3M Clean-Trace NGi Luminometer User manual

UserManual
3M™Clean-Trace™ NGi Luminometer
and
Docking Station

UserManual 1
3M™ Clean Trace™ NGi Luminometer and DockingStation
INDEX
1.0PRODUCTDESCRIPTION
1.01How the 3M™Clean-Trace™ NGi Luminometer works to MeasureContamination
1.02ATP and Surface HygieneTesting.
1.03ATP and WaterTesting.
2.0INSTRUMENTOVERVIEW
2.01Parts of the 3M™Clean-Trace™ NGiLuminometer
2.02GeneralFeatures
2.03Charging theBattery
2.04Switching the InstrumentON
2.05MeasurementModes
2.06Inserting aTest
2.07SwabSensor
2.08Low PowerMode
2.09PCCommunications
3.0ACCESSORIES
3.013M™Clean-Trace™ NG Luminometer DockingStation
3.023M™Clean-Trace™ NG Luminometer Soft CarryCase
4.0MEASURESAMPLE
4.01What is MEASURESAMPLE
4.02How to Test Using MEASURESAMPLE
5.0PROGRAMMODE
5.01What is ProgramMode
5.02How to Use ProgramMode
5.03Retests
5.04Partial ResultSets
5.05Reviewingresults
6.0SETUPMENU
6.01What the SETUP MENUallows
6.02How to find the SETUPMENU
6.03Setting DateFormat
6.04Setting theDate
6.05Setting theTime
6.06Setting theLanguage
6.07Performing a SelfCheck
6.08Setting the BacklightTimeout
7.0SYNCHRONISE NGi
7.01What is SYNCHRONISE NGi
7.02How to SYNCHRONISE NGi
8.0MANAGING MULTIPLEORGANISATIONS
9.0SCHEDULEDSAMPLING
9.01Using ScheduledSampling
9.02Retests
9.03Partial ResultSets
9.04ReassigningSchedules
10.0FLEXIBLEFIELDS
10.01Creating a New ResultSet
10.02Updating an Existing ResultSet
11.0USERGUIDELINES
12.0CARE OF THE 3M™CLEAN‑TRACE™ NGiLUMINOMETER
12.01Cleaning
12.02Temperature
12.03Battery Guidelines
13.0TROUBLESHOOTING
13.01General
13.02Battery LowScreen
13.03Communication ErrorScreens
13.04Self Check ErrorScreen
13.05Other ErrorScreens
13.06Inadvertent Changing of LanguageSetting
14.0APPENDICES
14.01TechnicalSpecifications
14.02Terms andConventions

1.0PRODUCTDESCRIPTION
1.01How the 3M™Clean‑Trace™ NGi Luminometer works to MeasureContamination
1.02ATP and Surface HygieneTesting
1.03ATP and Water Testing
1.01How the 3M™Clean‑Trace™ NGi Luminometer works to MeasureContamination
The 3M™Clean-Trace™ NGi Luminometer is a luminometer used in conjunction with 3M reagent kits to measure levels of contamination
on surfaces and in water samples. The technology used is Adenosine Triphosphate (ATP) bioluminescence. ATP is a substance present in all
animal and vegetable matter, including most food debris, bacteria, fungi and other micro-organisms. Measurement of ATP is performed using
an enzyme reaction which occurs naturally in the tail offireflies.
Luciferin/Luciferase (firefly reagent) + ATP =Light
The light emitted is in proportion to the amount of ATP. The intensity of the light emitted from a sample is measured by the
3M™Clean-Trace™ NGi Luminometer and is displayed in Relative Light Units(RLU).
1.02ATP and Surface HygieneTesting.
The 3M™Clean-Trace™ NGi Luminometer, together with the appropriate test kit, is an effective method for monitoring the hygiene status of
surfaces. ATP from both micro-organisms and product residues are detected. Measurement of “total ATP” provides an important insight to the
overall cleanliness. Product residues remaining on surfaces after cleaning will be a source of nutrients for any remaining micro-organisms
and may also protect the micro-organisms from disinfectant action. Because of the fast results, use of the 3M™Clean-Trace™ NGi
Luminometer allows immediate remedial action to be taken if an unacceptable result is obtained. Surfaces can be re-cleaned before use,
ensuring good sanitation and contaminationcontrol.
The test kit instructions should be read carefully beforeuse.
1.03ATP and WaterTesting.
Test kits are available for the assessment of contamination in liquid samples. Measurement of the total ATP in final rinse water samples is an
indicator of the effectiveness of Clean in Place (CIP) processes. The total ATP in process and waste water, water tanks and cooling towers can
be used to determine and monitor the efficiency of the biocide treatmentprogram.
Measurement of free or non microbial ATP can be performed in addition to total ATP for further information on the nature of
thecontamination.
The test kit instructions should be read carefully beforeuse.

2.0INSTRUMENTOVERVIEW
2.01Parts of the 3M™Clean‑Trace™ NGiLuminometer
2.02GeneralFeatures
2.03Charging theBattery
2.04Switching the InstrumentON
2.05MeasurementModes
2.06Inserting aTest
2.07SwabSensor
2.08Low PowerMode
2.09PC Communications
2.01Parts of the 3M™Clean‑Trace™ NGiLuminometer
The purpose of each of the Function Buttons can change according to the menu screen being displayed. Once a menu is selected the button
functions will be displayed at the bottom of the screen directly above eachbutton.
Figure 1 Figure 2
Chamber cap
Cap release
RS 232to PC
Battery level indicator USB to PC *
Display screen
Power from Charger
Navigation buttons
Cursor control
* USB is Inactive in
Version 1instruments
On/Off switch
2.02GeneralFeatures
The 3M™Clean-Trace™ NGi Luminometer is a light-weight, fully portable instrument with alphanumeric LCD display. The instrument is for
use in production areas, healthcare settings and laboratories and can be supplied with a carrying pouch. A separate internal Lithium Battery
is used to retain information in the instruments memory, with regard to sample plans, test points, data and calibrationsetting.
RS 232to PC is not functional on NGiLuminometer.
2.03Charging theBattery
The instrument runs on a removable, rechargeable Lithium Ion battery. The sealed battery pack is supplied with the 3M™Clean-Trace™ NGi
Luminometer and needs to be charged prior touse.
It is necessary to attach the battery pack to the instrument. Locate the battery pack into the aperture in the rear of the 3M™Clean-Trace™
NGi Luminometer by inserting the bottom of the battery pack first and pushing the battery into place. Secure the battery pack by turning the
fixing screw clockwise using a small screwdriver. Do notover‑tighten.
The 3M™Clean-Trace™ NGi Luminometer may now be charged using the charger supplied. Plug the connector from the charger into the
3M™Clean-Trace™ NGi Luminometer as shown in Figure 2, then connect the charger to an AC power outlet and switch on. Alternatively,
the optional docking station is used, insert the 3M™Clean-Trace™ NGi Luminometer into the docking station, plug the connector from the
charger into the docking station then connect the charger to an AC poweroutlet.
The battery in the 3M™Clean-Trace™ NGi Luminometer will charge fully in approximately 2hours and will then be ready for use. The
battery level indicator shows the state of charge. Remove the AC power and unplug the charger from the instrument. If a docking station is
used, the charger can remain attached to the docking station permanently. In this way, the instrument is always being charged ready for use
when it is in the dockingstation.
The 3M™Clean-Trace™ NGi Luminometer will display this screen when it is switched off
andcharging.

If the 3M™Clean-Trace™ NGi Luminometer is turned on and operating with the charger
connected, the arrow next to the battery symbol will beshown.
When switched on, the 3M™Clean-Trace™ NGi Luminometer will display a continuous indication of battery status at the top right of the
displayscreen.
The removable battery pack is held in place by a singlescrew.
To remove the battery pack, turn screw counter-clockwise untilloose
(it is a captive screw so will remain attached to the battery pack) then remove thebattery.
This shows the battery pack removed. Note the battery contacts. Do not touch or
contaminate them as this could lead to unreliable operation. The battery pack is only
available from 3M or localdistributor.
2.04Switching the InstrumentON
Ensure that no samples have been left in the chamber from previous tests before switching the instrument on. The power switch is the red
button on the instrument. Pressing and holding the On/Off button will switch the 3M™Clean-Trace™ NGi Luminometer on. Pressing and
holding the On/Off button again will switch the 3M™Clean-Trace™ NGi Luminometer off. When switching on, the 3M™Clean-Trace™
NGiLuminometer performs a series of self-checks, including an initial instrument background measurement and a calibration check.
Thefollowing screens will be displayed during initial start-up and system selfcheck.
The following screen is displayedfirst:
It is possible to pause the 3M™Clean-Trace™ NGi Luminometer at this point in order to note
the hardware and software versions to aid diagnosis in the event of a malfunction. Press any
button while this screen is displayed and the instrument willpause.
The version numbers are displayed at the bottom of thescreen.
Press any button to allow the 3M™Clean-Trace™ NGi Luminometer tocontinue.
Select
MAIN MENU
Measure Sample
Program Mode
Review Results
Setup Menu
Wed 04 Jan 2012 10:09
Battery removal screw
START-UP OK
Press any button to Pause
H:1.6 S:4.5 B:2.3 F:1.0
3M Clean-Trace NGi

The 3M™Clean-Trace™ NGi Luminometer will now perform an initial background and
calibrationcheck.
If the initial background and calibration check are within limits, the screen on the right will
bedisplayed.
Pressing the OK button will display the main systemmenu.
Should the 3M™Clean-Trace™ NGi Luminometer fail the start-up self check, one of
several screens may be displayed. If this occurs, see the troubleshooting guide in section 13
of thismanual.
2.05MeasurementModes
The 3M™Clean-Trace™ NGi Luminometer instrument offers three ways to measure asample.
MEASURE SAMPLE allows as many readings as required, but does not store the data for lateruse.
PROGRAM MODE allows data to be saved on the 3M™Clean-Trace™ NGi Luminometer and then transferred to the 3M™Clean-Trace™
Online Software. To use the Program Mode, sample plans will need to have been created and downloaded to the 3M™Clean-Trace™
NGiLuminometer. (See section 5for furtherdetails).
SCHEDULED SAMPLING allows the user to access predefined sampling schedules. It lists the dates, times and sample plans that have been
allocated to the user who has logged on to the NGi Luminometer. See Section 9for furtherinformation.
2.06Inserting aTest
Open chamber cap by sliding the cap release button to the left. The cap will springopen.
Insert the 3M test device into the chamber. Lift the cap back over the top of the device and
then push down until the caplatches.
Performing Self–check
Please Wait........
Performing Self Check
PASS
OK OK
Wed 04 Jan 2012 10:09
Select
MAIN MENU
Measure Sample
Program Mode
Review Results
Setup Menu
Wed 04 Jan 2012 10:09

2.07SwabSensor
It is recommended that as soon as the measurement of a test has been completed, the test device should be removed from the chamber
of the 3M™Clean-Trace™ NGi Luminometer. Should a 3M™Clean-Trace™ ATP test device be left in the chamber and the instrument
be inverted or laid on its side, there could be leakage of the contents of the test into the luminometer causing damage to the sensitive
components of theinstrument.
The 3M™Clean-Trace™ NGi Luminometer has a sensor to determine if a 3M™Clean-Trace™ ATP test device has been left in the chamber
for more than 30seconds following the completion of a measurement. If this situation is detected, the instrument will display a “Swab Left In
Chamber” error message on the screen and provide an audible beep to remind the user to remove the test following themeasurement.
Similarly, if a measurement is attempted in Program Mode without a 3M™ Clean-Trace™ ATP test in the chamber of the instrument, the test
sensor will notify the user that the chamber is empty with a “No Swab Detected” error message and an audible beep. This message will not
be displayed when tests are attempted in Measure Samplemode.
2.08Low PowerMode
If the 3M™Clean-Trace™ NGi Luminometer is switched on but left unattended for some
time during operation, the instrument will switch to Low Power Mode in order to conserve
battery power. During normal operation this will occur approximately 30seconds after the
backlight has switchedoff.
The Low Power Mode screen is displayed opposite. Low Power Mode can also be entered by
pressing and holding the top left green function button when in “MainMenu”display.
Pressing any green button (except red power button) on the 3M™Clean-Trace™ NGi
Luminometer will return the instrument to the screen displayed before the Low Power Mode
wasactivated.
2.09PCCommunications
In order to communicate between the 3M™Clean-Trace™ NGi Luminometer and
3M™Clean-Trace™ Online Software ensure that the luminometer is connected to the PC
via the USB cable or through the 3M™Clean-Trace™ Docking Station, which in turn is
connected to the PC via the USBcable.
The screen shown on the right is displayed when the PC is communicating with the
3M™Clean-Trace™ NGi Luminometer during the transfer of sample plans orresults.
In Progress
. . . .
Data Synchronisation

3.0ACCESSORIES
3.013M™Clean‑Trace™ NG Luminometer DockingStation
3.023M™Clean‑Trace™ NG Luminometer Soft Carry Case
3.013M™Clean‑Trace™ NG Luminometer DockingStation
Available as an optional accessory, the 3M™Clean-Trace™ NG Luminometer Docking Station is an
instrument holder which provides instant connection to the charger and PC communications. It offers the
following usefulfacilities:
• Ability to pick instrument up and walk away without unpluggingcables
• Convenient chargingcradle
• Provides extra desktopstability
• Docking station holds the 3M™Clean-Trace™ NGi Luminometer at an optimum working and viewing
angle for desktopuse
• User can either plug cables directly into instrument or dockingstation
Connector to Clean‑Trace™ NG Printer RS232 USB* Power from
To use, simply locate the 3M™Clean-Trace™ NGi Luminometer into the aperture and apply light downward pressure. The connectors to the
printer, PC and charger can be left permanently in place ifrequired.
RS232is not functional on NGiLuminometer.
Printer is not functional on NGiLuminometer.
*USB is inactive in version 1instruments
3.023M™Clean‑Trace™ NG Luminometer Soft CarryCase
The 3M™Clean-Trace™ NGi Luminometer is a light-weight, fully portable instrument for use in
production areas, healthcare settings and laboratories and can be supplied with an optional carry
pouch for added mobility and protection. The instrument may be removed from the carry pouch
when taking measurements but it is also possible to perform a measure with instrument still in its
pouch. Samples may be inserted after folding back the flap at the top of the carry pouch. To prevent
the 3M™Clean-Trace™ NGi Luminometer being dropped there is a lanyard attached to the pouch
which fixes onto theinstrument.
The illustration on the left shows a convenient method of use. Adjust the straps for the correct
height and degree ofsecurity.

4.0MEASURESAMPLE
4.01What is MEASURESAMPLE
4.02How to Test using MEASURE SAMPLE
4.01What is MEASURESAMPLE
MEASURE SAMPLE allows as many readings as required but does not store the data for lateruse.
NOTE: To store the data, refer to Section 5PROGRAMMODE.
4.02How to Test Using MEASURESAMPLE
Turn the 3M™Clean-Trace™ NGi Luminometer on. When initialized the screen will display
the MAINMENU:
The cursor will be pointing at MEASURE SAMPLE. Insert a 3M test device into the
3M™Clean-Trace™ NGi Luminometer (see section2.06)
MeasuringSamples
To measure a sample, press the SELECT key. The screen will display the word MEASURING
and a series of “count-up” blocks willappear.
The screen will then display the result in Relative Light Units(RLU).
To measure another sample, remove the previous test, insert the next one and press the MEASUREkey.
When all measurements are completed press the Back key to return to MAINMENU.
If the light in the chamber exceeds the range of the instrument, an Overload message will
bedisplayed.
Press any key tocontinue.
Overloads occur if the rate of rise of the RLU units exceeds a pre-determined factory set
rate. Accidental overload results may occur, e.g. if the sample chamber is opened while a
measurement takes place. In the case of an overload a repeat measurement can be made
once the cause has beenfound.
Select
MAIN MENU
Measure Sample
Program Mode
Review Results
Setup Menu
Wed 04 Jan 2012 10:09
Measuring
Please Wait........
52,827 RLU
This result has not been
Stored in the database
Back Measure
Wed 04 Jan 2012 10:09
Overload!
Count exceeded maximum
Back Measure
Wed 04 Jan 2012 10:09

5.0PROGRAMMODE
5.01What is PROGRAMMODE
5.02How to Use ProgramMode
5.03Retests
5.04Partial ResultSets
5.05Reviewing results
5.01What is ProgramMode
The PROGRAM MODE of the 3M™Clean-Trace™ NGi Luminometer is used to carry out testing and save results against specific test
points within a saved sample plan. Refer to the 3M™Clean-Trace™ Online Software User Guide to download the service driver to allow
communication between 3M™Clean-Trace™ Online Software and the 3M™Clean-Trace™ NGi Luminometer via the USBconnection.:
1. Ensure the 3M™Clean-Trace™ NGi Luminometer is connected to the PC via the USB cable or 3M™Clean-Trace™ NG Luminometer
DockingStation.
2. Create a sample plan using the 3M™Clean-Trace™ OnlineSoftware.
3. Transfer the sample plan from the 3M™Clean-Trace™ Online Software to the 3M™Clean-Trace™ NGiLuminometer.
4. Transfer stored results from the 3M™Clean-Trace™ NGi Luminometer to the 3M™Clean-Trace™ OnlineSoftware.
Once the sample plans have been transferred to the 3M™Clean-Trace™ NGi Luminometer, it is ready to takemeasurements.
5.02How to Use ProgramMode
From the MAIN MENU, use the navigation (▲/▼) buttons to highlight PROGRAM MODE and
pressSelect.
The PROGRAM MODE menu will display the current list of sample plans as shown on
the right. Use the navigation (▲/▼) buttons to highlight the required sample plan and
pressSelect.
The count of the Result Sets already taken against each Sample Plan is displayed on the
right-hand-side of thescreen.
You may be asked to identify yourself to the NGi Luminometer. If you are asked to identify
yourself then you will see a screen that allows you to select your UserName.
If your Website Administrator has allowed it, then there may be a <No User Name> option
at the end of the list of names. If you select this option then you will log onanonymously.
If you select a user name other than <No User Name> then you may be asked to enter a
4-digit PIN.
If you do not know your PIN then logon to the website and go to MyAccount to view or
change your NGi UserPIN.
Measure Sample
Program Mode
Review Results
Setup Menu
Select
MAIN MENU
Wed 04 Jan 2012 10:09
Result Sets↴
Ward1
Ward2
Ward3
Ward4
0
0
0
0
Back Select
PROGRAM MODE
Wed 04 Jan 2012 10:09
User 1
User 2
User 3
<No User Name>
Wed 04 Jan 2012 10:09
SELECT USER
Enter PIN
0 1 2 3 4 5 6 7 8 9 ←
Back Select
Wed 04 Jan 2012 10:09

At this point it is possible to use the 3M™Clean-Trace™ NGi Luminometer to perform a new series of tests for the selected sample plan, or
to perform retests on existing stored results which are shown in date order on thescreen.
The following symbols will appear against existing ResultSets:
Measurements have been taken against all Test Points in the SamplePlan.
Measurements have been taken against some Test Points in the Sample Plan.
These measurements will NOT be uploaded the next time the NGi Luminometer
isSynchronised.
Measurements have been taken against some Test Points in the Sample Plan and the
Result Set has been marked as Complete. These measurements will be uploaded the
next time the NGi Luminometer isSynchronised.
To make and store new measurements ensure NEW RESULTS is highlighted and pressSelect.
At this point you may be asked to enter a RESULT LABEL if your Website Administrator has
enabled this feature in the My Organisationsection.
This is a free text field that can be used to capture additional information about the
Measurements that you are about to take. This identifier can be used in the reports
produced by thewebsite.
To enter a Result Label move the cursor to the required character(s) and press Select. The
selected character(s) will appear at the top of the screen. When you have finished move the
cursor to DONE and pressSelect.
Additional screens may be displayed at this point if the Flexible Fields feature has been
configured by your website administrator and the appropriate software has been loaded
onto your NGi Luminometer. See Section 10‑ Flexible Fields for moreinformation.
To make and store new measurements, use the navigation (▲/▼) buttons to highlight the
test point required and pressMeasure.
The “►” symbol means that the Test Point Name is too long to be displayed on the screen.
You can view the rest of the Test Point Name by pressing the right hand side of the cursor
controlbutton.
The type of test that should be used for the selected Test Point is displayed to the right of the
Test Point name. In this case ATP‑S (Surface ATP Test) isrequired.
The 3M™Clean-Trace™ NGi Luminometer will then perform a measurementcycle…
…and display the result. Press OK to accept and store theresult.
NOTE: if a measurement is attempted without a 3M™ Clean-Trace™ ATP Test in the chamber of the 3M™Clean-Trace™ NGi Luminometer,
an error message stating “No Swab Detected” will be displayed and an audible beep will sound to notify the user that there is not a test in
thechamber.
19 Oct 02 09:25:11
19 Oct 02 13:47:02
21 Oct 02 09:05:44
22 Oct 02 08:55:17
22 Oct 02 14:30:22 ◒
◒
●
●
●
Back Select
Ward 1
New Results
Wed 04 Jan 2012 10:09
Done
A B C DE F G H I J K L M @
N O P Q R S T U V W X Y Z .
a b c d e f g h i j k l m !
n o p q r s t u v w x y z $
Cancel Select
RESULT LABEL
Wed 04 Jan 2012 10:09
Centre of overbed Ta
Top cot or side bed
Test 1:
ReTest:
Test 1:
ReTest:
-
►ATP-S
►ATP-S
-
-
-
Back Measure
Ward 1
New Results
Wed 04 Jan 2012 10:09
Please Wait........
Measuring
Ward 1
Swab:ATP-Surface
Centre of overbed table
Swab:ATP-Surface
X FAIL
OK OK
Ward 1
52,827 RLU
Wed 04 Jan 2012 10:09
Centre of overbed table

The result is now stored. The cursor will move to the next test point automatically, or use the
navigation (▲/▼) buttons to highlight the next test pointrequired.
Pressing the Back key will exit thetesting.
5.03Retests
It is possible to retest a previously failed test point at any time after remedial action has
been taken. Select the correct test point as shown above andmeasure.
Note that the measurement screen now displaysReTest.
The new result is stored in the ReTest area of the screen. This time the result is apass.
It is possible to retest the point several times until a pass is obtained. The screen will only
show the first test and most recent retest but all retests are actuallystored.
NOTE: It is possible to exit from a set of results and re-open it at a later time. For example, if not all measurements were taken at that time
or if there is a need to go back into set of results to perform retests, this may require returning to an unfinished set of results to collect
moredata.
5.04Partial ResultSets
A Partial Result Set is a Result Set that does not have Measurements against all of its TestPoints.
If you are collecting Measurements or Retests and you decide to leave the Result Set (by
pressing Back ) then, if there are no measurements against all of the Test Points (i.e. it is a
Partial Result Set), the INCOMPLETE RESULTS screen will bedisplayed.
If you select Yes then when you next synchronise the Result Set will be uploaded to the
server and removed from the NGiLuminometer.
If you select No, then when you next synchronise the Result Set will NOT be uploaded to the
server. After synchronisation the Result Set will remain on the NGi Luminometer so that you
can continue to capture measurements andretests.
5.05Reviewingresults
It is possible to review and inspect previousresults.
From the MAIN MENU use the navigation (▲/▼) buttons to highlight REVIEW RESULTS and
press the Selectbutton.
Back Measure
Ward 1
Wed 04 Jan 2012 10:09
Centre of overbed Ta
Top cot or side bed rail
Test 1:
ReTest:
Test 1:
ReTest:
52,847 RLU X
►ATP-S
-
-
-
New Results
ReTest
Ward 1
Swab:ATP-Surface
Measuring
Please Wait........
Centre of overbed table
Back Measure
Wed 04 Jan 2012 10:09
Ward 1
Centre of overbed Ta
Top cot or side bed ►ATP-S
Test 1:
ReTest:
Test 1:
ReTest:
52,847 RLU X
►ATP-S
108 RLU
-
-
New Results
- Test Points: 3
- Measurements: 2
Ward1
New Results
Mark results as Complete?
No Yes
INCOMPLETE RESULTS
Wed 04 Jan 2012 10:09
Select
MAIN MENU
Measure Sample
Program Mode
Review Results
Setup Menu
Wed 04 Jan 2012 10:09

The list of sample plans stored on the 3M™Clean-Trace™ NGi Luminometer will be
displayed. Use the navigation (▲/▼) buttons to highlight the plan to be reviewed and
pressSelect.
Use the navigation (▲/▼) buttons to highlight the required result set to review and
pressSelect.
The following symbols will appear against existing ResultSets:
Measurements have been taken against all Test Points in the SamplePlan.
Measurements have been taken against some Test Points in the Sample Plan. These measurements will NOT be uploaded the next time
the NGi Luminometer isSynchronised.
Measurements have been taken against some Test Points in the Sample Plan and the Result Set has been marked as Complete. These
measurements will be uploaded the next time the NGi isSynchronised.
The screen will now display all results for the result set. Use the navigation (▲/▼) buttons to
scroll through theresults.
Press Select to view the test point in moredetail.
Scroll down to viewRetests.
If UserName is being captured then the user who took the measurement or retest will be
displayed on thescreen.
Back Select
REVIEW RESULTS
Result Sets↴
Ward1
Ward2
Ward3
Ward4
1
0
0
0
Wed 04 Jan 2012 10:09
Back Select
Wed 04 Jan 2012 10:09
19 Oct 02 09:25:11
19 Oct 02 13:47:02
21 Oct 02 09:05:44
22 Oct 02 08:55:17
22 Oct 02 14:30:22 ◒
◒
●
●
●
●
Ward 1
22 Oct 02 14:46:22
T R
-
X !
X
X X
Back Select
Wed 04 Jan 2012 10:09
Ward 1
21 Oct 02 09:05:44
Test Point: Centre
Floor Under patients bed
Centre of overbed table
Top cot or side bed rail
Bed Comms unit keyboard
Top Bedside cabinet
Bed Comms unit keyboard
ATP-Surface
Back
Wed 04 Jan 2012 10:09
Result:
Ward 1
21 Oct 02 09:05:44
4112 RLU
22 Oct 02, 13:52:11
UserName

6.0SETUPMENU
6.01What the SETUP MENUallows
6.02How to find the SETUPMENU
6.03Setting DateFormat
6.04Setting theDate
6.05Setting theTime
6.06Setting theLanguage
6.07Performing a SelfCheck
6.08Setting the Backlight Timeout
6.01What the SETUP MENUallows
The SETUP MENUallows:
• Format the date as shown below and format the time to 12/24hourclock
• Set the current time anddate
• Change the displayedlanguage
• Check to see if the instrument performs within calibrationlimits
• Set the display backlight timeoutperiod
6.02How to find the SETUPMENU
From the MAIN MENU use the navigation (▲/▼) buttons to select SETUP MENU and Press
the Selectbutton.
The SETUP MENU can now beseen.
6.03Setting DateFormat
While in the SETUP Menu, use the navigation (▲/▼) buttons to select DATE & TIME and
press the Select button. The DATE & TIME menu can now beseen.
The choice of date formatsare:
DOW D M Y e.g. Wed 04Jan 201212:31
DOW M D Y e.g. Wed Jan 04201212:31
Y M D DOW e.g. 2012Jan 04Wed12:31
Use the navigation (▲/▼) buttons to select DATE FORMAT and press the Change button
repeatedly until the required format isshown.
Press the Back button to return to the SETUP Menu or use the navigation (▲/▼) buttons to
select another item in the DATE & TIMEmenu.
Select
MAIN MENU
Measure Sample
Program Mode
Review Results
Setup Menu
Wed 04 Jan 2012 10:09
Back Select
SETUP MENU
Date & Time
Language
Self–check
Backlight Timeout
Wed 04 Jan 2012 10:09
Back Select
DATE & TIME
Set Date
Set Time
Date Format
04-Jan-12
14:28:56
DOW D M Y
Wed 04 Jan 2012 10:09

6.04Setting theDate
While in the SETUP Menu, use the navigation (▲/▼) buttons to select DATE & TIME and
press the Select button. The DATE & TIME menu can now beseen.
While in the DATE & TIME menu, use the navigation (▲/▼) buttons to select SET DATE.
Day is highlighted only after the Select function button is pressed. Use the navigation (▲/▼)
buttons to change the Day until correct. Press the OKButton.
The Month will now be highlighted on the display. Set the Month using the navigation (▲/▼)
buttons then pressOK.
The Year will now be highlighted on the display. Set the Year using the navigation (▲/▼)
buttons then pressOK.
Now use the navigation (▲/▼) buttons to select another item in the DATE & TIME menu or
press Back to return to the SETUPmenu.
6.05Setting theTime
While in the SETUP Menu, use the navigation (▲/▼) buttons to select DATE & TIME and
press the Select button. The DATE & TIME menu can now beseen.
While in the DATE & TIME menu, use the navigation (▲/▼) buttons to select SET TIME, press
Select and Hours is highlighted on the display. Use the navigation (▲/▼) buttons to change
the Hour until correct. Press the OKButton.
The Minutes will now be highlighted on the display. Set the Minute using the navigation
(▲/▼) buttons then pressOK.
The Seconds will now be highlighted on the display. Set the Second using the navigation
(▲/▼) buttons then pressOK.
Use the navigation (▲/▼) buttons to select another item in the DATE & TIME menu or press
Back to return to the SETUPmenu.
NOTE: The date & time will automatically be corrected whenever the NGi isSynchronised.
6.06Setting theLanguage
Contact your local distributor for a current list of languagesavailable.
While in the SETUP Menu, use the navigation (▲/▼) buttons to select LANGUAGE and press
the Selectbutton.
The LANGUAGE menu can now beseen.
Use the navigation (▲/▼) buttons to select the Language required. Press either of the OK
buttons to return to the SETUPmenu.
NOTE: The language changes automatically as navigation (▲/▼) buttons select a
differentlanguage.
6.07Performing a SelfCheck
Self Check performs both a background reading and a calibration check. Before selecting Self Check, ensure the sample chamber is empty,
the chamber cap is closed and the instrument has been at a stable temperature for at least 15minutes.
While in the SETUP Menu, use the navigation (▲/▼) buttons to select SELF‑CHECK.
Back Select
DATE & TIME
Set Date
Set Time
Date Format
04-Jan-12
14:28:56
DOW D M Y
Wed 04 Jan 2012 10:09
Back Select
DATE & TIME
Wed 04 Jan 2012 10:09
Set Date
Set Time
Date Format
04-Jan-12
14:28:56
DOW D M Y
Back Select
SETUP MENU
Wed 04 Jan 2012 10:09
Date & Time
Language
Self–check
Backlight Timeout
Deutsch
Italiano
Back Select
Language
English
Francais
Espanol
Portugues
Wed 04 Jan 2012 10:09
Back Select
SETUP MENU
Wed 04 Jan 2012 10:09
Date & Time
Language
Self–check
Backlight Timeout

Press Select to start the Self Check, The screen will display the words
PERFORMING SELF‑CHECK and a series of “count-up”blocks.
If the instrument is performing within calibration limits and the background reading is
acceptable, PASS will bedisplayed.
If the instrument is not performing within calibration limits, one of the three screens below will be displayed. If this is the case the instrument
should not be used as performance may be compromised. Contact 3M or local 3M Distributor for service andcalibration.
Calibration checkfailed
Backgroundfailed
Calibration checkfailed
Performing Self–check
Please Wait........
Performing Self Check
PASS
OK OK
Wed 04 Jan 2012 10:09
OK OK
Wed 04 Jan 2012 10:09
SELF CHECK
Calibration Required –
Please contact nearest
Service Centre
OK OK
High Background:
1,047 RLU
SELF CHECK
Wed 04 Jan 2012 10:09
OK OK
Wed 04 Jan 2012 10:09
SELF CHECK
High Background:
1,047 RLU
Calibration Required –
Please contact nearest
Service Centre

6.08Setting the BacklightTimeout
The display on the 3M™Clean-Trace™ NGi Luminometer is fitted with a backlight to improve visibility in dark or dim lighting conditions.
The backlight may be set permanently off, permanently on; or may be set to remain on for a preset time interval after which the backlight
extinguishes. Pressing any button re-starts the backlight time interval. This interval may be set from 5seconds to 60seconds in 5second
increments. It should be noted that the instrument consumes more power when the backlight isilluminated.
Under normal ambient lighting conditions, there should be no need to use thebacklight.
If the backlight is not required, set the Backlight Timeout permanently to OFF to ensure
maximum battery life. By default, the Backlight Timeout is set to 15seconds.
While in the SETUP Menu, use the navigation (▲/▼) buttons to select BACKLIGHTTIMEOUT.
Select desired timeout value using the navigation (◄/►)buttons.
Press the Back button to return to the MAIN MENU or use the navigation (▲/▼) buttons to
select another item in the SETUPmenu.
Back Select
SETUP MENU
Date & Time
Language
Self–check
Backlight Timeout
Wed 04 Jan 2012 10:09
◄15 ►Change

7.0SYNCHRONISE NGi
7.01What is SYNCHRONISE NGi
7.02How to SYNCHRONISE NGi
7.01What is SYNCHRONISE NGi
Synchronisation allows for the transfer of results stored in the 3M™Clean-Trace™ NGi Luminometer to 3M™Clean-Trace™ Online
Software and the download of sample plans from 3M™Clean-Trace™ Online Software to the Clean-Trace NGi luminometer. In order to
UPLOAD RESULTS, the 3M™Clean-Trace™ NGi Luminometer must be connected via the USB cable or 3M™Clean-Trace™ Docking
Station to a PC, on which the 3M™Clean-Trace™ Online Software driver has been installed. The 3M™Clean-Trace™ Online Software
driver allows communication between the 3M™Clean-Trace™ NGi Luminometer and 3M™Clean-Trace™ Online Software. Refer to the
3M™Clean-Trace™ Online Software User Guide for moreinformation.
NOTE: A USB symbol will only appear on the 3M™Clean-Trace™ NGi Luminometer screen when it is connected to the 3M™Clean-Trace™
Online Software via the USB cable or DockingStation.
7.02How to SYNCHRONISE NGi
To start the synchronisation process you need to right-click on the 3M Synchronisation Manager icon on the System Tray of your PC and
select Open 3MSynchronisationManager .
If you cannot find the icon in the system tray then start it by goingto
Start\All Programs\3M CleanTrace Hygiene Management Solution\ 3M NGiSync.
Next you need to click on the Synchronise button at the bottom of the 3MSynchronisation Managerscreen.
You will then be asked to enter your Username and Password. Once you have entered them then click onOK.
The 3M Synchronisation Manager will validate your Username andPassword.
If they are incorrect then a Login Failed message will be displayed. Re-enter your Username & Password and
tryagain.
The 3M™Clean-Trace™ NGi Luminometer will begin communicating with
3M™Clean-Trace™ Online Software. The screen opposite will be displayed on the NGi.
Data Synchronisation
In Progress

At the end of a successful synchronisation, a synchronisation completed message will be
displayed on the NGi. Stored results will have been uploaded to 3M™Clean-Trace™ Online
Software and will have been removed* from the 3M™Clean-Trace™ NGi Luminometer. Any
adjustments to Sample Plans will be updated on the 3M™Clean-Trace™ NGi Luminometer.
Select OK to return to the MAINMENU.
* NOTE: Partial Result Sets that are not Marked Complete will NOT have been uploaded to the server or removed from the NGi.
SeeSection5.04Partial Result Sets for furtherinformation.
If you experience any problem during the synchronisation phase, please refer the Troubleshooting Section of this instruction manual or the
3M™Clean-Trace™ Online Software UserGuide.
8.0MANAGING MULTIPLEORGANISATIONS
If you have access to more than one organisation then, during synchronisation, you will
be asked which organisation(s) you wish to download to the NGi. If you download more
than one organisation then a new CHANGE ORGANISATION menu item will appear in the
MAINMENU and the organisation that you are currently using will appear at the top of the
MAINMENUscreen.
Selecting CHANGE ORGANISATION will display a screen listing the organisations that are on
the NGi. To select an organisation, scroll to the ORGANISATION and pressSelect.
When you change organisations the NGi only displays information for that organisation.
However, information gathered for other organisations is still stored in the NGi memory and changing back to those organisations will make
the stored information available for you touse.
Synchronisation complete
OK OK
Wed 04 Jan 2012 10:09
Select
Organisation1
Measure Sample
Program Mode
Review Results
Setup Menu
Change Organisation
Wed 04 Jan 2012 10:09
Cancel Select
SELECT ORGANISATION
Organisation1
Organisation2
Organisation3
Wed 04 Jan 2012 10:09

9.0SCHEDULEDSAMPLING
9.01Using ScheduledSampling
9.02Retests
9.03Partial ResultSets
9.04Reassigning Schedules
Scheduled Sampling allows you to specify how often measurements should be taken in a specific location and optionally specify who should
take the measurements. If the Scheduled Sampling system is set up on the website then, when you synchronise, Scheduled Sampling data
will be loaded onto your NGi.
See 3M™Clean-Trace™ Online Software Help File for further Information about configuring the system for
ScheduledSampling.
9.01Using ScheduledSampling
If Scheduled Sampling data has been loaded to your NGi then you will see a new option
SCHEDULED SAMPLING in the MAINMENU.
When you select SCHEDULED SAMPLING then you may be asked to identify yourself to the
NGi. If you are asked to identify yourself then you will see a screen that allows you to select
yourUser Name.
If your Website Administrator has allowed it, then there may be a <No User Name> option at the end of the list of names. If you select this
option then you will log on anonymously and you will only see schedules that have not been assigned to a specificuser.
If you select a user name other than <No User Name> then you may be asked to enter a
4-digit PIN. If you do not know your PIN then logon to the website and go to MyAccount to
view or change your NGi UserPIN.
Organisation1
Measure Sample
Program Mode
Review Results
Wed 04 Jan 2012 10:09
Select
Setup Menu
Scheduled Sampling
User 1
User 2
User 3
<No User Name>
Wed 04 Jan 2012 10:09
SELECT USER
Enter PIN
0 1 2 3 4 5 6 7 8 9 ←
Back Select
Wed 04 Jan 2012 10:09
Other manuals for Clean-Trace NGi Luminometer
1
Table of contents
Other 3M Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Turn Key
Turn Key iGAS operating instructions

Spot-on
Spot-on LW600PRO operating instructions

Tek-Trol
Tek-Trol Tek-Clamp1200A-100H quick start guide

POTOMAC INSTRUMENTS INC.
POTOMAC INSTRUMENTS INC. AA-51A user guide

FRONIUS
FRONIUS Smart Meter 50kA-3 operating instructions

FRONIUS
FRONIUS Smart Meter TS 5kA-3 operating instructions