MF5600 Series User Manual
5.2 Installation Instructions
The product at the time of shipment is fully inspected for product quality and meets all safety
requirements. Additional safety measures during the installation should be applied. This includes,
but is not limited to leakage verification procedures, standard EDS (electrostatic discharge)
precautions, DC voltage precautions, and heavy duty precautions. Other tasks such as calibration,
part replacement, repair, and maintenance must only be performed by trained personnel. Upon
requests, manufacturer will provide necessary technical support and/or training of the personnel.
Do not open the product cover or alter any part of the product. Any such actions will forfeit the terms
of the warranty and cause the liability to any damages thereafter.
The product is preferably to be installed horizontally. Flow direction should be aligned with the
arrow mark on the meter body. If the flow fluid may have particles or debris, a filter is strongly
recommended to be installed upstream of the meter.
Please follow the following steps to complete the installation:
a) Upon opening the package, the product physical integrity should be inspected to ensure no visual
damage.
b) Before installation of the product, please ensure that the pipe debris or particles or any other
foreign materials are completed removed.
c) Cautions during installation:
(i) It is preferably to first install the inlet end of the meter and then the outlet end of the meter; To
ensure the measurement accuracy, an upstream straight pipe of length no less than 10DN and a
downstream straight pipe of length no less than 5DN should be in place.
(ii)During installation, please make sure no any foreign materials (such as water, oil, dirty,
particles, etc.) falling into the pipe.
d) Connect electrical wires for LCD, and then electrical wires for inputs/outputs. Please pay special
attention to power supply range (i.e., +12~+24 VDC) and power supply polarization (see the
description on Electrical Interfaces in this manual).
e) When connect the communication wires, please make sure that the wires are correctly connected
to the proper ports on your data device/equipment.
f) Turn on the power supply, and make sure that the LCD works correctly.
g) Slowly open the valves at the both ends of the pipeline, and the meter should then start to measure
the flow in the pipeline
h) Completion of the installation.
5.3 Cautions
a) Don't try to loose any build-in part of the product.
b) Ensure electrical wires for the inputs/outputs to be reliably connected.
c) Release all the installation stresses so that no stresses will be exerted on the product.
d) The product should avoid strong electromagnetic interference sources nearby or periodic
mechanical shocks to its body or pipeline.
e) Slowly open/close valves to prevent abrupt pulse flow impact, which may damage the product.