Page
4. Installation
Donotopenoralteranypartoftheproductwhichwouldleadtomalfunctionandirrecoverable
damage.Itwillalsoforfeitthetermsofthewarrantyandcauseliability.
Theproductatthetimeofshipmentisfullyinspectedforitsqualityandmeetsallsafetyrequirements.
Additionalsafetymeasuresduringtheinstallationshouldbeapplied.Thisincludes,butisnotlimited
totheleakageverificationprocedures,standardEDS(electrostaticdischarge)precautions,DC
voltageprecautions.Othertaskssuchascalibration,partreplacement,repair,andmaintenance
mustonlybeperformedbytrainedpersonnel.Uponrequests,themanufacturerwillprovide
necessarytechnicalsupportand/ortrainingofthepersonnel.
Therearenopreferreddirectionsfortheinstallation.Flowdirectionshouldbealignedwiththearrow
markonthemeterbody.Iftheflowingfluidmayhaveparticlesordebris,afilterisstrongly
recommendedtobeinstalledupstreamofthemeter.
Pleasefollowthefollowingstepstocompletetheinstallation:
a)Uponopeningthepackage,theproduct'sphysicalintegrityshouldbeinspectedtoensureno
visualdamage.
b)Beforeinstallationoftheproduct,pleaseensurethatthepipedebrisorparticlesoranyother
foreignmaterialsarecompletedremoved.
c)Closethemetervalvecompletely.
d)Duringinstallation,pleasemakesurenoforeignmaterials(suchaswater,oil,dirt,particles,etc.)
enteringintotheinstallationpipeline.
e)ConnectelectricalwiresperthewiredefinitioninTable3.2.Pleasebesureofthepowersupply
range(i.e.,8~24VDC)andpowersupplypolarization.Ifanadapterotherthantheonesupplied
bythemanufacturer,makesuretheadaptermeetsindustrialstandardsandhasallsafety
certifications.
f)Forthedatacommunicationwireconnection,pleasefollowthedescriptioninTable3.2and
makesurethatthewiresarecorrectlyconnectedtotheproperportsonyourdata
device/equipment.Pleasemakesurethedatacablemeetsindustrialstandardswithproper
shielding.
g)Oncetheexternalpowerissuccessfullyconnected,theLEDshouldbelightedupwiththeproper
informationdisplayedworkscorrectly.
h)Slowlyopenthevalve(s)ofthegassupplyifany,upstreamordownstream,orbothofthe
pipeline.Andthenslowlyopenthevalveofthemeter,themetershouldthenstarttomeasure
theflowinthepipeline.Note:becausethemeterhasalargedynamicalmeasurementrange,it