Solinst 102 User manual

Instructions
For Mk2 Water Level Meters (Push-Release Cable Connection) 7. Unscrew the two screws holding the circuit board (and leads
from light if you have an older style light – see photo below)
to the Sonalert (watch for proper connection/polarity).
8. If you have an older style light (see photo below) push the
light out of the faceplate.
Back of Mk2 102 Faceplate with Older Style Light
Note: Instructions for Mk1 Water Level Meters are on page 3 (Molex
cable connection, soldered connections to battery tray).
1. Place the reel on a flat surface, with the faceplate up.
Remove the battery from the Meter.
2. Use the Phillips screwdriver to undo the three screws holding
the faceplate to the hub. Remove the faceplate.
3. To disconnect the cable from the circuit board, press
down on the white terminals of the push-release fittings on
the circuit board and pull out to remove the cable leads.
Remember how they were connected.
4. Unscrew the nut from the test button and push the button
out of the faceplate.
5. Use the small flat screwdriver to unscrew the small brass
screw on the side of the sensitivity knob. Remove the
sensitivity knob. Use the pliers to unscrew the nut holding
the sensitivity switch and remove the switch from the
faceplate.
6. Use the Phillips screwdriver to undo the two screws holding
the battery tray to the faceplate. Remove the black and red
wires by pulling the quick-connect fittings off of the terminals
on the battery tray. Remove the battery tray.
Sonalert
Sensitivity
Switch
Light
Battery
Tray
Test Button
102 Replacement Faceplate Assembly
indicating the location of the components
(comes with Handle Assembly attached and 3 Phillips screws)
Tools and Materials Needed
1. Required Replacement Assembly
• Faceplate and Handle (without electronics) (#104251)
2. Phillips or Robertson (Square Head) Screwdriver
3. Small Flat Screwdriver
4. Needle Nose Pliers
Light
Test
Button
Sonalert
Sensitivity
Switch
Battery
Tray
Handle
Push-Release
Fittings
Back of Mk2 102 Faceplate with New Style Light
Sonalert
Sensitivity
Switch
Light
Battery
Tray
Test Button
Push-Release
Fittings
9. Remove the circuit board assembly with connected
wires, test button, sensitivity switch, and new style light if
connected (see photo above).
10. If you have a new style light, unscrew the red lens from the
faceplate.
11. Unscrew the Sonalert retaining ring from the front of the
faceplate and remove the Sonalert.
12. Put the Sonalert through the new faceplate, and secure it
using the retaining ring. Position it as shown in the photos
above, depending on which light you have.
13. If you have a new style light, install the red lens in the new
faceplate and secure it using the white nut.
Leads
From Light
(Page 1 of 3) High Quality Groundwater and Surface Water Monitoring Instrumentation
Water Level Meter Faceplate Replacement
Model 102
14. If you have an older style light (see photo below), thread
the leads of the light assembly through the front of the
faceplate and snap the light in place.
Back of Mk2 102 Faceplate with Older Style Light
Sonalert
Sensitivity
Switch
Light
Battery
Tray
Test Button
Push-Release
Fittings
Back of Mk2 102 Faceplate with New Style Light
Leads
From Light
15. Place the circuit board assembly back in position on top of
the Sonalert, ensuring the sides marked +ve and -ve are
lined up with the corresponding terminals on the Sonalert
and if you have a new style light, it is seated in the lens (see
photo below).
16. Use the two screws to connect the wires from the circuit
board (and light if you have the older style) to the Sonalert.
Make sure the wires are connected to the correct terminals
(+ve and -ve) on the Sonalert.
17. Insert the test button through the opening in the faceplate,
and secure it in place using the nut.
18. Insert the sensitivity switch through the faceplate with the
three connection prongs positioned towards the battery
drawer, slightly towards the centre of the reel. Tighten the
nut. Fasten the sensitivity knob on the front of the switch
using the brass screw.
19. Insert the battery tray through the faceplate and secure
using the screws. Connect the black wire with quick-
connect fitting to the negative terminal on the battery
tray. Connect the red wire with quick-connect fitting to the
positive terminal.
20.To reconnect the cable to the circuit board, press down on
the white terminals and insert the cable leads. Release the
terminals and the leads should be secured. The positive lead
is inserted into the terminal with a square etched below it on
the circuit board. The positive lead has the pin connected to
insulated wire, the negative pin is connected to the braided
wire (see illustration below).
Insulated Wire
(Positive Lead)
Braided Wire
(Negative Lead)
21.Replace the battery. With the probe in a glass of tap water,
turn the Water Level Meter to the ‘ON’ position. If the
connections are correct the buzzer and light will activate. If
the buzzer or light do not activate, check connections and
the polarity of the battery.
22.Screw the new faceplate to the reel using the new Phillips
screws.
Sonalert
Sensitivity
Switch
Light
Battery
Tray
Test Button
Push-Release
Fittings
(Page 2 of 3) High Quality Groundwater and Surface Water Monitoring Instrumentation
Water Level Meter Faceplate Replacement
102 Replacement Faceplate Assembly
indicating the location of the components
(comes with Handle Assembly attached and 3 Phillips screws)
Instructions
For Mk1 Water Level Meters (Molex Cable Connection)
6. Use the small flat screwdriver to unscrew the small brass
screw on the side of the sensitivity knob. Remove the
sensitivity knob. Use the pliers to unscrew the nut holding
the sensitivity switch and remove the switch from the
faceplate.
7. Use the Phillips screwdriver to undo the two screws holding
the battery tray to the faceplate. Cut the red and black wires
from the back of the battery tray close to the terminals (other
end of black wire is connected to the sensitivity switch, and
red wire connects to the circuit board). Remove the battery
tray. Strip the wires about 1/4" (6 mm).
8. Unscrew the Sonalert retaining ring from the front of the
faceplate and remove the Sonalert.
9. Put the Sonalert through the new faceplate, and secure it
using the retaining ring.
10.Replace the sensitivity switch, light, and test button in the
new faceplate.
11.Reconnect the two wires, from the light, using the two
screws through the circuit board and onto the Sonalert.
12.Insert the battery tray through the new faceplate. Re-solder
the red and black wires to the correct terminals labeled on
the back of the battery tray. Red wire from the circuit board
to the positive terminal and black wire from sensitivity switch
to the negative terminal. Use the two screws to secure the
tray to the faceplate.
13.Connect the Molex connector from the faceplate to the cable.
14.Replace the battery. With the probe in a glass of tap water,
turn the Water Level Meter to the ‘ON’ position. If the
connections are correct the buzzer and light will activate. If
the buzzer or light do not activate, check the polarity of the
battery and Molex connector, and the soldered connections.
15.Screw the new faceplate to the reel using the new Phillips
screws.
Tools and Materials Needed
1. Required Replacement Assembly
• Faceplate and Handle (without electronics) (#104251)
2. Phillips or Robertson (Square Head) Screwdriver
3. Small Flat Screwdriver
4. Needle Nose Pliers
5. Wire Strippers and Cutters
6. Soldering Iron and Wire
Light
Test
Button
Sonalert
Sensitivity
Switch
Battery
Tray
Handle
Back of Mk1 102 Faceplate showing wiring connections
and the location of the components
Sonalert
Battery
Tray
Sensitivity
Switch
Test Button
Molex
Connector
Light
1. Place the reel on a flat surface, with the faceplate up.
Remove the battery from the Meter.
2. Use the Phillips or Robertson screwdriver to undo the three
screws holding the faceplate to the hub.
3. Remove the faceplate and disconnect the Molex connector
that attaches the faceplate electronics to the cable.
4. Unscrew the test button and push it out of the faceplate.
5. Remove the two screws which connect the wires to the
circuit board (watch for proper polarity). Push the light out
of the faceplate.
Note: Instructions for Mk2 Water Level Meters are on page 1 (cable
connection on circuit board and quick-connect fittings to
battery tray).
Printed in Canada
March 11, 2018
(#109743)
(Page 3 of 3)
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023
35 Todd Road, Georgetown, Ontario Canada L7G 4R8
Web Site: www.solinst.com E-mail: [email protected]
®Solinst is a registered trademark of Solinst Canada Ltd.

Other Solinst Measuring Instrument manuals

Solinst 101 P2 Operating and safety instructions

Solinst

Solinst 101 P2 Operating and safety instructions

Solinst 101M User manual

Solinst

Solinst 101M User manual

Solinst 101 Operating and safety instructions

Solinst

Solinst 101 Operating and safety instructions

Solinst 122M User manual

Solinst

Solinst 122M User manual

Solinst 615 ML User manual

Solinst

Solinst 615 ML User manual

Solinst 122 User manual

Solinst

Solinst 122 User manual

Solinst 101 Operating and safety instructions

Solinst

Solinst 101 Operating and safety instructions

Solinst 107 Guide

Solinst

Solinst 107 Guide

Solinst 107 Mk2 User manual

Solinst

Solinst 107 Mk2 User manual

Solinst 101 Parts list manual

Solinst

Solinst 101 Parts list manual

Solinst 101B User manual

Solinst

Solinst 101B User manual

Solinst 102 Operating and safety instructions

Solinst

Solinst 102 Operating and safety instructions

Solinst 101 User manual

Solinst

Solinst 101 User manual

Solinst LevelSender 5 User manual

Solinst

Solinst LevelSender 5 User manual

Solinst 101 P2 User manual

Solinst

Solinst 101 P2 User manual

Solinst 9700 SolSat 5 User manual

Solinst

Solinst 9700 SolSat 5 User manual

Solinst 101 P7 User manual

Solinst

Solinst 101 P7 User manual

Solinst 122M User manual

Solinst

Solinst 122M User manual

Solinst 102 Operating and safety instructions

Solinst

Solinst 102 Operating and safety instructions

Solinst 101 Operating and safety instructions

Solinst

Solinst 101 Operating and safety instructions

Solinst 101 User manual

Solinst

Solinst 101 User manual

Solinst 201 User manual

Solinst

Solinst 201 User manual

Solinst 102 User manual

Solinst

Solinst 102 User manual

Solinst 107 Mk2 User manual

Solinst

Solinst 107 Mk2 User manual

Popular Measuring Instrument manuals by other brands

Doculus Lumus Doculus Lumus+ quick guide

Doculus Lumus

Doculus Lumus Doculus Lumus+ quick guide

Triplett MS475 user manual

Triplett

Triplett MS475 user manual

GE DigitalFlow GS868 Service manual

GE

GE DigitalFlow GS868 Service manual

PCB Piezotronics 352A21/NC Installation and operating manual

PCB Piezotronics

PCB Piezotronics 352A21/NC Installation and operating manual

Hach BioTector B7000i user manual

Hach

Hach BioTector B7000i user manual

SSI DPL4000 Operation manual

SSI

SSI DPL4000 Operation manual

vissonic VIS-CRS02-B user manual

vissonic

vissonic VIS-CRS02-B user manual

MICRO-EPSILON surfaceCONTROL 3D SC3200-80 operating instructions

MICRO-EPSILON

MICRO-EPSILON surfaceCONTROL 3D SC3200-80 operating instructions

Precision Digital Corporation Helios PD2-6001 instruction manual

Precision Digital Corporation

Precision Digital Corporation Helios PD2-6001 instruction manual

Hach DR 5000 instruction sheet

Hach

Hach DR 5000 instruction sheet

LaserLiner Quadrum OneTouch Green manual

LaserLiner

LaserLiner Quadrum OneTouch Green manual

Fluke 435-II Service manual

Fluke

Fluke 435-II Service manual

RHS 14ESS quick start guide

RHS

RHS 14ESS quick start guide

Smart Sensor AR856 instruction manual

Smart Sensor

Smart Sensor AR856 instruction manual

CAISSON LVT 15 quick start guide

CAISSON

CAISSON LVT 15 quick start guide

Hitachi L44K instruction manual

Hitachi

Hitachi L44K instruction manual

Flowserve IPS-Beacon VB-105 Installation operation & maintenance

Flowserve

Flowserve IPS-Beacon VB-105 Installation operation & maintenance

Extech Instruments HDV7C-A2-39-HD-1 user manual

Extech Instruments

Extech Instruments HDV7C-A2-39-HD-1 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.