manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. OmniPod
  6. •
  7. Medical Equipment
  8. •
  9. OmniPod 5 User manual

OmniPod 5 User manual

Other manuals for 5

3

Other OmniPod Medical Equipment manuals

OmniPod 5 User manual

OmniPod

OmniPod 5 User manual

OmniPod ENT450 User manual

OmniPod

OmniPod ENT450 User manual

OmniPod DASH System User manual

OmniPod

OmniPod DASH System User manual

OmniPod 5 Manual

OmniPod

OmniPod 5 Manual

OmniPod PODDER Firmware update

OmniPod

OmniPod PODDER Firmware update

OmniPod PDM CAT45E User manual

OmniPod

OmniPod PDM CAT45E User manual

OmniPod PDM Series Firmware update

OmniPod

OmniPod PDM Series Firmware update

OmniPod DASH SYSTEM PODDER Firmware update

OmniPod

OmniPod DASH SYSTEM PODDER Firmware update

OmniPod DASH System User manual

OmniPod

OmniPod DASH System User manual

OmniPod Pod User manual

OmniPod

OmniPod Pod User manual

OmniPod 5 User manual

OmniPod

OmniPod 5 User manual

OmniPod DASH System User manual

OmniPod

OmniPod DASH System User manual

OmniPod PODDER Firmware update

OmniPod

OmniPod PODDER Firmware update

Popular Medical Equipment manuals by other brands

Itamar Medical WatchPAT Operation manual

Itamar Medical

Itamar Medical WatchPAT Operation manual

PROTEOR 1P650 KNEE Assembling and adjustment instructions

PROTEOR

PROTEOR 1P650 KNEE Assembling and adjustment instructions

Liberate Medical VentFree user manual

Liberate Medical

Liberate Medical VentFree user manual

Dicarre DA20 user guide

Dicarre

Dicarre DA20 user guide

Stryker Medical M Series Operation manual

Stryker Medical

Stryker Medical M Series Operation manual

Invacare ScanBed 750 user manual

Invacare

Invacare ScanBed 750 user manual

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Getinge

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics Sonicator 730 Maintenance manual

Mettler Electronics

Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care R1100 Mounting instruction

Pressalit Care

Pressalit Care R1100 Mounting instruction

Denas MS DENAS-T operating manual

Denas MS

Denas MS DENAS-T operating manual

bort medical ActiveColor quick guide

bort medical

bort medical ActiveColor quick guide

AccuVein AV400 user manual

AccuVein

AccuVein AV400 user manual

agiliti NP Trio user manual

agiliti

agiliti NP Trio user manual

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Rocket Medical

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Monteris Medical Neuroblate Instructions for use

Monteris Medical

Monteris Medical Neuroblate Instructions for use

Otto Block 8E33 7 manual

Otto Block

Otto Block 8E33 7 manual

ADVANCED HOME CARE Helios Guide

ADVANCED HOME CARE

ADVANCED HOME CARE Helios Guide

NeuLog NUL-208 Guide

NeuLog

NeuLog NUL-208 Guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Omnipod®5 Automated Insulin Delivery System
HCP Quick Glance Guide
CGM VALUES & TRENDS
How CGM Values
& Trends Affect
Automated Insulin
Delivery
Every 5 minutes, SmartAdjust™
technology receives a CGM value
and predicts where glucose will
be 60 minutes into the future.
It then automatically increases,
decreases or pauses insulin
delivery based upon the user’s
chosen Target Glucose value.
Adjustable Settings
AUTOMATED MODE
Target Glucose is the ONLY adjustable
setting that directly impacts automated
insulin delivery.
• Customizable and can include up to
8 time segments.
• Ranges from 110-150 mg/dL in
10 mg/dL increments.
• Used as the target for BOTH
automated insulin delivery AND
correction bolus dosing.
MANUAL MODE
Basal Programs for basal insulin delivery.
Adjusting basal rates ONLY impacts
Manual Mode basal insulin delivery.
EXAMPLE OF DAILY READINGS
TIME 12 AM 12 PM
AUTOMATED INSULIN DELIVERY
USER-INITIATED BOLUS
TARGET GLUCOSE 120 mg/dL
TARGET GLUCOSE 110 mg/dL
GLUCOSE VALUES
CUSTOMIZABLE TARGET GLUCOSE LEVEL
STANDARD GLUCOSE RANGE (70 - 180 MG/DL)
SMARTBOLUS CALCULATOR
AUTOMATED & MANUAL MODE
SmartBolus Calculator settings impact
suggested bolus doses in BOTH modes.
These include:
• Target Glucose & Correct Above
• Insulin to Carb Ratio
• Correction Factor
• Reverse Correction
• Duration of Insulin Action
• Minimum Glucose for Calculations
NOTE:
• Insulin to Carb Ratios can be
programmed in 0.1 g carb/U
increments
Omnipod 5 Automated Insulin Delivery System
HCP Quick Glance Guide
The Omnipod®5 Automated Insulin Delivery System is indicated for use by individuals with Type 1 diabetes mellitus in persons 2 years of age and older. The Omnipod 5 System is intended for single patient,
home use and requires a prescription. The Omnipod 5 System is compatible with the following U-100 insulins: NovoLog®, Humalog®, and Admelog®.
Refer to the Omnipod 5 Automated Insulin Delivery System User Guide at www.omnipod.com/safety for complete safety information including indications, contraindications, warnings, cautions, and instructions.
©2022 Insulet Corporation. Omnipod, the Omnipod 5 logo and SmartAdjust are trademarks or registered trademarks of Insulet Corporation in the United States of America and other various jurisdictions. All
rights reserved. All other trademarks are the property of their respective owners. The use of third party trademarks does not constitute an endorsement or imply a relationship or other aliation.
INS-OHS-11-2021-00012 V.2.0
HOW TO EDIT TARGET GLUCOSE
•Tap the Menu button
on the Home screen
•Tap Settings to
expand list
•Tap Bolus
•Tap on the segment
you want to edit. Edit
time, Target Glucose
and/or Correct Above.
•Tap NEXT to add/edit
more segments as
needed.
•Tap NEXT •Review changes and
tap SAVE.
•Tap Target Glucose &
Correct Above
HOW TO VIEW HISTORY & CGM GRAPH
•Tap the Menu button
on the Home screen
•Tap History Detail •Tap AUTO EVENTS to
see automated event
details.
•The SUMMARY section
includes CGM, Insulin,
and Carb details.
•Tap View on the
Home screen
CGM GRAPH
The CGM graph shows most recent CGM
readings as well as insulin delivery history.
This includes times that SmartAdjust
technology paused insulin, maximum basal
insulin was delivered, or a bolus was given. •Tap the
question mark
icon to view the
Graph legend
NOTE:
Follow these steps
to adjust settings for:
• Insulin to Carb
Ratio
• Correction Factor
• Reverse Correction
• Duration of Insulin
Action
• Minimum Glucose
for Calculations