
3-5
In most cases the user will be interested in the dominating component of the CUT, displayed in the
RCL AUTO mode, if any other parameter shall be displayed the user may select it from the front
panel menu by activating the stepping key Vor A.
The RCL meter primarily determines the series reactance and resistance of the CUT. From these two
quantities the selected CUT parameter is calculated. The algorithm used by the instrument including
series/parallel and parallel/series transformation formulas and phasor diagrams of the various CUT
types are presented in Appendix 1.
3.4.5. Special user instructions
As pointed out in the preceding chapter in RCL AUTO mode the instrument identifies the dominant
component of the CUT and display it. It must be considered that the decision, if the reactive or the
ohmic component is dominating, generally depends on the frequency. In PM 6303 an 1kHz test fre-
quency is applied. This must be taken into account especially if low-ohmic inductors and capacitors
or high-ohmic resistors are measured:
Lossy inductors: When testing small lossy inductances often the series loss resistance is identified as
dominant component and displayed, because at 1kHz the series reactance will be very low. Hence,
for Ls or Lp display this parameter must be selected from the front-panel menu.
Lossy capacitors with high capacitance, e.g. electrolytic capacitors:
When testing capacitors the user normally will be interested in the value of the capacitance. As the
reactance of large capacitors is very low, the series resistance can be dominant resulting in Q<1and
indication of Rp. Hence, for Cs or Cp display these parameters must be selected..
Small capacitances measured with the SMD test fixture PM 9542 SMD:
When testing very small capacitances up to some pF with PM 9542 SMD, the alteration of the zero -
capacitance depending on the contact opening must be taken into account, see Fig. 36a. For accurate
measurements in that range the adjustment of the zero -capacitance should be done at contact opening
corresponding to the length of the component to be measured; for all normal measurements the adjust-
ment should be done at max. contact opening.
High-ohmic resistors: When testing resistors in the higher Mf2 range the reactance of the parasitic parallel
capacitance may be lower than the resistance, resulting in aCp display. For indication of Rs or Rp
these parameters must be selected.
Additional user instructions:
In the Cs (2 VBIAS) mode capacitors can be tested with 2Vdc bias voltage.
For large capacitors some time is needed for stable display due to the charging process (approx.
0.5 s/mF).
For the parameter Cs (2 VBIAS) overrange is indicated for Q<0.1 or if an inductance is identified.
The resonant frequency of alarger inductance paralleled by aparasitic capacitance can be below the
test frequency. Then, of course, the CUT represents acapacitance at 1kHz which is displayed.
When testing large inductors especially in the kH range relative small parasitic parallel capacitances will
effect the measurement result. Thus special attention shall be paid on careful Co compensation.
When testing inductors with ferromagnetic cores normally due to saturation effects the inductance
will decrease with higher current or voltage amplitudes. At PM 6303 these amplitudes are resulting
from the 2Vrms open-circuit voltage and the internal 400 fi resistance of the instrument and the
CUT impedance. For lower amplitudes an additional resistor 3* 71.5 Q. may be connected between
GUARD and the centre 4mm socket (marked with a-sign). For Rp =71.5 ft fig. 6shows the CUT
voltage and current versus impedance relationship. In the shown impedance range the measurement
error is increased to about 0.5 %maximum by the load resistor.