination
algorithm the device becomes easier to use, while the measurement gets
more comfortable and more accurate.
Arrhythmia indicator is a special icon on the display that informs on the
irregular heartbeat, while the measurement result is correct.
PARTS AND COMPONENTS
1. CUFF
2. AC ADAPTOR
3. BATTERIES
4. ELECTRONIC UNIT
A. AIR PLUG
B. BATTERY COMPARTMENT
C. AC ADAPTOR JACK
D.
SET BUTTON (SETTINGS)
E. BUTTON (MEMORY 1)
F. BUTTON (MEMORY 2)
G. START/STOP BUTTON (START/STOP )
H. LCD DISPLAY
I. AIR CONNECTOR
K. AIR TUBE

3
ENG
an increase in pressure values. Therefore, blood pressure measured at home is often
dierent from that measured in the clinic. Since blood pressure increases at low tem-
peratures, measurements should be made at room temperature (about 20°C). If the
device was stored at low temperature, prior to using it should be kept for at least 1
hour at room temperature, otherwise the measurement result can be erroneous. Dur-
ing the day, the dierence in the readings in healthy people may attain 30-50 mm Hg
for systolic (upper) pressure and up to 10 mm Hg for diastolic (lower) pressure. De-
pendence of blood pressure on various factors is individual for each person. There-
fore it is recommended to keep a special recording of blood pressure readings. ONLY
A DOCTOR MAY ANALYZE TRENDS IN CHANGING YOUR BLOOD PRESSURE BASED ON
CORRESPONDING RECORDINGS.
4. In case of cardiovascular diseases and a number of other diseases that require
the blood pressure monitoring, measurements should be carried out in the hours
specied by a doctor. REMEMBER THAT THE DIAGNOSTICS AND ANY TREATMENT OF
ARTERIAL HYPERTENSION SHOULD BE CARRIED OUT ONLY BY A DOCTOR BASED ON
BLOOD PRESSURE READINGS OBTAINED BY A DOCTOR. MEDICAL DRUG ADMINIS-
TRATION OR CHANGE OF DOSAGES SHOULD BE MADE ONLY BY PRESCRIPTION OF
AN ATTENDING DOCTOR.
Blood pressure variations during a day
Systolic
Diastolic
Arterial pressure (mmHg)
Time of day
Fig.1
5. In case of disorders such as deep vascular sclerosis, weak pulse wave and break
in rhythm of heart contractions, the correct blood pressure measurement can be
complicated. IN THIS CASE, A DOCTOR SHALL PROVIDE RECOMMENDATIONS IN RE-
LATION TO USE OF THIS DEVICE.
6. KEEP QUIET DURING THE MEASUREMENT TO OBTAIN THE CORRECT BLOOD PRES-
SURE READING WHEN USING THE ELECTRONIC DEVICE. The blood pressure meas-
Touch control – is control of the device by slight touching it.
Detection of interference noises – indicator informs on the occurrence
of noises that can aect the measurement result.
Control of correct xing the cu – shows if the cu is fastened too tight
or too loose.
Indication of reliability – this symbol is displayed when all stages of
proper measurement procedure are observed.
Pulse pressure – along with the measurement result the device displays
the pulse pressure value. Pulse pressure is a dierence between systolic
and diastolic pressure.
WARNING! Using a cu that is dierent from that supplied with this device is not allowed.
COMPLETE SET
The complete set includes
- Electronic unit - 1 pc.
- Cu (including air hose and air hose plug) - 1 pc.
- Batteries - 4 pcs.
- Power supply source - 1 pc.
- Bag - 1 pc.
- Instruction Manual - 1 pc.
- Warranty - 1 pc.
- Packaging - 1 pc.
RECOMMENDATIONS ON CORRECT MEASUREMENTS
1. If treated with hemodialysis or anticoagulants, antiplatelets or steroids, refer to
your doctor about the blood pressure measurement.
2. Malfunctions are possible when the device is used near working mobile phones,
microwave ovens and other equipment generating electromagnetic radiation.
3. For correct measurement it is necessary to know that the BLOOD PRESSURE IS
SUBJECT TO SHARP FLUCTUATIONS EVEN IN SHORT TIME INTERVALS. The blood pres-
sure level depends on many factors. It is commonly lower in summer and higher in
winter. Blood pressure varies along with atmospheric pressure and depends on the
physical exertion, emotional excitability, stress and diet. Medical drugs, alcohol and
smoking exert great inuence as well. Occasionally, measurements in the clinic cause

4
ENG
urement should be carried out in a quiet comfortable atmosphere at room tem-
perature. Exclude meal an hour before the measurement, and exclude smoking, soft
drinks, and alcohol 1.5-2 hours before the measurement.
7. Accuracy of the blood pressure measurement depends on matching the device
cu and size of your hand. THE CUFF SHOULD NOT BE TOO SMALL OR TOO BIG.
8. Repeated measurements are carried out at 5-minute intervals to recover the
blood circulation. However, persons suering from severe atherosclerosis, due to a
signicant loss of elasticity of blood vessels, need longer intervals between meas-
urements (10-15 minutes).
This also concerns patients suering from long-term diabetes. For more accurate deter-
mination of blood pressure it is recommended to carry out a series of three consecutive
measurements and to calculate the average value of measurement results
.
POWER SUPPLY OF THE DEVICE
INSTALLATION OF BATTERIES
1.
Open the battery compart-
ment (Fig. 2).
2. Install four “AA” batteries in
the compartment.
Make sure that polarity corre-
sponds to signs (+) and (-) shown
inside the compartment (Fig. 3).
Batteries are readily installed
by pressing the end “-“ on the
spring.
Allowed to rechargeable batteries. To charge the batteries, use the special charger
(not included).
3. Close the battery compartment. Do not use excessive force when removing the cover.
Do not use excessive force when removing the cover.
Battery Replacement Indicator
Replace all the batteries when the battery replacement indicator is ashing on
the display during the measurement. If upon the device turning on the indicator
is steadily ashing, the measurement will not be possible until all the batteries are
replaced. The battery replacement indicator does not show a discharge degree.
Use alkaline batteries to increase the device operation duration. Ordinary zinc-carbon
batteries require more frequent replacement. The enclosed batteries are meant for
testing the sold device, and their operation period can be less than that of batteries
acquired in the trade network.
Since neither the device nor the batteries are the waste that can be utilized at home,
follow your national/local regulations for waste recycling and take them to corre-
sponding collection facilities.
USE OF THE DEVICE WITH THE POWER SOURCE
Socket for the power source is arranged on the left side of the device.To use the device
with the power source, connect the power source connector to the device, install the
power source plug into the socket outlet, and press the «START/STOP» button.
When nished, turn o the device by pressing the «START/STOP» button, unplug the
power source from the socket outlet and disconnect the power source connector
from the device.
NOTE!
If there is no battery in the device, turning o the power source will result in zeroing
of measurement results stored in the device memory and set date and time. If you do
not want to make the data erased, do not remove the batteries from the device when
using the power source.
SETTING DATE AND TIME
УСТАНОВКА ДАТЫ И ВРЕМЕНИ
Date and time can be set after installing batteries. Setting the date and time guarantees
the preservation of measurement results with indicated correct date and time. The de-
vice can be used without setting the date and time.
Press and hold the SET button until the display ashes the value of the year.
Fig.2 Fig.3

5
ENG
Date and time are set in the following order: year, month, day, hour and minute.
1. Setting the Year
Use the button to increase and button to decrease the year value. Press the SET
button to conrm and to pass to the next step.
2. Setting the Month
Use the button to increase and button to decrease the month value. Press the
SET button to conrm and to pass to the next step.
3. Setting the Day
Use the button to increase and button to decrease the day value. Press the SET
button to conrm and to pass to the next step.
4. Setting the Time
Watch uses a 12-hour time format of day. Use the button to increase and button
to decrease the hour or minute value. Press the SET button to conrm the settings.
To stop the setting, press the «START/STOP» button.
IMPORTANT! If the date and time are set, then when turned o the device display will
show the current time.
CORRECT POSITION DURING MEASUREMENT
Sit down at the table so that
during the blood pressure
measurement your hand is on
its surface. Make sure that the
place where the cu is put on
the upper arm is about the
same level as the heart and
the hand freely lies on the ta-
ble and does not move (Fig. 4).
You can also measure your
blood pressure when lying on your back. Look up, stay calm and do not move during
the measurement. Make sure that the place where the cu is put on the upper arm is
about the same level as the heart (Fig. 6).
Measured values may vary slightly, depending on the position during the
measurement. If the cu is above/below the level of the heart, resulting reading
may be incorrect (lower/higher)
CUFF PREPARATION
Fig.7
1. Apply the cu to your left upper arm so that the air
tube is directed to your palm
(Fig. 7)
. If the measurement
on your left arm is dicult, you may use your right arm. In
this case remember that the readings may dier by 5-10
mmHg and even more.
Fig.8
2.
Wrap the cu around your upper arm so that the bottom
of the cu is approximately 2-3 cm above your elbow. Air
tube should be directed towards the palm (Fig. 8).
Fig.9
3. Fix the cu so that it ts tightly to the arm, but see that
it is not overtight (Fig. 9). Too tight or too free placement
of the cu may give inaccurate readings.
Fig.10
4. If the hand is cone-shaped, it is recommended to put
the cu spirally, as shown in the gure (Fig. 10).
Fig.11
5. If the rolled-up sleeve squeezes the arm interfering
with free blood ow the Device may give inaccurate
gures not corresponding to your actual blood pressure
(Fig. 11).
Fig.4 Fig.5 Fig.6

6
ENG
MEASUREMENT PROCEDURE
IMPORTANT! The device has touch-sensitive buttons that
are pressed by slight touching. Moisture, dirt and extraneous
objects between the nger and the device panel can aect
the ability of buttons to respond to touching.
1. Insert the Air Tube Plug into the Cu Jack.
Before measurement, take a few breaths and relax. During the
measurement, do not talk and do not move.
2. Press the START/STOP button. The deating symbol « » will
ash on the display, and the device will release the remaining air
from the cu (Fig. 12).
3. Beep will be heard, and air will be pumped into the cu.
The symbol « » will ash, and the displayed value will
increase (Fig. 13). Ination will stop at optimum level owing
to the Fuzzy Ination algorithm.
4. The symbol « » will disappear, and the measurement will
start. The cu pressure will decrease slowly.
Noise Interference Detection
This device detects the presence of extraneous sounds or noises that can af-
fect the measurement results. When such interference is detected, the symbol « ».
appears on the display. In this case, it is recommended to repeat the measurement to
eliminate the interference.
Control of Correct Fixing the Cu
The symbol « » indicates that the cu is applied
correctly, the symbol « » indicates that the cu is fastened
too tight or too loose. In this case, it is recommended to repeat
the measurement after proper xing the cu.
Press the START/STOP button to stop forcedly the measure-
ment: the device will stop ination and quickly release the air.
Automatic Additional Ination.
If on starting the measurement, the operating pressure appears to be insucient due to
hand movements or due to excessive tension in a hand, the cu will be lled with air again
to the level of about 30 mm Hg higher than the original level. Automatic additional ination
will be repeated until the measurement is completed successfully. This is not a malfunction.
5. The symbol «» will start ashing synchro-
nously with a sound signal, as soon as the
sensor catches the pulse pressure (Fig. 14).
6. When the measurement is completed, the
display will show the values of blood pressure,
pulse pressure, WHO scale and heart rate (Fig.
15). The device will automatically release the
air from the cu.
7. Press the button or button and the result
will remain in the selected memory unit
.
8. Press the START/STOP button to turn o
the device.
If you forget to turn o the device, it will do so automatically after 3 minutes.
Indication of Reliability
Important aspects in measuring the blood pressure imply the correct
application of the cu and proper posture during the measurement. These aspects
may inuence the measurement result. The device displays the symbol « », when
the measurement procedure is carried out properly. If this symbol is not displayed,
it is recommended to repeat the measurement observing the guidelines for proper
measurement described in this manual.
Do not perform several measurements in a row.
This will cause numbing the hand and can aect the measurement result. Give your
hand a break for at least 5 minutes.
Pulse, stroke/min.
The Cuff Symbol
The Reliability Symbol
Systolic pressure,
mmHg
Diastolic pressure,
mmHg
Puls pressure, mmHg
Fig.15
Fig.14
Fig.12
Fig.13

7
ENG
Systolic blood pressure continues to increase
with aging, while diastolic blood pressure tends
to decrease since the age of about 50 years.
It is considered that a high pulse pressure is
related to poor elasticity of arteries and is
one of the risk factors for circulatory diseas-
es. Pulse pressure is considered to be normal
when it attains 35±10 mm Hg.
MEMORY FUNCTOIN
The measured values are automatically saved for later viewing in the selected
memory unit. These memory units can be used to save separately the measurement
results of two dierent persons or to save separately the results of morning and
evening measurements.
Each unit can keep up to 60 measurement results and the average value thereof.
When the number of kept values exceeds 60, the oldest readings are deleted to
record new values.
Saved values are kept with the measurement date and time, if set. If you want to
save time and date together with the measurement values, date and time should be
set before the measurement.
When an error occurs (ERR), the measurement results are not saved.
Viewing the Saved Data
1. Turn o the device by pressing the START/STOP
button. To view the results stored in the memory
unit 1, press , and to view the results stored in the
memory unit 2, press . The selected memory unit
will be underlined on the display.
The display shows the average value of stored
results indicated with index « » (Fig. 19). Average
value is displayed when the selected memory unit
contains two or more saved measurement results.
2. Each time you press the button or button,
saved measurement results will be displayed in a
sequential manner.
INDICATION OF ARRHYTHMIA
The ashing symbol « », appeared on the
display reports on an irregular heartbeat
rhythm (Fig. 16). If this symbol appears
from time to time, refer to your doctor. The
arrhythmia indicator occurrence can be also
caused by body movements in the course of
the measurement.
INDICATION OF READINGS BY THE WHO SCALE
In addition to pressure numerical values, the
result is also displayed as a graphic scale. This
scale makes it possible to classify the obtained
blood pressure value in accordance with the
World Health Organization recommendations.
The scale appears along with the numerical
value of blood pressure in the lower right cor-
ner of the display (Fig. 17).
Table of Results by the WHO Scale
Indication WHO Classication SYS DIA
Hypertension (severe) ≥180 ≥110
Hypertension (moderate) 160-179 100-109
Hypertension (slight) 140-159 90-99
Increased normal pressure 130-139 85-89
Normal pressure 120-129 80-84
Optimal <120 <80
Indication of Pulse Pressure
This device calculates and displays the pulse pressure value (Fig. 18). Pulse pressure – a
dierence between the systolic and diastolic pressures – tends to increase with aging.
N
orma
l
pu
l
se
Pulse with arrhythmia
Fig.16
WHO indicator
Fig.17
Selected memory
indicator Average
value
index
Number of saved measurements
Fig.19
Pulse pressure
mmH
g
Fig.18

8
ENG
3. Memory unit number, measurement date and time will be displayed one by one
in the upper right corner of the display.
4. The result saved in the memory unit 1 is the most recent among the saved data in
the selected memory. The higher memory number, the older the result.
The memory data are displayed for about 30 seconds and after that, if no button is
pressed, the device is automatically turned o.
Pressing the button switches the display from records in the memory unit 1 to
records in the memory unit 2, while pressing the button results in returning to
data recorded in the memory unit 1.
5. Press the START/STOP button to turn o the device.
Deletion of Saved Data
Readings can be deleted unit by unit: all together or separately. The memory may be
cleared when the selected memory unit contains two or more results.
1. Select a value from the memory unit to be deleted or the average value (unit with
index « ») to clear the whole memory unit.
2. Press and hold the button or until the sign «---» appears on the display.
INFORMATION ABOUT ERRORS
INDICATION LIKELY CAUSE METHODS OF CORRECTION
Blood pressure is too
low or too high. The cu is not at the heart level.
The cu is put on incorrectly.
During the measurement, a person
was talking or moving.
Put on the cu at the heart level.
Check the cu position on the hand.
Be calm and quiet during the
measurement.
Measurement results
are dierent each time.
Eect of measurement conditions,
physical or mental state. Take measurements under the same
conditions.
Measurement results
are dierent in clinic
and at home.
Eect of relaxed state at home and
tension in clinic. Show the pressure records made at
home to your doctor for advice.
Ination is repeated. Pressure can not be measured due to
movement or talking. Re-ination is not a malfunction of
the device. Do not talk and do not
move during the measurement.
Maximum allowable pressure:
pressure can not be measured due
to movement or talking during
the measurement, in spite of the
fact that the cu was pumped at
maximum.
Do not talk and do not move during
the measurement.
or
Pressure can not be measured due to
movement or talking. Do not talk and do not move during
the measurement.
Cu is not securely connected to
the device.
Cu is not put on properly.
Check the connection.
Make sure that the cu is put on
correctly.
No time indication on
the display.
Time was not set.
Note: there is no time indication when
batteries and/or power source are not
provided.
Set time and date.
Install batteries and connect the
power source.
Date and time are
displayed as «--/--». Time was not set or measurements
were carried out before setting the
time.
Set date and time.
Date and time can not be saved
without setting.
Batteries are discharged. Replace all batteries with new ones.
Nothing is shown on
the display. Batteries are discharged.
Batteries are installed incorrectly.
Connecting terminals are
contaminated.
Power source is not connected.
Insucient contact when touching
the button.
Replace all batteries with new ones.
Install batteries properly.
Wipe connecting terminals with a
dry cloth.
Connect the power source.
Ensure stronger pressing the button.

9
ENG
You touched the START/STOP button
when installing the batteries. Turn o the device by pressing the
START/STOP button and perform the
measurement again.
If, despite the above-given recommendations, you fail to obtain the right measurements, stop
the operation and contact the service center (addresses and telephone numbers of authorized
organizations are provided in the warranty certicate). Do not attempt to adjust the device
internal mechanism on your own.
WARRANTY
1. The manufacturer guarantees the warranty period of 5 years for the device from
the sale date provided that the consumer observes operation, transportation and
storage requirements. The warranty period for the cu and the power source is 12
months from the sale date.
2. Warranty liabilities are documented with the warranty certicate upon selling the
device to the buyer. The guarantee is valid provided that the device has not been
opened or damaged by the buyer.
3. Addresses of organizations engaged in the warranty service are specied in the
warranty certicate.
TECHNICAL SPECIFICATIONS
Operating Principle Oscillometric method
Indicator 15 digits liquid crystal display
Pressure Indicating Range , mmHg 3-300
Measuring Range :
cu pressure, mmHg
pulse rate, bpm 40-250
40-160
Accuracy:
cu pressure, mmHg
pulse rate, %
±3
±5
Ination Automatic ination (air pump, Fuzzy Ination algorithm)
Deation Automatic Deation (electric valve)
Electric supply voltage, V 6
Electric supply type Four 1.5 volt LR6 (AA alkaline) batteries or AC adaptor
Power Consumption , W 4
Memory 2 units, 60 values per each + average value
ADP-W5 Power Source
Output voltage, V
Maximum load current, A
Input voltage, V/Hz
6
0,5
100-240/50
Operating Condition
temperature, ˚C
humidity, % Rh
atmospheric pressure, hPa
+10 to +40
15 - 85
700-1060
Transportation / Storage Condition
temperature, ˚C
humidity, % Rh
-20 to +60
10 - 95
Cu Size adult (circumference of the upper arm 22-42 cm)
Overall dimensions:
Size (electronic unit), mm
Weight (without packaging,
bags, batteries and power
source), g.
115 x 115 х 65
406
Manufacture year: is specied on the device housing (in the battery
compartment) in the device serial number after the
letters «SN»
Protection level IP20: Protected against solid foreign particles with a
diameter of more than 12.5 mm, no protection against
water.
Key to Symbols Type BF applied part
Refer to instruction manual/booklet
Keep dry
When utilizing the waste, refer to current rules
applicable in your region

10
ENG
CARE, STORAGE, REPAIR AND DISPOSAL
1. This device should be protected from excessive moisture, direct sunlight, strokes,
and vibration. THE DEVICE IS NOT WATERPROOF.
2. Do not keep or do not use the device in close proximity to heaters and open ame.
3. If the Device was stored at a temperature below the freezing point, keep it at least
for 1 hour in some warm place before use.
4. If the device has not been used for a long time, remove the batteries. Leaking of
batteries can cause damage to the device and terminate the warranty. KEEP BAT-
TERIES AWAY FROM CHILDREN!
5. Do not contaminate the device and protect it from dust. The device can be
cleaned with a dry, soft cloth.
6. Do not allow the contact between the device and its parts with water, solvents,
alcohol, and gasoline.
7. Keep the cu away from sharp objects, and do not try to pull out the cu.
8. Do not expose the device to strong strokes and do not throw it.
9. The device does not contain any adjustment controls for settings. Unauthorized
opening of the electronic device is forbidden. If needed, repair the device only in
specialized organizations.
10.
On the expiry of the specied operation term, refer to specialists (specialized repair
organizations) on a periodic basis to check the technical condition of the device.
11. When utilizing the waste, refer to current rules applicable in your region. No spe-
cial utilization conditions are specied by the manufacturer for this device.
12 . The cu is resistant to repeated sanitation. The cu internal fabric surface (being
in contact with a patient’ hand) can be treated with a cotton swab moistened in a
3% solution of hydrogen peroxide. Partial discoloration of the cu covering tissue
is possible if used for a long time. Do not wash the cu and do not treat it with a
hot iron.
13. Do not leave unattended the device plugged into the network.
CERTIFICATION AND STATE REGISTRATION
The production of devices is certied pursuant to international standards such as
ISO 9001, ISO 13485, ISO 14001:2004.
The device meets international standards IEC60601-1:2005+CORR.1:2006+
CORR.2:2007, IEC60601-1-2:2001+A1:2004/IEC60601-1-2:2007 CISPR 11:2009+A1:2010.
Complaints and requests should be addressed to:
Little Doctor Europe Sp. z o.o.
57G Zawila Street, 30-390, Krakow, Poland
Service phone: +48 12 2684748, 2684749
Produced by Nihon Seimitsu Sokki Co., Ltd.
Address: 2508-13 Nakago Shibukawa Gunma 377-0293 Japan
Exporter: Little Doctor International (S) Pte. Ltd.
Address of Exporter: 35 Selegie Road #09-02 Parkline Shopping Centre, Singapore
188307

56
TECHNICAL DESCRIPTION
TECHNICAL DESCRIPTION
DSK-1031 complies with the EMC, electromagnetic compatibility, standard, IEC60601-1-2. Refer to the following tables for specic information regarding compliance to
the standard.
DSK-1031, as a medical electrical equipment, needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information
provided below.
• Portable and mobile RF communications equipments can aect the device.
• The use of accessories other than those specied in this manual may result in increased emissions or decreased immunity of the device.
• DSK-1031 should not be used adjacent to or stacked with other equipment.
Table 201. Guidance and manufacturer’s declaration - electromagnetic emissions
DSK-1031 is intended for use in the electromagnetic environment specied below. The customer or the user of DSK-1031 should assure that it is used in such
an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR11 Group 1 DSK-1031 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B DSK-1031 is suitable for use in all establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions IEC 61000-3-2 Class A
Voltage uctuations/icker
emissions IEC 61000-3-3
Complies

57
TECHNICAL DESCRIPTION
Table 202. Guidance and manufacturer’s declaration - electromagnetic immunity
DSK-1031 is intended for use in the electromagnetic environment specied below.The customer or the user of DSK-1031 should assure that it is used in such an environ-
ment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If oors are
covered with synthetic material, the relative humidity should
be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Mains power quality should be that of a typical commercial
or hospital environment.
Surge IEC 61000-4-5 ±1 kV dierential mode
±2 kV common mode
±1 kV dierential mode
±2 kV common mode
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines IEC 61000-4-11
<5% UT(>95% dip in UT) for 0,5
cycle
40% UT(60% dip in UT) for 5
cycles
70% UT(30% dip in UT) for 25
cycles
<5% UT(>95% dip in UT) for 5 sec
<5% UT(>95% dip in UT) for 0,5
cycle
40% UT(60% dip in UT) for 5
cycles
70% UT(30% dip in UT) for 25
cycles
<5%UT(>95% dip in UT) for 5
sec
Mains power quality should be that of a typical commercial
or hospital environment. If the user of DSK-1031 requires
continued operation during power mains interruptions,
it is recommended that DSK-1031 is be powered from an
uninterruptible power supply or a battery.
Power frequency (50/ 60 Hz)
magnetic eld IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
NOTE: UTis the a.c. mains voltage prior to application of the test level.

58
TECHNICAL DESCRIPTION
Table 204. Guidance and manufacturer’s declaration - electromagnetic immunity
DSK-1031 is intended for use in the electromagnetic environment specied below. The customer or the user of DSK-1031 should assure that it is used in such an envi-
ronment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of DSK-1031,
including cables, than the recommended separation distance calculated from the equation applicable to
the frequency of the transmitter. Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms, 150 kHz to
80 MHz
3 V d=[3.5/V1]√P
Radiated RF
IEC 61000-4-3
3 V/m, 80 MHz to
2.5 GHz
3 V/m d=[3.5/E1]√P, 80MHz-800MHz d=[7/E1]√P, 800MHz-2.5GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from xed RF transmitters, as determined by an electromagnetic site survey,1should be
less than the compliance level in each frequency range.2
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from structures, objects
and people.
1Field strength from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radios broad-
cast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an electromagnetic
site survey should be considered. If the measured eld strength in the location in which DSK-1031 is used exceeds the applicable RF compliance level above, DSK-
1031 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating
DSK-1031.
2
Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.

59
TECHNICAL DESCRIPTION
Table
206. Recommended separation distances between portable and mobile RF communications equipment and DSK-1031
DSK-1031 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of DSK-1031 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and DSK-1031 as recom-
mended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter, W
Separation distance according to frequency of transmitter, m
150 kHz to 80 MHz, d=[3.5/V1]√P 80 MHz to 800 MHz, d=[3.5/E1]√P 800 MHz to 2.5 GHz, d=[7/V1]√P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation ap-
plicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from
structures, objects and people.

E473/1602/04
Nissei Healthcare (UK) Ltd. Rede House, New Barn Lane Henfield West Sussex BN5 9SJ UK
Copyright Nihon Seimitsu Sokki Co., Ltd., 2014.
Registered Trade Mark.
NIHON SEIMITSU SOKKI CO., LTD.
2508-13 Nakago Shibukawa Gunma 377-0293 Japan
Other manuals for DSK-1031
3
Table of contents
Other Nissei Blood Pressure Monitor manuals

Nissei
Nissei WS-820 User manual

Nissei
Nissei WSK-1011 User manual

Nissei
Nissei DSK-1011J User manual

Nissei
Nissei DS-1873 User manual

Nissei
Nissei DS-10 User manual

Nissei
Nissei WS-1011 User manual

Nissei
Nissei DSK-1011 User manual

Nissei
Nissei WS-820 User manual

Nissei
Nissei DS-400 User manual

Nissei
Nissei WSK-1011 User manual

Nissei
Nissei ds-1902 User manual

Nissei
Nissei DSK-1031 User manual

Nissei
Nissei DS-1011 User manual

Nissei
Nissei DS-10 User manual

Nissei
Nissei DSK-1031 User manual

Nissei
Nissei DS-B10 User manual

Nissei
Nissei DS-137 User manual

Nissei
Nissei DS-10a User manual

Nissei
Nissei DS-S10 User manual

Nissei
Nissei DS-500 User manual