Seca 703 User manual


• 35
English
English
CONTENTS
1. Description of device . . . . . . . . . . .36
1.1 Congratulations! . . . . . . . . . . . . 36
1.2 Intended use . . . . . . . . . . . . . . 36
2. Safety information. . . . . . . . . . . . . .37
2.1 Basic safety precautions. . . . . .37
2.2 Safety information in this
manual . . . . . . . . . . . . . . . . . . .37
2.3 Handling (rechargeable)
batteries . . . . . . . . . . . . . . . . . .38
3. Overview . . . . . . . . . . . . . . . . . . . . 39
3.1 Controls/Features. . . . . . . . . . .39
3.2 Symbols in display . . . . . . . . . .40
3.3 Information on rating plate . . . .40
3.4 Menu structure . . . . . . . . . . . . .42
4. Before you get started … . . . . . . . . 43
4.1 Installation of column . . . . . . . .43
4.2 Create power supply . . . . . . . .44
Insert batteries . . . . . . . . . . . . . 44
Connect the power supply unit
(optional). . . . . . . . . . . . . . . . . . 44
4.3 Set up scales . . . . . . . . . . . . . .45
5. Operation . . . . . . . . . . . . . . . . . . . .45
5.1 Weighing . . . . . . . . . . . . . . . . .45
Start weighing. . . . . . . . . . . . . .45
Weigh babies/infants (2 in 1) . . .45
Continuous display of measured
result (HOLD) . . . . . . . . . . . . . . 46
Determine and evaluate body
mass index (BMI) . . . . . . . . . . .47
Calculate and print out BMI
automatically. . . . . . . . . . . . . . .48
Transmit measured results to a
radio receiver . . . . . . . . . . . . . .48
Switch off scales. . . . . . . . . . . . 48
5.2 Additional functions (menu). . . . 49
Navigate in the menu . . . . . . . . 49
Clear stored values automatically
(AClr) . . . . . . . . . . . . . . . . . . . . 50
Permanently store additional
weight (Pt) . . . . . . . . . . . . . . . . 50
Activate Autohold function
(Ahold) . . . . . . . . . . . . . . . . . . . 51
Activate acoustic signals
(BEEP) . . . . . . . . . . . . . . . . . . . 52
Select attenuation (Fil). . . . . . . . 52
Switch weight unit (Unit) . . . . . . 52
Restore factory settings
(RESET) . . . . . . . . . . . . . . . . . . 53
6. seca 360° wireless Network . . . . . . 54
6.1 Introduction . . . . . . . . . . . . . . . 54
seca wireless groups . . . . . . . . 54
Channels . . . . . . . . . . . . . . . . . 54
Detection of equipment devices 55
6.2 Operate scales in a wireless
group (menu) . . . . . . . . . . . . . . 55
Activate wireless module (SYS) . 56
Set up wireless group (Lrn) . . . . 56
Activate automatic transmission
(ASend). . . . . . . . . . . . . . . . . . . 58
Select print option (APrt) . . . . . . 58
Set time (Time) . . . . . . . . . . . . . 58
7. Cleaning. . . . . . . . . . . . . . . . . . . . . 59
8. What to do if …? . . . . . . . . . . . . . . 60
9. Maintenance. . . . . . . . . . . . . . . . . . 61
9.1 Information about maintenance 61
10. Technical data . . . . . . . . . . . . . . . . 62
11. Accessories . . . . . . . . . . . . . . . . . . 63
12. Disposal. . . . . . . . . . . . . . . . . . . . . 64
12.1 Disposal of device . . . . . . . . . 64
12.2 Batteries (including
rechargeables) . . . . . . . . . . . . . 64
13. Warranty . . . . . . . . . . . . . . . . . . . . 64

36 •
1. DESCRIPTION OF DEVICE
1.1 Congratulations!
By purchasing the electronic personal scales seca 703
you are now in possession of an extremely accurate
device that is also robust at the same time.
seca has put its experience at the service of health care
for over 170 years and as a market leader in many
countries, is constantly setting new standards with its
innovative developments for weighing and measure-
ment.
1.2 Intended use
The electronic personal scales seca 703 are mainly
used in hospitals, doctors' surgeries and inpatient care
facilities in accordance with national regulations.
Some versions of these scales allow you to switch the
weight display between kilos (kg), pounds (lbs) and
stones (sts). The result of weighing is available within a
few seconds.
Besides conventional weight measurement the
seca 703 also features a function to calculate the body
mass index. The keypad is used to enter the height,
and this is followed by automatic calculation of the
body mass index from the weighing value.
The seca 360° wireless network allows the measured
results to be transmitted wirelessly to a seca wireless
printer or a PC equipped with the seca analytics 105
software and the
seca 360° wireless USB adapter 456.
The seca 703 is provided with rollers to allow it to be
moved about and thanks to its low power consump-
tion, it will offer hours of battery operation for mobile
applications.

Safety information • 37
English
2. SAFETY INFORMATION
2.1 Basic safety precautions
•Pleasetakenoteoftheinformationinthisuser
manual.
• Keep the user manual with the declaration of
conformity in a safe place.
•Makesurethatthescalesarepositioned
securely on a flat and stable surface.
•Makesureyoudonotdropthescales.
Protect the scales from violent impacts.
•Whenusingthescaleswithapowersupply
unit, make sure you route the mains cable so
that no-one can trip over it.
•Havethescalesservicedregularlysee
"Maintenance" on page 61.
• Make sure that maintenance and repair are
only carried out by an authorised service part-
ner. You can find your local service partner at
www.seca.com or just send an e-mail to
service@seca.com.
• Make sure you only use genuine seca acces-
sories and spare parts. Otherwise the war-
ranty provided by seca will become null and
void.
•MakesureRFequipmentsuchasmobile
phones is kept at a minimum distance of
approx. 1 metre to prevent incorrect meas-
urements or interference with radio transmis-
sion.
2.2 Safety information in this manual
DANGER!
Used to identify an extremely hazardous situa-
tion. If you fail to take note of this information,
serious irreversible or fatal injuries will result.
WARNING!
Used to identify an extremely hazardous situa-
tion. If you fail to take note of this information,
serious irreversible or fatal injuries may result.

38 •
CAUTION!
Used to identify a hazardous situation. If you fail
to take note of this information, minor to moder-
ate injuries may result.
ATTENTION!
Used to identify possible incorrect usage of
device. If you fail to take note of this information,
you may damage the device or the measured
results may be incorrect.
NOTE:
Includes additional information about use of the
device.
2.3 Handling (rechargeable) batteries
The device is supplied with 6 batteries, type AA
(Mignon). This type of battery is not rechargeable.
Please take note of the following safety information.
WARNING!
Personal injury with improper handling
Batteries contain harmful substances which
may explode if not handled properly.
–Donottrytorechargesuchbatteries.
– Do not expose (rechargeable) batteries to
heat.
– Do not burn (rechargeable) batteries.
– If acid is leaking out, avoid contact with the
skin, eyes and mucous membranes. Rinse
affected areas with plenty of clean water and
seek medical help at once.
ATTENTION!
Damage to device and malfunctions with
improper handling
–Onlyusethetypeof(rechargeable)battery
specified see "Insert batteries" on page 44.
– When replacing (rechargeable) batteries,
always replace a complete set at a time.
–Donotshort-circuit(rechargeable)batteries.
– If you do not use the device for a long period
of time, remove the batteries (incl.
rechargeables). This will prevent acid from
leaking into the device.

Overview • 39
English
3. OVERVIEW
3.1 Controls/Features
1
2
4
3
5
6
7
9
8
No. Control/Feature Function
1To switch scales off and on
2
Arrow key
• During weighing:
-Pressbriefly:ActivateHoldfunction
-Holddown:Activate2in1function
•Inmenu:
-Selectsubmenu,selectmenuitem
-Increasevalue
3
Arrow key
• During weighing:
-Pressbriefly:ActivateBMIfunction
-Holddown:Callupmenu
•Inmenu:
-Selectsubmenu,selectmenuitem
- Reduce value
4
Enter key
•Duringweighing(ifwirelessnetworkissetup):
- Press briefly: Send measured result to receive-ready
equipment (PC with USB wireless module)
-Holddown:Printoutmeasuredresult(wireless
printer)
•Inmenu:
- Confirm menu item selected
-Storesetvalue
5Display Display element for measured results and for device
configuration
6Battery
compartment Designed to take 6 batteries, type AA (Mignon), 1.5 V

40 •
3.2 Symbols in display
3.3 Information on rating plate
7Power connec-
tion Used to connect an optional power supply unit
8Transport rollers The scales can be moved about with these rollers
9Foot screw 4 screws, for precise alignment of device
No. Control/Feature Function
A
B
C
Symbol Meaning
ABatteries are weak
BOperation with power supply unit
CNon-calibratable function active
Text/Symbol Meaning
Model Model number
Type Type designation
Ser. No. Serial number
Refer to user manual
Type B electromedical device
Class II totally insulated appliance
FCC ID
For USA:
Device licensing number issued by US authority FCC
(Federal Communications Commission)
IC
For Canada:
Device licensing number issued by authority Industry
Canada
Device complies with EC standards and directives.
0123

Overview • 41
English
Symbol for FCC (USA)
Only operate device with DC
Do not discard with household waste
Text/Symbol Meaning
-+

42 •
3.4 Menu structure
Other functions are available in the menu of the device.
This enables you to configure the device perfectly to
suit the conditions in which you use it (details from
page 49).
Reg. Devices (Mo)
•Channel 1 (C1)
•Channel 2 (C2)
•Channel 3 (C3)
StoprF* Group (ID)Learn
Autosend
•on
•off
Autoprint
•HI
•MA
•HI_MA
•off
Time
•Year
•Month
•Day
•Hour
•Minute
Reset
Autohold
•on
•off
Fil
•0
•1
•2
Beep •on
•off
•on
•off
Press
Hold
Unit
•kg
•lbs
•sts
Erkannte Geräte (Mo):
-1: Personenwaage
-2: Längenmessstab
-3: Funkdrucker
-4: PC mit USB-Funkmodul
-7: Babywaage
Raum (ID):
Max. drei seca Funkräume: 0, 1, 2
Maximal-Konfiguration pro Funkraum:
-1 Babywaage
-1 Personenwaage
-1 Längenmessstab
-1 Funkdrucker
-1 PC mit USB-Funkmodul
Kanal (C1, C2, C3):
-Drei Kanäle pro Funkraum ( insg. 9 Kanäle)
-Kanalzahlen: 0-99
-Jede Kanalzahl für nur jeweils einen Kanal
-Empfohlener Abstand: 30
Beispiel-Konfiguration:
-Raum 0: C1__0, C2_30, C3_60
-Raum 1: C1_10, C2_40, C3_70
-Raum 2: C1_20, C2_50, C3_80
(Hinweis: Im Display keine Leerzeichen)
System •on
•off
PT Weight
PT Net Weight
*Funknetzwerk seca 360° Wireless:
•on
•off
Autoclear
Navigation:
Call up menu:
*Wireless network seca 360°
wireless:
Group (ID):
Max. 3 seca wireless groups: 0, 1, 2
Maximum configuration per
wireless group:
-1setofbabyscales
-1setofpersonalscales
- 1 length measuring device
- 1 wireless printer
- 1 PC with USB wireless module
Channel (C1, C2, C3):
-Threechannelsperwireless
group (in total 9 channels)
-Channelnumbers:0-99
- Only use each channel number
once
- Recommended spacing: 30
Example of configuration:
-Group0:C10,C230,C360
-Group1:C110,C240,C370
-Group2:C120,C250,C380
(Note: no spaces in display)
Devices detected (Mo):
-1:Personalscales
-2:Lengthmeasuringdevice
- 3: Wireless printer
-4:PCwithUSBwirelessmodule
-7:Babyscales

Before you get started … • 43
English
4. BEFORE YOU GET STARTED …
4.1 Installation of column
The column with the LCD display can be fitted in two
positions, so allowing you to determine the direction in
which the display is legible - facing either towards or
away from the platform.
1. Fit the column cover to the column.
2. Place the column in the recess, facing it in the de-
sired direction.
3. Screw the column to the weighing platform from
underneath.
If necessary, lay the scales down carefully on their
side to tighten up the screws.
4. On the underside of the weighing platform connect the
plugs of the display cable as shown in the diagram.
ATTENTION!
Malfunction with any error in installation
If cables are strained during fitting, e.g. bent
tight or with plugs kinked, this may result in mal-
functions and failure of the display.
–Makesurethatallcablesarelaidsoasto
prevent them from being bent too tight or the
plugs kinked.
–Ensurestrainreliefbylayingallcablesinthe
relevant holders.
5. Secure cables in their holders.
6. Then place the scales upright again.
7. Slide the column cover right down until it engages
on the weighing platform.

44 •
4.2 Create power supply
Power is supplied to the scales either using batteries or
with a power supply unit available as an accessory.
Insert batteries You need 6 batteries, type AA (Mignon), 1.5 V (included
in scope of supply).
1. Press on latch of battery compartment.
2. Move down lid of battery compartment.
3. Remove battery holder.
4. Insert batteries in holder.
NOTE:
Observe correct polarity of batteries (markings
on battery holder). Should the display read
bAtt, you have inserted one of the batteries
incorrectly or they are flat.
5. Replace the holder with the batteries in the battery
compartment.
6. Close the battery compartment.
Connect the power
supply unit (optional)
The scales can be operated with a power supply unit
available as an accessory.
WARNING!
Personal injury and damage to device if incor-
rect power supply unit is used
The voltage provided by standard power supply
units may be higher than their indicated rating.
This may cause the scales to overheat, catch
fire, melt or short-circuit.
– Only use genuine seca plug-in power supply
units with 9V or a controlled 12 V output
voltage.
1. Insert the power plug necessary for your power
supply in the power supply unit.
2. Insert the connector plug of the power supply unit
in the connecting socket of the scales.
3. Plug the power supply unit into a mains socket.

Operation • 45
English
4.3 Set up scales
ATTENTION!
Incorrect measurement with force shunt
If the housing of the scales is resting on some-
thing, e.g. a towel, the weight measurement will
not be correct.
–Whensettingupthescales,ensurethatthey
only make contact with the ground via the
foot screws.
1. Place the scales on a flat, stable surface.
2. Align the scales by turning the foot screws.
5. OPERATION
5.1 Weighing
Start weighing 1. Check that there is no load on the scales.
2. Press the Start key.
The display will read se[A and then quickly run
through all elements of the display.
The scales are ready for operation when the display
reads 0.00.
If the device is being operated with a power supply
unit, the symbol will appear in the display.
3. Step onto the scales and try not to move.
4. Read off the measured result.
Weigh babies/infants
(2 in 1)
You can use the 2 in 1 function to weigh babies and
infants. Here the child is held by an adult during weigh-
ing. Proceed as follows:
1. Check that there is no load on the scales.
2. Ask the adult to step onto the scales.
3. Read off the measured result.
4. Hold down the arrow key (hold/2 in 1) until NET
appears in the display.
The weight is stored.
0.00 appears in the display.
Foot screws

46 •
ATTENTION!
Incorrect measurement due to change in initial
weight
If the child is weighed with a different initial
weight, the child's weight will not be calculated
correctly.
– Make sure that the child is always weighed
with the adult used to determine the initial
weight..
– Check that the weight of this adult does not
change, e.g. by taking off a garment.
5. Ask the adult to step onto the scales with the child.
The child's weight is displayed.
The symbol, the symbol and the messages
"HOLD" and "NET" are displayed.
6. Ask the adult to step off the scales with the child.
7. To disable the 2 in 1 function, press the arrow key
(hold/2 in 1) until the symbol, the symbol and
and the messages "HOLD" and "NET" are no longer
displayed or switch the scales off.
Continuous display of
measured
result (HOLD)
If you activate the HOLD function, the weighing value
will continue to be displayed after the load has been
removed. This allows you to attend to the patient
before noting down the weight.
1. Check that there is no load on the scales.
2. Ask the patient to step onto the scales.
3. Briefly press the arrow key (hold/2 in 1).
The display will flash until a stable weight is meas-
ured. The weighing value will then be displayed
continuously. The symbol (non-calibratable func-
tion) and the message "HOLD" will be displayed.
4. To disable the HOLD function, briefly press the ar-
row key (hold/2 in 1).
The symbol and the message "HOLD" will no
longer be displayed.
NOTE:
If the Autohold function is activated, continuous
display of the weighing value will take place
automatically. see "Activate Autohold function
(Ahold)" on page 51.

Operation • 47
English
Determine and evaluate
body mass index (BMI)
The body mass index compares height and weight, so
resulting in more accurate figures than with Broca's for-
mula for ideal weight. It includes a tolerance range
which is considered ideal in health terms.
1. Check that there is no load on the scales.
2. Briefly press the arrow key (bmi/menu).
The message "BMI" appears.
Arrows can be seen flashing in the display.
The height last entered is displayed.
3. You can either accept this height or select another
value using the arrow keys.
4. Confirm your selection with the Enter key (send/
print).
5. Ask the patient to step onto the scales and try not
to move.
6. Read off the BMI and compare it with the catego-
ries given below.
7. To disable the BMI function, briefly press the Enter
key (send/print).
BMI Evaluation
Under 18.5
The patient is underweight. There might be a tendency to
anorexia. It is advisable that the patient puts on weight to
improve well-being and performance. If in doubt, a medical
specialist should be consulted.
Between 18.5 and
24.9 The patient's weight is normal. No change is necessary.
Between 25 and 30
(pre-obesity)
The patient is slightly to moderately overweight and should
lose weight if already suffering from disease (e.g. diabetes,
high blood pressure, gout, disorders of the fat metabolism).
Over 30
The patient urgently needs to lose weight as this is putting a
strain on the metabolism, circulation and skeleton. The
recommendation is for careful adherence to a diet, plenty of
exercise and behavioural training. If in doubt, a medical speci-
alist should be consulted.

48 •
Calculate and print out
BMI automatically
If you use these scales with a wireless printer and a
length measuring device from the seca 360° wireless
system, you can calculate and print out the BMI auto-
matically.
NOTE:
The prerequisite for this function is that the
equipment devices are logged into a wireless
group (see "seca 360° wireless Network" on
page 54).
1. Carry out weighing.
2. Briefly press the Enter key (send/print) on the
scales.
The measured value is sent to the wireless printer
but not printed out.
3. Measure the height.
4. Hold down the Enter key (send/print) on the length
measuring device.
The measured value is sent to the wireless printer.
The BMI is calculated.
The height, weight and BMI will be printed out.
Transmit measured
results to a radio
receiver
If the scales are integrated in a seca 360° wireless net-
work, you can transmit the measured results to
receive-ready equipment (wireless printer, PC with
wireless USB module) at the touch of a button.
Press the Enter key (send/print).
– Press key briefly: Send measured results to
all receive-ready equipment
–Holddown:Printoutmeasuredresultusing
wireless printer
Switch off scales Press the Start key.
NOTE:
When in battery mode, the scales will switch off
automatically after a short period if no load is
placed on them.

Operation • 49
English
5.2 Additional functions (menu)
Additional functions are available in the menu for the
scales. This allows you to configure the scales perfectly
to your own needs.
* The menu item "rF" is described in section “Operate
scales in a wireless group (menu)" on page 55.
Navigate in the menu 1. Switch on the scales.
2. Hold down the arrow key (bmi/menu) until the
menu is called up.
The last menu item selected is shown in the display
(here: Autohold "Ahold").
3. Keep pressing one of the arrow keys until the re-
quired menu item appears in the display (here: at-
tenuation "Fil").
rF* ….
Reset
Autohold
•on
•off
Fil
•0
•1
•2
Beep •on
•off
•on
•off
Press
Hold
Unit
•kg
•lbs
•sts
PT Weight
PT Net Weight
•on
•off
Autoclear

50 •
4. Confirm your selection with the Enter key (send/
print).
The current selection for the menu item or a sub-
menu is displayed (here level "0").
5. To change your selection or call up another sub-
menu, keep pressing one of the arrow keys until the
required selection is displayed (here: level "2").
6. Confirm the selection with the Enter key (send/
print).
You will leave the menu automatically.
7. To make further selections, call up the menu again
and proceed as described above.
NOTE:
If no key is pressed for approx. 24 seconds, you
will leave the menu automatically.
Clear stored values
automatically (AClr)
To avoid storing obsolete measured results in the
device memory and so calculating the BMI incorrectly,
you can set the scales to ensure measured values are
automatically cleared after 5 minutes.
NOTE:
With certain models this function is already acti-
vated at the factory. If required, you can disable
this function.
1. Select the item "AClr" in the menu.
2. Confirm the selection.
3. Select the setting you require:
–On
–Off
4. Confirm the selection.
You will leave the menu automatically.
Permanently store
additional weight (Pt)
You can use the Pre-Tare function to permanently store
an additional weight and automatically deduct it from
the measured result. You can for example store a flat-
rate figure as the weight of shoes and clothing and then
always deduct it from the measured result when a
patient is weighed fully dressed.
1. Select the item "Pt" in the menu.
The message "Pt" appears.

Operation • 51
English
2. Confirm your selection.
Arrows can be seen flashing in the display.
The additional weight last stored is displayed.
3. You can accept the stored value or adjust it using
the arrow keys.
NOTE:
If you enter the value "0", the function will be
switched off. The message "Pt" will no longer
be displayed.
4. Confirm your selection.
5. Ask the patient to step onto the scales.
The patient's weight is displayed.
The additional weight stored has been automati-
cally deducted.
6. To disable this function, select the item "Pt" in the
menu again.
7. Confirm your selection.
The function is disabled.
You will leave the menu automatically.
Activate Autohold
function (Ahold)
If you activate the Autohold function, the measured
result for each weighing operation will continue to be
displayed after the load has been removed. This means
you no longer have to manually activate the Hold func-
tion for every weighing operation.
NOTE:
•Withcertainmodelsthisfunctionisalready
activated at the factory. If required, you can
disable this function.
•Whateverselectionismadehere,inthe2in1
function the child's weight is always deter-
mined with Autohold.
1. Select the item "Ahold" in the menu.
2. Confirm the selection.
The current selection is displayed.
3. Select the setting you require:
–On
–Off
4. Confirm your selection.
You will leave the menu automatically.

52 •
Activate acoustic
signals (BEEP)
You can select whether an acoustic signal is to be
emitted whenever a key is pressed or a stable weighing
value has been attained. The latter is relevant for the
Hold/Autohold function.
NOTE:
The function "Acoustic signal with stable weigh-
ing value" is activated at the factory. If required,
you can disable this function.
1. Select the item "BEEP" in the menu.
2. Confirm the selection.
3. Select a menu item:
–Press:Acousticsignalwhenever a key is
pressed
–Hold:Acousticsignalwith a stable weighing
value.
4. Confirm your selection.
The current selection is displayed.
5. Select the setting you require:
–On
–Off
6. Confirm your selection.
You will leave the menu automatically.
7. If you also wish to activate the acoustic signal for
the second function, repeat this procedure.
Select attenuation (Fil) You can use attenuation (Fil = filter) to reduce any inter-
ference during weighing (e.g. caused by patient mov-
ing).
1. Select the item "Fil" in the menu.
2. Confirm the selection.
The current selection is displayed.
3. Select an attenuation level.
– 0: low attenuation
–1:moderateattenuation
–2:highattenuation
4. Confirm the selection.
You will leave the menu automatically.
Switch weight unit
(Unit)
With uncalibrated scales you can select the unit you
wish to use for displaying the weight.
1. Select the item "Unit" in the menu.

Operation • 53
English
2. Confirm the selection.
The current selection is displayed.
Select the unit you wish to use for displaying the
weight:
– kilos (kg)
–pounds(lbs)
–stones(sts)
3. Confirm the selection.
You will leave the menu automatically.
Restore factory settings
(RESET)
You can restore the factory settings for the following
functions:
NOTE:
The wireless module is switched off when
restoring the factory settings. Data for existing
wireless groups remains in the memory. These
groups do not need to be set up again.
1. Select the item "Reset" in the menu.
2. Confirm the selection.
You will leave the menu automatically.
3. Switch the scales off.
The factory settings are restored and will be availa-
ble when the scales are next switched on.
Function Factory setting
Autohold (Ahold) depending on
model
Acoustic signal (Press) off
Acoustic signal (Hold) on
Attenuation (Fil) 0
Autoclear (Aclear) depending on
model
Pre-Tare (Pt) 0 kg
Height for body mass index
(BMI) 170 cm
Weight unit kg
Wireless module (SYS) off
Autosend (ASend) off
Autoprint (APrt) off
Other manuals for 703