SYMPTOMSCAUSES&REMEDY
Nodisplay lightson1.AJDPboardis defective.
2.ACadaptoris defective.
3.Wrongconnectionofconnection cords inside.
4.Built-inbattery(option)is discharged.
q-nor-n
appears selftestof
segments
1.Wrongweighingpanis applied..
2.Tuning-fork sensorormechanismunitis defective.
3.AJDPboardis defective.
4.Settingofaddress data hasmistake.
5.Coefficientmemories(address data)have changed by
noisesorstatic electricity.Adjustlinearity
Display dose notget
settled down.
Display dose not
repeatcorrectly.
Zeropointdrifts.
1.Someparts such as stoppertouchesothers.
2.Weighingpantouchesotherparts.
3.Foreignsubstance is inthe scale.
4.Tuning-fork sensorormechanismunitis defective.
5.Affected by a windordisturbingoscillation.Check
environmentorworkingbase.Checkalsosettingof
stabilizationtime.See page**ofoperationmanual.
-n appears
withanetloadless
thanspecified
capacity(F.S.+9d).
1.Gross weight appliedtothe scalepan(netweight+tare
value)exceedsthe scale capacity.
2.Coefficientmemories(address data)have changed by
noisesorstatic electricity.Adjustlinearity.
3.Settingofaddress data hasmistake.
4.Calibrationweightis notcorrect.
Spanisoutof
specifiedrange.1.Tuning-fork sensorormechanismunitis defective.
2.AJDPboardis defective.
3.Settingofaddress data hasmistake.
4.Coefficientmemories(address data)have changed by
noisesorstatic electricity.
Linearityisoutof
specifiedrange.1.Tuning-fork sensorormechanismunitis defective.
2.Settingofaddress data hasmistake.
3.Coefficientmemories(address data)have changed by
noisesorstatic electricity.
4.Calibrationweightis notcorrect.
3-2 TROUBLESHOOTINGTABLE