Set up Solis on the Kymera 193i
MAN-09416
SR193QSG 1021 R1
1. The 193i should be connected to the camera by either:
• I2C cable (supplied, if applicable), or
• USB cable (provided).
2. If your spectrograph has a shutter, then this is controlled through the I2C cable (if used for the
connection), or through the BNC connection via a cable from the camera (supplied), or for simple open/
close operation no connection is required.
3. The 193i accessories - motorised slits, lter wheel, shutter - are all ‘plug and play’ and do not require
any setup.
Connect the Camera and Spectrograph
Refer also to the Kymera 193i user guide Conguring Solis for the Kymera 193i
1. Power on the Kymera 193i and if appropriate
the detector, wait until the short start-up tone
signals it has nished system checks.
2. Start Solis.
3. Solis will auto detect the Kymera and show a
dialog on startup alerting the user of this.
4. Once fully loaded the Spectrograph
control UI will show with the auto detected
peripheral controls.
5. Please refer to your camera and Kymera
193i user manuals for any further setup or
calibration steps required before using your
system.
Connect the Camera to the PC
5
Troubleshoot
Spectrograph is not recognised by Solis
• Ensure that the Kymera 193i has nished its system checks before attempting to connect with it.
Check the connections between the camera, PC and spectrograph.
Camera is not recognised by PC
• Check that the camera power and USB cable are connected securely and the mains power cable is inserted.
• Check that the drivers have been installed properly
(Control Panel /System/Hardware /Device Manager /Imaging Devices).
Shutter is not opening
• The shutter should always be controllable open/closed in the Solis GUI (when shutter is present). When the I2C
cable or the BNC input is used, the ‘BNC’ option should be selected in the software to enable operation of all
shutter modes.
• Check opening/closing times have been properly setup.
Turret lid dicult to open
• Rotate the lid lock from the locked to the unlocked position.
• The turret lid can then be lifted up using the handle of the turret lid.
Communication error when selecting side port
• When shipping, a locking screw is used to secure the ipper mirror in dual exit systems (KYMERA-193i-Bx). This
needs to be loosened fully or removed before use.
Please see the Hardware Guide for further information on Troubleshooting.
Online access to technical documentation is available at: andor.oxinst.com/downloads
?
6
I2CUSB
Shutter (BNC)
Accessory
Connector
(e.g. lter
wheel)
Connection Panel of the Kymera 193i
Note: If your 193i is a dual exit system (model KYMERA-193i-Bx), then you must loosen fully or remove the
locking grub-screw used during shipping to hold the ipper mirror in place.
Power
Input
Power Switch
Flipper
mirror
locking
screw
Above Accessories are automatically detected and
displayed in the Solis software.
Connect the Camera to the PC
Depending on the camera model supplied there are three types of possible connectivity between the camera and
the PC:
• USB 2.0 or 3.0
• USB 2.0 or 3.0 and I2C
• Cameralink
Note for Newton, iDus, iVac and iStar models the I2C connection is strongly recommended.
Please refer to the Quick Start Guide provided with the Andor camera for full installation directions applicable to
the specic camera model.
Note that if camera is ordered for the second output port, it will arrive detached and attach-
ment procedure as per user manual must be followed.
Ensure that the camera AND the PC are switched OFF.
A pre-aligned camera and spectrograph package comes pre-tted with a ange. For a standalone spectrograph,
the unit is supplied with a camera ange, the camera must be attached to the ange and the ange attached to
the spectrograph body with the screws supplied. Please see the spectrograph user guide for mounting information.
4