AEMC SR759 User manual

99-MAN 100168 v5 08/07
OPERATION
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model SR759
•Connect the black and red terminals to the 2VAC range of your DMM or voltage measuring instrument.
Select the appropriate current range (1000/1V). Clamp the probe around the conductor to be tested
with the arrow pointed toward the load. Read the value display on the DMM and multiply it by the probe
ratio selected. (e.g. If reading = 185mV, the current flowing through the probe is 185 x 1000 =
185,000mA = 185A when 1mV/A range is selected.)
•For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
•When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
•Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
•For maintenance use only original factory replacement parts.
•To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
•To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents
into the probe.
Cleaning
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you
want a standard calibration, or a calibration traceable to N.I.S.T. (includes calibration certificate plus
recorded calibration data).
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive •Dover, NH 03820 USA
Tel: (800) 945-2362 (Ext. 360)
(603) 749-6434 (Ext. 360)
Fax: (603) 742-2346 or (603) 749-6309
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or
application of this instrument, please call our technical hotline:
(800) 343-1391 • (508) 698-2115 • Fax (508) 698-2118
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
www.aemc.com
AC Current Probe
Model SR759
User Manual
DESCRIPTION
The SR759 (Catalog #2116.33) is designed for use in industrial environments. The ergonomic design
allows itm to easily attach to cables or small bus bars. The “circular” jaws guarantee a very good
accuracy and low phase shift. The probe has four measurement ranges with the highest range measuring
to 1000Arms continuous and is compatible with any AC Voltmeter, multimeter, or other voltage
measurement instrument that has the following features: Range and resolution capable of displaying 1mV
of output per amp of measured current; Voltmeter accuracy of 0.75% or better to take full advantage of
the accuracy of the probe; Input impedance of ≥1MΩ.
WARNING
These safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument.
• Read the instruction manual completely and follow all the safety information before attempting
to use or service this instrument.
• Use caution on any circuit: Potentially high voltages and currents may be present and may
pose a shock hazard.
• Read the Safety Specifications section prior to using the current probe. Never exceed the
maximum voltage ratings given.
• Safety is the responsibility of the operator.
• ALWAYS connect the current probe to the display device before clamping the probe onto the
sample being tested.
• ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use.
Replace any defective parts immediately.
• NEVER use the current probe on electrical conductors rated above 600V in overvoltage
category III (CAT III). Use extreme caution when clamping around bare conductors or bus bars.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before
using the instrument.
This is a type A current sensor. This symbol signifies that application around and removal from
HAZARDOUS LIVE conductors is permitted.
DEFINITION OF MEASUREMENT CATEGORIES
Cat. I: For measurements on circuits not directly connected to the AC supply wall outlet such as
protected secondaries, signal level, and limited energy circuits.
Cat. II: For measurements performed on circuits directly connected to the electrical distribution system.
Examples are measurements on household appliances or portable tools.
Cat. III: For measurements performed in the building installation at the distribution level such as on
hardwired equipment in fixed installation and circuit breakers.
Cat. IV: For measurements performed at the primary electrical supply (<1000V) such as on primary
overcurrent protection devices, ripple control units, or meters.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately
with the carrier and notify your distributor at once, giving a detailed description of any damage.

ELECTRICAL SPECIFICATIONS
Current Range:
0.1 to 1000AAC, continuous cycle
with frequency ≤500Hz
Output Signal: 1mVAC/AAC (1VAC at 1000A)
Accuracy and Phase Shift*:
1A Range
Primary Current Accuracy % Phase Shift
0.001 to 0.01A 3% + 1mV Not Specified
0.01 to 0.1A ≤3% + 1mV Not Specified
0.1 to 1A ≤0.7% + 1mV ≤10°
1 to 1.2A ≤0.7% + 1mV ≤10°
10A Range
Primary Current Accuracy % Phase Shift
0.01 to 0.1A 1% + 0.2mV Not Specified
0.1 to 1A ≤0.5% + 0.2mV ≤5°
1 to 10A ≤0.5% ≤2°
10 to 12A ≤0.5% ≤2°
100A Range
Primary Current Accuracy % Phase Shift
0.1 to 1A 1% + 0.2mV Not Specified
1 to 10A ≤0.5% + 0.2mV ≤2°
10 to 100A ≤0.3% ≤1°
100 to 120A ≤0.2% ≤1°
1000 A Range
Primary Current Accuracy % Phase Shift
1 to 10A 1% + 0.2mV Not Specified
10 to 100A ≤ 0.5% + 0.2mV ≤ 2°
100 to 1000A ≤ 0.2% ≤ 1°
1000 to 1200A ≤ 0.2% ≤ 1°
*Reference conditions: 23°C ± 3°K, 20 to 75% RH, 48 to
65Hz, external magnetic field < 40A/m, no DC component,
no external current carrying conductor, test sample
centered. Load impedance 1MΩ
Overload: 1200A for 15 min on, 30 min off
Accuracy: Per IEC 185
Frequency Range: 30Hz to 3kHz; current derating
above 500Hz using the formula:
Load Impedance:
100kΩmin
Working Voltage: 600V Cat. III
Common Mode Voltage: 600V Cat. III
Influence of Adjacent Conductor: < 1mA/AAC
Influence of Conductor in Jaw Opening:
0.3% of Reading
Influence of Frequency:
1mV/A Range: < 1% R from 10Hz to 1kHz
10mV/A Range: < 1.5% R from 10Hz to 3kHz
100mV/A Range: < 2% R from 10Hz to 3kHz
1000mV/A Range: < 2% R from 30Hz to 1kHz
Open Voltage: 8V peak max.
MECHANICAL SPECIFICATIONS
Operating Temperature:
-14° to 122°F (-10° to 50°C)
Storage Temperature:
-4° to 158°F (-20° to 70°C)
Influence of Temperature:
< 0.15% per 10°K
Influence of Humidity:
From 10 to 90%: 0.1%
Jaw Opening: 2.25" (57mm) max
Maximum Conductor Size:
2.05" (52mm)
Envelope Protection: IP 40 (IEC 529)
Drop Test: 1 m (IEC 68-2-32)
Mechanical Shock:
100 g (IEC 68-2-27)
Vibration:
5 to 15Hz, 0.15mm (IEC 68-2-6)
15 to 25Hz, 1mm
25 to 55Hz, 0.25mm
Polycarbonate Material:
Handles: ABS Grey and
Lexan 500R, Red: UL94V0
Jaws: Lexan 500R, Red : UL94V0
Dimensions:
4.37 x 8.50 x 1.77" (111 x 216 x 45mm)
Weight:
1.21 lbs (550g)
Output:
SR759: 5 ft (1.5m) lead with safety
4mm banana plug
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation
between the primary or secondary and the outer
case of the handle conforms to IEC 1010-2-032.
Common Mode Voltage:
600V Category III, Pollution Degree 2
Dielectric Strength:
5550V, 50/60Hz between primary, secondary
and the outer case of the handle
Electromagnetic Compatibility:
EN 50081-1 Class B
EN 50082-2 Electrostatic discharge
IEC 1000-4-2
Radiated field IEC 1000-4-3
Fast transients IEC 1000-4-4
Magnetic field at 50/60 Hz IEC 1000-4-8
ORDERING INFORMATION
AC Current Probe SR759 ...............Cat #2116.33
Includes a user manual and a product warranty and
registration card
Accessories:
Leads, set of 2, 5 ft safety (1000V)...Cat #2111.29
Banana plug adapter
(to nonrecessed plug) ......................Cat #1017.45
4.37"
(111 mm)
3.98"
(101 mm)
2.25"
(57 mm)
1.22"
(31 mm)
8.50"
(216 mm)
1.71"
(43.5 mm)
1.38"
(35 mm)
1.77"
(45 mm)
3.90"
(99 mm)
2.05"
(52 mm)
E192383
1000 A x 1
2 x F
(in kHz)

ELECTRICAL SPECIFICATIONS
Current Range:
0.1 to 1000AAC, continuous cycle
with frequency ≤500Hz
Output Signal: 1mVAC/AAC (1VAC at 1000A)
Accuracy and Phase Shift*:
1A Range
Primary Current Accuracy % Phase Shift
0.001 to 0.01A 3% + 1mV Not Specified
0.01 to 0.1A ≤3% + 1mV Not Specified
0.1 to 1A ≤0.7% + 1mV ≤10°
1 to 1.2A ≤0.7% + 1mV ≤10°
10A Range
Primary Current Accuracy % Phase Shift
0.01 to 0.1A 1% + 0.2mV Not Specified
0.1 to 1A ≤0.5% + 0.2mV ≤5°
1 to 10A ≤0.5% ≤2°
10 to 12A ≤0.5% ≤2°
100A Range
Primary Current Accuracy % Phase Shift
0.1 to 1A 1% + 0.2mV Not Specified
1 to 10A ≤0.5% + 0.2mV ≤2°
10 to 100A ≤0.3% ≤1°
100 to 120A ≤0.2% ≤1°
1000 A Range
Primary Current Accuracy % Phase Shift
1 to 10A 1% + 0.2mV Not Specified
10 to 100A ≤ 0.5% + 0.2mV ≤ 2°
100 to 1000A ≤ 0.2% ≤ 1°
1000 to 1200A ≤ 0.2% ≤ 1°
*Reference conditions: 23°C ± 3°K, 20 to 75% RH, 48 to
65Hz, external magnetic field < 40A/m, no DC component,
no external current carrying conductor, test sample
centered. Load impedance 1MΩ
Overload: 1200A for 15 min on, 30 min off
Accuracy: Per IEC 185
Frequency Range: 30Hz to 3kHz; current derating
above 500Hz using the formula:
Load Impedance:
100kΩmin
Working Voltage: 600V Cat. III
Common Mode Voltage: 600V Cat. III
Influence of Adjacent Conductor: < 1mA/AAC
Influence of Conductor in Jaw Opening:
0.3% of Reading
Influence of Frequency:
1mV/A Range: < 1% R from 10Hz to 1kHz
10mV/A Range: < 1.5% R from 10Hz to 3kHz
100mV/A Range: < 2% R from 10Hz to 3kHz
1000mV/A Range: < 2% R from 30Hz to 1kHz
Open Voltage: 8V peak max.
MECHANICAL SPECIFICATIONS
Operating Temperature:
-14° to 122°F (-10° to 50°C)
Storage Temperature:
-4° to 158°F (-20° to 70°C)
Influence of Temperature:
< 0.15% per 10°K
Influence of Humidity:
From 10 to 90%: 0.1%
Jaw Opening: 2.25" (57mm) max
Maximum Conductor Size:
2.05" (52mm)
Envelope Protection: IP 40 (IEC 529)
Drop Test: 1 m (IEC 68-2-32)
Mechanical Shock:
100 g (IEC 68-2-27)
Vibration:
5 to 15Hz, 0.15mm (IEC 68-2-6)
15 to 25Hz, 1mm
25 to 55Hz, 0.25mm
Polycarbonate Material:
Handles: ABS Grey and
Lexan 500R, Red: UL94V0
Jaws: Lexan 500R, Red : UL94V0
Dimensions:
4.37 x 8.50 x 1.77" (111 x 216 x 45mm)
Weight:
1.21 lbs (550g)
Output:
SR759: 5 ft (1.5m) lead with safety
4mm banana plug
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation
between the primary or secondary and the outer
case of the handle conforms to IEC 1010-2-032.
Common Mode Voltage:
600V Category III, Pollution Degree 2
Dielectric Strength:
5550V, 50/60Hz between primary, secondary
and the outer case of the handle
Electromagnetic Compatibility:
EN 50081-1 Class B
EN 50082-2 Electrostatic discharge
IEC 1000-4-2
Radiated field IEC 1000-4-3
Fast transients IEC 1000-4-4
Magnetic field at 50/60 Hz IEC 1000-4-8
ORDERING INFORMATION
AC Current Probe SR759 ...............Cat #2116.33
Includes a user manual and a product warranty and
registration card
Accessories:
Leads, set of 2, 5 ft safety (1000V)...Cat #2111.29
Banana plug adapter
(to nonrecessed plug) ......................Cat #1017.45
4.37"
(111 mm)
3.98"
(101 mm)
2.25"
(57 mm)
1.22"
(31 mm)
8.50"
(216 mm)
1.71"
(43.5 mm)
1.38"
(35 mm)
1.77"
(45 mm)
3.90"
(99 mm)
2.05"
(52 mm)
E192383
1000 A x 1
2 x F
(in kHz)

99-MAN 100168 v5 08/07
OPERATION
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model SR759
•Connect the black and red terminals to the 2VAC range of your DMM or voltage measuring instrument.
Select the appropriate current range (1000/1V). Clamp the probe around the conductor to be tested
with the arrow pointed toward the load. Read the value display on the DMM and multiply it by the probe
ratio selected. (e.g. If reading = 185mV, the current flowing through the probe is 185 x 1000 =
185,000mA = 185A when 1mV/A range is selected.)
•For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
•When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
•Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
•For maintenance use only original factory replacement parts.
•To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
•To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents
into the probe.
Cleaning
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you
want a standard calibration, or a calibration traceable to N.I.S.T. (includes calibration certificate plus
recorded calibration data).
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive •Dover, NH 03820 USA
Tel: (800) 945-2362 (Ext. 360)
(603) 749-6434 (Ext. 360)
Fax: (603) 742-2346 or (603) 749-6309
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or
application of this instrument, please call our technical hotline:
(800) 343-1391 • (508) 698-2115 • Fax (508) 698-2118
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
www.aemc.com
AC Current Probe
Model SR759
User Manual
DESCRIPTION
The SR759 (Catalog #2116.33) is designed for use in industrial environments. The ergonomic design
allows itm to easily attach to cables or small bus bars. The “circular” jaws guarantee a very good
accuracy and low phase shift. The probe has four measurement ranges with the highest range measuring
to 1000Arms continuous and is compatible with any AC Voltmeter, multimeter, or other voltage
measurement instrument that has the following features: Range and resolution capable of displaying 1mV
of output per amp of measured current; Voltmeter accuracy of 0.75% or better to take full advantage of
the accuracy of the probe; Input impedance of ≥1MΩ.
WARNING
These safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument.
• Read the instruction manual completely and follow all the safety information before attempting
to use or service this instrument.
• Use caution on any circuit: Potentially high voltages and currents may be present and may
pose a shock hazard.
• Read the Safety Specifications section prior to using the current probe. Never exceed the
maximum voltage ratings given.
• Safety is the responsibility of the operator.
• ALWAYS connect the current probe to the display device before clamping the probe onto the
sample being tested.
• ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use.
Replace any defective parts immediately.
• NEVER use the current probe on electrical conductors rated above 600V in overvoltage
category III (CAT III). Use extreme caution when clamping around bare conductors or bus bars.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before
using the instrument.
This is a type A current sensor. This symbol signifies that application around and removal from
HAZARDOUS LIVE conductors is permitted.
DEFINITION OF MEASUREMENT CATEGORIES
Cat. I: For measurements on circuits not directly connected to the AC supply wall outlet such as
protected secondaries, signal level, and limited energy circuits.
Cat. II: For measurements performed on circuits directly connected to the electrical distribution system.
Examples are measurements on household appliances or portable tools.
Cat. III: For measurements performed in the building installation at the distribution level such as on
hardwired equipment in fixed installation and circuit breakers.
Cat. IV: For measurements performed at the primary electrical supply (<1000V) such as on primary
overcurrent protection devices, ripple control units, or meters.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately
with the carrier and notify your distributor at once, giving a detailed description of any damage.
Table of contents
Other AEMC Test Equipment manuals
Popular Test Equipment manuals by other brands

Martindale Electric
Martindale Electric NC4 instruction manual

Senior Technologies
Senior Technologies WanderGuard 18029 operating instructions

Mills
Mills C00-6435 user guide

Pall
Pall Palltronic Flow Check II Instructions for use

Finest
Finest 401 user manual

Keysight
Keysight InfiniiVision 4000G X Series Service guide

Agilent Technologies
Agilent Technologies BERT Serial Getting started

Changzhou Tonghui Electronic
Changzhou Tonghui Electronic TH8601 Operation manual

MTS Systems
MTS Systems Landmark 370 Series Product information

Tool it
Tool it PBT 812 operating manual

Oakton
Oakton ORPTestr 30 operating manual

RZproducts
RZproducts DOZ30 Operation manual