ADE M500020 Series User manual

M500020-210322-Rev009-UM-e
Operating Manual
Electronic Wheelchair and Multipurpose Scales
Series
M500020

2 M500020-210322-Rev009-UM-e
Content
Introduction ............................................................................................................................................ 3
Equipment supplied 4
Assembly and set up ............................................................................................................................. 4
Attachment o the indicator 4
Power supply 5
Replacing rechargeable batteries 5
Tips or handling the batteries 6
Quality eatures 6
Display .................................................................................................................................................... 7
Description 7
Connecting additional devices 7
Housing and connections 7
Sa ety ...................................................................................................................................................... 8
Sa ety instructions 8
Settings ................................................................................................................................................... 8
Setting date and time 9
Operation ................................................................................................................................................ 9
Modes o operation 9
Weighing 9
Tare unction 10
Hold unction 10
Determining the Body Mass Index (BMI) 10
Evaluating Body Mass Index 11
Print out a record 11
Care, servicing and recalibration ....................................................................................................... 11
Cleaning 11
Error messages 11
Waste disposal 12
Storage and transport conditions 12
Warranty 12
Recalibration 12
Technical speci ications ..................................................................................................................... 13
Con ormity declaration ....................................................................................................................... 14

M500020-210322-Rev009-UM-e 3
Introduction
This symbol means: "Consult operating instructions" To achieve accurate results, read the
instructions for use carefully before use and follow all instructions provided therein Keep the
instructions for use for future reference
Your ADE electronic scale is a quality, sturdy and durable product manufactured under the high standards and
regulations for calibrated class III scales in accordance with national regulation in Germany The scale is de-
signed and manufactured for the daily practice in hospitals, doctors’ surgeries and in-patient nursing facilities
The scale is used to evaluate the weight of patients sitting in a wheelchair, able to walk on their own or by
using any kind of walking aid or lying on an emergency litter
ADE instruction manuals have been prepared to provide the user with all the necessary knowledge about the
safe use of ADE scales and their related accessories
For further questions about how to use the products or about this or other service manuals, feel free to contact
your local ADE representative or visit our web page at www ade-hamburg de
This operation instruction applies to the following models of electronic scales of the M500020-series with the
different adaptations (see table 1) In order to achieve the best possible results kindly read and follow the op-
eration manual in all respects
Multipurpose and wheelchair scales
Article number and size
o plat orm in mm:
M500020
M500020-A
M500020-B
M500020-C
M500020-D
M500020
-
E
800 x 800
1000 x 800
1200 x 800
1400 x 800
800 x 800
1000
x
1000
Con iguration o the basic
version M500020:
1 platform
1 ramp
1 spirit level
1 mains adapter
1 indicator (display)
Reference number Description of supplementary equipment to the standard version
M500020-01 2 x
s
ide boundaries
without wheels and hand grip
M500020-02 1 foldable column
M500020-03 1 foldable hand rail
M500020-04 2 x side boundaries and a folding
column without wheels and hand grip
M500020-A-01 large platform plus two side bounda-
ries without wheels and hand grip
M500020-A-02 large platform plus foldable column
M500020-B-01 large platform plus foldable column
M500020-C-01 large platform plus foldable column
M500020-D handrails on three sides optional with wheels
[table 1]
M500020
M500020-A
M500020-B
M500020-C
M500020-01
M500020-A-01
[picture 1]

4 M500020-210322-Rev009-UM-e
M500020-03
M500020-E-01
M500020-02
M500020-A-02
M500020-B-01
M500020
-
C
-
01
[picture 2]
M500020-04
M500020-D
[picture 3]
Equipment supplied
All wheel chair scales and multipurpose scales are being delivered in an already assembled condition with
handrails, side boundaries or columns The equipment supplied includes:
1) One indicator including the correspondent set of screws
2) Two guard plates for attachment of the indicator Model numbers: M500020-03; M500020-E-01; M500020-
02; M500020-A-02; M500020-B-01; M500020-C-01; M500020-04 and M500020-D
3) One wall mounting device and one cable extension for model numbers: M500020; M500020-A; M500020-B;
M500020-C; M500020-01 and M500020-A-01
4) One mains adapter
5) Operation manual including declaration of conformity
Assembly and set up
The indicator/ display of your ADE scale is able to be mounted on walls, on sockets, handrails or on stand-on
devices for the table depending on the model you have selected
Please insure that your scale is being placed vibration free at a dry place Please take the parts out of the
packaging and follow the instruction for assembly
Attachment o the indicator
The following mounting instructions are valid for all models The Wheelchair and multipurpose scales of ADE
are coming in an already assembled condition The only part which needs to be attached is the indicator The
indicator is being placed on top of a column, on the supporting plate at the handrail or to be used with the wall
mounting device
1) Place the indicator together with the two guard plates on top of the column or the handrail
2) Fix them with the corresponding screws
[picture 4] Attachment of the indicator onto
the foldable column of the M500020-02
[picture 5] Attachment of the indicator onto the handrail of
the M500020-03 and/or the M500020-D

M500020-210322-Rev009-UM-e 5
Alignment o the scale
1) Place the scale on a plain, dry and level surface
2) Undo the locknuts of the four feet of the scale and unscrew them until they got contact to the surface
3) The air bubble in the spirit level must be right in the centre of the circle
4) Please ensure that the scale has no other contact to the floor than by the four feet
4) For all scales with a foldable column or a foldable Handrail:
(M500020-03 M500020-E-01 M500020-02 M500020-A-02 M500020-B-01 M500020-C-01 and M500020-04)
a
) Release the handle (pic 4a) at the bottom of the column or the handrail
b) Upright the column or the handrail
c) Finally re-tighten the handle (pic 4b) at the bottom of the column or the handrail again
Please ensure that the oldable handrail and column o the scale is ixed and stable in order to
avoid any unexpected moving o same. A ter transport or movement o the scale ollow step 1 to 4
again.
Power supply
Power is supplied either by the rechargeable battery in the display unit or by the mains power via the supplied
power unit The imprinted voltage value at the mains adapter must match the local mains voltage Connect the
mains cable to the indicator and to a socket
The wall socket must be located near to the unit and be easily accessible Please use only the mains adapter
supplied by ADE The mains adapter is charging the battery pack at the same time
When the battery is being charged for the first time a sufficient charging period of approx 24 hours is neces-
sary to achieve a full battery load
If the unit is in battery mode and the display indicates – Lo, the battery needed to be recharged with the mains
adapter For a full battery load a charging period of 12-16 hours is required with the scale being switched off
If–Lo still appears at the display inform your service partner
Replacing rechargeable batteries
The battery compartment is located on the front of the unit
1) Open the battery compartment
2) Remove the battery cables carefully from the housing, connect the wires (red to red, black to black) and
insert the battery into the housing
3) Close the battery compartment again
Only use original rechargeable batteries and charger units rom ADE. The warranty does not
cover any damage caused by using other types o batteries or charger units.
(Pic 4a)
(4-b)

6 M500020-210322-Rev009-UM-e
Tips or handling the batteries
The life of a battery depends essentially on the load, frequency of use, temperatures, settings and the acces-
sories being used
– New batteries or batteries that have been stored for long periods require a longer charging period
– Charge the batteries at room temperature
– Never expose the batteries to temperatures below -10°C (14°F) or above 45°C (113°F)
– Rechargeable batteries lose their charging capacity over the time and require a longer charging period
the older they are If the battery performance drops considerably, then it is time to replace the battery
– New batteries are supplied in partial charged condition Batteries achieve their full capacity after several
complete charge and discharge cycles
– Do not throw old rechargeable batteries into the domestic waste Ensure that the valid legal and municipal
regulations of your country are being followed
– Never dispose of batteries in fire as they can leak or explode
Quality eatures
ADE products correspond to the applicable EC Directives and European standards to manufacture
and distribute world-wide medical devices as per the latest technical standards ensuring a long life
time and continuous reliability
This symbol stands for compliance of the European Directive covering medical
devices and declares the conformity as per the Medical Device Law
Approved scales classes III respecting the standards of the European guideline
covering approved scales are carrying this sign
Scales carrying this symbol are approved and accredited within the European
Community and fulfils first class demands for calibrated scales
Products carrying this symbol meet the requirements as per:
The valid EC Directive for non automatic scales
in the latest version
M21

M500020-210322-Rev009-UM-e 7
Display
Description
Fig. 1-1 Indicator and operating controls
No.
Name Description
1 Display LCD display for weight, mode, settings and configuration information
2 Battery charge level Status indicator for the current battery charge level
3 Switching points 1 & 2 In program mode: Status indicator for switching points 1 & 2
4 BMI status indicator
(HI, OK, LO) Status indicator of the automatic BMI (Body Mass Index) evaluation
5 Net status indicator LCD on - net weight is being displayed
LCD off – gross weight is being displayed
6 Status indicator The scale is at zero level
7 Hold – button Activates the hold function
8 On/off, zero – button On and off switch and re-set of the display
9 Tare- button Activate/ deactivate the tare function
10 Preset tare – button Press briefly to access the tare function
Press for a long period to enter and save the tare value
11 BMI – Button Enter to proceed with BMI mode
12 Units – Button Switch button between primary and secondary units Not available for CE
approved class III scales
13 Number pad
Figures 0-9, Clear, Enter and Navigation functions: Key 6 menu direction
to the right; Key 4 menu direction to the left; Key 8 menu direction down;
Key 2 menu direction up For better understanding see Pic 1 1 the blue
arrows Used to navigate through menus, to select numerical values and
to increase or lower the value
Special function:
"0" key
In weight mode, you can change between gross and net weight using
figure "0" (if any weight have been tared off)
"Enter" key Function in weight mode: "Print" if printer has been connected
14 Standstill status indica-
tor (14) :
The scale is at a standstill The weight is stable with no movement This
is mandatory for functions like reset, print and tare
15 “Hold” If displayed is the hold function is activated
17 “kg” If displayed is the weight is displayed in kilograms
“g” If displayed is the weight is displayed in grams
Connecting additional devices
Printer, PC or remote-display could be connected by the RS-232 interface For connection of additional devic-
es kindly ask your local representative or authorized ADE dealer for further instructions
Housing and connections
The connections to mains adapter, serial port and to the load cells are located at the back of the ADE display
The set-up switch to start the configuration mode is underneath the housing It is protected by a cover plate
and secured by a screw (not shown in Fig 2-1)
17
16
15
14
1
2
3
4
5
6
7
8
9
10
11
12
13

8 M500020-210322-Rev009-UM-e
Do not use the set-up button as otherwise the calibration will be lost. This procedure is only al-
lowed to be done by authorized service personnel appointed by ADE (GmbH & Co.). Any unau-
thorized manipulations made to the con iguration mode will a ect the warranty agreement.
Sa ety
Sa ety instructions
The ADE-display unit (MAK) corresponds to the latest state of the art technology and complies with standard
regulations and directives Kindly follow the safety instructions in this operating manual, the country-specific
installation standards as well as the applicable safety conditions and accident prevention regulations in all
aspects Electronic medical devices are restricted to special safety regulations in respect of their electromag-
netic compatibility For positioning and operation follow the instructions of this Operating Manual
– Modell M500020-D: The pre-installed seat is only allowed to be used by persons up to 100kgs
– Portable and mobile HF-communication systems may influence electronic medical devices
– The mains voltage and power type stated on the type plate should match the mains voltage and power
type at the place of use
– Make sure that the mains cable does not pose any potential trip hazard
– Do not operate the unit in hazardous or unstable conditions
– Do not submerge the unit in water or other liquids
– Set up the unit as far away as possible from equipment that generates electromagnetic or other interfer-
ence
– Only use original rechargeable batteries and power adapters from ADE The Warranrty does not cover
any damage caused by using other types of batteries or charger units
A safe operation is no longer guaranteed:
- If the power unit exhibits visible damage
- If the power unit is no longer working
- After longer periods of storage in damp rooms
In these cases inform your service partner
Settings
The following functions are available in setting mode:
– Configuration of the serial interface
– Set up of date, time and consecutive printout numbers
– Set up of print format
– Set up of switching points
– Test functions
The setting mode is not made for metrological parameters Access the function menu by pressing the CLEAR
(13) key until TIME appears on the display Use the key pad (13) to navigate within the menus, to select nu-
merical values and to increase or lower the values Press the ENTER key to save the values entered and to
return to the super ordinate menu level

M500020-210322-Rev009-UM-e 9
Menus structure o setting mode:
After entering the clear button following options are available and appear on the screen when moving with the
navigation buttons:
FILTER SERIAL PROGR PFORMT
SETPTS
TEST TIME DATE VERS
Setting date and time
– Enter the CLEAR key for approx 4 sec; FILTER appears on the display
– Press key "6" six times and TIME appears on the display
– Press ENTER key and SHOW appears on the display
– Press key "6" one time and HOUR appears on the display
– Press ENTER key and set the hours using the arrow keys "2", "4", "6" and "8"
– Press the ENTER key and set the minutes and seconds
– Go back by pressing key "2"; TIME appears on the display
– Press key "6" one time and DATE appears on the display like the time entry
– Press key "2" two times and EXIT Y appears on the display
– Confirm with ENTER and date & time are set
Operation
Modes o operation
Weighing mode
The "weighing mode" is of course the standard mode of your scale After initial configuration and sealing (by
attaching a calibration sticker to the back of the unit), the ADE-scale can only be operated in this mode
Setting mode
Date, time and consecutive numbers for print commands are being set in this mode To start setting mode,
hold the GROSS/NET key pressed until TIME appears on the display
Con iguration mode (set-up mode)
The values for different parameters including the calibration could be set in the configuration mode This func-
tion is reserved for authorized service personnel only
Test mode
There are different diagnostic functions for the ADE-MAK in test mode This is reserved for service personnel
only
Manipulations to test and con iguration modes can revoke the calibration status and will result
in loss o warranty!
Weighing
Press the ON key for at least 3 seconds to turn the scales on An automatic function test will be conducted
afterwards The test is done when the display shows zero The scale is now ready for the weighing procedure
Resetting the scales back to zero: Wait until the standstill status indicator (14) appears By entering the
ZERO key (8), measuring inaccuracies after switching on the scale are being compensated The zero key
restarts the scale The status indicator (6) appears when the scale is being resetted
The modell M500020-D is being supplied with a seat The seat is designed to take a weight upto
100kg Take care that the patient, who uses this seat is weighing not more than max 100kgs

10 M500020-210322-Rev009-UM-e
Tare unction
The TARE key (9) is used when the patient is being weighed with other items (e g wheelchairs, clothing,
shoes or other items) which should be ignored by the scale To do this, weigh these items first Weigh these
items first After the weight is displayed, press the TARE key (9) to reset the display back to zero The display
changes to net mode (NET status indicator (5) is on) and shows zero
Entering manually tare - values:
The user is able to insert manually the gross weight, which is for example mentioned at the identification plate
of the wheelchair Enter the appropriate tare value just by using the key pad under normal weighing mode
The value will be stored after use of the ENTER key (or TARE) The ADE-display is now ready for weighing
and changes to net mode [NET status indicator (5) is on] Enter the tare button to leave this mode
Preset tara
Press and hold the PRESET TARE key (10) for approx 3 seconds PRO PT X appears on the display Select
the memory location 1-10 Now enter the TARE VALUE which you would like to be ignored by the scale and
confirm same with ENTER The value is displayed as a negative figure and will be deducted when a load is
being placed on the scale Enter the on/off, zero – button (8) to leave this mode
Restore the preset Preset tara value:
Press the PRESET TARE key (10) "Pt x" appears on the display Enter a number from 0-9 (memory location
for the tare value) If no new number will be inserted to change the already displayed value within five sec-
onds, the unit will turn back to weighing mode and displays the last stored value as negative figure as above
Enter the on/off, zero – button (8) to leave this mode
Displaying the gross weight, net weight and tare value:
Press the "0" key to change between the net weight (gross weight minus the tared value) and the gross
weight
Hold unction
The determined weight could be continued being displayed even when the load has already been removed
That enables taking care of the patient first before reading the results
To use the hold function the scale must be at standstill The standstill status indicator (14) is on and
Hold appears at the digital display Deactivate the HOLD function by pressing the key again
Determining the Body Mass Index (BMI)
The BMI figure is a index which displays the ratio between height and weight
BMI=
body weight
(height)²
To determine this value you need the height of the person Best way to get the height is of course to use an
ADE height measuring rod; model MZ10023
Start the scale and weigh the person Then press the BMI key (11) to start BMI mode The last value entered
appears on the display Use the numerical key pad to enter a new value for the height (in cm) or retain the
value which is already displayed
Confirm the value entered with the ENTER key or delete same by using the CLEAR key After the entry has
been confirmed by the ENTER key, the BMI value will be calculated using the weight on the scales The dis-
play changes between weight value and BMI value "b xx x" (every two seconds) for a period of 3 minutes
Press the BMI key (11) to cancel before this time has expired

M500020-210322-Rev009-UM-e 11
E
valuating Body Mass Index
Print out a record
The scales must be at standstill, e g standstill status display (14) appears at the screen
Press the ENTER key (13) to send the data to the serial interface
Note: Printing is not possible if the net weight is 0 The print function only produces one print-out A reset is
required to print out another copy
Care, servicing and recalibration
Cleaning
Remove the mains plug before cleaning the unit Do not use any aggressive cleaning materials for cleaning;
only use a damp cloth or conventional disinfectant Ensure that no liquids or any fluid gets into the scale
Error messages
Error mes-
sage
Description Problem solution
Overload display Weight exceeds max capacity of the
scales
Overflow indicator (negative) Negative weight indicated – weight too
low to be displayed (<=99999)
Overflow indicator (positive) Positive weight indicated – weight too
high to be displayed (>=999999)
AD LOW A/D above negative range Check if the scales are correctly as-
sembled or if they are damaged
AD HI A/D above positive range Check if the scales are correctly as-
sembled or if they are damaged
EE SUM Parameter or calibration error Recalibrate – please contact your sup-
plier or service partner:
EE WR EEPROM indicates an error Please contact your supplier or service
partner
PM SUM Internal program indicates a check error Please contact your supplier or service
partner
HOFSET Switch on – above zero Please remove the weight from the
scales and restart
LOFSET Switch on – below zero Check scales for damage and inform
your service partner:
UOFSET Irregular
Do not touch the scales after they have
been switched on
BMI evaluation according to WHO
BMI
< 16 Severely underweight
16 - 17 Moderately underweight
17 - 18 5
Slightly underweight
BMI
18 5 - 25 Normal weight
BMI
25 - 30 Overweight
30 - 35 Obese Class I
35 - 40 Obese Class II
> 40 Obese Class III

12 M500020-210322-Rev009-UM-e
Waste disposal
Used electrical equipment and batteries do not belong in the domestic waste.
Please dispose o de ective or used equipment and batteries according to legal regulations o
your speci ic country!
Please make sure that only discharged batteries or batteries with insulated "poles" are dis-
posed o as otherwise they may short-circuit!
Storage and transport conditions
Keep all packaging in case you need to return the item, in order to avoid any potential transport damage
Disconnect all connected cables before dispatch This is how you can avoid unnecessary damage
Warranty
ADE will repair or replace this product within a period of 2 years from date of delivery applying to defects oc-
curred due to poor material or workmanship (presentation of invoice is necessary) All moveable parts – bat-
teries, cables, mains units, rechargeable batteries etc – are excluded Warranty does not include natural
wear, damages caused by incorrect or negligent handling, overuse or unsuitable or improper storage The cost
of transport will be borne by the customer should the equipment be located anywhere other then the custom-
er’s premises In the event of transport damage, claims under warranty can be accepted subject to using the
complete original packaging and the equipment secured the way it was packed and secured originally All the
packaging should therefore be retained
Any warranty claim will be rejected if the equipment is opened by persons not authorized expressively by ADE
to do so
Customers abroad are kindly requested to contact their local dealer for any warranty claim, customer service
and spare parts
Recalibration
Calibration to be done only by an authorised company or authorised person in accordance with national legis-
lative regulations The CE symbol shows the notified body (0122) and the year of initial calibration The scale
must be recalibrated if:
- the calibration mark has been damaged
- after calibrated scales have been repaired
- or after expiry of a period for recalibration as stipulated by national, legal regulations

M500020-210322-Rev009-UM-e 13
Technical speci ications
Power supply
Mains Adapter:
Mains voltage 115 or 230VAC
Frequency 50 or 60HZ
Power Consumption 11W in-charge, 2W off-charge
Main power circuit
Input Voltage 5 0VDC
Max Current (in-charger) 1000mA
Battery Requirement: Rechargeable Lithium Ion Cylindrical (18650x2) Cell 3 7V 4000mAH
Battery Operation:
Charger 7~10 Hours
Saving (350ohm, Backlight-ON 30Sec,
Shutdown 5Min) 200Hours
Serial Communications
EDP Port Full duplex RS-232
38400,19200,9600, 4800, 2400, 1200 bps
Printer Port Output-only RS-232 or active 20 mA current loop,
9600, 4800, 2400, 1200 bps
Both Ports 7 or 8 data bits; even, odd, or no parity
Operator Inter ace
Display 6-digit LCD display 7-segment, 0 815 in (21 2 mm) digits
LCD annunciator net, centre of zero, standstill, kg/lb/g/oz/ units,
Battery, Set points, Hold, HI/OK/LO
Keyboard 24-key flat membrane panel
Environmental
Operating Temperature –10 to +40°C (legal);
Storage Temperature –25 to +70°C
Humidity 0–95% relative humidity
Altitude 2000 m (6500 ft) maximum
Housing
Dimensions 18 6 cm x 9 5 cm x 10 cm
Weight 0 4 kg (without battery)
Material ABS plastic, white

14 M500020-210322-Rev009-UM-e
Con ormity declaration
Manu acturer
:
ADE Germany GmbH
Neuer Hoeltigbaum 15
22143 Hamburg / Germany
℡ +49 40 432 776-0
+49 40 432 776-10
info@ade-germany de
Noti ied body: TÜV NORD CERT GmbH
NB 0044 (93/42/EEC)
Langemarckstraße 20
D - 45141 Essen
℡ +49 201 825-0
+49 201 825-2517
info tncert@tuev-nord de
Declaration o con ormity by the manu acturer
This product has been manufactured in accordance with the harmonised European standards They con-
form to the regulations of the EC-directives listed below:
NAWI Directive 2014/31/EC
Medical Devices Directive 93/42/EEC
EMC Directive 2014/30/EC
Low Voltage Directive 2014/35/EC
RoHS Directive 2011/65/EC
in the currently valid versions
This declaration loses its validity if modifications are made to the scale without our approval
Hamburg, March 2021
ADE Germany GmbH
Neuer Hoeltigbaum 15
22143 Hamburg / Germany

M500020-210322-Rev009-UM-e 15

16 M500020-210322-Rev009-UM-e