• 43
English
English
TABLE OF CONTENTS
1. Signed and sealed . . . . . . . . . . . . 45
2. Device description . . . . . . . . . . . 46
2.1 Congratulations! . . . . . . . . . . . 46
2.2 Intended use . . . . . . . . . . . . . . 46
2.3 Description of function . . . . . . 46
2.4 User qualification . . . . . . . . . . 47
Assembly . . . . . . . . . . . . . . . . 47
Operation . . . . . . . . . . . . . . . . 47
3. Safety information . . . . . . . . . . . . 47
3.1 Safety information in these
instructions for use . . . . . . . . . 47
3.2 Basic safety instructions . . . . . 48
Handling the instrument . . . . . 48
Preventing electric shock . . . . 49
Avoiding infections . . . . . . . . . 49
Avoiding injuries . . . . . . . . . . . 50
Preventing device damage . . . 50
Handling measuring results . . . 51
Handling packaging material . . 52
4. Overview . . . . . . . . . . . . . . . . . . . 53
4.1 Controls . . . . . . . . . . . . . . . . . 53
4.2 Symbols in the display . . . . . . 54
4.3 Marking on the device and
on the rating plate . . . . . . . . . . 54
4.4 Marking on the packaging . . . 55
4.5 Menu structure . . . . . . . . . . . . 56
5. Before you really get started… . 57
5.1 Scope of delivery . . . . . . . . . . 57
5.2 Making display housing ready
for use (desktop display) . . . . . 58
Fitting the angle adapter . . . . . 58
Connecting the connecting
cable to the platform . . . . . . . 58
5.3 Making display housing ready
for use (wall-mounted display) 59
Fitting the angle adapter . . . . . 59
Connecting the connecting
cable to the platform . . . . . . . 60
Installing the wall bracket . . . . 61
Placing the display housing
on the wall bracket . . . . . . . . . 61
5.4 Establishing power supply . . . 62
5.5 Setting up the scale . . . . . . . . 63
Aligning the scale . . . . . . . . . . 63
6. Operation . . . . . . . . . . . . . . . . . . . 64
6.1 Weighing . . . . . . . . . . . . . . . . 64
Switching on the scale . . . . . . 64
Weighing a patient . . . . . . . . . 65
Taring off extra weight
(TARE) . . . . . . . . . . . . . . . . . . 65
Keeping the measurement
result in the display (HOLD) . . . 66
Determining and rating
body mass index (BMI) . . . . . . 67
Switching off the scale . . . . . . 68
6.2 Other functions (menu) . . . . . . 69
Navigating in the menu . . . . . . 69
Deleting saved values
automatically (AClr) . . . . . . . . . 70
Permanently saving the
additional weight (Pt) . . . . . . . 70
Activating the auto hold
(Ahold) function . . . . . . . . . . . 71
Activating beeps (BEEP) . . . . . 72
Setting filtering (Fil) . . . . . . . . . 72
Switching unit of weight (Unit) . 73
Restoring factory settings
(RESET) . . . . . . . . . . . . . . . . . 74

44 •
7. Hygiene treatment . . . . . . . . . . . 75
7.1 Cleaning . . . . . . . . . . . . . . . . . 75
7.2 Disinfecting . . . . . . . . . . . . . . . 75
7.3 Sterilizing . . . . . . . . . . . . . . . . 76
8. Function check . . . . . . . . . . . . . . 76
9. Troubleshooting . . . . . . . . . . . . . 77
10. Servicing . . . . . . . . . . . . . . . . . . 78
11. Technical data . . . . . . . . . . . . . 78
12. Optional accessories . . . . . . . . 79
13. Spare parts . . . . . . . . . . . . . . . . 79
14. Disposal . . . . . . . . . . . . . . . . . . . 80
15. Warranty . . . . . . . . . . . . . . . . . . 80

Signed and sealed • 45
English
1. SIGNED AND SEALED
With seca products, you are not only purchasing tech-
nology which has been perfected over 100 years, but
also quality with official legal certification confirmed by
institutions. seca products comply with European
directives, standards and national laws. When you buy
seca, you buy the future.
Products bearing this symbol meet the applicable regu-
latory requirements of the European Community, espe-
cially the following:
• Directive 93/42/EEC for medical devices
seca’s professionalism is also recognized by official
testing agencies. TÜV Süd Product Service, the
appointed office for medical devices, confirms with this
certificate that as a medical devices manufacturer,
seca consistently complies with the strict legal require-
ments. seca’s quality assurance system includes the
areas of design, development, production, sales and
service of medical scales and length measuring sys-
tems as well as software and measuring systems for
assessing state of health and nutrition.
seca helps the environment. Saving natural resources
is very important to us. We therefore make every effort
to save on packaging materials wherever it makes
sense and whatever is left over can be conveniently
disposed of on site via the recycling system.
0123

46 •
2. DEVICE DESCRIPTION
2.1 Congratulations!
With the seca 674 electronic multi-functional scale, you
have purchased a highly precise and simultaneously
robust device.
For over 170 years, seca has used its experience in the
service of healthcare and, as a market leader, it has
always set standards in many countries of the world
with innovative developments in weighing and
measuring.
2.2 Intended use
Electronic multi-functional scale seca 674 is mainly
used in hospitals, medical practices and in-patient care
facilities in accordance with national regulations.
The scale is for conventional determination of weight
and establishment of general state of nutrition; it assists
the physician supervising treatment in making a diag-
nosis or deciding on a course of treatment.
To make an accurate diagnosis, however, other spe-
cific examinations have to be ordered by the physician
and their results taken into account, in addition to
determining a weight value.
2.3 Description of function
On some variants of this scale, the weight display can
be switched between kilograms (kg), pounds (lbs) and
stones (sts). Weight is determined within a few
seconds.
In addition to conventional determination of weight, the
seca 674 provides a function for determining body
mass index. To this end, height is entered on the key-
pad and the body mass index associated with the
weight value is calculated automatically.
The seca 674 can be moved on casters.
Use the scale only for the purpose named in the
section entitled "Intended use" on page 46.

Safety information • 47
English
2.4 User qualification
Assembly Devices that are shipped partially assembled may only
be mounted by sufficiently qualified persons such as
specialist dealers, hospital technicians or seca service
technicians.
Operation The device may only be operated by healthcare
professionals.
3. SAFETY INFORMATION
3.1 Safety information in these instructions for use
DANGER!
Used to identify an extremely hazardous situa-
tion. If you fail to take note of this information,
serious irreversible or fatal injuries will occur.
WARNING!
Used to identify an extremely hazardous situa-
tion. If you fail to take note of this information,
serious irreversible or fatal injuries may result.
CAUTION!
Used to identify a hazardous situation. If you fail
to take note of this information, minor to moder-
ate injuries may result.
ATTENTION!
Used to identify possible incorrect usage of the
device. If you fail to take note of this information,
you may damage the device, or the measured
results may be incorrect.
NOTE
Includes additional information about use of the
device.

48 •
3.2 Basic safety instructions
Handling the instrument •Pleasetakenoteoftheinformation in these instruc-
tions for use.
• Keep the instructions for use in a safe place. The
instructions for use are a component of the device
and must be available at all times.
DANGER!
Risk of explosion
Do not use the device in an environment in
which one of the following gases has accumu-
lated:
–oxygen
– flammable anesthetics
– other flammable substances/air mixtures
CAUTION!
Patient hazard, damage to device
•Additionaldeviceswhichareconnectedto
electrical medical devices must provide evi-
dence of compliance with the relevant IEC or
ISO standards (e.g. IEC 60950 for data-
processing devices). Furthermore, all configu-
rations must comply with the requirements of
standards for medical systems (see IEC
60601-1-1 or Section 16 of the 3rd edition of
IEC 60601-1 respectively). Anyone connect-
ing additional devices to electrical medical
devices is considered a system configurer
and is therefore responsible for ensuring that
the system complies with the requirements of
standards for systems. Your attention is
drawn to the fact that local laws take prece-
dence over the above-mentioned require-
ments of standards. In the event of any
queries, please contact your local specialist
dealer or Technical Service.
•Haveservicingcarriedoutregularlyas
described in the relevant section of this docu-
ment.
•Technicalmodificationsmaynotbemadeto
the device. The device does not contain any
parts for servicing by the user. Only have serv-
icing and repairs performed by an authorized
seca service partner. You can find service
partners in your area at www.seca.com or by

Safety information • 49
English
•Onlyuseoriginalsecaaccessoriesandspare
parts, otherwise seca will not grant any war-
ranty.
CAUTION!
Patient hazard, malfunction
• Keep other electrical medical devices, e.g.
high-frequency surgical devices, a minimum
distance of approx. 1 meter away to prevent
incorrect measurements or wireless transmis-
sion interference.
•KeepHFdevicessuchascellphonesamini-
mum distance of approx. 1 meter away to
prevent incorrect measurements or wireless
transmission interference.
•TheactualtransmissionoutputofHFequip-
ment may require minimum distances of more
than 1 meter. Details can be found at
www.seca.com.
Preventing electric
shock WARNING!
Electric shock
•Setupthedevicesothatthepowersupply
socket is easy to reach and the device can be
disconnected from the power supply quickly.
•Ensurethatyourlocal power supply matches
the information on the power supply unit.
•Donottouchthepowersupply unit with wet
hands.
• Do not use extension cables or power strips.
• Ensure that the power cable is not pinched or
damaged by a sharp edge.
•Donotoperatethedeviceatanaltitudeof
more than 3000 m above sea level.
Avoiding infections WARNING!
Risk of infection
• Hygienically reprocess the scales regularly as
described in the respective section in this
document.
• Make sure that the patient has no infectious
diseases.
•Makesurethatthepatienthasnoopen
wounds or infectious skin alterations, which
may come into contact with the device.

50 •
Avoiding injuries
WARNING!
Risk of falling
•Ensurethatthedeviceispositioned firmly and
level.
•Routethemainscablesothat there is no risk
of tripping for the patient or user.
•Thedeviceisnotdesignedasastandingaid.
Assist people with limited motor skills when
they are getting up, e.g. from a wheelchair.
•Makesurethatthepatientdoesnotsteponto
the weighing platform directly at the edges.
•Makesurethatthepatientstepsontothe
weighing platform slowly and securely.
WARNING!
Risk of slipping
•Ensurethattheweighing platform is dry
before the patient steps onto it.
•Ensurethatthepatientsfeet are dry before he
or she steps onto the weighing platform.
•Makesurethatthepatientstepsontothe
weighing platform slowly and securely.
Preventing device
damage ATTENTION!
Damage to device
• Ensure that no liquids enter the device. They
can damage the electronics.
•Switchoffthedevicebeforedisconnectingthe
power supply unit from the mains socket.
•Disconnectthepowersupplyunitfromthe
mains socket if you intend to not use the
device for a longer period of time. Only this
way it can be ensured that the device is cur-
rentless.
• Make sure not to drop the device.
•Donotexposethedevicetoanyimpactsor
vibrations.
•Performfunctioncontrols regularly as
described in the relevant section in this docu-
ment. Do not operate the device if it is dam-
aged or not working properly.

Safety information • 51
English
•Ensurethatthereisnoheat source in the
immediate vicinity. Do not expose to direct
sunlight. The excessive temperature could
damage the electronics.
•Avoidrapidtemperature fluctuations. When
the device is transported so that a tempera-
ture difference of more than 20 °C occurs, it
must stay turned off for at least 2 hours before
it can be turned on again. Otherwise, conden-
sation water will form which can damage the
electronics.
•Causticdetergentsmaydamagethesurfaces.
Only use a soft cloth dampened with mild
soapsuds to clean the surfaces of the device.
•Onlyusedisinfectantssuitable for sensitive
surfaces. Suitable disinfectants are available
from specialist dealers.
Handling measuring
results
WARNING!
Patient hazard
This device is no diagnostic device. It simply
assists the treating physician in establishing a
diagnosis.
–Inordertomakeaprecisediagnosisand
initiate therapeutic measures, besides
determination of the weight, further targeted
examinations must be set up by the
physician, and their results must be
considered.
–Theresponsibilityfordiagnosis and treatment
lies with the treating physician.
CAUTION!
Patient hazard
In order to avoid misinterpretations, test results
for medical use must be displayed and used in
SI units only. Some devices offer the ability to
display test results in other units. This is only an
additional function.
– Use the results exclusively in SI units.
–Theuseofmeasurementresultsinnon-SI
units is the sole responsibility of the user.

52 •
ATTENTION!
Loss of data
•Beforeyousavemeasurementsaquiredwith
this device and process them for further use
(e.g. in the PC software seca analytics or in a
hospital information system), make sure the
measurement values are plausible.
•Ifmeasurementvaluesaretransmittedtothe
seca analytics PC software or a hospital
information system, make sure prior to further
use that the measurement values are plausi-
ble and are assigned to the correct patient.
Handling packaging
material
WARNING!
Risk of suffocation
Packaging material made of plastic foil (bags) is
a choking hazard.
–Keeppackagingmaterialoutofreachof
children.
–Intheeventthattheoriginalpackingmaterial
may not be available anymore, only use
plastic bags with security holes in order to
reduce the risk of suffocation.
NOTE
Keep the original packing material for future use
(e.g. returning for maintenance service).

Overview • 53
English
4. OVERVIEW
4.1 Controls
6
2
3
5
4
7
9
10
11
1
8
No. Control Function
1Display housing •Centralcontrolanddisplayelement
•Canbesetuponaworksurfaceorfittedtoawall
2Switch scale on and off
3
Arrow key
• During weighing:
- press briefly: activate hold function
-pressandhold:activatetarefunction
•Inthemenu:
- select submenu, select menu item
- increase value
tare

54 •
4.2 Symbols in the display
4.3 Marking on the device and on the rating plate
4
Arrow key
•Duringweighing:
- press briefly: activate BMI function
-pressandhold:callupmenu
•Inthemenu:
-selectsubmenu,selectmenuitem
-reducevalue
5
Enter key
•Inthemenu:
-confirmselectedmenuitem
-savesetvalue
6Display Display element for measuring results and to configure
the device
7Transport
casters The scale can be moved on these casters.
8Ramps Make it easier to access the scales with a wheelchair
9Spirit level Shows whether the device is horizontal
10 Power supply
connection Serves to connect the power pack supplied
11 Foot screw 4 pcs, for precise alignment
No. Control Function
enter
2
A
B
C
Symbol Meaning
AOperation with power pack
BNon-calibratable function active
CMemory location currently in use
2
1
3
Text/symbol Meaning
Model Model number
Ser. no. Serial number

Overview • 55
English
4.4 Marking on the packaging
Follow instructions for use
Electrical medical device, type B
Insulated device, protection class II
dValue which gives the difference between two consecutive
display values
Device conforms to EC directives
•0123: appointed office for medical devices: TÜV Süd
Product Service
Rating plate on the power supply connection socket
•x-y V: required supply voltage
•max xxx mA: maximum current consumption
•:notepolarityofdeviceplug
• : operate device with direct current
Do not dispose of device with household waste
Text/symbol Meaning
0123
use compatible
seca adapter only
x-y V
max. xxx mA
Protect from moisture
Arrows indicate top of product.
Transport and store in an upright position.
Fragile
Do not throw or drop.
Permitted min. and max. temperature for transport and
storage
Permitted min. and max. moisture for transport and
storage
Packaging material can be disposed of through recycling
programs.

56 •
4.5 Menu structure
Other functions are available to you in the menu for the
device. This enables you to configure the device per-
fectly to suit your needs (details from page 69).
Reset
Autohold
!on
!off
Fil
!0
!1
!2
Beep
!on
!off
!on
!off
Press
Hold
Unit
!kg
!lbs
!sts
Sto 2 (Weight)
PT Net Weight
Sto 1 (Weight)
Sto 3 (Weight)
Menü aufrufen:
Navigation:
enter
Navigation:
Call up menu:

Before you really get started… • 57
English
5. BEFORE YOU REALLY GET STARTED…
5.1 Scope of delivery
a
bcdef
USA/Japan
Euro Australia
UK
g
No. Component Pcs.
aWeighing platform 1
bWall bracket 1
cAngle adapter 1
dScrews, 3 x 35 mm 2
eWall plugs Ø 5 mm 2
fDisplay housing with connecting cable 1
gPower pack with adapters
(depending on model: power pack with Euro connector) 1
Instructions for use, not shown 1

58 •
5.2 Making display housing ready for use (desktop display)
ATTENTION!
Damage to device
The angle adapter cannot be removed once
fitted.
– Ensure before fitting that the display housing
is to be set up on a desktop.
Fitting the angle
adapter
1. Hold the angle adapter against the display housing
in the position shown.
2. Ensure that the connecting cable is running in the
guide groove of the angle adapter.
3. Push the angle adapter onto the display housing
until all four lugs engage audibly.
Connecting the
connecting cable to the
platform
NOTE
When selecting the setup location, take
account of the length of the connecting cable
so that you can set up the display housing on a
work surface.
1. Place the scales on a firm, level surface.
2. Connect the connecting cable to the electronics
box of the scales.
=> =>
Click!

Before you really get started… • 59
English
ATTENTION!
Incorrect measurement due to force shunts
If the cable of the display housing contacts the
weighing platform, the patient's weight cannot
be measured correctly.
– Route the cable in such a way that it cannot
contact the weighing platform.
3. Set up the display housing on a work surface.
5.3 Making display housing ready for use (wall-mounted
display)
ATTENTION!
Damage to device
The angle adapter cannot be removed once
fitted.
–Ensurebefore fitting that the display housing
is to be wall-mounted.
Fitting the angle
adapter
1. Hold the angle adapter against the display housing
in the position shown.
2. Ensure that the connecting cable is running in the
guide groove of the angle adapter.
3. Push the angle adapter onto the display housing
until all four lugs engage audibly.
=>
=>
Click!

60 •
Connecting the
connecting cable to the
platform
NOTE
When selecting the setup location, take
account of the length of the connecting cable
so that you can place the display housing on
the wall bracket.
1. Place the scales on a firm, level surface.
2. Connect the connecting cable to the electronics
box of the scales.
ATTENTION!
Incorrect measurement due to force shunts
If the cable of the display housing contacts the
weighing platform, the patient's weight cannot
be measured correctly.
– Route the cable in such a way that it cannot
contact the weighing platform.
Other manuals for 674
2
Table of contents
Other Seca Scale manuals
Popular Scale manuals by other brands

Silvercrest
Silvercrest 100926 Operation and safety notes

Silvercrest
Silvercrest SPWB 180 A1 Operating instructions and safety instructions

SRScales
SRScales SR520 owner's manual

Silvercrest
Silvercrest Z31123 Operation and safety notes

Trans-Tek
Trans-Tek GBF-1319-B6 user manual

health sense
health sense PS 121 user manual

German pool
German pool ESB-120 operating instructions

My Weigh
My Weigh iBalance 211 Operation manual

TScale
TScale ATW series Operation manual

Cambridge Scale Works, Inc.
Cambridge Scale Works, Inc. SS660-PT-R Installation, Set-Up and Operation Manual

Silvercrest
Silvercrest SNAW 1000 B1 operating instructions

My Weigh
My Weigh iBalance i500 user manual