Instructions for Operation and Care of Thermo-Therapy Dry Heat Therapy
9
Please refer the photos on page 4 for TT-101, page 5 for
TT-101L and page 6 for TT-202 .
Thermo-Therapy Units will only operate with the power
switch on and the treatment timer activated.
Rotate the Celstim agitation or speed control(s) clockwise
to increase air speed and ow. Rotate the controls counter
clockwise to decrease the air speed and ow. On the
TT-202, speed or agitation is normally balanced between
the two blowers- but can be adjusted differently to effect
varying treatment modes.
Heat is also increased by rotating the heat control knob
clockwise- from “A” lowest heat to “K” highest heat.
Treatment time is indicated as minutes with 20 minutes
highlighted as being the usual treatment cycle.
All Thermo-Therapy Units are temperature safety switched
at approximately 125° F (51.7°C) and if an operating
thermometer ever indicates temperature higher than this,
the unit must be referred to a qualied repair service or
returned to Whitehall Manufacturing.
In an effort to best prolong the motor life, the unit must
be turned off between treatments. DO NOT LEAVE
THE UNIT ON CONTINUOUSLY. Once the Celstim is
at operating temperature and the blowers are turned
off, heat loss is approximately 2-3°F (0.55-1.65°C)
per hour.
Make sure the body part being treated is as dry as possible
as moisture in the Celstim will cause “caking”.
Also, be careful at all times, to tightly close the sleeves at
the treatment ports so minimal amount of Celstim escapes
from the unit.
MAINTENANCE OF THE UNIT
The most important maintenance function on all
Thermo-Therapy products is to make sure to keep all
the incoming air lters clean and unclogged. If this is
not done, airow will be restricted resulting in less agitation,
over-heated motors, and premature motor failure. We have
enclosed a spare lter so one lter can always be cleaned
and on standby, and a proper wrench to remove the lter
cover.
The housing can be cleaned with a mild household cleaner
without damage as its surface is similar to that found on
kitchen counters and cabinets.
The treatment port sleeves can be washed and replaced.
To remove the sleeves, carefully pull the bottom of the rims
toward the center away from the side and at the same time
pull or push them up and out of the housing. Then pull the
sleeves from the Velcro found underneath the rim. Reverse
the procedure to replace the sleeves. If the sleeve or rims
become worn, replacements are readily available from
Whitehall Manufacturing or your therapy products dealer.
On the TT-101 and TT-101L models, remove the outer
cover and remove and clean the primary lter. If necessary,
remove the four screws to access the secondary lters.
Replacements are readily available.
On the TT-202, unscrew and remove the lter covers located
on the lower rear panel and remove the lters. Again, the
lter media can be washed and dried and reinstalled or a
replacement can be obtained. A hex or Allen wrench has
been provided for removing the lter cover.
Pay particular attention to the air inlet on the TT-101 and
TT-101L since it is on the units’ bottom and close to whatever
surface it is placed upon. Also make sure this surface is
rigid as the blower will pull up any exible material and
cause the blower to labor.
The Celstim, after prolonged use, may become too ne
for effective therapy and can be replaced by removing the
top sleeve and “digging” it out or by removing the rear top
panel. If the panel is removed, be prepared to contain any
spillage. Also, make sure the panel is tightly reinstalled to
prevent any Celstim from escaping. A “hex” wrench has
been provided for removing and replacing the screws from
the top panel.
Sterilized Celstim, with nes removed, for replacement or
addition is readily available in a 10 lb. (4.5 kg) box containing
two (2) 5 lb. (2.25 kg) plastic “jugs” for easy adding.
Too much Celstim in the treatment chamber can cause a
“rolling” action- remove the excess Celstim through a top
port.
Too little Celstim results in a perceived loss of agitation.
Normal capacity is as follows:
TT-101 – 10 lbs. (4.5 kg)
TT-101L – 12 lbs. (5.4 kg)
TT-202 – 40 lbs. (18 kg)
INSTRUCTIONS FOR USE AND
OPERATING CONTROLS