
EN 9
pyrometer TP10
Setting the Max/Min function
1. Repeatedly press the Modebutton(3) until the MxMn Save
Avg menu is displayed:
2. Press the F1button(4) to select MxMn.
ðThe maximum value is displayed in the Upper measured
value indication(18).
ðThe minimum value is displayed in the Lower measured
value indication(20).
ðThe discrepancy between the current measured value
and the minimum value is indicated as deflection on the
scale(21).
Setting the Avg/Dif function
Please bear in mind that the average value is continually
determined anew during a measurement, which means that
there can be deviations of the displayed Difvalue in the range of
hundredths of a °C.
1. Repeatedly press the Modebutton(3) until the MxMn Save
Avg menu is displayed.
2. Press the F3button(2) to select Avg.
ðThe average value is displayed in the Upper measured
value indication(18).
ðThe difference between the current measured value and
the average value is displayed in the Lower measured
value indication(20).
ðThe difference between the current measured value and
the average value is indicated on the scale(21).
Using the Save function
1. Repeatedly press the Mode button(3) until the MxMn Save
Avg menu is displayed.
2. Press the F2button(5) to select Save.
ðThe Yes and Esc menus are shown on the display.
3. Press the F1button (4, Yes) to save the measured value.
ðIt takes a short moment until the measured value will be
saved. The Savingindication(17) lights up on the
display. After that the last measured value will be
displayed again.
4. Press the F3button(2, Esc) to cancel the operation.
ðThe last measurement will be displayed.
Using the Memory function
Using the Memory function you can call up values previously
saved with the Save function. The device comes with 30 internal
storage locations.
1. Repeatedly press the Modebutton(3) until the Unit Mem Ɛ
menu is displayed.
2. Press the F2button(5).
ðThe last saved value appears on the display.
ðThe Set menu will be displayed.
3. Either press the F3button(2) or the F1button(4) to view
the saved values.
4. Press the F2button(5) to select Del.
ðThe Yes All Esc menu is displayed.
5. If you want to delete a measured value (Yes), press the
F1button(4).
ðThe currently displayed measured value will be deleted.
6. If you want to delete all measured values (All), press the
F2button(5).
ðAll saved measured values will be deleted.
7. If you want to cancel the operation (Esc), press the
F3button(2).
ðThe Del menu is shown on the display.
8. Press the Modebutton(3) to return into measuring mode.
Setting the temperature unit
1. Repeatedly press the Modebutton(3) until the Unit Mem Ɛ
menu is displayed.
2. Press the F1button(4).
ðThe Unit °C °F menu is shown on the display.
3. Press the F2button(5) for °C or the F3button(2) for °F.
Setting the emissivity
1. Repeatedly press the Modebutton(3) until the Unit Mem Ɛ
menu is displayed.
2. Press the F3button(2).
ðThe last saved value for the emissivity appears on the
display.
ðThe Tab menu will be displayed.
3. Either press the F3button(2) or the F1button(4) to
change the emissivity.
ðThe value range is between0.10 and1.00.
ðThe selected setting will be saved automatically.
4. Press the F2button(5) to select a specific emissivity.
ðThe Tab menu will be displayed.
ðA list with different emissivities will be displayed:
Default 0.95
0x Aluminium 0.30
0x Brass 0.50
0x Copper 0.60
Paint 0.93
5. Either press the F3button(2) or the F1button(4) to select
a specific emissivity.
6. Confirm your selection with the F2button(5).
ðThe new emissivity is set.
7. Press the Modebutton(3) to exit the menu.
Info
A set specific emissivity will not be stored in the device
after it has been switched off. When being switched on
again, the device will always return to the emissivity
previously set by hand.