Intended use
Medical Purpose
This product is intended to be used for inhaling medication for respiratory
disorders.
Intended User
• Legally certified medical experts, such as doctor, nurse and therapist.
• Caregiver or patient under the guidance of qualified medical experts for
home treatment.
• The user should also be capable of understanding general operation of
NE-C28P and the content of instruction manual.
Intended Patients
This product should not be used on patients, who are unconscious or are
not breathing spontaneously.
Environment
This product is intended for use in a medical facility, such as Hospital, clinic
and doctor’s office, and in a room of general household.
Durable period
Durable periods are as follows, provided the product is used to nebulize
saline 3 times a day for 10 minutes each time at room
temperature (23°C).
Durable period may vary depending on usage environment.
Frequent use of the product may shorten the durable period.
Compressor (Main unit) 5 years
Nebulizer Kit 1 year
Mouthpiece 1 year
Nosepiece 1 year
Air Tube (PVC, 200cm) 1 year
Air Filter 60 days
Masks (PVC)/(SEBS) 1 year
Precautions for use
Warnings and cautions described in the instruction manual should be
observed.
Know your unit
Important safety instructions
Read all the information in the instruction manual and any other literature included in
the box before using the device.
Warning:
Indicates a potentially hazardous situation which, if not avoided, could result in
serious injury.
(Usage)
• For type, dose, and regime of medication follow the instructions of your doctor or
respiratory therapist.
• If you experience any allergic reactions or other difficulties during use, stop using the
device immediately and consult your doctor.
• Do not use only water in the nebulizer for inhaling purposes.
• Keep the device out of the reach of unsupervised infants and children. The device
may contain small pieces that can be swallowed.
• Do not store the air tube while there is moisture or medication remaining inside.
• Do not use or store the device where it may be exposed to noxious fumes or volatile
substances.
• Do not use the device where it may be exposed to flammable gas.
• Do not cover the compressor with a blanket, towel, or any other type of cover during
use. The compressor could get hot and there is a possible risk of a burn if touched.
• Always dispose of any remaining medication after use, and use fresh medication each
time.
• Do not use in anaesthetic or ventilator breathing circuits.
• Do not block the ventilation slots. Never place the device where the ventilation slots
may be obstructed during operation. The compressor could get hot and there is a
possible risk of a burn if touched.
• Store the parts in a clean location to avoid infection.
(Risk of electrical shock)
• Never plug or unplug the device with wet hands.
• The compressor and power cord are not waterproof. Do not spill water or other liquids
on these parts. If liquid does spill on these parts, immediately unplug the
power cord
and wipe off the liquid with gauze or other absorbent material.
• Do not immerse the compressor in water or other liquid.
• Do not use or store the device in humid locations.
• Do not operate the device with a damaged power cord or plug.
• Keep the power cord away from heated surfaces.
(Cleaning and disinfecting)
Observe the rules below when cleaning or disinfecting parts. Failure to observe these
rules may result in damage, inefficient nebulization or infection. For the instructions,
refer to “Cleaning and disinfecting” section.
• Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before using
them when as follows:
- the first time after purchase
- if the device has not been used for a long period of time
- if more than one person uses the same device
• Be sure to wash or wipe the parts after use, and ensure that they are thoroughly
disinfected and dried, and stored in a clean location.
• Do not leave the cleaning solution to dry on the parts.
Caution: Indicates a potentially hazardous situation which if not avoided, may result in
minor or moderate injury, or physical damage.
(Usage)
• In order to prevent strangulation by the power cord or air tube, or injuries caused by
mishandling, please provide close supervision when this device is used by, on, or
near children or invalids.
• Make sure that the parts are attached correctly.
• Make sure that the air filter is correctly attached.
• Make sure that the air filter is clean. If the air filter has been used for more than
60 days, replace it with a new one.
• Do not tilt the nebulizer kit at an angle of greater than 45 degrees in all directions or
shake it while in use.
• Do not use or store the device while the air tube is creased.
• Use only original nebulizing parts, air tube, air filter and air filter cover.
• Do not add more than 7 ml of medication to the medication tank.
• Do not carry or leave the nebulizer kit while the medication tank contains medication.
• Do not leave the device unattended with infants or persons who cannot express their
consent.
• Do not subject the device or the parts to any strong shocks such as dropping on the
floor.
• Do not distort the vaporizer head, or jab the nozzle of the medication tank with a pin
or any sharp object.
• Do not insert fingers or objects inside the compressor.
• Do not disassemble or attempt to repair the compressor or power cord.
• Do not leave the device or its parts in extreme hot or cold temperature, or under direct
sunlight.
• Do not block the air filter cover.
• Do not use the device while sleeping or if drowsy.
• Approved for human use only.
• The compressor may become hot during operation.
• Do not touch the compressor for other than necessary operation such as turning off
the power while nebulizing.
• To avoid the medication residue on the face, be sure to wipe the face after removing
the mask.
• To avoid injury to the nose mucosa, do not squeeze the nosepiece into the back of
the nose.
• Do not block the slit between the cap and the inhalation air inlet.
• Make sure that the vaporizer head is correctly installed before use.
• Do not use a damaged nebulizer kit, mouthpiece or nosepiece.
• While using this device, make sure that no mobile phone or any other electrical
devices that emit electromagnetic fields is within 30cm. This may result in
degradation of performance of the device.
(Risk of electrical shock)
• Always unplug the device from the power outlet after use and before cleaning.
• Plug the device into the appropriate voltage outlet. Do not overload power outlets or
use extension cords.
• Do not misuse the power cord.
• Do not wind the power cord around the compressor.
• Changes or modifications not approved by OMRON HEALTHCARE will void the user
warranty.
• Do not pull the power cord strongly.
Optional Medical Accessories
Product description Model
V.V.T. Nebulizer Kit N-01 & Mouthpiece Set NEB-NSET3-81E
Adult Mask (PVC) NEB-MSLP-E
Child Mask (PVC) NEB-MSMP-E
Adult Mask (SEBS) NEB-MSLS-E
Child Mask (SEBS) Set NEB-MSSS-E
Mouthpiece NEB-MP-E
Nosepiece NEB-NP-E
Air Tube (PVC, 200cm) NEB-TP-L20E
Air Tube (Silicone, 100cm) NEB-TSC-L10E
Other Optional/Replacement Parts
Product description Model
Air Filters x 5 NEB-AFR-28E
Thank you for purchasing the NE-C28P.
This product was developed in conjunction with respiratory therapists
for the successful treatment of asthma, chronic bronchitis, allergies
and other respiratory disorders.
This is a medical device. Operate the device only as instructed by
your doctor and/or respiratory therapist.
How the nebulizer kit works
The medication that is pumped up
through the medication channel is
mixed with compressed air which is
generated by a compressor pump.
The compressed air mixed with medi-
cation is turned into fine particles and
sprayed when in contact with the baffle.
Nebulizer Kit with Virtual Valve
Technology
The high efficiency Virtual Valve
Technology (V.V.T.) reduces drug wastage while breathing out and
maximizes drug availability when breathing in.
Manufacturer OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, KYOTO,
617-0002 JAPAN
EU-representative OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp,
THE NETHERLANDS
www.omron-healthcare.com
Importer in EU
Production facility OMRON DALIAN Co., Ltd.
No. 3, Song Jiang Road,
Economic and Technical Development Zone,
Dalian 116600, China
OMRON Subsidiary OMRON HEALTHCARE UK LTD.
Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, UK
www.omron-healthcare.com
Made in China
(Cleaning and disinfecting)
Observe the rules below when cleaning or disinfecting parts. Failure to observe these
rules may result in damage, inefficient nebulization or infection. For the instructions,
refer to “Cleaning and disinfecting” section.
• Do not use a microwave oven, dish dryer or hair dryer to dry the device or the parts.
• Do not use an autoclave, EOG gas sterilizer or low temperature plasma sterilizer.
• When disinfecting parts by boiling, make sure that the container does not boil dry.
Otherwise it may also result in a fire.
General Safety Precautions:
• Inspect the device and parts before using them each time, and check that there are
no problems. In particular, be sure to check the following:
- That the vaporizer head, nozzle or air tube are not damaged.
- That the nozzle is not blocked.
- That the compressor operates normally.
• When using this device, there will be some noise and vibration caused by the pump in
the compressor. There will also be some noise caused by the emission of compressed
air from the nebulizer kit. This is normal and does not indicate a malfunction.
• Operate the device only as intended. Do not use the device for any other purpose.
• Do not use the device at temperatures greater than +40°C.
• Make sure that the air tube is securely attached to the compressor and nebulizer kit,
and does not come loose. Twist the air tube plug slightly when inserting it into the air
tube connectors to avoid the air tube disconnecting during use.
• To completely isolate the device from the power source unplug the plug from the
power source.
Keep these instructions for future reference.
Important information regarding Electromagnetic Compatibility (EMC)
NE-C28P-UK manufactured by OMRON HEALTHCARE Co., Ltd. conforms to
EN60601-1-2:2015 Electromagnetic Compatibility (EMC) standard. Further documentation
in accordance with this EMC standard is available at OMRON HEALTHCARE EUROPE
at the address mentioned in this instruction manual or at www.omron-healthcare.com.
Refer to the EMC information for NE-C28P-UK on the website.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it should not be
disposed of, with other household wastes at the end of its working life. To prevent
possible harm to the environment or human health from uncontrolled waste disposal,
please separate this product from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this product,
or their local government office, for details of where and how they can return this item
for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions of
the purchase contract. This product should not be mixed with other commercial waste
for disposal.
How to use
Warning:
• Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before using
them when as follows:
- the first time after purchase
- if the device has not been used for a long period of time
- if more than one person uses the same device
For the instructions, refer to “Cleaning and disinfecting” section.
1. Make sure that the power switch is in the off ( ) position.
2. Plug the power plug into a power outlet.
Note: Do not place the device in a location
where it is difficult to disconnect the
power cord.
3. Remove the cap and the inhalation top from
the medication tank.
4. Add the correct amount of prescribed
medication to the medication tank.
5. Make sure that the vaporizer head is securely
attached to the inhalation top.
6. Put the inhalation top back inside the
medication tank.
7. Attach the cap and mask, mouthpiece
or nosepiece to the nebulizer kit tightly.
Note:
It is not necessary to attach the cap when
nebulizing with Adult mask. If the amount of
aerosol is too great, attach the cap.
8. Attach the air tube.
Twist the air tube plug slightly and push it firmly into
the air tube connector.
9. Hold the nebulizer kit as
indicated on the right.
Follow the instructions of your
doctor or respiratory therapist.
Caution:
Do not tilt the nebulizer kit at an
angle of greater than 45 degrees
in all directions. Medication may
flow into the mouth or it may result
in not nebulizing effectively.
10.Press the power switch to the on ( ) position. As the compressor
starts, nebulization begins and aerosol is generated.
Inhale the medication.
11. When treatment is complete, turn the power off and unplug the
compressor from the power outlet.
TM
1
2
2
1
2
1
2
EN
Compressor Nebulizer
NE-C28P (NE-C28P-UK)
Instruction Manual
Baffle
Aerosol
Medication
Channel
Medication
Aerosol
Nozzle
Medication
Compressed Air
IM-NE-C28P-UK-EN-04-03/2020
1679307-2D
Issue Date: 2020-04-08
Rear View
Power Cord
Ventilation Slots
Power Plug
Ventilation Slots
Power Switch
Air Filter Cover
Air Tube Connector
Nebulizer Kit Holder
Cap
Vaporizer Head
Compressor (Main Unit) x 1
Nebulizer Kit x 1 Nosepiece x 1Mouthpiece x 1
Adult Mask (PVC) x 1 Child Mask (PVC) x 1
Air Tube
(PVC, 200 cm) x 1
Carrying Bag x 1
Instruction Manual
Warranty card
Spare Air Filters x 5
Inhalation Air Inlet
Inhalation Top
Nozzle
Medication Tank
Air Tube Connector
Right angle