AEMC GROUNDFLEX 6474 User manual


Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
certifies that this instrument has been calibrated using
standards and instruments traceable to international
standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
NIST traceable calibration is only possible when
the Model 6474 is connected to the Ground Tester
Model 6472.
To obtain NIST calibration (for a nominal charge), the
instrument would need to be returned to our facility,
along with your Ground Tester Model 6472.
The recommended calibration interval is 12 months
and begins on the date of receipt by the customer. For
recalibration, please use our calibration services. Refer
to our repair and calibration section at www.aemc.com.
Serial #: ________________________________
Catalog #: ______________________________
Model #: 6474
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.
d.b.a AEMC®Instruments
www.aemc.com

2
GroundFlex™ Adapter Model 6474
Table of Contents
1. INTRODUCTION............................................................................... 3
1.1 International Electrical Symbols...........................................................4
1.2 DenitionofMeasurementCategories.................................................4
1.3 ReceivingYourShipment.....................................................................4
1.4 OrderingInformation ............................................................................5
1.4.1 AccessoriesandReplacementParts......................................5
2. GROUNDFLEX™ADAPTER ............................................................... 6
2.1 ControlFeatures ..................................................................................6
2.2 GroundFlex™Sensor............................................................................7
2.2.1 CalibrationofGroundFlex™Sensors.......................................7
2.3 MeasurementsinAUTOMode...........................................................10
2.3.1 Preparation ...........................................................................10
2.3.2 MakingaMeasurement(ResistanceofLegs) ......................11
2.3.3 OtherMeasurements ............................................................13
2.3.4 MANUALModeandSWEEPMode ......................................13
3. SPECIFICATIONS........................................................................... 15
3.1 Electrical.............................................................................................15
3.2 Mechanical.........................................................................................17
3.3 Environmental ....................................................................................17
3.4 Safety.................................................................................................17
4. MAINTENANCE ............................................................................. 18
3.1 Maintenance.....................................................................................18
3.2 Cleaning...........................................................................................18
RepairandCalibration........................................................................................19
TechnicalandSalesAssistance .........................................................................19
LimitedWarranty ................................................................................................20
WarrantyRepairs................................................................................................20

GroundFlex™ Adapter Model 6474
3
CHAPTER 1
INTRODUCTION
WARNING
Thesesafetywarningsareprovidedtoensurethesafetyofpersonnel.
Please read and complywiththeseprecautions:
Thisinstrumentisprotectedfromaccidentalvoltagesofnotmorethan•
50VwithrespecttoearthinmeasurementCat.IV.Theguaranteedlevel
ofprotectionofthisequipmentmaybecompromisedifusedinamanner
notspeciedbythemanufacturer.
Safetyistheresponsibilityoftheoperator.•
Allmetalobjectsorwiresconnectedtotheelectricalsystemshouldbe•
assumedtobelethaluntiltested.Groundingsystemsarenoexception.
Never exceed the maximum rated voltage and current, and the mea-•
surementcategory.
Neverexceedtheprotectionlimits,andalwayscomplywiththecondi-•
tionsandplaceofuse,indicatedinthespecications.
Donotusetheinstrumentoritsaccessoriesiftheyappeardamaged.•
Useaccessories thathave overvoltagecategory andservice voltages•
greaterthanorequaltothoseoftheinstrument(Cat.IV50V).Useonly
accessoriesthatcomplywithsafetystandards(IEC61010-2-031&32).
Weartheappropriateprotectivegear(insulatingbootsandgloves).•
CheckthatnoterminalisconnectedandtheswitchissettoOFFbefore•
openingthedevice.
Useonlythechargingunitsuppliedwiththeinstrumenttorechargethe•
battery.
Troubleshootingandmetrologicalvericationproceduresmustonlybe•
performedbyqualied,approvedpersonnel,orthefactory.
NOTE:• Thepotentialsonthevariousrodsusedforanearthmeasurement
maybedifferentifanearbyelectricalinstallationisdefectiveorcertain
weather conditions prevail (thunderstorms). It is up to the operator
to decide whether to continue or postpone measurements in these
situations.

4
GroundFlex™ Adapter Model 6474
1.1 International Electrical Symbols
Signifies that the instrument is protected by double or reinforced insulation.
This symbol on the instrument indicates a WARNING that the operator must
refer to the user manual for instructions before operating the instrument. In
this manual, the symbol preceding instructions indicates that if the instructions
are not followed, bodily injury, installation/sample and/or product damage may
result.
Risk of electric shock. The voltage at the parts marked with this symbol may be
dangerous.
Ground/Earth symbol
In conformity with WEEE 2002/96/EC
1.2 Definition of Measurement Categories
Cat. I: For measurements on circuits not directly connected to theAC supply wall
outletsuchasprotectedsecondaries,signallevel,andlimitedenergycircuits.
Cat. II: Formeasurementsperformedoncircuitsdirectlyconnectedtotheelectrical
distributionsystem.Examplesaremeasurementsonhouseholdappliancesor
portabletools.
Cat. III:
Formeasurementsperformedinthebuildinginstallationatthedistributionlevel
suchasonhardwiredequipmentinxedinstallationandcircuitbreakers.
Cat. IV: Formeasurementsperformedattheprimaryelectricalsupply(<1000V)such
asonprimaryovercurrentprotectiondevices,ripplecontrolunits,ormeters.
1.3 Receiving Your Shipment
Uponreceivingyourshipment,makesurethatthecontentsareconsistentwiththe
packinglist.Notifyyourdistributorofanymissingitems.Iftheequipmentappears
tobedamaged,leaclaimimmediatelywiththecarrierandnotifyyourdistributor
atonce,givingadetaileddescriptionofanydamage.Savethedamagedpacking
containertosubstantiateyourclaim.

GroundFlex™ Adapter Model 6474
5
1.4 Ordering Information
NOTE: The GroundFlex™ Adapter Model 6474 is a factory supplied product
and only operates with the Ground Tester Model 6472 Kit - 500 ft.
GroundFlex™ Adapter Model 6474...............................................Cat. #2136.01
Includes the GroundFlex™ Adapter Model 6474, four GroundFlex™ sensors (5m/16ft) with
twelve color-coded rings, connection lead, two extension leads on H reel (black/green) with
color-coded alligator clips, six BNC extension leads, calibration loop, three C-clamps, carrying
case for meter and a user manual.
GroundFlex™ Field Kit (Model 6472 and 6474) ...........................Cat. #2136.03
Includes the GroundFlex™ Adapter Model 6474 (Cat. #2136.01 above), Ground Resistance
Tester Model 6472, rechargeable NiMH batteries, optical USB cable, power adapter 110/240V
with power cord 115V US, two 500 ft color-coded leads on spools (red/blue), two 100 ft color-
coded leads (hand-tied, green/black), one 30 ft lead (green), four T-shaped auxiliary ground
electrodes, set of ve spaded lugs, one 100 ft AEMC®tape measure, DataView®software,
ground tester workbook CD, carrying bag for meter, carrying bag for kit, hard carrying case,
product warranty and registration card and a user manual.
1.4.1 Accessories and Replacement Parts
Bag#6-CarryingBagforGroundKits............................................. Cat. #2119.82
TapeMeasure-AEMC®(100ft) .......................................................Cat. #2130.60
GroundTesterDVD/WorkbookSet...................................................Cat. #2130.64
Setoftwo,14.5"T-shapedAuxiliaryGroundElectrodes..................Cat. #2135.39
OpticalUSBCable............................................................................Cat. #2135.41
Replacement-OneGroundFlex™(5m /16ft)Sensor.......................Cat. #2135.73
Replacement-ConnectionLeadforModel6472/6474....................Cat. #2135.75
Replacement-OneBNC(15m/50ft)ExtensionLead......................Cat. #2135.76
Replacement-Setof12,GroundFlex™Rings
(red,yellow,brown,orange) .............................................................Cat. #2135.77
Replacement-GreenExtensionLead .............................................Cat. #2135.78
Replacement-BlackExtensionLead...............................................Cat. #2135.79
Replacement-Setof3,C-clamps....................................................Cat. #2135.80
Replacement-CalibrationLoop.......................................................Cat. #2135.82
Replacement-CarryingCase ..........................................................Cat. #2135.83
Replacement-SafetyAlligatorClip(black) ......................................Cat. #2140.53
Replacement-SafetyAlligatorClip(green) .....................................Cat. #2140.69
Order Accessories and Replacement Parts Directly Online
Check our Storefront at www.aemc.com/store for availability

6
GroundFlex™ Adapter Model 6474
CHAPTER 2
GROUNDFLEX™ADAPTER
2.1 Control Features
50V CAT IV
1-2-3-4
1-2-3
1-2
1
2
3
4
For Use with the
Ground Tester Model 6472
GroundFlex™
Sensor
SENSOR TURNS
SENSITIVITY
To Model 6472
Overload
INPUT
1
2
3
4
GroundFlex™Adapter
MODEL 6474
2
4
1
6
5
3
1. Inputs1through4forfourGroundFlex™sensors
2. ConnectorforconnectionleadtotheGroundTesterModel6472
3. OVERLOAD indicator
4. SelectorforNUMBEROFTURNS:1,2,3or4
5. SelectorforSENSITIVITY:x10,x1orx1/10
6. INPUTSELECTION:1,2,3,4,1-2,1-3or1-4

GroundFlex™ Adapter Model 6474
7
2.2 GroundFlex™Sensor
TheGroundFlex™sensorsareplacedaroundaconductor(pylonlegorpole)and
closedusingascrewconnector.Thisclosedloopisthenusedtomakeacontact-
freemeasurement of the currentowingin the conductor itis wrapped around.
These sensors are like current measurement clamps but are able to enclose
objectsthataremuchlarger.GroundFlex™sensorsareavailableinlengthsofup
toapprox16ft(5m)andcanbeplacedaroundobjects(suchaspylonlegs)upto
inapprox5ftdiameter(1.5m).
2.2.1 Calibration of GroundFlex™Sensors
GroundFlex™sensors need to be calibrated before the rst use. Calibration is
necessaryforanewsensor,onethatisnolongeridentied,orwhenasensorhas
beenreplaced.ThecalibrationcoefcientsarethenrecordedintheModel6474.
NOTE: Do not move or touch the GroundFlex™sensor or the calibration loop
during the calibration. This could cause errors.

8
GroundFlex™ Adapter Model 6474
Startbyidentifyingeachsensorusingtheidenticationrings(brown,red,orange,
yellow)suppliedwiththedevice.
ConnectthecalibrationloopbetweenterminalsHandEoftheModel6472.•
ConnecttheModel6472and6474togetherusingtheconnectioncable.•
ConnecttheGroundFlex• ™sensorwiththebrownringtoinput1oftheModel
6474,theonewiththeredringtoinput2,theonewiththeorangeringtoinput
3,andtheonewiththeyellowringtoinput4.
Coil the GroundFlex• ™ sensor four times around in the calibration loop and
closeitusingitsscrewconnector.Placetheconnectorasfaraspossiblefrom
thecalibrationdevice.
NOTE: • The sensors have arrows to indicate a direction.
• All sensors should face the same direction.
(clockwise or counter-clockwise)
• All sensors should have the same number of wraps.
• Sensors should encircle both leg AND ground system.
6472 6474

GroundFlex™ Adapter Model 6474
9
1. Settheswitchofthe6472toSET-UP.
2. SettheSENSOR TURNSswitchto4.
3. SettheSENSITIVITYswitchtox1.
4. SettheINPUT SELECTIONswitchto1.
1-2-3-4
1-2-3
1-2
1
2
3
4
SENSOR TURNS
SENSITIVITY INPUT
5. PresstheHz/OPTIONSbutton5times.
ThedevicestartsbycomparingthecurrentIIntinjectedbythedeviceintothe
calibrationlooptothecurrentIAmeasuredbytheGroundFlex™sensor.Ifthey
aredifferent,the"CAL"symbolblinkstorecommendacalibration.
x 5
4
H
E
S
x
11
AUTO
mA
m A
NOTE: To avoid measurement errors, do not touch the GroundFlex™sensor or
the calibration loop during the calibration.
6. Tostartthecalibration,press2nd + START.

10
GroundFlex™ Adapter Model 6474
7. ThedevicecalculatesandstoresacalibrationcoefcientfortheGroundFlex™
sensorconnectedtochannel1.
4
H
E
S
x
11
AUTO
mA
m A
Thisoperationmustberepeatedforchannels2,3and4;theINPUT SELECTION
switchmustbesettothecorrespondingpositioneachtime.
Thecalibrationvaluesarenowstoredintomemory.
NOTE: Always connect the same sensor to the same input.
2.3 Measurements in AUTO Mode
2.3.1 Preparation
TheGroundFlex™AdapterModel6474isdesignedtobeusedonlywiththeModel
6472device.The twodevices must be connected using thespecial connection
cablesupplied.TheModel6474hasnoON/OFFbutton,andispoweredonlyby
theModel6472,viathisconnectioncable.
TheModel6474isintendedprimarilytomeasurethecurrentthatowsthrougha
high-voltagepylontotheground.Youcanwrapupto4legsofthepylonseachwith
aGroundFlex™sensorandmeasurethecurrentowingtoearththrougheachof
thelegsorthroughseverallegs.
Aftercalibration,theGroundFlex™sensorsmustalwaysbeconnectedtotheirre-
spectivechannels.
1. SettheswitchtoGroundFlex™.
WhenyouconnecttheGroundFlex™AdapterModel6474tothe6472,aself-
testwillbegin.Duringthistest,theOVERLOADchecklightonthe6474islit.
2. Connecttherequirednumberofsensorstochannels1,2,3,and/or4.
NOTE: The direction of coiling of the GroundFlex™sensors around a conductor
has no effect on the measurement, but all of the GroundFlex™sensors must
be coiled in the same direction and have the same orientation (check the
arrow on the connector of the sensor) and they must all have the same
number of turns.

GroundFlex™ Adapter Model 6474
11
6474
To the Model 6472
NOTE: To increase measurement sensitivity, we recommend when possible,
wrapping a maximum number of turns (up to 4) around the structure.
3. SettheINPUT SELECTIONswitchaccordingly.Youcanmeasurethecurrent
owingthrough oneGroundFlex™ sensor (1,2, 3,or 4),or through several
sensors(1,2or1,2,3),orthroughallofthesensors(1,2,3,4).
4. Select how many times the sensor is wrapped around the conductor to be
measured(from1to4turns)usingtheSENSOR TURNSswitch.
5. SettheSENSITIVITYswitchtothedesiredsensitivity:x1/10,x1,orx10.The
choiceofsensitivitydependsontheexpectedcurrentvalue.Asaprecaution,
alwaysstartwiththelowestsensitivity,x1/10,andincreaseitasneeded.
2.3.2 Making a Measurement (Resistance of Legs)
1. PlaceauxiliaryelectrodesH (Z) and S (Y)onoppositesidesofthepylon,as
farawayaspossible(100-150ft)and,ifpossible,perpendiculartothehigh-
voltageline(depthisnotimportant).Thiswilleliminateinterferencewiththe
measurementbystray voltages or currentsinducedunder the high-voltage
line.
2. In order to avoid electromagnetic interference, it is best to unwind the full
lengthofeachcablefromthereel,tokeepthecablesasfarapartaspossible
ontheground,takingcarenottoformloops,andtoavoidplacingthecables
nearorparalleltometallicconductors(cables,rails,fences,etc.).Effectsof
interferencearemeasuredinmilli-ohms.
3. ConnecttheseauxiliaryelectrodestoterminalsH (Z) and S (Y).
4. TheterminalsES (Xv) and E (X)mustbeconnectedtoaconductingmetallic
partofthepylonlocatedabovetheGroundFlex™sensors.Thisensuresmea-
surementofthecurrentowingdownwardandtowardstheearth(otherwise,
youwouldmeasurethecurrentowingupward,fromtheearthtothetipofthe
pylon).

12
GroundFlex™ Adapter Model 6474
5. Connect the required number of GroundFlex™ sensors to channels 1, 2, 3,
and/or4(thechannelsforwhichtheyarecalibrated)andplacetheGround-
Flex™sensorsaroundthelegsofthepylon.Theorientationhasnoeffecton
themeasurement,butalloftheGroundFlex™sensorsmustbeorientedinthe
same directionandtheymust allhavethesame number of turns.
E
S
H
RS
RH
ES
1
2
3
4
6474
6472
Overhead earth wire
High-voltage
lines
4 GroundFlex™ sensors
connected with the
same orientation
WARNING: If the GroundFlex™Adapter Model 6474 is connected to the earth
tester, the earth potential connected to the terminals E (X) and ES (Xv) will
also be present at the BNC connectors of the GroundFlex™sensors and at the
connecting cable between the adapter and tester. Whenever any doubt exists
about this earth potential, the user should proceed to a voltage measurement
using the S (Y) and ES (Xv) terminals of the tester.
1. Setthe3switchesoftheGroundFlex™Adapteraccordingtotheconnection
made.
2. StartthemeasurementbypressingtheSTART/STOPbutton.
1
AUTO
k
7
mA
V
H
S
ES
E
S
x
12

GroundFlex™ Adapter Model 6474
13
Todisplaythemeasurementparameters,pressDISPLAY several times.
Thedevicedisplaysthefollowingparameters:
R
SEL,US-ESanditsfrequency,ISEL,R-Act(RPASS),U-Act(US-ESanditsfrequency,
UH-Eanditsfrequency),I-Act(ISELanditsfrequency).
TomeasuretheresistancesofelectrodesH (Z) and S (Y),oriftheresistanceof
theelectrodesistoohigh,startthemeasurementbyalongpressoftheSTART/
STOPbutton.
R
H,RS,UH-E,IH-Ewillbedisplayed.
2.3.3 Other Measurements
Asexplainedabove,terminalsES (Xv) and E (X)mustbeconnectedtoametallic
partofthepylonabovetheGroundFlex™sensorinordertomeasurethecurrent
owing ISEL downward through the sensor, therefore to the earth to measure
groundresistanceoftheleg.IfconnectionsES (Xv) and E (X)areplacedbelow
theGroundFlex™sensor,thelattermeasuresthecurrentowingtowardsthetop
ofthepylonviatheoverheadearthwireatitstip,andyoucaninthiswayevaluate
thequality of this counter poise.Youcanevenmakethismeasurementwithout
movingthecontactsofES (Xv) and E (X) onthepylonbysimplypassingthe2con-
nectingcablesthroughtheGroundFlex™sensor.ThecurrentinconnectingcableE
(X) andthecurrentthroughthelegofthepylonthencanceleachotherout.
Youcanalsomeasureany leakage currents(I-Act)owingthrougheachlegof
thepyloninnormaloperationofthehigh-voltageline.Ifyounddifferentleakage
currentsineachleg,buttheycanceloutwhenaddedtogether,youcanconclude
thatthesecurrents,inducedbytherotatingeldofthehigh-voltagelines,indicate
an earthing fault of the pylon via the overhead earth wire at the tip of the
mast.
2.3.4 MANUAL Mode and SWEEP Mode
Use the Hz/OPTIONS button and ► button to change to MANUAL mode. The
followingparameterscanbemodied:
128Hzblinks• Changethetestfrequency
Testvoltageblinks• Switchbetween16and32V
Youcanalsoperformafrequencysweepandasmoothing.
Inthesweepmode,themeasurementscanbemadeatfrequenciesupto5kHz.
Onasinglepylon,frequenciesoftheorderof5kHzdonotalwaysentailinductive
effects,butthe"currentloops"formedbytheadjacentpylonsandthereturnchan-
nelthroughthegroundmayresultininductancesthatcanbedetectedonlyathigh
frequencies.

14
GroundFlex™ Adapter Model 6474
Anequivalentdiagramofthecomponentsformingthiscircuitisasfollows:
Point of injection
of the current
Overhead
earth wire
LLL
Rpylon
Next pylon
to the left
Pylon
measured
Next pylon
to the right
A measurement in sweep mode (SWEEP) of an installation like this having 10
pylonsgivesthefollowing resultsassuminganearthresistance Rpylonof10Ωfor
eachpylonandaninductanceLof550µHforthelinkviatheoverheadearthwire
atthetipofthepylons:
Impedance of the pylons
0
2
4
6
8
10
12
14
16
0100020003000 4000 5000
F (Hz)
Z (ohm)
MeasurementofapylonearthimpedancewithGroundFlex™
sensors
Conventional4polemeasurement
Measurementintothetopofapylon(iterativenetworkimpedance)

GroundFlex™ Adapter Model 6474
15
CHAPTER 3
SPECIFICATIONS
3.1 Electrical
Measuring Method:
Voltage/CurrentmeasurementwitharectangularACsignal
Short Circuit Current: > 200mAAC
Noise Suppression: >80dBatfrequenciesdifferingby20%ormorefromthe
testfrequency
Max. Overload: 250Vrms
Max. Value for RH& RS: 100kΩ
Measuring Time: Short push on START: approx.7sforrstvalueofREat
128Hz,then3measurementspers.
Long push on START:approx.15sforrstvalueofREat
128Hz,then3measurementspers.
Pylon Measurement with GroundFlex™
Measurement
Range
0.067 to
9.999Ω
10.00 to
99.99Ω
100.0 to
999.9Ω
1.000 to
9.999kΩ
10.00 to
99.99kΩ
Resolution 0.001Ω0.01Ω0.1Ω1Ω10Ω
Test Voltage 16 or 32V user selectable
Frequency From 41 to 128Hz selectable
Resistance Measurement
Frequency From 41 to 5078Hz selectable
Accuracy ±5% of Reading + 1ct @ 128Hz
Theintrinsic error specied refers toreference conditions with atestvoltage of
32V,testfrequencyof128Hz,RH,andRS=1kΩ,noexternalvoltage.
TheoperatingerrorformeasurementsofRH,RSand REisthesameasthatspecied
for4-Poleearthresistancemeasurement.
TheoperatingerrorofACresistancemeasurementscanbelessthanthatspecied
forvoltageorcurrentbecausefrequencycharacteristicsofthevoltagechannelare
matchedtothoseofthecurrentchannel.

16
GroundFlex™ Adapter Model 6474
For test frequencies between 41 and 5087 Hz, between 1 and 4 turns of the
GroundFlex™ sensor, and a selection of 1 to 4 measurement channels, the
operatingaccuraciesarethefollowing:
SENSITIVITY and minimum ISEL minimal Operating accuracy for RS-ES
S x 1/10 ISEL > 10mA ± (10% + 4cts)
S x 1 ISEL > 5mA ± (5% + 4cts)
S x 10 ISEL > 5mA ± (5% + 4cts)
5mA > ISEL > 0.5mA ± (15% + 10cts)
Current Measurement with GroundFlex™
Measurement Range 0.0 to
99.9mA (1)
100 to
999mA
1.00 to
9.99A
10.0 to
99.9A
Resolution 0.1mA (1) 1mA 0.01A 0.1A
Frequency 16 to 400Hz
Accuracy <5% of Reading @ 50/60Hz
(1): Valid only in the SENSITIVITY x 10 setting
ThecurrentmeasureddependsonthenumberofturnsoftheGroundFlex™sensor
aroundtheconductortobemeasured.(e.g.if the GroundFlex™sensor is wrapped
4 times around a conductor carrying a current of 1A, the input signal will be the
same as if the GroundFlex™sensor were wrapped once around a conductor car-
rying a current of 4A. The operating error is therefore indicated for an input signal
current in A.tr (Ampere.turns).
TheminimumcurrentmeasureddependsonthesettingoftheSENSITIVITYswitch
asindicatedbythetablebelow:
SENSITIVITY IMIN (A.tr) Points additional error (pt-er)
x 10 0.01 5
x 1 0.04 2
x 1/10 0.16 2
Accuracy
Current (A * turn) 16 to 49Hz 50 to 99Hz 100 to 400Hz
IMIN to 0.399 ± (20% + pt-er) ± (5% + pt-er) ± (3% + pt-er)
0.4 to 39.9 ± (10% + 2ct) ± (3% + 2ct) ± (3% + 2ct)
40 to 99.9 ± (10% + 2ct) ± (3% + 2ct) ± (20% + 2ct)

GroundFlex™ Adapter Model 6474
17
3.2 Mechanical
Dimensions:10.7x9.76x5.12"(272,248x130mm)
Weight: 7lbs(3.2kg)approx
Case Material:ULV0Polypropylene
Terminals: 4mmrecessedbananajacks
Case Protection: EN60529-IP53(coverclosed)
Drop Test: Per EN 61010-1
Vibration Test: Per EN 61557-1
3.3 Environmental
Operating Temperature: 32°to113°F(0°to45°C);0to90%RH
SpeciedOperatingTemperature(1):0°to95°F(0°to35°C);0to75%RH
Storage Temperature:-40°to158°F(-40°to70°C);0to90%RH
Altitude: <3000m
(1): This range corresponds to the one defined by standard EN 61557, for which an operating error including the
quantities of influence is defined. When the device is used outside this range, 1.5%/10°C and 1.5% between
75 and 90% RH must be added to the operating error.
3.4 Safety
Electromagnetic Compatibility
ThisinstrumentsatisestheEMCandLVDdirectivesrequiredfortheCEmarking
andproductstandardIEC61326-1(Ed.97)+A1(Ed.98).
Immunityinindustrialenvironment•
Emissions in residential environment.•
ElectricalsafetyaccordingtoEN61010-1(Ed.2of2001)
MeasurementaccordingtoEN61557(Ed.2of2007)parts1,4and5.
Cat.IV,50V
PollutionDegree2
*All specications are subject to change without notice

18
GroundFlex™ Adapter Model 6474
CHAPTER 3
MAINTENANCE
3.1 Maintenance
Please make sure that you have read and fully understand the WARNING
section on page 3.
Toavoid electrical shock,do not attemptto perform anyservicing unless•
youarequaliedtodoso.
Toavoidelectricalshockand/ordamagetotheinstrument,donotgetwater•
orotherforeignagentsintothecase.
Turn the instrument OFF and disconnect the unit from all circuits before•
openingthecase.
Usespeciedsparepartsonly.•
3.2 Cleaning
Disconnect the instrument from any source of electricity.
Useasoftclothlightlymoistenedwithsoapywater.Wipewithamoistclothand
thendrywithadrycloth.Neverusealcohol,solventsorhydrocarbons.
Other manuals for GROUNDFLEX 6474
1
Table of contents
Other AEMC Test Equipment manuals

AEMC
AEMC 6416 User manual

AEMC
AEMC Simple Logger AmpFlex AL24-2500 User manual

AEMC
AEMC JM845A User manual

AEMC
AEMC 3620 User manual

AEMC
AEMC MN213 User manual

AEMC
AEMC 6470 User manual

AEMC
AEMC 4620 User manual

AEMC
AEMC 6240 User manual

AEMC
AEMC FlexProbe 24-3001 User manual

AEMC
AEMC 6400 series User manual