D e t e c t i on
www.watchgas.eu
5
3. Operating the PDM Mono Dock
Powering the PDM Mono Dock
The PDM Mono Dock can be powered by 6X 1.5V AA alkaline batteries or by applying a 12V dc external
power supply.
Tuning on the PDM Mono Dock
Turn on the PDM Mono Dock by pressing either the Calibration or Bump button. The PDM Mono Dock
will automatically wake up out of sleep mode which gets activated when not used. Through the use of
this sleep mode up to 1600 tests can be performed on 6X AA batteries.
Calibration Gas
Use the PDM Mono Dock with certied valid calibration gas. Use a demand ow regulator connect the
regulator with the gas sampling hose to the gas inlet right side of the case.
By default the PDM Mono Dock is programmed to use default gas concentration, as span gas for your
detector. If you have a bottle with a dierent concentration please change this setting to the right
values. Chapter Load PDM Mono Dock Conguration
Default value span gas:
O2: 18%
CO: 50ppm
HS: 10ppm
SO: 10ppm
NH3: 50ppm
4. Preform Bump/Calibration
Bump test
Place an turned on warmed up detector into the detector bay.
Ensure that Gas concentration in gas bottle matches with the gas bottle setting in the cong.ini
Press the Bump button.
Now the LED around the button will glow blue, when nished the LED will glow green.
If an error occurs the LED will blink the error code in RED. (see Error Codes for explanation)
If error stays existing after a retry try to determinate your error with chapter troubleshooting.
Calibration
Place an turned on warmed up detector into the detector bay.
Ensure that Gas concentration in gas bottle matches with the gas bottle setting in the cong.ini
Press the Cal button.
Now the LED around the button will glow blue, when nished the LED will glow green.
If an error occurs the LED will blink the error code in RED. (see Error Codes for explanation)
If error stays existing after a retry try to determinate your error with chapter troubleshooting.
SD Log
After testing an detector successful or fail, the PDM Mono Dock writes the data onto the SD card.
Bump, Calibration and event log of the tested instrument will be saved.