manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Solinst
  6. •
  7. Measuring Instrument
  8. •
  9. Solinst 101 Operating and safety instructions

Solinst 101 Operating and safety instructions

Water Level Meter Circuit Board Replacement
Model 101/102
Tools and Materials Needed
1. Mk2 Circuit Board Assembly (#102977) Includes:
• Wiring, Sensitivity Switch and Knob, Light and Red Lens, and Test Button
2. Phillips Screwdriver
3. 10 mm (3/8”) Wrench
4. Wire Cutters (if required)
5. Small Flat Screwdriver (3 mm (1/10”)
6. Pliers
Back of Mk2 Water Level Meter Faceplate with New Style Light
Instructions
For Mk2 Water Level Meters (Push-Release Tape Connection)
Note: Instructions for Mk1 Water Level Meters (Molex tape
connection, soldered connections to battery tray) are on Page 2.
1. Remove the battery. Unscrew the three screws and remove
the faceplate.
2. Press down on the white terminals of the push-release
fittings on the circuit board and pull out to remove the tape
leads.
3. Pull the quick-connect fittings off of the terminals on the
battery tray.
4. From the front of the faceplate, use the wrench to unscrew
the nut holding the test button and push the button back
out of the faceplate.
Back of Mk2 Water Level Meter Faceplate
with Older Style Light
Light
5. If you have an older style light (leads connected to Sonalert
terminals – see photo above), cut the red and black wires
connected to the light, and push the light out through the
front of the faceplate.
6. Use the small flat screwdriver to unscrew the small brass
screw on the side of the sensitivity knob. Remove the
sensitivity knob. Use the pliers to unscrew the nut holding
the sensitivity switch and remove the switch from the
faceplate. Save the knob and screw to be used with the
new sensitivity switch.
7. Remove the two screws to disconnect the old circuit board
assembly from the Sonalert. Remove the old circuit board.
8. Loosen the Sonalert retaining nut on the front of the
faceplate, so you can reposition the Sonalert as shown in
the photo below (negative terminal next to the battery tray).
9. If you had an older style light (or the lens needs replacing)
insert the new red lens through the opening in the faceplate
and secure using the white nut.
10. Place the circuit board assembly in position on top of the
Sonalert, ensuring the sides marked +ve and -ve are lined
up with the corresponding terminals on the Sonalert and
the light seated in the lens.
11. Install the two screws into the Sonalert terminals. Do not
tighten. Wrap the two bare wires from the circuit board
around the screws, between the screw head and Sonalert
terminals. Start by bringing the wires underneath the
screws (see photo below). Tighten the two screws.
(Page 1 of 3)
Leads From
Light Connected
To Sonalert
Light
Battery
Tray
Sonalert
High Quality Groundwater and Surface Water Monitoring Instrumentation
12. Insert the test button through the opening in the faceplate,
and secure it in place using the nut.
13. Insert the sensitivity switch through the faceplate with the
three connection prongs positioned towards the battery
drawer, slightly towards the centre of the reel. Tighten the
nut. Fasten the sensitivity knob on the front of the switch
using the brass screw.
14. Connect the black wire with quick-connect fitting to the
negative terminal on the battery tray. Connect the red wire
with quick-connect fitting to the positive terminal.
Back of Mk1 Water Level Meter Faceplate
Note: Instructions for Mk2 Water Level Meters are on Page 1
(tape connection using push-release fittings on circuit board
and quick-connect fittings to battery tray).
Back of Fully Assembled Water Level Meter Faceplate
(Page 2 of 3)
15. Connect the tape to the new circuit board assembly by
pressing down on the white terminals on the circuit board
and inserting the tape leads. Release the terminals and the
leads should be secured. The lead on the bottom of the tape
(numbers facing up, may be black mark on lead) is inserted
into the terminal with a white square below it on the circuit
board.
16. Install the battery.
17. With the probe in a glass of tap water, turn the Water Level
Meter to the ‘ON’ position. If the connections are correct
the buzzer and light will activate. If the buzzer or light do
not activate, check all connections and the polarity of the
battery.
18. Reattach the faceplate to the reel using the three screws.
Instructions
For Mk1 Water Level Meters (Molex Tape Connection)
Tools and Materials Needed
1. Mk2 Circuit Board Assembly (#102977) Includes:
• Wiring, Sensitivity Switch and Knob, Light and Red
Lens, and Test Button
2. Phillips Screwdriver
3. 10 mm (3/8") Wrench
4. Wire Cutters
5. Small Flat Screwdriver (3 mm (1/10")
6. Pliers
Sensitivity
Switch
Sonalert
Light
Battery
Tray
Test Button
Molex
Connector
1. Remove the battery. Unscrew the three screws and remove
the faceplate.
2. Disconnect the Molex connector that connects the circuit
board to the tape.
3. Cut the red and black wires from the battery tray.
4. From the front of the faceplate, use the wrench to unscrew
the nut holding the test button and push the button back
out of the faceplate.
5. Cut the red and black wires connected to the light, and
push the light out through the front of the faceplate.
Sonalert
Sensitivity
Switch
Light
Battery
Tray Test Button
Push-Release
Fittings
Square
High Quality Groundwater and Surface Water Monitoring Instrumentation
Water Level Meter Circuit Board Replacement
6. Use the small flat screwdriver to unscrew the small brass
screw on the side of the sensitivity knob. Remove the
sensitivity knob. Use the pliers to unscrew the nut holding
the sensitivity switch and remove the switch from the
faceplate. Save the knob and screw to be used with the
new sensitivity switch.
7. Remove the two screws to disconnect the old circuit board
assembly (and the wires connected to the light) from the
Sonalert.
8. Insert the new red lens through the opening in the faceplate
and secure using the white nut.
9. Use the pliers to bend the terminals on the Sonalert, so
they are straight up from the Sonalert (to accommodate
the new circuit board in between the terminals). Loosen
the Sonalert retaining nut on the front of the faceplate, so
you can reposition the Sonalert as shown in the photo on
Page 2 (negative terminal next to the battery tray).
10. Place the circuit board assembly in position on top of the
Sonalert, ensuring the sides marked +ve and -ve are lined
up with the corresponding terminals on the Sonalert and
the light seated in the lens.
11. Install the two screws into the Sonalert terminals. Do not
tighten. Wrap the two bare wires from the circuit board
around the screws, between the screw head and Sonalert
terminals. Start by bringing the wires underneath the
screws (see photo on Page 1). Tighten the two screws.
12. Insert the test button through the opening in the faceplate,
and secure it in place using the nut.
13. Insert the sensitivity switch through the faceplate with the
three connection prongs positioned towards the battery
drawer, slightly towards the centre of the reel. Tighten the
nut. Fasten the sensitivity knob on the front of the switch
using the brass screw.
14. Connect the black wire with quick-connect fitting to the
negative terminal on the battery tray. Connect the red wire
with quick-connect fitting to the positive terminal.
15. Remove the Molex connector by pushing out the two pins.
Cut each pin in half (see image at top right for correct
location). Use pliers to flatten the remainder of the pins
against the tape lead, so it fits easily into the terminals on
the circuit board.
Remove the Pins from the Tape Molex Connector
Cut the Pins on Each Tape Lead in Half (as shown)
16. Connect the tape to the new circuit board assembly by
pressing down on the white terminals on the circuit board
and inserting the tape leads. Release the terminals and the
leads should be secured. The lead on the bottom of the
tape (numbers facing up, may be black mark on lead) is
inserted into the terminal with a white square below it on
the circuit board.
17. Install the new battery.
18. With the probe in a glass of tap water, turn the Water Level
Meter to the ‘ON’ position. If the connections are correct
the buzzer and light will activate. If they do not activate,
check all connections and the polarity of the battery.
19. Reattach the faceplate to the reel using the three screws.
Square
Printed in Canada
March 19, 2018
(#113984)
(Page 3 of 3)
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023
35 Todd Road, Georgetown, Ontario Canada L7G 4R8
Web Site: www.solinst.com E-mail: [email protected]
®Solinst is a registered trademark of Solinst Canada Ltd.

This manual suits for next models

1

Other Solinst Measuring Instrument manuals

Solinst Levelogger 3001 Use and care manual

Solinst

Solinst Levelogger 3001 Use and care manual

Solinst 122M User manual

Solinst

Solinst 122M User manual

Solinst 101 P7 User manual

Solinst

Solinst 101 P7 User manual

Solinst 102 Operating and safety instructions

Solinst

Solinst 102 Operating and safety instructions

Solinst 401 User manual

Solinst

Solinst 401 User manual

Solinst 107 Guide

Solinst

Solinst 107 Guide

Solinst 101 P7 SC1000 User manual

Solinst

Solinst 101 P7 SC1000 User manual

Solinst 101 P2 Operating and safety instructions

Solinst

Solinst 101 P2 Operating and safety instructions

Solinst 101 User manual

Solinst

Solinst 101 User manual

Solinst 101 Operating and safety instructions

Solinst

Solinst 101 Operating and safety instructions

Solinst 101 User manual

Solinst

Solinst 101 User manual

Solinst 102M User manual

Solinst

Solinst 102M User manual

Solinst 104 User manual

Solinst

Solinst 104 User manual

Solinst 122 User manual

Solinst

Solinst 122 User manual

Solinst 101M User manual

Solinst

Solinst 101M User manual

Solinst 122M User manual

Solinst

Solinst 122M User manual

Solinst 101 Operating and safety instructions

Solinst

Solinst 101 Operating and safety instructions

Solinst 107 Mk2 User manual

Solinst

Solinst 107 Mk2 User manual

Solinst 102M User manual

Solinst

Solinst 102M User manual

Solinst 102 User manual

Solinst

Solinst 102 User manual

Solinst 102 Operating and safety instructions

Solinst

Solinst 102 Operating and safety instructions

Solinst 107 Mk2 User manual

Solinst

Solinst 107 Mk2 User manual

Solinst Levelogger Series User manual

Solinst

Solinst Levelogger Series User manual

Solinst 101 Parts list manual

Solinst

Solinst 101 Parts list manual

Popular Measuring Instrument manuals by other brands

Hanna Instruments HI 96813 instruction manual

Hanna Instruments

Hanna Instruments HI 96813 instruction manual

KALAS AK-1097 user manual

KALAS

KALAS AK-1097 user manual

PEAK INSTRUMENTS E-1000 Series User's manual and maintenance guide

PEAK INSTRUMENTS

PEAK INSTRUMENTS E-1000 Series User's manual and maintenance guide

Unipro UniGo One kit 1 user guide

Unipro

Unipro UniGo One kit 1 user guide

Danfoss SonoMeter 30 installation guide

Danfoss

Danfoss SonoMeter 30 installation guide

Amprobe VPC-10 instruction manual

Amprobe

Amprobe VPC-10 instruction manual

Amprobe MAN02-A user manual

Amprobe

Amprobe MAN02-A user manual

TQC POWDERSAVE II user manual

TQC

TQC POWDERSAVE II user manual

Rohde & Schwarz SMB100A operating manual

Rohde & Schwarz

Rohde & Schwarz SMB100A operating manual

V&A M266 Series user manual

V&A

V&A M266 Series user manual

Macherey-Nagel PF-3 manual

Macherey-Nagel

Macherey-Nagel PF-3 manual

ST DELTA SOLAR user manual

ST

ST DELTA SOLAR user manual

CSX P301 instruction manual

CSX

CSX P301 instruction manual

Thermo Scientific Scientific AquaSensors DataStick user guide

Thermo Scientific

Thermo Scientific Scientific AquaSensors DataStick user guide

Gima SP-10 user manual

Gima

Gima SP-10 user manual

ETS-Lindgren EMCenter user manual

ETS-Lindgren

ETS-Lindgren EMCenter user manual

ENMET ISA-M Operation & maintenance manual

ENMET

ENMET ISA-M Operation & maintenance manual

Vortex Ranger 1000 manual

Vortex

Vortex Ranger 1000 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.