TriBrer TB-6 Series User manual

Optical Fiber Instrument/OTDR
TB-6XX
User's Guide

Any undefined changeor modification ofthis manual willdeprive you ofthe right tooperate
the equipment.
Toreduce the riskof fire orelectric shock, donot expose theequipment to rainor humidity.
Toprevent electric shock,please do notopen the shell,and it mustbe repaired byqualified
personnel.
Please ensure nosignal in fiberbefore testing,active fibermay damage thedevice and not
in warranty range.
NOTENOTE
As the laseris harmful tothe eyes, don'tlook directly atthe laser outletand don't attemptto
disassemble the cabinet.
PRECAUTIONS FOR USE
The content of this manual is for reference only, and everything is based on the
actual product.
Using the battery:
The device can be recharged with a special battery, and can not be mixed with
different types or capacity batteries.
Avoiding condensation:
Sudden changes in temperature should be avoided. Do not use the device
immediately after moving the device from the cold area to the hot area, or when
the room suddenly heats up, because the device may have condensation
phenomenon. If the temperature changes abruptly , stop using it and take out
the battery, and the power can be switched on after at least an hour.
Storage:
When the device is not used for a long time, please take out the battery to avoid
the damage caused by battery leakage .
WARNINGWARNING
1

567
1 2 3 4
8
9
2
1OTDR
2VFL
3TF(Micro SD) card slot
4OPM
5Charge indicator
83.2 inch LCD
6Reset key
9Function buttons
7USB interface
Host

SETTING MARKER SAVE/FILE
ESC
ENTER
+ESC=FULL
SCAN
3
Setting:In OTDR and iOLA interface, press to set up.
Zoom control:Operate with direction and ESC key for waveform.
Power:Long press to turn on, or choose to turn off.
Direction:Select up, down, left, and right.
Marker:Switch cursor in OTDR interface.
Save/File:In OTDR and iOLA interface, press to save or enter the file.
Scan:In OTDR and iOLA interface, press to measure.
Esc:Return to the previous menu.
Enter:Enter the next level page or confirm the operation.
Buttons

4
Enter the main menu after booting. There are up to five modules, select by the
direction buttons, and then press the " " button to enter the corresponding
functional.
Mini OTDR
OPM/VFL
System
2018-03-28 WED 21:25
OLS
USB TF card
Bluetooth(optional)
2018-03-28 WED 21:25
OPM/VFL
System
Mini OTDR
2018-03-28 WED 21:25
OLS
Title Bar
Highlight when the function module is selected
Date and time
Battery(charging)
ENTER
Main Menu
iOLA iOLA

ENTER
Event list
Measurement info
Waveform region
5
OTDR
Mini OTDR
Select Mini OTDR module in the main menu,and press to enter the
OTDR interface.
Scale
File name
(Measurement warning)
The distance
and attenuation
File Info: 1550nm 20km 250ns
1234_0001.sor
NO. Dis.
Loss
Ref.
dB/km C.Loss
1/3
3/3
0.000
14.154
0.00
--.--
-34.04
-20.95 0.18
--.-- 0.00
5.20
A-B: 10.057km 1.81dB --.--dB/km
A 0.000km
5.00dB/div
XY
2.000km/div
2/3 10.057 2.65 -36.08 0.18 1.81
1
2018-03-28 WED 21:25
B 10.057km
23

+ESC=FULL
ESC
+Full-screen display
+ESC=FULL
+ESC=FULL
+
+Vertical reduction
Horizontal reduction
+ESC=FULL
+ESC=FULL
+
+
Vertical magnification
Horizontal magnification
6
A 0.000km
B 10.057km
23
The current waveform can be scaled after the measurement or the files is
opened.
OTDR Waveform Zoom in/out
File Info: 1550nm 20km 250ns
1234_0001.sor
NO. Dis. Loss Ref.
dB/km
C.Loss
1/3
3/3
0.000
14.154
0.00
--.--
-34.04
-20.95 0.18
--.-- 0.00
5.20
A-B: 10.057km 1.81dB --.--dB/km
A 0.000km
5.00dB/div
XY
2.000km/div
2/3 10.057 2.65 -36.08 0.18 1.81
1
2018-03-28 WED 21:25
B 10.057km
23
A 0.000km
1
B 10.057km
23
Vertical scaling
Horizontal scaling

7
Measure Info: 1550nm 20km 250ns
*
A-B: 13.820km 12.04dB 0.87dB/km
A 0.000km
B 13.820km
4.00dB/div
XY
0.200km/div
2018-03-28 WED 21:25
INVISIBLE LASER RADIATION
AVOID EYE OR SKIN EXPOSURE
TO DIRECT OR SCATTERED RADIATION
EXCIMER LASER
10mW MAX PULSE at 800-1700nm
DANGER Laser radiation
Do not look directly
to the laser
Real time scanning...
Loss
Ref.
dB/km C.Loss
NO. Dis.
Measure Info: 1550nm 20km 250ns
*
A-B: 13.820km 12.04dB 0.87dB/km
A 0.000km
B 13.820km
4.00dB/div
XY
0.200km/div
2018-03-28 WED 21:25
10s
INVISIBLE LASER RADIATION
AVOID EYE OR SKIN EXPOSURE
TO DIRECT OR SCATTERED RADIATION
EXCIMER LASER
10mW MAX PULSE at 800-1700nm
DANGER Laser radiation
Do not look directly
to the laser
Loss
Ref.
dB/km C.Loss
NO. Dis.
OTDR Measure
Press the " " button torun real-time measurement. If the Real time
analyse function isactivated, press again to run average measurement,the
progress bar isdisplayed at the bottom.Press the " " button inOTDR
Interface,Scan type canbe switched from "Real time" to "Average" .
SCAN
SETTING

8
OTDR Event List
The event list shows up to 5 events, which can be viewed by scrolling up and down.
File Info: 1550nm 20km 250ns
1234_0001.sor
NO. Dis. Loss Ref.
dB/km C.Loss
1/3
3/3
0.000
14.154
0.00
--.--
-34.04
-20.95 0.18
--.-- 0.00
5.20
A-B: 10.057km 1.81dB --.--dB/km
A 0.000km
5.00dB/div
XY
2.000km/div
2/3 10.057 2.65 -36.08 0.18 1.81
1
2018-03-28 WED 21:25
B 10.057km
23
Optic fiber start
Reflection event
Attenuation event
Five event types:
Gain event
Optic fiber ending

Press key ENTER to edit settings.
Wavelength
Pulse width
Average time
Reflection threshold
Scatter coefficient
End threshold
Optical detector
Refractive rate
Real time analyse
1550nm
Manual
250ns
15s
65.0dB
52.1
3.0dB
1.46832
Measure Settings
Real time
Scan mode
Scan type
2018-03-28 WED 21:25
Range 2.5km
Splice loss 0.05dB
9
OTDR Measure Settings
Press the " " button to pop up the setting box, and the switch setting will
switch directly. Multi-digit setting frame, after positioning the cursor with the left and
right buttons, select up and down.
ENTER
Press key ENTER to edit settings.
Wavelength
Pulse width
Average time
Reflection threshold
Scatter coefficient
End threshold
Optical detector
Refractive rate
Real time analyse
1550nm
Manual
250ns
15s
65.0dB
52.1
3.0dB
1.46832
Measure Settings
Real time
Scan mode
Scan type
2018-03-28 WED 21:25
Range 2.5km
Splice loss 0.05dB
.00
5
116
994

10
OTDR Save
After finishing measurement, press the " " button to save the file and
the keyboard will pop up. If the Auto Name function is turned on, the file name is
automatically generated when saving.
SAVE/FILE
: 1550nm 20km 250ns
1234_0001.sor
1/3
3/3
0.000
14.154
0.00
--.--
-34.04
-20.95 0.18
--.-- 0.00
5.20
A-B: 10.057Km 1.81dB --.--dB/Km
A 0.000km
5.00dB/div
XY
2.000Km/div
2/3 10.057 2.65 -36.08 0.18 1.81
1
2018-03-28 WED 21:25
B 10.057km
23
Save SOR
2
w
-
3
e
#
4
r
5
t
6
y
7
u
8
i
9
o
0
p
as
z
d
x
f
c
g
v
h
b
j
n
k
m
l
1
q
.Space Enter
025404

11
OTDR Files
When there is no measurement, press the " " button to enter the file list.
Press the " " button to open the folder or file, and press the " " button to pop
up the operation item. The waveform thumbnail is displayed at the bottom.
SAVE/FILE
ENTER
Press key RIGHT to operate.
2018-03-25
2018-03-25
2018-03-25
2018-03-25
2018-03-25
2018-03-25
2018-03-25
2018-03-25
0325_0001.sor
0325_0002.sor
0325_0004.sor
0325_0005.sor
0325_0006.sor
0325_0007.sor
0325_0008.sor
0325_0009.sor
. .
Files
Delete Rename
2018-03-25
0325_0003.sor
2018-03-28 WED 21:25
Wavelength: 1550nm
Range: 2.5km
Pulse width: 250ns
Press key RIGHT to operate.
2018-03-25
2018-03-25
2018-03-25
2018-03-25
2018-03-25
2018-03-25
2018-03-25
2018-03-25
0325_0001.sor
0325_0002.sor
0325_0004.sor
0325_0005.sor
0325_0006.sor
0325_0007.sor
0325_0008.sor
0325_0009.sor
. .
Files
Delete Rename
2018-03-25
0325_0003.sor
2018-03-28 WED 21:25
Wavelength: 1550nm
Range: 2.5km
Pulse width: 250ns
Press key RIGHT to operate.
20180327
20180326
20180324
20180323
20180322
20180328
Files
20180325
2018-03-28 WED 21:25
Press key RIGHT to operate.
20180327
20180326
20180324
20180323
20180322
20180328
Files
20180325
2018-03-28 WED 21:25

12
iOLA(optional)
iOLA
Select iOLA module in the main menu,and press to enter the iOLA
interface,then press to run measurement,the progress bar is displayed
in the middle .The measurement end interface is shown as follows.
ENTER
Ref.
0.0102
0.0000
0.0306
0.0409
Total loss
1310nm Pass or not
0.000 Not Pass
0.0102 0.0306A0.0204 0.0409
AB
0.0621
2018-03-28 WED 21:25
Type Dis.
No. Loss
1/5
2/5
4/5
5/5
AB
0.0409km
0.0204
-62.012
-53.680
-35.194
-49.564
-62.950
2.099
2.235
---.---
0.000
2.345
2018-03-28 WED 21:25
3/5 Event list
Linear view of
event list
(Measurement warning)
Visual event range
Total loss
SCAN

13
Press " " in the iOLA interface to enter the iOLA parameter setting
interface,and press the " " button to pop up the setting box,Multi-digit setting
frame, after positioning the cursor with the left and right buttons, select up and down.
ENTER
iOLA Measure Settings(optional)
Press key ENTER to edit settings.
Loss threshold
Reflection threshold
20.000dB
0.300dB
40.000dB
iOLA Settings
15.000dB
Totalloss threshold
ORL threshold
2018-03-28 WED 21:25
Pass/Not Pass Settings
Splice loss
Reflection threshold
End threshold
Refractive rate
Scatter coefficient
0.05dB
65.0dB
3.0dB
1.46770
49.6
Measure Settings
Reflection threshold 0.750dB
Press key ENTER to edit settings.
Loss threshold
Reflection threshold
20.000dB
0.300dB
40.000dB
iOLA Settings
15.000dB
Totalloss threshold
ORL threshold
2018-03-28 WED 21:25
Pass/Not Pass Settings
Splice loss
Reflection threshold
End threshold
Refractive rate
Scatter coefficient
0.05dB
65.0dB
3.0dB
1.46770
49.6
Measure Settings
Reflection threshold 0.750dB
.0+7
50
1861
9-649
SETTING

14
After finishing measurement, press the " " button to save the file and
the keyboard will pop up. If the Auto Name function is turned on, the file name is auto
matically generated when saving.
SAVE/FILE
iOLA Save(optional)
Total loss
1310nm Pass or not
0.000 Not Pass
AB
2018-03-28 WED 21:25
AB
0.0409km
2018-03-28 WED 21:25
2
w
-
3
e
#
4
r
5
t
6
y
7
u
8
i
9
o
0
p
as
z
d
x
f
c
g
v
h
b
j
n
k
m
l
1
q
.Space Enter
025404

15
iOLA Files(optional)
When there is no measurement, press the " " button to enter the file list.
Press the " " button to open the folder or file, and press the " " button to pop
up the operation item.
SAVE/FILE
ENTER
Press key RIGHT to operate.
2018-03-25
2018-03-25
2018-03-25
2018-03-25
2018-03-25
2018-03-25
2018-03-25
2018-03-25
0325_0001.sor
0325_0002.sor
0325_0004.sor
0325_0005.sor
0325_0006.sor
0325_0007.sor
0325_0008.sor
0325_0009.sor
. .
Files
Delete Rename
2018-03-25
0325_0003.sor
2018-03-28 WED 21:25
Wavelength: 1550nm
Range: 2.5km
Pulse width: 250ns
Press key RIGHT to operate.
20180327
20180326
20180324
20180323
20180322
20180328
Files
20180325
2018-03-28 WED 21:25
Press key RIGHT to operate.
180325_0001.isor
180325_0002.isor
180325_0004.isor
180325_0005.isor
180325_0006.isor
180325_0007.isor
180325_0008.isor
180325_0009.isor
180325_0010.isor
180325_0011.isor
. .
Files
Delete Rename
180325_0003.isor
2018-03-28 WED 21:25
Press key RIGHT to operate.
20180327
20180326
20180324
20180323
20180322
20180328
Files
20180325
2018-03-28 WED 21:25

16
OPM / VFL
Operate according to the prompts.
dB
nw
0.61
-2.78
Laser radiation! Do not stare into beam!
1550nm Auto
Switching
wavelength
VFL on or off
View REF
REF setting
CAUTION
DO NOT STARE INTO THE BEAM OR VIEW
DIRECTLY WITH OPTICAL INSTRUMENTS
LASER RADIATION
635-650nm LASER
CLASS 1 LASER PRODUCT
Blink
2018-03-28 WED 21:25
REF:-70.00dBm
2KHz
OPM/VFL
Note:when the wavelength recognition function is on and the carrier
wavelength is displayed,the button cannot successfully switch the wavelength
(return after switching for 2 seconds)
Press key ENTER to edit settings.
Wavelength recognition
OPM setting
Frequency identification
2018-03-28 WED 21:25
Press the key to enter the OPM
setting interface.
SETTING
Select OPM/VFL module in the main menu,and press to enter the
OPM/VFL interface.
ENTER

17
OLS
Operate according to the prompts.
Wavelength:1310nm
Modulation:1kHz+Blink
Switching
wavelength
Change
modulation
ON/OFF
2018-03-28 WED 21:25
CLASS 1M LASER PRODUCT
DO NOT VIEW DIRECTLY
WITH OPTICAL INSTRUMENTS
LASER RADIATION
OLS
Wavelength:1310nm
Modulation:1kHz+Blink
2018-03-28 WED 21:25
Blink
CLASS 1M LASER PRODUCT
DO NOT VIEW DIRECTLY
WITH OPTICAL INSTRUMENTS
LASER RADIATION
ON/OFF
Switching
wavelength
Change
modulation
Operate according to the prompts.
Select System module in the main menu,and press to enter the
System interface.When the Change modulation is selected as 1kHz+Blink or
2kHz+Blink ,if the light source is turnded on,the light source flag will flicker.
TB600 is a single wavelength,and Switching wavelength on the interface will be gray.
ENTER

Press key ENTER to edit settings.
Auto power off
About
Factory settings
10Min
System Settings
80
Auto name
Brightness
2018-03-28 WED 21:25
Voice
Bluetooth
USB connection
OTG
Date 2018-03-28
Time 21:25
Format storage device
Time:
Format Storage Device:
Delete all files in the current TF card.
Date:
Bluetooth:
Voice:
Set brightness of LCD,settings range
Brightness:
Name automatically when file is saving,
Auto Name:
Set time of automatic off or cancel this
Auto Power Off:
USB Connection:
Connect USB status to view TF card or
OTG:
Connect USB status settings to view TF
System
About:
View the device hardware/software version
Factory Settings:
Is used in resetting to default ,and has no
effect on the set time and stored data.
18
System
Select System module in the main menu,and press to enter the
System interface.
ENTER
number,CPU number and Language.
internal storage File.
Function.
can cancel this function.
from 10 to 100.
Set whether to open or not.
Set whether to open or not.
Set year, month, day.
Set hour, minute.
card files or not.

PD_007#0004 Rev.200420
WWW.TRIBRER.COM
Made In China
Table of contents
Other TriBrer Measuring Instrument manuals

TriBrer
TriBrer AOU400 User manual

TriBrer
TriBrer EPN70 User manual

TriBrer
TriBrer EVA50-40 User manual

TriBrer
TriBrer AOP100T User manual

TriBrer
TriBrer AOP110 User manual

TriBrer
TriBrer AOF500 User manual

TriBrer
TriBrer FOT-100 User manual

TriBrer
TriBrer AOP100 User manual

TriBrer
TriBrer Lemon 5 User manual

TriBrer
TriBrer EPN90 User manual
Popular Measuring Instrument manuals by other brands

Keller
Keller CellaTemp PK Series quick start guide

EUTECH INSTRUMENTS
EUTECH INSTRUMENTS CYBERSCAN PH 1100 2100 PHION METER instruction manual

General tools
General tools DAF80PWM Operation manual

S+S Regeltechnik
S+S Regeltechnik THERMASGARD RGTM1-I Operating Instructions, Mounting & Installation

PQ Plus
PQ Plus UMD 705 Quick start manual

HT
HT HTA103 user manual