manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Solinst
  6. •
  7. Measuring Instrument
  8. •
  9. Solinst 101 Operating and safety instructions

Solinst 101 Operating and safety instructions

Water Level Meter Circuit Board Replacement
Model 101/102
Tools and Materials Needed
1. Mk2 Circuit Board Assembly (#102977) Includes:
• Wiring, Sensitivity Switch and Knob, Light and Red Lens, and Test Button
2. Phillips Screwdriver
3. 10 mm (3/8”) Wrench
4. Wire Cutters (if required)
5. Small Flat Screwdriver (3 mm (1/10”)
6. Pliers
Back of Mk2 Water Level Meter Faceplate with New Style Light
Instructions
For Mk2 Water Level Meters (Push-Release Tape Connection)
Note: Instructions for Mk1 Water Level Meters (Molex tape
connection, soldered connections to battery tray) are on Page 2.
1. Remove the battery. Unscrew the three screws and remove
the faceplate.
2. Press down on the white terminals of the push-release
fittings on the circuit board and pull out to remove the tape
leads.
3. Pull the quick-connect fittings off of the terminals on the
battery tray.
4. From the front of the faceplate, use the wrench to unscrew
the nut holding the test button and push the button back
out of the faceplate.
Back of Mk2 Water Level Meter Faceplate
with Older Style Light
Light
5. If you have an older style light (leads connected to Sonalert
terminals – see photo above), cut the red and black wires
connected to the light, and push the light out through the
front of the faceplate.
6. Use the small flat screwdriver to unscrew the small brass
screw on the side of the sensitivity knob. Remove the
sensitivity knob. Use the pliers to unscrew the nut holding
the sensitivity switch and remove the switch from the
faceplate. Save the knob and screw to be used with the
new sensitivity switch.
7. Remove the two screws to disconnect the old circuit board
assembly from the Sonalert. Remove the old circuit board.
8. Loosen the Sonalert retaining nut on the front of the
faceplate, so you can reposition the Sonalert as shown in
the photo below (negative terminal next to the battery tray).
9. If you had an older style light (or the lens needs replacing)
insert the new red lens through the opening in the faceplate
and secure using the white nut.
10. Place the circuit board assembly in position on top of the
Sonalert, ensuring the sides marked +ve and -ve are lined
up with the corresponding terminals on the Sonalert and
the light seated in the lens.
11. Install the two screws into the Sonalert terminals. Do not
tighten. Wrap the two bare wires from the circuit board
around the screws, between the screw head and Sonalert
terminals. Start by bringing the wires underneath the
screws (see photo below). Tighten the two screws.
(Page 1 of 3)
Leads From
Light Connected
To Sonalert
Light
Battery
Tray
Sonalert
High Quality Groundwater and Surface Water Monitoring Instrumentation
12. Insert the test button through the opening in the faceplate,
and secure it in place using the nut.
13. Insert the sensitivity switch through the faceplate with the
three connection prongs positioned towards the battery
drawer, slightly towards the centre of the reel. Tighten the
nut. Fasten the sensitivity knob on the front of the switch
using the brass screw.
14. Connect the black wire with quick-connect fitting to the
negative terminal on the battery tray. Connect the red wire
with quick-connect fitting to the positive terminal.
Back of Mk1 Water Level Meter Faceplate
Note: Instructions for Mk2 Water Level Meters are on Page 1
(tape connection using push-release fittings on circuit board
and quick-connect fittings to battery tray).
Back of Fully Assembled Water Level Meter Faceplate
(Page 2 of 3)
15. Connect the tape to the new circuit board assembly by
pressing down on the white terminals on the circuit board
and inserting the tape leads. Release the terminals and the
leads should be secured. The lead on the bottom of the tape
(numbers facing up, may be black mark on lead) is inserted
into the terminal with a white square below it on the circuit
board.
16. Install the battery.
17. With the probe in a glass of tap water, turn the Water Level
Meter to the ‘ON’ position. If the connections are correct
the buzzer and light will activate. If the buzzer or light do
not activate, check all connections and the polarity of the
battery.
18. Reattach the faceplate to the reel using the three screws.
Instructions
For Mk1 Water Level Meters (Molex Tape Connection)
Tools and Materials Needed
1. Mk2 Circuit Board Assembly (#102977) Includes:
• Wiring, Sensitivity Switch and Knob, Light and Red
Lens, and Test Button
2. Phillips Screwdriver
3. 10 mm (3/8") Wrench
4. Wire Cutters
5. Small Flat Screwdriver (3 mm (1/10")
6. Pliers
Sensitivity
Switch
Sonalert
Light
Battery
Tray
Test Button
Molex
Connector
1. Remove the battery. Unscrew the three screws and remove
the faceplate.
2. Disconnect the Molex connector that connects the circuit
board to the tape.
3. Cut the red and black wires from the battery tray.
4. From the front of the faceplate, use the wrench to unscrew
the nut holding the test button and push the button back
out of the faceplate.
5. Cut the red and black wires connected to the light, and
push the light out through the front of the faceplate.
Sonalert
Sensitivity
Switch
Light
Battery
Tray Test Button
Push-Release
Fittings
Square
High Quality Groundwater and Surface Water Monitoring Instrumentation
Water Level Meter Circuit Board Replacement
6. Use the small flat screwdriver to unscrew the small brass
screw on the side of the sensitivity knob. Remove the
sensitivity knob. Use the pliers to unscrew the nut holding
the sensitivity switch and remove the switch from the
faceplate. Save the knob and screw to be used with the
new sensitivity switch.
7. Remove the two screws to disconnect the old circuit board
assembly (and the wires connected to the light) from the
Sonalert.
8. Insert the new red lens through the opening in the faceplate
and secure using the white nut.
9. Use the pliers to bend the terminals on the Sonalert, so
they are straight up from the Sonalert (to accommodate
the new circuit board in between the terminals). Loosen
the Sonalert retaining nut on the front of the faceplate, so
you can reposition the Sonalert as shown in the photo on
Page 2 (negative terminal next to the battery tray).
10. Place the circuit board assembly in position on top of the
Sonalert, ensuring the sides marked +ve and -ve are lined
up with the corresponding terminals on the Sonalert and
the light seated in the lens.
11. Install the two screws into the Sonalert terminals. Do not
tighten. Wrap the two bare wires from the circuit board
around the screws, between the screw head and Sonalert
terminals. Start by bringing the wires underneath the
screws (see photo on Page 1). Tighten the two screws.
12. Insert the test button through the opening in the faceplate,
and secure it in place using the nut.
13. Insert the sensitivity switch through the faceplate with the
three connection prongs positioned towards the battery
drawer, slightly towards the centre of the reel. Tighten the
nut. Fasten the sensitivity knob on the front of the switch
using the brass screw.
14. Connect the black wire with quick-connect fitting to the
negative terminal on the battery tray. Connect the red wire
with quick-connect fitting to the positive terminal.
15. Remove the Molex connector by pushing out the two pins.
Cut each pin in half (see image at top right for correct
location). Use pliers to flatten the remainder of the pins
against the tape lead, so it fits easily into the terminals on
the circuit board.
Remove the Pins from the Tape Molex Connector
Cut the Pins on Each Tape Lead in Half (as shown)
16. Connect the tape to the new circuit board assembly by
pressing down on the white terminals on the circuit board
and inserting the tape leads. Release the terminals and the
leads should be secured. The lead on the bottom of the
tape (numbers facing up, may be black mark on lead) is
inserted into the terminal with a white square below it on
the circuit board.
17. Install the new battery.
18. With the probe in a glass of tap water, turn the Water Level
Meter to the ‘ON’ position. If the connections are correct
the buzzer and light will activate. If they do not activate,
check all connections and the polarity of the battery.
19. Reattach the faceplate to the reel using the three screws.
Square
Printed in Canada
March 19, 2018
(#113984)
(Page 3 of 3)
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023
35 Todd Road, Georgetown, Ontario Canada L7G 4R8
Web Site: www.solinst.com E-mail: [email protected]
®Solinst is a registered trademark of Solinst Canada Ltd.

This manual suits for next models

1

Other Solinst Measuring Instrument manuals

Solinst 101 P7 SC1000 User manual

Solinst

Solinst 101 P7 SC1000 User manual

Solinst 122 User manual

Solinst

Solinst 122 User manual

Solinst 102 Operating and safety instructions

Solinst

Solinst 102 Operating and safety instructions

Solinst 102 User manual

Solinst

Solinst 102 User manual

Solinst 101 User manual

Solinst

Solinst 101 User manual

Solinst 101 Operating and safety instructions

Solinst

Solinst 101 Operating and safety instructions

Solinst 615 ML User manual

Solinst

Solinst 615 ML User manual

Solinst 102 Operating and safety instructions

Solinst

Solinst 102 Operating and safety instructions

Solinst 104 User manual

Solinst

Solinst 104 User manual

Solinst Levelogger 3001 Use and care manual

Solinst

Solinst Levelogger 3001 Use and care manual

Solinst Levelogger Series User manual

Solinst

Solinst Levelogger Series User manual

Solinst 101 Operating and safety instructions

Solinst

Solinst 101 Operating and safety instructions

Solinst 9700 SolSat 5 User manual

Solinst

Solinst 9700 SolSat 5 User manual

Solinst 122M User manual

Solinst

Solinst 122M User manual

Solinst 101 SC2000 Operating and safety instructions

Solinst

Solinst 101 SC2000 Operating and safety instructions

Solinst 102M User manual

Solinst

Solinst 102M User manual

Solinst 101 P2 Operating and safety instructions

Solinst

Solinst 101 P2 Operating and safety instructions

Solinst 107 Mk2 User manual

Solinst

Solinst 107 Mk2 User manual

Solinst 101 User manual

Solinst

Solinst 101 User manual

Solinst 101D User manual

Solinst

Solinst 101D User manual

Solinst 101 User manual

Solinst

Solinst 101 User manual

Solinst LevelSender 5 User manual

Solinst

Solinst LevelSender 5 User manual

Solinst 107 Mk2 User manual

Solinst

Solinst 107 Mk2 User manual

Solinst 101 Operating and safety instructions

Solinst

Solinst 101 Operating and safety instructions

Popular Measuring Instrument manuals by other brands

Hach 5500sc Installation

Hach

Hach 5500sc Installation

Apogee Instruments MS-100 owner's manual

Apogee Instruments

Apogee Instruments MS-100 owner's manual

Bender ISOMETER isoCHA425 manual

Bender

Bender ISOMETER isoCHA425 manual

ATAGO PAL-61S instruction manual

ATAGO

ATAGO PAL-61S instruction manual

Parker RK 11-1669 Instruction

Parker

Parker RK 11-1669 Instruction

YSI 2900 series Operation and maintenance manual

YSI

YSI 2900 series Operation and maintenance manual

Saker XFVP-L user manual

Saker

Saker XFVP-L user manual

KROHNE OPTIMASS 7400 quick start

KROHNE

KROHNE OPTIMASS 7400 quick start

DeFelsko PosiTector UTG manual

DeFelsko

DeFelsko PosiTector UTG manual

Kolida DL-200 user manual

Kolida

Kolida DL-200 user manual

Novus NP400 instruction manual

Novus

Novus NP400 instruction manual

Boneco X210 manual

Boneco

Boneco X210 manual

Mark-10 7i user guide

Mark-10

Mark-10 7i user guide

Saluki S2106S Series user manual

Saluki

Saluki S2106S Series user manual

Dewetron TRIONet3 Technical reference

Dewetron

Dewetron TRIONet3 Technical reference

Acer AM100 user manual

Acer

Acer AM100 user manual

Thermo Scientific Sarasota FD910 user guide

Thermo Scientific

Thermo Scientific Sarasota FD910 user guide

enmco PT16FM Operating	 instruction

enmco

enmco PT16FM Operating instruction

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.