Calibration TK100C
The output voltage of the TK100C can be adjusted.
On the bottom side of the device is a trimmer potentiometer located under the identification label.
It is marked by the "DC Adjust" symbol.
Please note, that any previous calibration is invalidated by adjustments you make to this trimmer.
Safety Precautions & Override
The TK100C has a safety circuit inside, wich enables output voltage only if a BNC Adapter is connected.
You can override this safety feature at your own risk following these steps:
1. Disconnect the TK100C from the 15 V DC supply voltage.
2. Wait for at least 20 seconds.
3. Hold down the Mode Button (5) while reconnecting the 15 V DC supply voltage.
The safety circuit is now circumvented.
You can reenable the safety circuit by repeating the steps above, once again.
Specifications
Squarewave output voltage : +100 V ± 0.5%
Risetime : < 6 ns
Overshoot : < 3%
Repetition Rate : 100 Hz
Pulse Width: 5 ms
Dimensions: 135 x 30 x 70 mm
Weight: 120 g
Environmental Specifications
Altitude operating up to 2000 m
non-operating up to 15000 m
Temperature Range operating 0° C to +50° C
non-operating -40° C to +71° C
Maximum Relative Humidity operating 80 % relative humidity for temperatures up to +31° C,
decreasing linearly to 40 % at +50° C