manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Diesse
  6. •
  7. Laboratory Equipment
  8. •
  9. Diesse MINI-CUBE User manual

Diesse MINI-CUBE User manual

July 2021
QUICK START GUIDE
REFER TO THE USER MANUAL FOR ALL DETAILS
PERFORMING AN ANALYSIS CYCLE
- Switch on the instrument
- Click START
Wait for the instrument to complete the initialization phase
Gently invert the tube 10-12 times for complete
inversion of the tube so that the air bubble that forms run through it
from one end to the other, immediately before starting the analysis.
From this moment on, it is possible to introduce the sample for testing.
To start a cycle, insert the tubes orienting them as shown in the
photo in one of the free positions. If a barcode reader (optional) is
present, read the sample identification code before inserting the
tube. After insertion, the instrument scans to determine the blood
level at zero sedimentation time.
During the first scan a “CHECK TUBE” pop-up window appears which determines the
display of the tube type (normal or pediatric) and the blood level, after which the
countdown will be activated.
Procedure for map TUBES (BD MICROTAINER®, MICROVETTE® APT
(SARSTEDT))
To analyze a MAP tube sample, touch the screen in one of the
empty positions to open the NEW SAMPLE window.
Select “Microtube MAP” YES (Y) and insert the tube in one of the
free positions.
The tube will be represented as a pediatric tube, with the letter
M on the cap, indicating that they are MAP tubes.
If most of the tubes are BD Microtainer® MAP, in the settings menu, MICROTUBE MAP
parameter, select ON.
July 2021
EXECUTION OF ANALYSIS CYCLE - RANDOM ACCESS
- During a cycle it will be possible at any time to insert a new sample in each of the free
positions present.
- Mix the sample thoroughly before inserting it into the instrument as described above
RESULTS
At the end of the 20 minutes (14 min for Pediatric) the instrument displays the result.
Green color indicates ESR within the normal range.
Red color indicates an ESR that has exceeded the normal level.
The results are expressed in mm / h (Westergren 1st hour).
The results can be printed via Bluetooth printer (Optional)
You will need to connect it to the instrument first from the menu.
Printing can take place automatically at the end of the analysis, or
- Press the ARCHIVE key
- Select the samples for consultation / printing of the results
QUALITY CONTROL
Entering the settings menu, under QC, you can enter the
characteristics of each control, just read the bar code of the QC
Diesse. Then, every time the control is inserted in any position of
the instrument, it will be recognized as QC, and displayed on the
screen, as shown in the image.
The tube with the white cap indicates the Normal control, the
one with the black cap the Pathological control.
In case you are using another trade control, it is recommended that you
create dedicated barcode labels and manually enter the data.
REFILL
- Press the SETTINGS button
- Select REFILL on the second page
- Insert the TEST DEVICE NEXT in the appropriate compartment
(as shown in the figure)
- Press the CHECK key
- Press OK to confirm the test number to reload
- Automatic update of the number of tests displayed on the
instrument screen (top left)
July 2021
TROUBLESHOOTING
ERROR MESSAGES
CAUSE
REMEDY
Remove all test tubes
A cycle was started with some
tubes already inserted.
Otherwise, there may be a
foreign body in front of the
sensors.
Remove the tubes and restart the run.
Turn off the instrument, turn it over and try to
clean it internally. If the problem persists,
contact technical assistance.
Low (Lo)
Blood level inside the tube too
low
If possible, add blood up to established levels.
High (HI)
- Blood level inside the tube too
high
-Interference caused by labels
- If possible, remove blood up to the
established levels
- Try to reposition the tube as described
previously
ERR
Generic error, pressing on the
error message a number will
appear that differentiates the
cause.
See complete troubleshooting
in the User Manual
- Check that the tube used is compatible with
the instrument or the possible presence of a
number of labels greater than 2 in case remove
the excess labels. Check for any lumps or clots.
- Try to repeat the test
Printer error, retry?
Printer not found.
Make sure the printer is turned on. Otherwise,
deselect the Bluetooth mode (OFF)
Exam not available. Need refill
The executable test counter is
equal to zero
Follow the instructions to recharge the
instrument with a new transponder described
in chapter 4.
USB pendrive not found
The instrument does not
recognize the USB stick.
Remove and reinsert the USB stick or change
it.
Warning! Remain n tests
The number of tests remaining
is less than 50
It is recommended to recharge the instrument
as soon as possible with a new transponder
CHECK DEVICE already used
The transponder used is empty.
Use a new transponder
CHECK DEVICE not found
The instrument does not
recognize the transponder used
Remove and reinsert the transponder. Use
another transponder
Enrica Gentile
Junior customer care specialist
DIESSE Diagnostica Senese SpA

Other manuals for MINI-CUBE

4

Other Diesse Laboratory Equipment manuals

Diesse CHORUS TRIO CHORUS EXPANDER UPGRADE KIT User manual

Diesse

Diesse CHORUS TRIO CHORUS EXPANDER UPGRADE KIT User manual

Diesse Chorus User manual

Diesse

Diesse Chorus User manual

Diesse VES-MATIC CUBE 30 User manual

Diesse

Diesse VES-MATIC CUBE 30 User manual

Diesse Chorus Trio User manual

Diesse

Diesse Chorus Trio User manual

Diesse CHORUS SPECIFIC IgE User manual

Diesse

Diesse CHORUS SPECIFIC IgE User manual

Diesse VES-MATIC 20 User manual

Diesse

Diesse VES-MATIC 20 User manual

Diesse MINI-CUBE User manual

Diesse

Diesse MINI-CUBE User manual

Diesse VES MATIC 5 User manual

Diesse

Diesse VES MATIC 5 User manual

Diesse cube 30 touch User manual

Diesse

Diesse cube 30 touch User manual

Diesse VES-MATIC 30 User manual

Diesse

Diesse VES-MATIC 30 User manual

Diesse cube 30 touch User manual

Diesse

Diesse cube 30 touch User manual

Diesse VES MATIC CUBE 80 User manual

Diesse

Diesse VES MATIC CUBE 80 User manual

Diesse MINI-CUBE User manual

Diesse

Diesse MINI-CUBE User manual

Diesse cube 30 touch User manual

Diesse

Diesse cube 30 touch User manual

Diesse cube 30 touch User manual

Diesse

Diesse cube 30 touch User manual

Diesse Ves-Matic Cube 200 User manual

Diesse

Diesse Ves-Matic Cube 200 User manual

Diesse CHORUS AUTOSAMPLER User manual

Diesse

Diesse CHORUS AUTOSAMPLER User manual

Diesse VES MATIC Original User manual

Diesse

Diesse VES MATIC Original User manual

Diesse CHORUS AUTOSAMPLER User manual

Diesse

Diesse CHORUS AUTOSAMPLER User manual

Diesse MINI-CUBE User manual

Diesse

Diesse MINI-CUBE User manual

Diesse cube 30 touch User manual

Diesse

Diesse cube 30 touch User manual

Popular Laboratory Equipment manuals by other brands

Teledyne Lecroy ZD500 Getting started manual

Teledyne Lecroy

Teledyne Lecroy ZD500 Getting started manual

Merck APC SmartTouch Operation manual

Merck

Merck APC SmartTouch Operation manual

Micromeritics Saturn II 5205 DigiSizer Installation Instructions and Checklist

Micromeritics

Micromeritics Saturn II 5205 DigiSizer Installation Instructions and Checklist

HTI HT-C600 Operation instructions

HTI

HTI HT-C600 Operation instructions

Parker Laboratories Thermasonic Gel Warmer user manual

Parker Laboratories

Parker Laboratories Thermasonic Gel Warmer user manual

Endress+Hauser analytikjena Biometra TOne Short manual

Endress+Hauser

Endress+Hauser analytikjena Biometra TOne Short manual

IKA IKATRON ETS-D6 operating instructions

IKA

IKA IKATRON ETS-D6 operating instructions

Thermo Scientific PikoReal user manual

Thermo Scientific

Thermo Scientific PikoReal user manual

THORLABS DC4100 Operation manual

THORLABS

THORLABS DC4100 Operation manual

Esco Streamline SCR-2A Service manual

Esco

Esco Streamline SCR-2A Service manual

Cleaver Scientific CSL-CELLAS instruction manual

Cleaver Scientific

Cleaver Scientific CSL-CELLAS instruction manual

Labconco Protector Echo 180 Series user manual

Labconco

Labconco Protector Echo 180 Series user manual

Clontech PT3139-1 user manual

Clontech

Clontech PT3139-1 user manual

Kasanova KGS000001NOC instruction manual

Kasanova

Kasanova KGS000001NOC instruction manual

BioFluidix Tek-Matic PipeJet NanoDispenser Hardware manual

BioFluidix

BioFluidix Tek-Matic PipeJet NanoDispenser Hardware manual

Dräger DrugTest 5000 Quick reference guide

Dräger

Dräger DrugTest 5000 Quick reference guide

Datalogic RIDA DBT6400 Quick reference guide

Datalogic

Datalogic RIDA DBT6400 Quick reference guide

Appleton appSTIR instruction manual

Appleton

Appleton appSTIR instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.