Description
The PDM is a portable multigas monitor designed to warn the user about the presence of dangerous
concentrations of oxygen, combustible, or toxic gases in the atmosphere. It is simple to operate. The
PDM alerts the workers to danger through acoustic alarm, ashing LEDs, and vibration when the
concentration exceeds the safety gas levels. The PDM shows the gas concentration in real time and
identies the maximum and minimum concentration. The settings values can be modied through the
optional WatchGas IR-Link.
Warning
• Any unauthorized attempt to repair or modify the product and any cause of damage beyond the
intended use voids any warranty and liability.
• Activate the product only if sensor, display, and beeper port are clear from contaminants, dirt, dust
and debris. Remove any debris on the surfaces of the sensor, LED or beeper port before use.
• Test the performance of the gas sensor regularly.
• Test the products functioning of LEDs, alarm and vibration on a regular basis.
• Use the product under the conditions instructed, including the temperature, humidity and
pressure range. Use in other conditions may cause malfunction or failure.
• The sensor inside the device may indicate the gas concentration dierently according to the
environment such as temperature, pressure and humidity. Calibrate the detector under the same
or similar conditions in which it will be used.
• Extreme changes in temperature may cause drastic changes of the gas concentration. (e.g. using
the detector where there is a huge gap between the inside and outside temperature). Wait for the
concentration to become stable before use.
• Severe pressure or impact may cause drastic changes of the gas concentration. Therefore, use the
device when the concentration is stable. Severe pressure or impact may cause malfunction in the
sensor or the device.
• The alarms are set according to international standards and must be changed by an authorized
expert.
• Do not attempt to replace the battery and sensor. The PDM is a disposable product.
• IR communication should be done in a safe area where there is no risk of explosion or re.
• Incorrectly changing settings in the available software with IR-link may pose a risk, which can
result in trauma or death.
Caution
• Only use the device after reading and understanding this manual!
• This product is not a measurement instrument, but a gas detector.
• Cease use and consult the manufacturer if the calibration or self-test fail continuously.
• Perform a fresh air calibration in clean air without gases at least every 30 days.
• Do not activate the product past the activation date.
• Clean the exterior of the device with soft cloth and do not clean it with chemical detergent.