520 Analog Output Card Installation Instructions
Analog Output Calibration
The following calibration procedure requires a
multimeter to measure voltage or current output from
the analog output module.
NOTE:
The analog output must be calibrated
after
the
indicator itself has been configured and calibrated.
1. Enter setup mode and go to the ALGOUT
menu (see Figure 4):
•Set OFFSET to 0% for 0-10 V output, 20%
for 4-20 mA output
•Set mode select jumper on analog output
board to “V” for 0-10 V output, or “I” for
4-20 mA output
•Set MIN to lowest weight value to be
tracked by the analog output
•Set MAX to highest weight value to be
tracked by the analog output
2. Connect multimeter to connector J1 on the
analog output board:
•For voltage output, connect voltmeter
leads to pins 3 and 4
•For current output, connect ammeter leads
to pins 1 and 2
3. Adjust zero calibration: Scroll to the TWZERO
parameter. Press to view zero value, then
check voltage or current reading on
multimeter. Press and hold or to adjust the
zero value up or down.
4. Adjust span calibration: Scroll to the TWSPAN
parameter. Press to view span value, then
check voltage or current reading on
multimeter. Press and hold or to adjust the
span value up or down.
5. Final zero calibration: Return to the TWZERO
parameter and verify that the zero calibration
has not drifted. Press and hold or to
re-adjust the zero value as required.
6. Return to normal mode. Analog output
function can be verified using test weights.
See the
520
Installation Manual
, PN 68973, for more
configuration information.
Figure 4. Analog Output Menu
Specifications
Resolution 16-bit, monotonicity over temperature
Linearity ±0.03% of full scale input
Current Output 0–20 mA or 4–20 mA (20% offset)
Maximum Load Resistance 840
Ω
Power Consumption 3.9W (max. load @ 20 mA)
Voltage Output 0–10 VDC
Minimum Load Resistance 1.2 K
Ω
Power Consumption 3.9W (max. load @ 10 VDC)
Input Protection Short circuit protection, 300W transient voltage suppression
Protection for ESD, EFT (electrical fast transients), tertiary lightning, and system-generated transients per
IEC 60001-4-2, 60001-4-4, and 60001-4-5; European Standards EN50082 and EN61000-4