version 1.1 rev 07 July 2022
6
Safety and Warning Information
PLEASE READ THIS INFORMATION FIRST BEFORE USING YOUR PRODUCT
CAUTION – USE OF CONTROLS OR ADJUSTMENTS OR PERFORMANCE OF PROCEDURES
OTHER THAN THOSE SPECIFIED HEREIN MAY RESULT IN HAZARDOUS RADIATION EXPO-
SURE.
1. IftheequipmentisusedinamannernotspeciedbyAndor,theprotectionprovidedbytheequipment
may be impaired.
2. WARNING: The BCU must be switched on when the HLE is emitting, otherwise serious damage will
occur to the BCU due to the power of the HLE laser radiation. It is important that the BCU interlock
connector is wired into the interlock system correctly as this will stop the HLE emitting to protect the
BCU.
3. Donotpositionthisproductsothatitisdiculttooperatethemainsdisconnectingdevice.SeeSECTION
4.1, “Emergency Mains Disconnection”.
4. Before using the product, please follow and adhere to all warnings, and safety, manual handling, and
operating instructions located either on the product, or in this manual.
5. IfusedinaDragonysystem,refertotheDragonyhardwareguide.
6. Keep this manual in a safe place for future reference.
7. Users must be authorised and trained personnel only; otherwise, this may result in personal injury, and/
or equipment damage and impaired system performance.
8. There are no user-serviceable parts inside the product and the enclosure must not be opened. Only
authorised service personnel may service this equipment.
9. IECTechnicalDocumentIECTR60825-14recommendsthepresenceofaLaserSafetyOcer(LSO);
however, national guidelines should be referred to.
10. Do not attempt to bypass any safety interlocks. They are provided to comply with the safety
requirements of various regulatory agencies and must be employed to protect the operator.
11. Protective earth is an integral part of the protection against electric shock in this product and is provided
via the earth pin of the external power supply. Ensure that this is plugged into the building earth system
via the mains socket. Do not tamper with any of the earthing measures.
12. AnyAC/DCPowerSupplyusedwiththisproductmustmeettherequirementsspeciedin“HLE-FL1
External AC/DC Power Supply Requirements” on page 48 and 49.
13. Nopartsshouldbereplacedbythecustomer,exceptforthemainscables,whichmustbeofthesame
typeandratingasthatsupplied(andasspeciedin“ElectricalPowerSpecications”onpage47)
andcertiedinaccordancewithyourregion’ssafetyregulations.
14. Make sure the power supply cord is located so that it will not be subject to damage.
15. While running an experiment, keep room temperature as stable as possible.
16. Performanceofthesystemmaybeadverselyaectedbyrapidlychangingenvironmentalconditionsor
operationoutsideoftheoperatingconditionsspeciedin“AppendixA:TechnicalSpecications”
17. Medical Diagnosis: This equipment has not been designed and manufactured for the medical diagnosis
of patients.
18. Electromagnetic Compatibility (EMC) – Caution: This product was designed for and tested using the
IEC/EN61326-1EMCstandardforClassAemissionsandaBasicimmunityenvironment.ClassA
means that it is not designed for a domestic or residential environment, and Basic immunity refers to
the fact that it is not designed for a typical industrial environment. This equipment is not intended for
use in residential environments and may not provide adequate protection to radio reception in such
environments.