scigiene RD0370 User manual

Digital RTD Thermometer
Part# RD0370
Instruction Manual

Table of Contents
Features......................................................................................
3
S
p
ecifications
....................................................................................
3-4
Instrument
Description
.....................................................................
5-6
Operating Instructions
......................................................................
7-9
Temperature
Measurement.............................................................
10
Battery Replacement.........................................................................11
Calibration Procedure..........................................................................12
Temperature Meter Calibration Setup.............................................12-14
Product
Maintenance.........................................................................15
Temperature
Probe.............................................................................15
2

Features
This digital RTD
thermometer
is to be
used
with
platinum-type
temperature sensors.
The
temperature indication
follows the IEC751
temperature
table for
PT-type
sensors.
•
IP67
waterproof housing
•
Platinum
resistor temperature sensor requires
less
temperature
compensation
than
thermocouple units
•
4-digit LCD
readout
with LED
backlight
•
°C/°F
switchable
•
Min/Max/Avg
functions
•
Relative
mode
and Data
hold
•
Low
battery indication
with auto power off
•
Includes
piercing type
temperature probe
(Pt-100, IEC751
Class
A
accuracy)
•
Ideal for food
service applications
Specifications:
Measurement Range:
-100°C ~ 300°C; -148°F ~
572°F
Resolution:
0.1°C;
0.1°F;
Sensor types:
Platinum
resistance temperature sensor
for
pt-100, pt-500, pt-1000 (selectable)
4
wir
es
.
ALPHA=0.003850
Measurement current:
Approx
0.53mA
Maximum Voltage at
Thermocouple
input: 60V DC, or 24Vrms AC
3

Envir
onmental:
•
Operating Temperature
and Humidity:
0°C ~50°C (32°F - 122°F); 0 - 80% RH
•
Storage Temperature
and Humidity:
-10°C to 60°C (14°F - 140°F); 0 - 80% RH
•
Altitude up to 2000
meters,
Accuracy
(at 23 ± 5°C):
Range
Accuracy
-100°C ~
300°C
±(0.1%
reading
+
0.4°C)
-148°F ~
572°F
±(0.1%
reading
+
0.8°F)
Temperature Coefficient:
For
ambient temperatures
from 0°C to 18°C and 28°C to 50°C, for
each
degree
celcium
ambient
below 18°C or
above
28°C add the following
tolerance
into the
accuracy specification.
0.01% of reading + 0.03°C ( 0.01% of reading + 0.06°F )
The
basic accuracy Specification does
not include the error of
the
probe,
refer to the
probe accuracy specifications
(on
page
16) for
additional details.
Sample
Rate: 2 times per
second
Water
resistance:
IP67
Power Supply: Battery 1.5V x 3 size AAA
Battery Life: Approx. 100 hours with alkaline
batteries
Dimensions:
150 × 66 ×
31mm
Weight: 175g
Appr
ox.
Accessories:
Pt-100 Probe
(class
A)
(model TP-R01),
batteries
and
instruction manual
Optional Accessories
:
Pt-100 Probe
(class A
)
(model
TP-R01),
soft carrying
case
(model CA-05A)
4

Instrument Description
1 — Pt type
temperature sensor connector
2 — LCD
display
3 — Power ON/OFF
button
4 — HOLD
button
5 — MAX/MIN/Average control
button
6 — Back light
button
7 — Relative
readout button
5

Display
Symbol definitions
and b
utton locations
-
This
indicates
that the minus
temperature
is
sensed
°C/°F
Celsius and
Fahrenheit indication
Pt xxx Platinum Type
Indication
HOLD
This
indicates
that the display data is being
held
MAX The Maximum value is being
displayed
MIN
The Minimum value is being
displayed
AVG
The Average value is being
displayed
REL The
reading
is under Relative
Mode
The
battery
power level
indication
This
indicates
Auto Power Off is
enabled
6

..
Operating Instructions
Turning the meter on
Press
the
button
to turn the
thermometer
ON or
OFF
.
Connecting the temperature probe
For
measurement,
plug the
temperature probe
into the input
connectors.
Selecting the Temperature Scale
When the meter is first
turned
on, the default scale
setting
is set to
the
Celsius (°C)
scale.
It can be
changed
to
Fahrenheit
(°F)
by
pr
essing
“
°C/°F
” button
and vice versa back to
Celsius.
Data-Hold Operation
To hold the
present reading
and keep it on the display, press
the
“
HOLD
”
.
b
utton.
When the hold data is no longer needed
, release
the
data-hold
operation
by
pressing
“HOLD”
button again.
When the meter is under Data Hold
operation,
the “ REL”,
and “°C/°F”
buttons
are
disabled.
Back light Operation
Press
the “Back Light”
button
to turn the back light on,
press i
t
once
again to turn it
of
f.
The meter will
turn the
back light off
automatically
if the “Back
Light
”
button
is not
pressed
after 10
seconds
7
AVG
MAX
MIN

AVG
MAX MIN
Relative Operation
When the “ REL”
button
is
pressed,
the meter will
memorize
the
pr
esent
reading
and the
difference between
the new
reading
and the
memorized
reading
and the result will
be
shown
on the display.
Press
the “
REL
”
button
again to exit the relative
operation.
MAX/MIN/AVG Operation
When
the
AVG
MAX
MIN
button
is
pressed,
the meter will
enter the
MAX/MIN/AVG
mode.
Under this
mode
the
maximum
value,
minimum
value and average
value
will
appear simultaneously.
The
average
value
is
calculated
using the last 8
readings
from the
memory,
it
will
update with
every new r
eading.
When the
MAX
symbol is
displayed,
the Maximum value is
shown
on
the
display.
Press AVG
MAX
MIN again and the NIN symbol is on the display and also
the
minimum r
eading.
Press AVG
MAX
MIN again and the
AVG
symbol is on the display as well as
the
average reading.
Press AVG
MAX
MIN again and the
MAX,
MIN and AVG
will
blink
together. This
means
that all
these readings
are
updated
in the memory and the
r
eading
is the
current
temperatur
e.
Press AVG
MAX
MIN to
circulate
the display mode
among these options.
When the meter is
under
functions
are
disabled.
AVG
MAX
MIN
operation,
the “ REL” and “
°C/°F
”
To exit the
MAX/MIN
mode,
press
and hold
8
for two seconds.

Auto Power Of
f
By
default,
the meter is in auto power off
mode.
The meter will
power itself
off after 30
minutes
without
operation.
To
disable
the auto power off,
press
and hold the “HOLD”
button
and turn
the meter on. There
will
be two
successive beeps
to
indicate
that the
auto
power off has been
disabled.
Low Battery
This meter indicates
the
battery
power level directly on the display.
Use
this
reference
guide to
establish battery
power level
.
Battery power level is FULL, you can still
take measurements
.
Battery power lever is LOW, the
battery
will
need
to be
replaced,
you can still
take
measur
ements.
Battery power lever is EMPTY, the
batteries
need
to
replaced
(size
AAA,
1.5V x 3).
Y
ou
can no longer take measurements
.
For
complete instructions
on how to
replace
the
batteries
see
page
11.
9

Temperature Measurement
Correct Measurement
Method
The
temperature sensor
is
located
at
the
end of the metal
sheath
of the
sheath
type temperature probe.
To
accurately
test internal
temperatures,
insert the
probe
to
a
distance
of at least 15 times the
diameter
of the
sheath
directly into the
item
you
want
to
measur
e.
Connector Configuration
Source-
Sense +
Source +
Sense -
NC
10

Battery Replacement
1)
Remove
all of the back cover screws
to be able to
remove
the
cover.
2) Verify the polarity of the
batteries
and install the new LR03
(AAA
size)
Alkaline
batteries
into the
instrument.
3)
Replace
cover properly onto the meter and tighten screws
.
The
unit’s
back cover is fitted with
rubber
rings. After
replacing
the
batteries, check
that the
rubber
rings
ar
e
properly
placed
before reinstalling the back cover.
Impr
oper
placement
of the
rubber
rings
will
compromise
the unit’
s
water-resistant structure,
and
possibly
result in
damage to
the
meter
.
.
11

Calibration Procedure
Step 1) Make sure you have a certified reference thermometer that is at least
twice the accuracy of the thermometer to be recalibrated. Eg. RD0370R
https://www.scigiene.com/Scigiene-Reference-Thermometer-NIST-Traceable
Step 2) Insert temperature probe into https://www.scigiene.com/SCCAL-2-
Calibrator or https://www.scigiene.com/SCCAL-9-Calibrator or
https://www.scigiene.com/TempCalibrator140 at 0°C in order to calibrate
precisely, and follow the steps in the calibration setup instructions. NOTE regular
ice bath methods are not accurate enough to recalibrate these devices in order to
meet stated specifications.
Temperature Meter Calibration Setup
1. Below table diagram shows each button function when user enter into
calibration mode.
Note:
Turn the Power “OFF” before attempting following SETUP.
Setup mode is cancelled during below procedure if “POWER” button is pressed.
12

Press and hold “Power” + “Light” buttons for 2
seconds
to enter setup mode
Press
“Hold” button to enter PT selection mode
Press “Max” button to select between PT probe type
(Pt100, Pt500, Pt1000 is circulate)
Press “°C/°F” button to confirm selection
Press and hold “Power” + “Light” buttons for
2 seconds to enter setup mode
Press “REL” button to enter calibration
mode.
Enter PT selection first.
Press “°C/°F” button to confirm selection
Press “REL” button to enter calibration mode
2. Pt type selection
3. O°C Calibration
13

Press “REL” button to confirm present calibration
value.
Press “°C/°F” button to confirm
PT selection
Press and hold “Power” + “Light” buttons for 2
seconds to enter setup mode.
Press “REL” button to enter calibration
mode.
Enter PT selection first.
Press “°C/°F” button to confirm PT selection
Press “Light” button to enter recall mode
Press “°C/°F” button to confirm revert back to
“Default factory setting value”
Press “°C/°F” button to confirm selection
4. Recall Default factory setting value
14

Product Maintenance
In order to
ensure
the
accuracy
of the
thermometer
for a long period
of
time it
should
be
calibrated once
a year. For
service
(repairs or
calibration)
or additional information
on this or any other
products
, contact Scigiene
at 416-261-4865 or visit our website at: www.scigiene.com.
Clean the
device
and the window of the display with a
clean, lint-fr
ee,
antistatic
and dry
cleaning cloth.
Do no use cleaning
agents
that contain carbon or
benzenes,
alcohol or anything similar to clean the meter as
these substances
will
damage
the surface of the
meter.
Do not use tools with sharp edges
, screwdrivers,
metal
brushes or
anything similar to clean the
meter.
Temperature Probe
Piercing type
temperature probe
Piercing type temperature probe Specification
Sensor Type Platinum
resistance thermometer
sensor Pt 100 (4 wires)
Accuracy IEC751, class A
±0.15°C ±0.002
measurement temperatur
e
Measurement
Range -100 to 400°C
Temp. Sensor Dim. Approx. 3.2mm
(
0.125”)
Temp. Sensor Length Approx. 120 mm (4.72”)
Cable Length Approx. 1100 mm (43.3”)
Water-resistant
EN60529:1991, IP67
15
Table of contents
Other scigiene Thermometer manuals

scigiene
scigiene RD0375 User manual

scigiene
scigiene RD0370R User manual

scigiene
scigiene RD0375 User manual

scigiene
scigiene RHT-Air User manual

scigiene
scigiene PTL572 User manual

scigiene
scigiene PTC572 User manual

scigiene
scigiene FI51P User manual

scigiene
scigiene FI50 User manual

scigiene
scigiene FP-180 User manual

scigiene
scigiene TK0305 User manual