5.5.8 Select the display language
This function is reserved for future upgrades. The current models are
configured to be English or Chinese based on the shipping
geographic regions. However, if one likes to switch between these
two default languages, following the above-mentioned steps, at the
MENU setting screen, use the “Up” or “Down” key to select F39 -
LAnG. Press the “M” key to confirm, and it will open the sub-MENU
showing the current language. Use the “Up” or “Down” and the “M”
confirming key to select the desired one, and press the “M” key again
to complete the task. The display will then return to the F39 - LAnG
screen, which indicates the task is completed. Use the “Up” or “Down” key to select F99 - qUIT and
the “M” key to exit the MENU and return to the Main Display screen.
5.5.9 Open/close the valve
This function is to open or close the valve. If one likes to open or close
the valve, following the above-mentioned steps, at the MENU setting
screen, use the “Up” or “Down” key to select F43 -VALVE. Press the
“M” key to confirm, and it will open the sub-MENU showing the
current valve status. Use the “Up” or “Down” and the “M” confirming
key to select the desired one, and press the “M” key again to
complete the task. The display will then return to the F43 - VALVE
screen, which indicates the task is completed. Use the “Up” or “Down”
key to select F99 - qUIT and the “M” key to exit the MENU and return
to the Main Display screen.
5.5.10 Set an alarm: upper instant flow rate limit
This function allows the user to set the instant flow rate above which
the meter will trigger the alarm function. The alarm will be a sharp
sound every 2 seconds whiling the screen/display will flash. The
triggered alarm will not stop the meter from continuous operation.
The alarm will be off once the metering value is below the set value.
The flow rate has two decimal points with a maximum not over the
full-scale flow rate at the order. Following the above-mentioned
steps, at the MENU setting screen, use the “Up” or “Down” key to
select F51 - ALm-H. Press the “M” key to confirm, and it will open the sub-MENU showing a default
flow rate of 999.99. Use the “Up” or “Down” and the “M” confirming key to enter the desired one,
and press the “M” key again to complete the task. The display will then return to the F51 - ALm-H