BIOS Diagnostics BD802 User manual

BIOS
Diagnostics
™
BD802
instruction Manual
ear/foreheaD
DiGital therMoMeter
DUAL

Table of Contents
1 Important Safety Instructions
2 Product Diagram
3 Inserting the battery
4 Real Time Clock Setting
5 Directions for Use - Ear Mode
6 Directions for Use - Forehead Mode
7 Fever Alarm
8 Memory Storage
9 Measuring Temperature Under Scan Mode
10 Stopwatch Mode
11 Measuring the Room Temperature
12 Changing from Celsius to Fahrenheit
13 Care and Cleaning
14 Battery Replacement
15 Specifications
16 Troubleshooting
17 Two Year Warranty
BD802 Dual Ear/Forehead Digital Thermometer
Instruction Manual
1

Dual Ear/Forehead Digital Thermometer
BD802
Thank you for purchasing the BIOS Diagnostics™ Dual Ear/Forehead
Thermometer.
The BIOS Diagnostics™ Dual Ear/Forehead Thermometer is a high quality
product incorporating the latest technology and tested in accordance with
international standards.
With its unique technology, this thermometer can provide safe, accurate and
reliable readings with each measurement on people over 2 years of age.
Features:
• 9 sets of memory with time and date
• Probe cover free
• Room temperature function
• Fever alarm
• °C/°F switchable
Please read these instructions carefully before using this thermometer and keep
both in a safe place.
2

1. IMPORTANT SAFETY INSTRUCTIONS
• Never use the thermometer for purposes other than those for which it
is intended. Please follow the general safety precautions when using
on children.
• Never immerse the BIOS Diagnostics™ Dual Ear/Forehead Thermometer
into water or other liquids (not waterproof). For cleaning and
disinfecting please follow the instructions in the ‘Care and Cleaning’
section.
• Keep the thermometer away from direct exposure to the sun and keep it
in a dust-free, dry area between -20°C to 50°C (-4°F to 122°F).
• Do not use the thermometer if there are any signs of damage on the
measuring tip or the instrument itself. If damaged, do not attempt to
repair the instrument.
• This BIOS Diagnostics™ Dual Ear/Forehead Thermometer consists of
high quality precision parts. Do not drop this instrument. Protect it from
severe impact and shock.
WARNING
• Use of this BIOS Diagnostics™ Dual Ear/Forehead Thermometer is not
intended as a substitute for consultation with your physician.
• Keep the forehead and protective probe cover out of the reach of
children. Small pieces may be ingested and present a choking hazard.
• It is not recommended to use infrared thermometers on children under 2
years of age.
• Infrared thermometers should not be used to manage important health
3

4
concerns, and therefore, parents should not rely solely on temperature
readings to evaluate the status of your child’s health. If you have any
concerns, consult with your doctor.
2. PRODUCT DIAGRAM
Ear Probe
Forehead Probe Scan Button
ON / MEM Button
Protective Probe Cover
LCD Display
FIGURE 1
Battery Cover Pinhole
Battery Cover
Ear Probe
Forehead ProbeProtective Probe Cover

3. INSERTING THE BATTERY
1. Open the battery cover located on the lower inner side of the
thermometer.
2. Insert the battery (included) under the metal hook on the left side and
press the battery down until you hear a click.
3. Replace the battery cover.
4. REAL TIME CLOCK SETTING
1. Power on: Press the ‘ON/MEM’ button. The thermometer is ready for
use after hearing two beeping sounds.
2. Press and hold the ‘ON/MEM’ button for 5 seconds until the <<SET>>
icon starts to flash in the upper right hand corner of the screen. This
indicates that you are in the setting mode screen. (see Figure 2 below)
3. Press the ‘SCAN’ button to make the ‘24’ icon
flash. This indicates that you are in the 12/24 hour
setting screen. Press the ‘SCAN’ button again to
change/set the time format from ‘24’ (24 hour) to
‘12’ (12 hour).
4. Press the ‘ON/MEM’ button to move through the setting sequence
shown below. Use the ‘SCAN’ button to change the hour, minutes, year,
month, and day settings while flashing.
FIGURE 2
SET
FIGURE 2
5

SETTING SEqUENCE
HOUR > MINUTES > YEAR > MONTH > DAY >Then leave the setting
screen.
NOTE: The device will automatically shut off if left idle for more than
one minute. The LCD will show the date, time, and room temperature
continuously.
5. DIRECTIONS FOR USE – EAR MODE
1. If Forehead Probe is attached, remove following the steps below:
Attaching or Removing Forehead Probe: On the backside of the
thermometer you will see a small arrow close to the head. When
removing or attaching the forehead probe line up the arrow on the
device head with the arrow on the probe. Only when the arrows
are aligned will the probe attach or remove. Always make sure the
probe is clean and not damaged.
2. Power On: Press the ‘ON/MEM’ button.
3. The default mode for the thermometer is the ear
mode when the forehead probe has been removed.
The thermometer is ready for use after you see the
icon on the display screen and hear two beeping
sounds (see Figure 3).
FIGURE 3
6

4. To measure ear temperature accurately, first
straighten the ear canal. To straighten the ear
canal for ages 2 years to adult: gently pull
the outer ear “up and back”. Fully insert the
thermometer gently into the ear.
5. Press and hold the ‘SCAN’ button until you hear a
beeping sound. Remove the probe from ear and the
temperature measurement is displayed on the screen.
• The key to an accurate reading is straightening the ear canal so
that measurements are taken directly from the eardrum.
• It is recommended that you measure the temperature 3 times
in the same ear. If the 3 measurements are different, select the
highest temperature.
• Always insert the thermometer in the same ear with the same
direction and depth.
• Always clean the probe accordingly (see Care and Cleaning)
after each use.
• Before measurements, remain in a stable environment and avoid
exercise and bathing for 30 minutes prior.
7
Ages 2 years to adult

6. DIRECTION FOR USE – FOREHEAD MODE
1. If probe is not attached, attach following the steps below.
Attaching or Removing Forehead Probe: On the backside of the
thermometer you will see a small arrow close to the head. When
removing or attaching forehead probe line up the arrow on the device
head with the arrow on the forehead probe. Only when the arrows are
aligned will the probe attach or remove. Always make sure the probe is
clean and not damaged.
2. Power On: Press the ‘ON/MEM’ button.
3. The default mode for the thermometer is the forehead mode when the
ear probe is attached. The thermometer is ready for use after you see
the icon on the display screen and hear two beeping sounds.
4. Place the thermometer against the required left or right temple location
(see Figure 5) .
5. Press the ‘SCAN’ button and gently scan around the temple area (see
Figure 5.1). While scanning, you will hear a beeping sound, which
indicates that the newest measurement is taking place. Measurement
has been completed after two short beeps are heard and the forehead
icon stops flashing.
8

6. Forehead temperature is displayed in oral mode. This mode converts the
forehead temperature to display its ‘oral equivalent’ value.
7. Before measurement, remain in a stable environment and avoid
exercise and bathing for 30 minutes prior.
8. Remember to keep the temple area clean and free from cosmetics, hair,
perspiration, and avoid any scarring while taking temperature.
9. The time needed for a measurement will vary depending on the device
obtaining an accurate forehead temperature. Normally 5 to 8 seconds
(up to 30 seconds).
FIGURE 5 FIGURE 5.1
9

7. FEVER ALARM
If the thermometer detects a body temperature greater than 37.5°C / 99.5°F
in either the Ear or Forehead modes it will make a long beeping sound followed
by 3 short beeps, indicating a potential fever.
8. MEMORY STORAGE
When in ‘power on’ mode, press ‘ON/MEM’ to view the stored temperatures.
The stored temperatures are indicated with the .
Keep pressing ‘ON/MEM’ to scroll through each stored temperature.
• The thermometer stores 9 sets of temperature measurements.
9. MEASURING TEMPERATURE UNDER SCAN MODE
1. Power On: Press the ‘ON/MEM’ button.
2. After the power is on, press and hold the ‘ON/MEM’
button and then at the same time press the ‘SCAN’
button once for the to appear on the display (see
Figure 6).
3. Hold the thermometer against the desired surface to be measured and
press the ‘SCAN’ button to obtain the real time temperature. If you
press and hold the ‘SCAN’ button the temperature will be continuously
updated.
FIGURE 6
10

• This mode shows the actual, unadjusted surface temperatures,
which are different from the body temperature.
10. STOPWATCH MODE
1. Power On: Press the ‘ON/MEM’ button.
2. After the power is on, press and hold the ‘ON/
MEM’ button and then at the same time press the
‘SCAN’ button twice until the appears on the
display (see Figure 7).
3. To start and stop counting press the ‘SCAN’ button.
To reset press the ‘ON/MEM’ button.
NOTE: The stopwatch will run to a maximum of 3 minutes.
11. MEASURING THE ROOM TEMPERATURE
The room temperature is continuously displayed when the unit has been left idle
for more than 1 minute. It is shown on the display screen with the icon.
The temperature will be updated every 15 minutes.
• For a more accurate room temperature reading, the thermometer should
be placed on a table and away from direct sunshine or air conditioner
flow.
FIGURE 7
11

Note: The real time clock must be set up first before this function is
available (see Section 3 for how to set up the time and date.)
12. CHANGING FROM CELSIUS TO FAHRENHEIT
1. In ‘Power Off; mode (The LCD only shows the date, time, and room
temperature). Press and hold the ‘SCAN’ button, then immediately press
and release the ‘ON/MEM’ button. After 3 seconds the °C icon will be
replaced with the °F icon.
2. Repeat steps to switch back and forth between Celsius and Fahrenheit.
13. CARE AND CLEANING
• After each use, use an alcohol swab or cotton swab moistened
with alcohol (70% Isopropyl) to clean and disinfect the probes and
thermometer casing.
• Keep the unit dry and away from any liquids and direct sunlight.
• Storage temperature range -20°C to 50°C (-4°F to 122°F).
12

14. BATTERY REPLACEMENT
1. Open the battery cover located on the lower inner side of the
thermometer.
2. Hold the device and insert a small screwdriver or pen to remove the
existing battery.
3. Insert the new battery under the metal hook on the left side and press
the battery down until you hear a click.
4. Replace the battery cover.
15. SPECIFICATIONS
Temperature Measurement Range:
• Forehead / Ear Mode: 34°C to 42.2°C (93.2°F to 108°F)
• Scan Mode: -22°C to 80°C (-7.6°F to 176°F)
Operating Temperature Range:
• 10°C to 40°C (50°F to 104°F)
Storage Temperature Range:
• -20°C to 50 °C (-4°F to 122°F), RH ≤85 %
• Transportation temperature shall be less than 70 °C, RH ≤95 %
13

Compliance with the following standards:
• ASTM E1965-98, EN12470-5:2003 Clinical thermometers-Part 5:
Performance of infra-red ear thermometers (with maximum device).
IEC60601-1-2 (EMC), EN60601-1 (Safety) standards.
Accuracy:
1. For Forehead / Ear mode:
• +/- 0.2°C (0.4°F) during 36°C to 39°C (96.8°F to 102°F)
• +/- 0.3°C (0.5°F) for other range
2. For Scan Mode:
• +/- 0.3°C (0.5°F) during 22°C to 42.2°C (71.6°F to 108°F),
others +/- 4% or +/- 2°C (4°F) whichever is greater.
14

16. TROUBLESHOOTING
Error Message Problem Solution
Error 5-9, the system is not
functioning properly.
Unload the battery, wait
for 1 minute and repower
it.
Measurement before
device stabilization.
Wait until all the icons
stop flashing.
The device showing a
rapid ambient temperature
change.
Allow the thermometer
to rest in a room for at
least 30 minutes at room
temperature: 10°C and
40°C (50°F and 104°F).
The ambient temperature
is not within the range
between 10°C and 40°C
(50°F and 104°F).
15

(1) In Ear / Forehead Mode:
Temperature taken is higher
than 42.2°C (108°F).
(2) In Scan Mode:
Temperature taken is higher
than 80°C (176°F).
Take measurement
within one of the target
zones specified. If the
temperature is still beyond
the specified range,
remove the battery and
replace after 1 minute.
Repeat the temperature
measurement.
(1) In Ear / Forehead Mode:
Temperature taken is lower
than 34°C (93.2°F).
(2) In Scan Mode:
Temperature taken is lower
than - 22°C (-7.6°F).
Device cannot be powered
on to the ready stage.
Change with a new
battery.
R
16

17. TWO YEAR WARRANTY
If this product proves to be defective in material or workmanship within two
years of purchase, it will be repaired or replaced without charge upon receipt of
the unit prepaid with $5.00 to cover handling, packaging and return postage to
the address below. Please include your full name, address and daytime phone
number or email address.
If you have any questions regarding the operation of your thermometer please
call the Help Line: 1-866-536-2289.
Thermor Ltd.
Repair Department
16975 Leslie Street
Newmarket, ON L3Y 9A1
www.biosexactly.com
17
TH1708_BD802_BD_IB
091109
Table of contents
Other BIOS Diagnostics Thermometer manuals

BIOS Diagnostics
BIOS Diagnostics PrecisionTemp1 240DI User manual

BIOS Diagnostics
BIOS Diagnostics 192DC User manual

BIOS Diagnostics
BIOS Diagnostics 240DI User manual

BIOS Diagnostics
BIOS Diagnostics TH4620 User manual

BIOS Diagnostics
BIOS Diagnostics 127DU User manual

BIOS Diagnostics
BIOS Diagnostics 33302 User manual

BIOS Diagnostics
BIOS Diagnostics 100DC User manual

BIOS Diagnostics
BIOS Diagnostics 120DC User manual

BIOS Diagnostics
BIOS Diagnostics 194DI User manual

BIOS Diagnostics
BIOS Diagnostics 282DI User manual