Compliance to UL60601-1
This medical equipment complies with UL60601-1 which is the standard for safety testing
electrical medical devices, under the following minimum conditions:
An ambient temperature range of +10°C to 43°C (+50°F to 110°F)
A relative humidity range of 30% to 75%
An atmospheric range of 700hPa (525mmHg) to 1060hPa (795mmHg)
Unpacking and Set-Up Instructions
Installation of the Platinum 6000™ System
The Platinum 6000™ Mattress can be placed on any conventional bed used in hospitals, nursing
homes or a home medical bed. The original bed mattress should be removed and stored in an
appropriate place. The following steps should be completed in installing the system:
1. Roll out mattress onto the bed frame with the air hoses exiting at the foot of the bed.
2. Strap the mattress base to the bed frame. Use caution when attaching the straps to make
sure there is no strap damage when the bed position is changed.
3. Place the Platinum 6000™ System Controller on the footboard as close to the center as
possible.
4. Make sure the hoses can reach the controller without undue strain.
5. Connect the air hoses to the controller.
6. Press the hose onto the controller connector until the hose connector clicks.
7. Pull back gently on the hose connector to verify the mechanical connection is secure.
a. The appropriate connections will be obvious as only certain mattress side
connectors will mate properly with the control unit side ports.
8. Examine each of the air cells in the mattress. Make sure each air cell is vertical and not
twisted in the mattress base.
9. Verify positive air connections by gently pulling the air manifold and air cell connections
gently apart.
a. If the connectors separate, press them back together until a click is heard.
10. Loose air cell connectors may prevent the mattress from inflating properly.
11. Check the snaps on the end of each air cell to make sure they are securely fastened to the
mattress base.
12. Route the power cord in a safe manner to the closest power outlet and plug in the unit.
a. When the unit is powered up, set the control unit mode to Auto Firm mode. This
will fill the mattress more rapidly in preparation for the patient.
13. Note the inflation of the air cells. Make sure all air cells are fully inflated and are not
twisted in the mattress base.
14. Change the mode from Auto Firm mode to the required mode of operation.
15. Adjust the firmness level for the patient by entering the patient’s height and weight on the
display control panel. This is accomplished by first depressing the patient Height button,
using the Up/Down buttons to increase or decrease the default height setting in
increments of 1” (2 cm). The same procedure is used when setting the Weight button to
set the patient’s weight in kilograms (pounds), using increments of 2 lbs. (1 kg).
a. Heavier patients are set to higher firmness levels. Lighter patients are set to lower
firmness levels. The Platinum 6000™ is preprogrammed to automatically select