Chino IR-SA Series User manual

INST.No.INE-510-P6 IR-SA Series
Infrared radiation thermometer for
Processing
Model : IR-SAI
IR-SAS
IR-SAH
Always keep this instruction manual with the unit.
Please be sure to deliver this instruction manual
with the unit to the end user.


INST.No.INE-510-P6
-C1-
1. Introduction.............................1
1.1 General ................................................... 1
2. Model........................................1
2.1 Model...................................................... 1
2.2 Belongings.............................................. 1
2.3 Accessories Model.................................. 2
3. Names and functions of
component parts.................3
3.1 Overview ................................................ 3
3.2 Serial No. Label...................................... 3
4. Installation...............................4
4.1 Precautions in Installation ...................... 4
4.2 Installation.............................................. 5
4.3 Measuring distance and measuring
diameter ....... 6
4.4 Targeting................................................. 9
4.4.1 Telescope (Model : IR-ZYTS)............. 9
4.4.2 Laser pointer (MODEL : IR-ZYLZ1) 10
5. Connections and wirings...... 11
5.1 Connections and wirings ...................... 11
5.1.1 Connection of the exclusive cable
IR-ZYRC to IR-SA............ 11
5.1.2 Connection of the exclusive cable
IR-ZYRC to IR-GZ............ 11
6. Operation...............................12
6.1 Self-diagnostic function........................ 12
7. Maintenance and check........13
7.1 Periodical checking… .......................... 13
7.2 Trouble shooting................................... 13
7.2.1 Measuring value not displayed or
displayed lower.......... 13
7.2.2 Measuring value displayed higher..... 13
7.2.3 Display fluctuated.............................. 13
8. Reference................................14
8.1 Emissivity table .................................... 14
8.1.1 Emissivity table (= 0.65µm)........... 14
8.1.2 Emissivity table (= 0.9µm).............. 15
8.1.3 Emissivity table (= 1.55µm)............ 15
8.1.4 Emissivity table (=2.0µm) ............. 16
9. General specifications...........17
9.1 Thermometer ........................................ 17
9.2 IR-SAIN, IR-SASN .................... 18
Outside dimensions.................... 18
9.3 Accessories outside dimensions ........... 18
9.3.1 Installation fittings
(MODEL : IR-ZYHAW :
BELONGINGS)................. 18
9.3.2 Connection cable
(MODEL : IR-ZYRC)........................ 18
9.3.3 Telescope (MODEL : IR-ZYTS) ....... 19
9.3.4 Laser pointer (MODEL : IR-ZYLZ1) 19
9.3.5 Fitting (MODEL : IR-ZYHG1).......... 19
9.3.6 Fitting (MODEL : IR-ZYHG2).......... 20
9.3.7 Universal head (MODEL : IR-ZMSS)20
CONTENTS

INST.No.INE-510-P6
-C2-
◆Request to the operator of the thermometer
This instruction manual describes the maintenance of the thermometer, too.
Keep this instruction manual with the thermometer.
If you have unclear points or need technical assistance, please contact your sales agent of
CHINO Corporation.
Notices
1. The information in this manual is subject to change without notice and does not represent a
commitment on the part of CHINO Corporation.
2. No part of this manual may be reproduced or transmitted in any form or by any means,
electronic or mechanical, including photocopying and recording, for any purpose other
than the purchaser's personal use without permission of CHINO Corporation.
3. CHINO Corporation shall not be liable for any operation results.
To use infrared radiation thermometer for processing (Model IR-SAI, IR-SAS, IR-SAH) correctly
and safely, please keep the following safety measures for the operation and storage of the
thermometer.
Working conditions and environment
∙The working temperature range of the thermometer is 0 to 90 °C. (No dew condensation)
∙Do not use the thermometer in dusty places, etc. Remove the dust after using it.
∙Be careful not to give vibration or impact to the thermometer. Install the thermometer as far as
possible from an inductive oscillator or electric power line.
Storage
∙Do not store the thermometer in hot and humid places.
∙For failures of the thermometer, don’t overhaul it by yourself, and contact your sales agent of CHINO
Corporation.
Symbols in this instruction manual
The symbols shown below are used depending on important degrees for using the thermometer safely and
avoiding unexpected situations.
Important
degree Symbols Contents
1This symbol is attached to a title for the sentence with .
2
For avoiding dangerous accidents (may cause death or serious injury)
like as electrical shock, fires, or troubles/damages of the thermometer.
3For avoiding injury or in physical damage to the thermometer.
4Information that we suggest to read carefully.
5Information that you can use as a reference.
Preface
Warning
Warning
!
!
1
2
3
Caution
Reference
Remarks
Request and notices

INST.No.INE-510-P6
-C3-
Warnings and Cautions
Please use the thermometer correctly by keeping the following items.
In addition, please read this instruction manual carefully and keep it at the place
where you can access easily.
The mark indicates prohibited operations.
Warning (May cause death or serious injury)
Don’t operate the thermometer in places where combustible or volatile gas is existed.
It is extremely dangerous to use the thermometer in such environment.
When connecting cable connecter, make sure that all mains is turned
off to prevent an
electric shock.
Don’t use the thermometer if it was broken, smoking or nasty smelling.
These may cause fire. For such abnormal c
ondition, turn off the power switch at once
and contact your sales agent of CHINO Corporation.
Using “telescope” (Model: IR-ZYTS), do not see the object exceeding 1500ºC
directly.
Using “Laser pointer” (Model: IR-ZYLZ1) may damage your eyes. Do
n’t stare into a
laser beam. Make sure to target the laser when you want to
decide the center of the
measuring object only and to go off it after the center of the measuring object is
decided.
Never take the thermometer apart or convert it. These may cause trouble and danger.
Caution (May cause injury or physical damage)
Be careful not to give vibration or impact to the thermometer. Install the thermometer
by keeping it as far as possible from an inductive oscillator or electric power line.
Do not use the thermometer in dusty places, noisy places and static electricity places.
Do not wire the connection cable near a noise occurrence resource, relay drive line,
high frequency line and power line.
Do not bundle the connection cable with the line that noise is doubling, and do not
store it into the same duct.
Read the entire contents in this instruction manual to have the thermometer function
perfectly.
!
!
!
!
!
!
!
!
!

INST.No.INE-510-P6
-1-
1. Introduction
1.1 General
The IR-SA series is an infrared radiation thermometer.
These thermometers has fixed focusing of long-distance factor, ability of working in 0 to 90ºC ambient
temperature, for free installation.
The radiation energy collected through the objective lens is transmitted to the element is converted into
an electrical signal. The element output is digitally converted and processed through emissivity
compensation, linealizer and modulation. The standardized final output is 4 to 20mA DC. Connected to
“Setting display unit” (Model: IR-GZ), make programming or selection of emissivity, signal modulation
and alarm function easy.
Various options and accessories are prepared for every kind of applications.
2. Model
2.1 Model
IR-SA N Model
I : Middle temperature type
S: High temperature type
H: High temperature of 2 color type
Measurement diameter and measurement distance
(at focal point of objective lens)
10: φ 5/ 500mm
11: φ10/1000mm
12: φ20/2000mm
15: φ50/5000mm (option)
1S: φ 3/ 200mm (option)
20: φ 3/ 500mm
21: φ 5/1000mm
22: φ10/2000mm
25: φ25/5000mm (option)
2S: φ 1/ 200mm (option)
2.2 Belongings
・Installation fittings 1piece
・Screw for Installation fittings 2piece
Installation fittings

INST.No.INE-510-P6
-2-
Make sure to light the laser when you want to target a measurement area only and
to go off it after targeting.
Accessories have many kinds of combinations for applications. Refer to a separate
instruction manual for each accessory.
2.3 Accessories Model
Accessories Model Reference
Setting display unit IR-GZ Setting Parameter, display measurement value,
and power supply to IR-SA
Connecting cable IR-ZYRC□□□ □□□:length (m)
002:2m,005:5m,010:10m,020:20m,100:100m
Telescope IR-ZYTS Insert to IR-SA, and bring the object into focus
Laser pointer IR-ZYLZ1
Protective case IR-ZYCH With water-cooling and air-purge
Sealing window IR-ZW□C
Water-cooling flange IR-VSW
Sighting tube IR-ZYSS
Setting plate IR-ZYHG1 L type
Setting plate IR-ZYHG2 L×2 type
Universal head IR-ZMSS
Caution
Reference
φ73
IR-SA IR-ZYLZ1
IR-ZYTS
IR-ZYRC
IR-GZ
IR-ZMSS
IR-ZYHG2
IR-ZYHG1
IR-ZYSS
IR-VSW IR-ZW□C IR-ZYCH

INST.No.INE-510-P6
-3-
3. Names and functions of component parts
3.1 Overview
(1) Cover glass (outer)
(2) Object lens (inner) (3) Case
(4) Inserting hall for telescope and laser pointer (5) Blind cap
(6) Connecter
(7) Serial No.label
CH INO
(8) Screw for instllatin fittings
(1) Cover glass An optical glass for protecting the objective lens from scratches,
smudges, etc.
(2) Object lens For focusing in a measuring area.
(3) Case φ50mm stainless tube
(4) Inserting hall for telescope,
laser pointer A hall of Optical pass for inserting telescope(model : IR-ZYTS),
Laser pointer (model:IR-ZYLZ1)
(5) Blind cap For protecting inserting hall from water, dust
(6) Connecter Connect the thermometer by using exclusive cable
(model : IR-ZYRC)
(7) Serial No. label A label for indication of Model No., Serial No.
For your inquiries, inform us of items.
(8) Screw for installation fittings Two screws (2-M3、depth 6mm) for mounting installation fittings
3.2 Serial No. Label
Serial No. label indicates Model No. and Serial No.
SI077S007
S 1 1 N
IR- SA
Model
No.

INST.No.INE-510-P6
-4-
4. Installation (refer to [3. Names and functions of component parts].)
4.1 Precautions in Installation
This IR-SA series radiation thermometer is a sensitive instrument.
Install it by referring to the following cautions.
●Vibration and impacts
Vibration or impacts reduces the liability of the thermometer and causes an
unstable measurement by the targeting shake to a measuring surface.
If you install the thermometer in the place where vibration or impacts exist, its
careful periodic inspection is requested.
Caution
●Induction
The thermometer is designed for anti-
induction but install it as far as possible
from an induction heating generator and power line distance factor.
Caution
●Working temperature
The working temperature of the thermometer is 0 to 90ºC. If the ambient
temperature is high, or if the temperature of the thermometer exceeds the
maximum working temperature by reflection from a high temperature
substance, etc., water-cooling of the thermometer is absolutely necessary.
When the temperature of the thermometer
does not exceed but closely reaches to
the maximum working temperature, water-
cooling of the thermometer is
recommended for maintaining of the reliability of the thermometer.
Caution
●Optical path
Select a place for installation of the thermometer, where water-
drops, dust,
smoke, steam, etc. would not enter between the thermometer and a measuring
surface.
If such place cannot be selected and the affection by existence of such substan
ces
cannot be ignored, blowing-out of such substance with air-purge is necessary.
Caution
●Disturbances to make indication of measured value higher
Select a place for installation of the thermometer, where sunlight, light of
incandescent lamp, flame, heat radiat
ion from a high temperature substance,
etc. is not reflected to the measured surface and the cover glass of the
thermometer. When such light is reflected, higher temperature than the exact
one will be indicated. (The affect by such reflection will be great
for the
measurement of low temperature.) If such place cannot be selected, shade the
thermometer or take similar precautionary measures.
Caution

INST.No.INE-510-P6
-5-
When you use protective case (model: IR-ZYCH), please mount the
installation fittings belonged to the protective case. In details, refer to
instruction manual of IR-ZYCH.
There is top and bottom on attached installation fittings, install the wide one of
the hole in the connector side.
4.2 Installation
Before install the thermometer, mount the
installation fittings to the thermometer by
using screw.
And install the thermometer by using
installation fittings.
Please read the operation manual for an
accessory when instilling it by using the
exclusive accessories.
CH INO
Screw (Belongings) 2-M3
Instllation fittings
IR-SA
Don’t use the thermometer in the following places.
1) Dusty place or a corrosive gas atmosphere.
2) Noisy and static electricity
3) Places where the ambient temperature is higher than 90°C or lower than 0°C.
4) Places where the ambient temperature changes abruptly, or high humid places.
5) Places where there are mechanical vibrations and impacts.
6) Places where combustible or volatile gas is existed.
Caution
Reference
Reference

INST.No.INE-510-P6
-6-
4.3 Measuring distance and measuring diameter
The relation of the measuring distance and the measuring diameter is indicated in figure below.
IR-SA□10N
IR-SA□11N
IR-SA□12N
IR-SA□15N (Option)
IR-SA□1SN (Option)
0
φ10
φ5
φ20
500
1000
φ30
2000
φ10
1000
φ10
0
φ10
0
φ20
2000
φ50
4000
φ10
0
φ50
5000
10000
φ110
φ10
φ2
φ14
200
400
[Unit mm]

INST.No.INE-510-P6
-7-
IR-SA□20N
IR-SA□21N
IR-SA□I22N
IR-SA□25N (Option)
IR-SA□2SN (Option)
500
1000 0
φ10
φ3
φ15
0
φ10
φ5
φ20
2000 1000
φ30 φ10 φ10
0
2000
4000
φ10
0
φ25
φ60
5000
10000
φ10
φ1
φ12
200
400
[Unit mm]

INST.No.INE-510-P6
-8-
Considering the dislocation of the optical axis, take the measuring area more than 1.5
times the measuring diameter.
At focusing position of object lens, the relation of measuring distance and the measuring diameter are
defined by the distance factor as shown in the following formula.
In case of IR-SA, distance factor and focusing position of object lens are shown in model name, the
number of next beginning of alphabet shows it, and next number shows focusing position of object
lens.
In position that does not focus, the relation of measuring distance and the measuring diameter are
different. In this case, the relation of measuring distance and the measuring diameter are defined as
shown in the following formula, assumed distance of focusing position of object lens from the
thermometer is A, measuring diameter at focusing position is B.
It is an upper figure to have shown these expressions of relation s to the figure. 10 of the whole
expression of relations is an object lens diameter.
Place the thermometer at the measuring distance, confirmed measuring diameter by an upper figure or
formula.
Measuring diameter (mm) = Measuring distance (mm)/distance factor
IR-SAS □ □ N
Distance factor 1=100, 2=200
Focusing position of object lens 0=500mm, 1=1000mm, 2=2000mm
5=5000mm, S=200mm
When measuring distance is nearer than the focusing position of object lens
Measuring diameter = ( B – 10 )×Measuring distance /A+ 10
When measuring distance is longer than the focusing position of object lens
Measuring diameter = (B + 10 ) ×Measuring distance /A – 10 Unit (mm)
Reference

INST.No.INE-510-P6
-9-
Telescope (model: IR-ZYTS) is a instrument to confirm the center of the
measurement object, and is not related with the measuring diameter to show by the
telescope directly. (Refer to 4.3 measuring distance and measuring diameter)
Please turn focus ring counterclockwise to focus it near, adversely clockwise to
focus it in the distance.
Waterproofing structure is not realized in the state that you used telescope. If it is
finished, remove telescope and install blind cap as before.
Please refer to an individual instruction manual for details.
4.4 Targeting
Before beginning the measurement, please point the thermometer to the measurement object and install it.
Telescope (model: IR-ZYTS (option)) and laser pointer (model: IR-ZYLZ1(option)) enable more correct
targeting.
You can confirm the center of the measurement object, by using telescope or laser pointer.
4.4.1 Telescope (Model : IR-ZYTS)
(2)Main body fixation screw
(1)Prevention of loser screw
(3)Focusing ring (5)Eye piece
(4)Body
1) Please remove blind cap of the thermometer counterclockwise, and insert telescope to insert hall for
telescope and laser pointer.
2) Screw prevention of loser screw clockwise by operating body.
3) Screw main body fixation screw clockwise, and fix it in the thermometer.
4) See measurement object from eye piece and focus it by adjustment focusing ring.
5) Adjust installing to be able to see the center of measurement object.
6) If it is finished, remove telescope and install blind cap as before.
●Using “telescope”(Model: IR-
ZYTS),do not see the object exceeding 1500 ºC
directly.
●When screw prevention of loser screw and main body fixation screw, do not
operate the focusing ring, operate body for avoiding trouble.
!
Warning
Reference
Reference
Reference
Reference
Caution

INST.No.INE-510-P6
-10-
Laser pointer (model: IR-ZYLZ1) is a instrument to confirm the center of the
measurement object. (Refer to 4.3 Measuring distance and measuring diameter)
Please turn focus ring counterclockwise to focus it near, adversely clockwise to
focus it in the distance. You can make laser beam a point.
Waterproofing structure is not realized in the state that you used laser.
If it is finished, remove laser pointer and install blind cap as before.
If it is finished, turn off the switch to avoid consumption of the battery.
The battery to use is CR2 lithium battery for camera.
Please refer to an individual instruction manual for details.
4.4.2 Laser pointer (MODEL : IR-ZYLZ1)
(1)Prevention of loser screw
(3)Focusing ring
(2)Main body fixation screw
(4)Body
(5)Switch
(6)Battery case
(7)Battery lid
1) Please remove blind cap of the thermometer counterclockwise, and insert laser to insert hall for
telescope and laser pointer.
2) Screw prevention of loser screw clockwise by operating body.
3) Screw main body fixation screw clockwise, and fix it in the thermometer.
4) Turn on the switch and bring the object into focus by adjustment focusing ring.
5) Adjust installing to be able to see the center of measurement object.
6) If it is finished, remove telescope and install blind cap as before.
●Using “Laser pointer”(Model:IR-ZYLZ1) may damage your
eyes. Don’t stare
into a laser beam. Make sure to target the laser when you want to decide the
center of the measuring object only and to go off it after the center of the
measuring object is decided.
●When screw prevention of loser screw and main body fixation screw, do not
operate the focusing ring, operate body for avoiding trouble.
Warning
!
Reference
Reference
Reference
Reference
Reference
Caution

INST.No.INE-510-P6
-11-
5. Connections and wirings
5.1 Connections and wirings
5.1.1 Connection of the exclusive cable IR-ZYRC to IR-SA
The connections are completed by simply connecting the connector of the cable (IR-ZYRC) to the
connector placed at the rear side of the thermometer.
(1) For the connection, align the slit in the connector of the cable and the key position in the connector of
the thermometer, and firmly insert the cable connector, and then turn the coupling ring clockwise until
it is locked.
(2)For the disconnection, turn the coupling ring counterclockwise for releasing its locked status, and pull
the connector of the cable forward.
5.1.2 Connection of the exclusive cable IR-ZYRC to IR-GZ
The other side of IR-ZYRC, connect to setting display unit (model: IR-GZ). For details, refer to
instruction manual of IR-GZ.
Make sure to turn off the power supply for
preventing
an
electric
shock when
connecting and wiring.
Warning
Connecter
Thermometer(IR-SA) side
Key position Slit
Coupling ring
Cable side
A+
B-
C+
D-
E SA
F SB
G SG
H Ground
24V DC power supply
Temperature output 4 to 20 mA DC
Serial Communication for IR-GZ
(normally SG is not used, empty terminal)
5
6
7
8
9
10
11
12
13
15
21
22
23
24
25
26
27
28
29
30
Rear view of IR-GZ
IR-ZYRC
Turminal No.
21
22
23
24
25
26
27
Signal +
Signal-
RS485 SA
RS485 SB
Earth
Power +
Power-
When the connecter of the cable (IR
-
ZYRC) is connected, and the coupling ring
is tightened, the waterproof of the IR-SA series is achieved. There is no
waterproof property in the connecter of the thermometer in the state that the cable
is not connected.
!
!
Caution

INST.No.INE-510-P6
-12-
6. Operation
In case of IR-SA, parameter setting, the measurements indication are performed entirely setting display
unit (model : IR-GZ). For details, refer to instruction manual of setting display unit (model : IR-GZ).
6.1 Self-diagnostic function
In IR-SA, self-diagnosis function is built-in.
On abnormal conditions, an ERR00 will be displayed in IR-GZ. ERR00 means inside error of
thermometer.
Inside error of thermometer
・The ambient temperature is abnormal
・E2PROM abnormal (Writing and reading impossible)
・Analog output correction data abnormal (Same data before correction existed.)
・Zero/Span adjustment abnormal.

INST.No.INE-510-P6
-13-
7. Maintenance and check
7.1 Periodical checking…
Check the followings periodically or if required.
•Lens: Check the objective lens for dust or dirty.
If dirt is present, remove it with a blower for camera lens. If the dirt cannot be removed with the
blower, wipe the lens gently with cotton ball soaked in alcohol.
•Connections and wirings: Check all connections and wirings.
7.2 Trouble shooting
7.2.1 Measuring value not displayed or displayed lower
Checking item Countermeasure
1) The power voltage is not in the allowable range. Supply the power with the correct voltage.
2) The Connection cable is not connected firmly. For the disconnection of the cable, replace it.
3) The view field is interfered. Make the view field not interfered by referring to
[4. Installation].
4) The measured temperature is lower than the
minimum value of the measuring range. It is necessary to use a thermometer with the
measuring range covering the measured temperature.
5) The emissivity value programmed is too high. Program the emissivity value by using setting display
unit (model : IR-GZ)
6) The atmosphere temperature is low, the optic
system is doing no dew . It uses it in the place where does not do dew.
7.2.2 Measuring value displayed higher
Checking items Countermeasure
1) The measured temperature is higher than the
maximum value of the measuring range. It is necessary to use a thermometer with the
measuring range covering the measured temperature.
2) The emissivity value programmed is too low. Program the emissivity value by using setting display
unit (model : IR-GZ)
3) The heat radiation of the high temperature is
shining to the measurement face or thermometer
from the outside.
Heat is prevented with the board that does not pass the
change or light of the place
7.2.3 Display fluctuated
Checking items Countermeasure
1) The radiation thermometer is not fixed firmly or
vibrated. Fix the radiation thermometer firmly and install it in a
place not vibrated.
2) The connector and terminal are not connected firmly. Connect the connector and terminal firmly.
3) The power voltage is not in the allowable rang Supply the power with the correct voltage.
4) The view field is interfered by steam. Purge the steam by air.
5)
The measured temperature is fluctuated exactly.
Emissivity value is changeabling.
Program the emissivity value and the modulation
degree by referring to the following paragraphs.
[Emissivity programming]
[Signal modulation mode selection]
[Modulation degree programming]
Refer to instruction manual of setting display unit
(model : IR
-
G
Z)

INST.No.INE-510-P6
-14-
8. Reference
The emissivity are values determined by the material of object, profile of its surface, surface roughness,
oxidized or not, measuring temperature, measuring wavelength and other factors.
They are represented by the thermal radiation ratio "" when a black body furnace at the same
temperature is measured in the same wavelength band.
The emissivity "" is generally known by a value at the wavelength of 0.65µm when an optical
pyrometer is used. The emissivity changes according to the above factors even in case of the same
material. Please use the following table as a reference.
8.1 Emissivity table
8.1.1 Emissivity table (= 0.65µm)
Metal Emissivity Oxide Emissivity
Solid Liquid
Zinc 0.42 ―Alumel(*)0.87
Alumel 0.37 ―Chromel(*)0.87
Aluminum 0.17 0.12 Constantan(*)0.84
Antimony 0.32 ―Ceramics 0.25 to 0.5
Iridium 0.30 ―Cast iron(*)0.70
Yttrium 0.35 0.35 55Fe. 37.5Cr. 7.5Al(*)0.78
Uranium 0.54 0.34 70Fe. 23Cr. 5Al. 2Co(*)0.75
Gold 0.14 0.22 80Ni. 20Cr(*)0.90
Silver 0.07 0.07 60Ni. 24Fe. 16Cr(*)0.83
Chromium 0.34 0.39 Stainless steel(*)0.85
Chromel P 0.35 ―Aluminum oxide 0.22 to 0.4
Cobalt 0.36 0.37 Yttrium oxide 0.60
Constantan 0.35 ―Uranium oxide 0.30
Zirconium 0.32 0.30 Cobalt oxide 0.75
Mercury ―0.23 Columbium oxide 0.55 to 0.71
Tin 0.18 ―Zirconium oxide 0.18 to 0.43
Carbon 0.8 to 0.9 ―Tin oxide 0.32 to 0.60
Tungsten 0.43 ―Cerium oxide 0.58 to 0.82
Tantalum 0.49 ―Titanium oxide 0.50
Cast iron 0.37 0.40 Iron oxide 0.63 to 0.98
Titanium 0.63 0.65 Copper oxide 0.60 to 0.80
Iron 0.35 0.37 Thorium oxide 0.20 to 0.57
Copper 0.10 0.15 Vanadium oxide 0.70
Thorium 0.54 0.34 Beryllium oxide 0.07 to 0.37
Nickel 0.36 0.37 Magnesium oxide 0.10 to 0.43
80Ni /20Cr 0.35 ―
60Ni / 024Fe / 16Cr 0.36 ―(*) : Oxidized on surfaces
Platinum 0.30 0.38
90Pt / 10Rh 0.27 ―
Palladium 0.33 0.38
Vanadium 0.35 0.35
Bismuth 0.29 ―
Beryllium 0.61 0.61
Manganese 0.59 0.59
Molybdenum 0.37 0.40
Rhodium 0.24 0.30

INST.No.INE-510-P6
-15-
Use the above tables as a reference.
8.1.2 Emissivity table (= 0.9µm) 8.1.3 Emissivity table (= 1.55µm)
Metal Emissivity
Aluminum 0.10 to 0.23
Gold 0.015 to 0.02
Chrome 0.36
Cobalt 0.28 to 0.30
Iron 0.33 to 0.36
Copper 0.03 to 0.06
Tungsten 0.38 to 0.42
Titanium 0.50 to 0.62
Nickel 0.26 to 0.35
Platinum 0.25 to 0.30
Molybdenum 0.28to 0.36
Alloy Emissivity
Inconel X 0.40 to 0.60
Inconel 600 0.28
Inconel 617 0.29
Inconel 0.85 to 0.93
Incoloy 800 0.29
Kanthal 0.80 to 0.90
Stainless steel 0.30
Hastelloy X 0.3
Semi conductor Emissivity
Silicon 0.69 to 0.71
Germanium 0.60
Gallium arsenic 0.68
Ceramics Emissivity
Silicon carbide 0.80 to 0.83
Titanium carbide 0.47 to 0.50
Silicon nitride 0.89 to 0.90
Other Emissivity
Carbon pigment 0.90 to 0.95
Graphite 0.87 to 0.92
Metal Emissivity
Aluminum 0.09 to 0.40
Chrome 0.34 to 0.80
Cobalt 0.28 to 0.65
Copper 0.05 to 0.80
Gold 0.02
Steel plate 0.30 to 0.85
Lead 0.28 to 0.65
Magnesium 0.24 to 0.75
Molybdenum 0.25 to 0.80
Nickel 0.25 to 0.85
Palladium 0.23
Platinum 0.22
Rhodium 0.18
Silver 0.04 to 0.10
Tantalum 0.20 to 0.80
Tin 0.28 to 0.60
Titanium 0.50 to 0.80
Tungsten 0.30
Zinc 0.32 to 0.55
Alloy Emissivity
Brass 0.18 to 0.70
Chromel, Alumel 0.30 to 0.80
Constantan, Manganin 0.22 to 0.60
Inconel 0.30 to 0.85
Monel 0.22 to 0.70
Nickel Chrome 0.28 to 0.85
Ceramics Emissivity
Alumina ceramics 0.30
Red brick 0.80
White brick 0.35
Silicon brick 0.60
Sillimanite brick 0.60
Ceramics 0.50
Other Emissivity
Asbestos 0.90
Asphalt 0.85
Carbon 0.85
Graphite 0.80
Soot 0.95
Cement, Concrete 0.70
Cloth 0.80
Reference
This manual suits for next models
3
Table of contents
Other Chino Thermometer manuals

Chino
Chino IR-FA Series User manual

Chino
Chino IR-CAK User manual

Chino
Chino IR-H Series User manual

Chino
Chino IR-BZ Series User manual

Chino
Chino IR-AH Series User manual

Chino
Chino IR-E Series User manual

Chino
Chino IR-CA series User manual

Chino
Chino IR-CZ Series User manual

Chino
Chino IR-BAT Series User manual

Chino
Chino TP-S Easy Thermo User manual