Diesse VES MATIC Original User manual

Rev. 0 –30/09/2022
1/3
–Quick Start Guide –Displayed on the instrument by pressing the key
ALWAYS REFER TO THE USER MANUAL FOR ALL THE DETAILS ON THE
OPERATION OF THE INSTRUMENT.
ANALYSIS CYCLE EXECUTION
-
Turn on the instrument
-
Press the START button
-
Open the top lid
-
Insert the test tubes until they are fully flush, orienting them
with the label facing the mark on the plate
-
Close the top lid
-
Press the button corresponding to the cycle
mode you want to start
-
Automatic reading of the barcodes and indication on the display of
the number of samples present
-
Automatic mixing of the samples
-
Blood level reading at time 0’
-
Sedimentation
-
Blood level reading after 30 minutes (1H) or after 60 minutes (2H), with any
intermediate readings for sedimentation kinetics (1H KYN or 2H KYN)
ANALYSIS CYCLE EXECUTION –RANDOM ACCESS
-
During a cycle it will be possible at any time (excluding the cycle modes in kinetics
and the mixing phase) to insert a new sample
-
Mix the sample thoroughly before inserting it into the instrument (at least 8 times per
complete inversion)
-
Press the ADD button
-
Wait for the indication authorizing the opening of the top lid
-
Open the top lid and insert the test tube (as described above) in the position
indicated in the warning window that appears on the screen
-
Close the top lid
-
Automatic reading of the barcode of the added sample
-
Blood level reading at time 0’
-
Sedimentation
-
Blood level reading after 30 minutes (1H) or after 60 minutes (2H)
RESULTS
At the end of the analysis, the instrument displays the result for each sample in mm/h:
-
It will be green for ESR within the normal range
-
It will be red for ESR outside the normal range (which can be defined in the settings
menu under "THRESHOLD VALUE")
At the end of the analysis, the results are printed automatically (default ON)
ARCHIVE
-
Press the ARCHIVE button
-
Select the samples for consultation/printing of the results and of the
sedimentation graphs
Rev. 0 –30/09/2022
2/3
QUALITY CONTROLS
Control materials can be processed like any blood sample.
QUALITY CONTROLS PROVIDED BY DIESSE DIAGNOSTICA SENESE ARE AUTOMATICALLY
RECOGNIZED BY THE INSTRUMENT.
RECORDING OF QUALITY CONTROLS NOT PROVIDED BY DIESSE
To carry out this procedure, the Quality Controls must be identified by a dedicated
barcode. If this is not present, it is necessary to print labels with a barcode, which, read
by the instrument, can be combined with the information introduced below.
-
Press the SETTING button
-
Press the QC icon
-
Introduce the Normal Control (Level 1) in position 1
-
Press the DETECT button. The instrument identifies the barcode and displays it in
the barcode window
-
Manually enter the data relating to name, lot, expiration date and limits of acceptability
-
Remove the Normal QC
-
Introduce the Abnormal Control (Level 2) in position 1
-
Repeat the procedure already performed for the Normal Control
Following this procedure, the instrument will be able to automatically identify the controls
during a cycle, by reading their barcode and displaying them on the screen:
-
Normal Control: test tube with white cup
-
Abnormal Control: test tube with black cup
SYMBOLS AND ICONS
If the barcode is not read, a black dot appears next to the sample, and a warning
window indicates all the unread positions.
Option 1
-
Open the top lid and reposition the tube correctly in the sample holder plate
-
Press the RETRY button to have the barcode read again
Option 2
-
Press on the text line
-
Enter the code manually (using the on-screen keyboard)
-
Press the OK button
Option 3
Do not enter any code for this sample and move to the next position by pressing
the SKIP button
Option 4
Do not enter codes for all unidentified samples: press the SKIP ALL button
If none of these options is undertaken within 30 seconds, the SKIP ALL choice is
activated and the cycle starts automatically.
Residual sedimentation time (in minutes) for all the samples inserted at the start
of the cycle
Residual sedimentation time (in minutes) relating to the sample added in random
access mode, in correspondence with the position in which it was inserted
Rev. 0 –30/09/2022
3/3
REFILL TEST NUMBER
-Press the SETTING button
-Select the REFILL item
-Insert the TEST DEVICE Original 4K in the dedicated
compartment (as shown in the figure)
-Press the CHECK button
-Press OK to confirm the number of tests to refill
-Automatic update of the number of tests displayed on
the instrument screen (top left)
TROUBLESHOOTING
ERROR
MESSAGES
CAUSE
HELP
Low (Lo)
Blood level inside the test
tube too low
If possible, add blood up to
established levels
High (HI)
Blood level inside the test
tube too high
If possible, remove blood up to
established levels
Interference caused
by labels
Try to remove the label and
repeat the test
ERR
Generic error, pressing the
error message will display a
number that differentiates
the cause. See the complete
TROUBLESHOOTING
in the User Manual
Verify that the tube used is
compatible with the instrument.
If there are more than 2 labels,
remove the excess labels.
Check for any lumps or clots.
Try to repeat the test
Exam not
available.
Need to
reload
The executable test
counter is equal to zero
Follow the instructions described
at the top of this page to recharge
the instrument with a new Test
Device
USB Pendrive
not found
The instrument does not
recognize the USB Pendrive
Remove and reinsert the device
or replace it
Attention!
“n” test remain
The number of tests
remaining is less than 50
It is recommended to recharge
the instrument as soon as
possible with a new Test Device
TEST DEVICE
Already used
The used Test Device is
empty
Repeat with a new Test Device

Other manuals for VES MATIC Original

1

Other Diesse Laboratory Equipment manuals

Diesse cube 30 touch User manual

Diesse

Diesse cube 30 touch User manual

Diesse VES-MATIC CUBE 30 User manual

Diesse

Diesse VES-MATIC CUBE 30 User manual

Diesse cube 30 touch User manual

Diesse

Diesse cube 30 touch User manual

Diesse MINI-CUBE User manual

Diesse

Diesse MINI-CUBE User manual

Diesse CHORUS AUTOSAMPLER User manual

Diesse

Diesse CHORUS AUTOSAMPLER User manual

Diesse VES-MATIC 20 User manual

Diesse

Diesse VES-MATIC 20 User manual

Diesse CHORUS AUTOSAMPLER User manual

Diesse

Diesse CHORUS AUTOSAMPLER User manual

Diesse MINI-CUBE User manual

Diesse

Diesse MINI-CUBE User manual

Diesse CHORUS TRIO CHORUS EXPANDER UPGRADE KIT User manual

Diesse

Diesse CHORUS TRIO CHORUS EXPANDER UPGRADE KIT User manual

Diesse CHORUS SPECIFIC IgE User manual

Diesse

Diesse CHORUS SPECIFIC IgE User manual

Diesse Chorus Trio User manual

Diesse

Diesse Chorus Trio User manual

Diesse MINI-CUBE User manual

Diesse

Diesse MINI-CUBE User manual

Diesse VES MATIC 5 User manual

Diesse

Diesse VES MATIC 5 User manual

Diesse VES-MATIC 30 User manual

Diesse

Diesse VES-MATIC 30 User manual

Diesse Ves-Matic Cube 200 User manual

Diesse

Diesse Ves-Matic Cube 200 User manual

Diesse cube 30 touch User manual

Diesse

Diesse cube 30 touch User manual

Diesse Chorus User manual

Diesse

Diesse Chorus User manual

Diesse cube 30 touch User manual

Diesse

Diesse cube 30 touch User manual

Diesse VES MATIC CUBE 80 User manual

Diesse

Diesse VES MATIC CUBE 80 User manual

Diesse cube 30 touch User manual

Diesse

Diesse cube 30 touch User manual

Diesse MINI-CUBE User manual

Diesse

Diesse MINI-CUBE User manual

Popular Laboratory Equipment manuals by other brands

Hanna Instruments HI 208 instruction manual

Hanna Instruments

Hanna Instruments HI 208 instruction manual

Tuttnauer 5050 Technician manual

Tuttnauer

Tuttnauer 5050 Technician manual

Nic MA-3000 instruction manual

Nic

Nic MA-3000 instruction manual

Merand ATOUPAIN MILLENIUM user manual

Merand

Merand ATOUPAIN MILLENIUM user manual

Tuttnauer 3840 ELVG-D Technician manual

Tuttnauer

Tuttnauer 3840 ELVG-D Technician manual

Gilson PIPETMAX 268 user guide

Gilson

Gilson PIPETMAX 268 user guide

Labconco Purifier Logic+ 30132 Technical Manual and Specifications

Labconco

Labconco Purifier Logic+ 30132 Technical Manual and Specifications

Waters ACQUITY UPLC H-Class Series System guide

Waters

Waters ACQUITY UPLC H-Class Series System guide

Bio-Optica UT200 User and service manual

Bio-Optica

Bio-Optica UT200 User and service manual

Diagon Coag2D user manual

Diagon

Diagon Coag2D user manual

VITACOM LAMP-UV-PR02 instruction manual

VITACOM

VITACOM LAMP-UV-PR02 instruction manual

Sutter Instrument MP-225A Operation manual

Sutter Instrument

Sutter Instrument MP-225A Operation manual

Unico PowerSpin C8624 Operation manual

Unico

Unico PowerSpin C8624 Operation manual

Kramer PSE-4 user manual

Kramer

Kramer PSE-4 user manual

Gamry Instruments Reference 30k Booster Operator's manual

Gamry Instruments

Gamry Instruments Reference 30k Booster Operator's manual

Thermo Scientific Finnpipette Instructions for use

Thermo Scientific

Thermo Scientific Finnpipette Instructions for use

Hettich ROTOFIX 32 A operating instructions

Hettich

Hettich ROTOFIX 32 A operating instructions

Lab-Line 3608 Operation manual

Lab-Line

Lab-Line 3608 Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.