Page 4
AURORA FOUR OWNER'S • REV7
Use of incompatible cleaning agents
willcausedamagetothexture.Avoid
the use of cleaning solutions which
contain high concentrations of alcohol,
ethylene glycol, phenol, iodophors, or
glutaraldehyde based disinfectants.
Staining, pitting, discoloration and
diusercrackingorpersonalinjurymay
occur if these are used.
CAUTION
CAUTION without the safety alert
symbol, is used to address practices
not related to personal injury but with
a possibility of damage to equipment.
Sterilizable focus/positioning handles
are subject to normal wear and tear.
Always examine the handles for wear
or damage to ensure proper and safe
operation with the surgical light.
Make sure the MAIN POWER switch is
OFFonthelightxturewallcontroland
the camera control unit before installing
or removing the camera.
DO NOT attempt to focus or position
the lighthead or camerahead using
the camera body or counter weight.
Damage to the camera rotation motor
may result.
DO NOT accidentally push in the camera
release button on the attachment ring
above the camera cover, as this will
cause the camera to disengage from
the lighthead or the camerahead.
Avoid pulling or attempting to move
the stand light by the focus/positioning
handle. The fixture may become
unstable and damage may occur.
Exercise caution when moving the stand
light to avoid obstacles such as power
cords or other items in the pathway. DO
NOT transport over rugged or unstable
ooringthatmaydamagethecasters.
Be sure the sterilizable handle is
properly secured before use. An
improperly installed handle could fall
out, causing possible injury to the patient
orsurgicalsta.
Make sure camera is securely locked in
lighthead or camerahead before moving
into use position.
DO NOT block vents. Position CCU
where it is not exposed to dripping or
splashing liquids and away from open
ameorheatproducingequipment.
To ensure patient safety, DO NOT
touch any component of the system
and patient simultaneously.
Always inspect product prior to use to
ensure safe and correct operation. Any
product deemed to be malfunctioning
should be removed from service and
labeled inoperable. Refer all service
to a qualified SKYTRON service
representative.
The AUA5S is equipped with locking
casters. Make sure the casters are
unlockedpriortomovingthextureto
avoid the risk of damage or personal
injury.
Risk of overbalancing! To avoid injuries
or equipment damage, DO NOT push
with excessive force, lean on, or rest
onthelightingxture.
DO NOT forcibly overcome a hurdle.
Ifneeded,usearampandhavearm
grasp on the support post handles to
maintain balance. A lack of balance
would cause a risk of falling.
Due to the high output capability of
the lightheads, it may be required to
decrease the intensity to reduce the risk
of excessive heat in the surgical site
when positioning multiple lightheads
together to form a single spot.