Ecom Ex-MP4 a User manual

1
Zone 1
Manual de instruçoes | Bruksanvisning
Intrinsically Safe temperature meter Ex-MP4 a
Operating Instructions

3
Note:
The current safety instructions, EC Declaration of Conformity and the Certicate of Conformity can be
downloaded from the relevant product page at ww.ecom-ex.com, or alternatively they can be requested
directly from the manufacturer
Content
1. Application .................................................................................................................................................... 4
2. Safety precautions ........................................................................................................................................ 4
3. Faults and damage ....................................................................................................................................... 4
4. Safety regulations......................................................................................................................................... 4
4.1 Special conditions for conducting measurements in zone 0..................................................................... 4
5. Safety precautions ........................................................................................................................................ 5
6. Ex-Data ......................................................................................................................................................... 5
7. Specifications................................................................................................................................................ 5
7.1 Table with type-approved batteries ........................................................................................................... 5
8. Functional description / operating instructions .......................................................................................... 5
8.1 Mode of operation....................................................................................................................................... 6
8.2 Structural overview..................................................................................................................................... 6
8.3 Optical diagram and field of sight .............................................................................................................. 6
8.4 Operation of the laser sight........................................................................................................................ 7
8.5 Operating the unit....................................................................................................................................... 7
8.6 Emission level............................................................................................................................................. 8
8.7 Battery replacement .................................................................................................................................. 8
8.8 Troubleshooting .......................................................................................................................................... 8
9. Repair............................................................................................................................................................ 8
11. Cleaning, maintenance and storage........................................................................................................... 9
12. Warranty and liability.................................................................................................................................. 9
13. EC Declaration of Conformity / Ec declaration of conformity ................................................................. 10

4
1. Application
The Ex-MP4 a is a non-contact temperature meter for industrial application in areas with an increased
risk of explosion in zones 2 and 1, in accordance with directive 1999/92/EC (ATEX) and 94/9/EG.
2. Safety precautions
These operating instructions contain information and safety regulations which are to be observed without
fail for safe operation in the described conditions. Non-observance of this information and these instruc-
tions can have serious consequences or it may violate regulations.
Please read these operating instructions carefully before starting to use the unit!
In case of any doubt (in the form of translation or printing errors) the German language operating
3. Faults and damage
If there is any reason to suspect that the safety of the unit has been impaired, it must be withdrawn from
use and removed immediately from the Ex-area.
Measures must be taken to prevent it from being restarted accidentally.
We recommend that the unit is returned to the manufacturer for inspection.
The safety of the unit may be compromised, if, for example:
- there is visible damage to the housing.
- the unit has been subjected to excessive loads.
- the unit has been improperly stored.
- the unit has been damaged in transit.
- unit markings or inscriptions are illegible.
- malfunctions occur.
- obvious measuring inaccuracies occur.
- it is not possible to conduct any more measurements with the unit.
- permitted limit values have been exceeded.
4. Safety regulations
Use of this unit assumes that the operator observes the conventional safety regulations and has read the
certicate in order to eliminate faulty operation of the unit.
The following safety regulations must also be complied with:
- The unit should not be opened inside the Ex-area.
- The battery should only be exchanged outside the Ex-area.
- Additional batteries must not be carried in the Ex-area.
- Only type-tested batteries should be used.
- The unit should only be operated in the explosive area while situated in the appropriate leather case
030280.
- Only accessories approved by ecom instruments GmbH may be used.
- Avoid use in aggressive acids or bases.
- It must be ensured that the unit is not taken into zone 0.
- The device is to be protected against water and dust ingress.
4.1 Special conditions for conducting measurements in zone 0
The Ex-MP4 a should only be used to measure the temperature in zone 0 after consideration of the fol-
lowing conditions.
1. The Ex-MP4 a itself should not be taken into the zone 0 area.
2. It is imperative to make sure that the unit is not able to inadvertently enter the zone 0 area either.
To ensure this the Ex-MP4 a unit must be used with the corresponding wrist strap.
Only the manufacturer‘s wrist strap should be used.

5
5. Safety precautions
- Caution! Laser radiation!
- Do not stare directly at the laser!
- Do not aim the unit at other people!
- Risk of eye damage!
- Laser class 2
- Power: < 1 mW
- Wavelength: 630 -670 nm
6. Ex-Data
EC type approval test certicate no.: EPS 10 ATEX 1 242 X
Ex-designation: † II 2 G Ex ia op is IIC T4
Approved for zone 1, device group II, gas group IIC (explosive gases, vapours or mist), temperature class T4.
7. Specifications
Measuring range -18 °C to + 400 °C
Ambient temperature Ta 0 ... +50 °C
Storage temperature -20 °C to +65 °C
Relative humidity 10- 95 % r.h. non-condensing at 30 °C
Emission level xed at 0.95
Smallest display unit 0.2 °C
Target sighting Laser
Accuracy -18 °C … -1°C ±3 °C
(at 23°C) -1°C … +400 °C : ±2 % of reading or ±2 °C, whichever is greater
Repeatability ± 2 % of reading or ±2 °C whichever is greater
Spectral response 7-18 µm
Response time (95%) 500ms
Batteries / power supply: 1 x 9V alkaline battery according to IEC 6LR61, type-tested
7.1 Table with type-approved batteries
Manufacturer Model
Panasonic EVOLTA
Panasonic Powerline
Duracell DURALOCK
Dimensions: 152 x 101 x 38 mm
Weight: approx. 200g
CE-designation: ‡0102
8. Functional description / operating instructions
Please read these operating instructions carefully so that you are familiar with, and are able to use, all of
the functions of your Ex-MP4 a. For your own safety and information please read through the instructions
on the following pages!

6
8.1 Mode of operation
All objects with a temperature above absolute zero radiate infrared energy, which extends in every direction
with the speed of light. If an infrared thermometer is trained on an object, the lens gathers the energy and
focuses it on an infrared sensor.
The sensor responds by emitting a voltage signal which is directly proportional to the absorbed energy. The
microprocessor-controlled electronics in the unit are able to ascertain and display the current temperature
(in regard to a wider parameter).
Objects with radiant or polished surfaces not only radiate energy but also reect a proportion of radia-
tion from their surrounding area. A factor between 0.1 and 1.0 designated as the emission level makes
allowances for this factual calculation so that only the true radiated energy, and not the reected energy, is
consulted for calculating the measured value. The majority of applications are measured with an emission
level of 0.95.
In the Ex-MP4 a the level is set at 0.95 and cannot be altered.
8.2 Structural overview
Display Laser
Battery compartment
Battery compartment cover
°C deg / °F deg switch
Operating trigger
8.3 Optical diagram and field of sight
Distance sensor – object [mm]
8

7
Ensure that the specific area or object to be measured is
larger than the laser spot size of the device. The smaller
the object the closer the equipment must be. For opti-
mum measuring accuracy the area/object should be at
least twice that of the spot size.
8.4 Operation of the laser sight
The laser sight is a component of the Ex-MP4 a.
Please read through the following section carefully:
Warning! Do not stare directly at the laser!
Risk of eye damage!
Operate the unit carefully!
Do not aim the unit at other people!
The laser sight enables the unit to be aimed accurately and is useful for sighting small or distant objects.
However it does not indicate the size of the spot. This value is distance-dependent and can be inferred from
the optical diagram in this manual.
The units are equipped with a robust solid state laser which is housed inside the unit itself.
8.5 Operating the unit
°C deg / °F deg Change-over
In order to switch between centigrade and
fahrenheit, the battery compartment has to be opened – having
first taken off the leather case.
It is then possible to select the temperature
indicator required by sliding Switch A from either degree °C or
degree °F.
After selecting the temperature indicator required the leather
case must again be fitted before operation of the equipment in the
hazardous area.
The LCD indicator shows the temperature in degrees °C or
degrees °F. After release of the trigger the temperature measu-
rement is displayed for a further 7 seconds and at the same time
the word „Hold” is displayed. The battery condition indicator sym-
bol shows the state of the battery.
(A)

8
Operation of the Device
For a temperature reading, the equipment is
pointed towards an object and the trigger is pressed. Consider the distance
in relation to the spot size at this point (optical diagram) as well as taking
the eld of sight into consideration. With increased distance from the object,
increase the surface of the measured area of equipment.
8.6 Emission level
The majority of organic materials, as well as coated and oxidised surfaces, possess an emission level of
0.95. For this reason the emission level of the Ex-MP4 a has been set to 0.95 and cannot be altered.
When determining the temperatures of shiny or highly polished metallic surfaces inaccurate measure-
ments are produced. To compensate this effect the object being measured can be covered with adhesive
tape or coated with matt black paint. Wait until the adhesive tape is the same temperature as the underly-
ing material. Then determine the temperature of the tape (which is equivalent to the material) or the coated
surface.
8.7 Battery replacement
If the battery is depleted and only a residual charge remains, the battery symbol appears on the display.
If this occurs the battery should be changed in order to ensure safe operation. The battery should only be
exchanged outside the Ex-area. When changing the battery make sure that only those batteries specied
in this manual are used (see point 7.1:Table with type-tested batteries). The use of different batteries is
strictly prohibited!
To open the battery compartment rst remove the leather case from the handle. The compartment can now
be opened. You can then remove and exchange the battery. When the battery has been replaced the leather
case must be retted over the handle.
- Pay attention to the correct polarity.
- Dispose of depleted batteries in an appropriate and environmentally friendly manner!
8.8 Troubleshooting
Code Malfunction Procedure
- - - (on the display) Target temperature above or
below the measuring range
Select target within the range
Battery symbol
appears
Battery virtually depleted Check or replace the battery
No display Battery discharged Check or replace the battery
Laser does not func-
tion
Weak or discharged battery Replace battery
9. Repair
Repairs are to be conducted by ecom instruments GmbH or by personnel or service centres authorised by
ecom instruments GmbH, because the safety of the unit needs to be tested after repairs are implemented.

9
10. Disposal
Old electrical devices as well as „historical“ old electrical devices from ecom instruments GmbH are
supplied for disposal at our expense and disposed of free of charge in accordance with European Directive
2002/96/EC and the German electrical law of 16/03/2005. The costs associated with returning devices to
ecom instruments GmbH are to be borne by the sender.
In accordance with Article 1, Section 18 and Article 2 of the Act revising the law of Waste-Related Product
Responsibility for Batteries and Accumulators (Gesetz zur Neuregelung der abfallrechtlichen Produktver-
antwortung für Batterien und Akkumulatoren) dated 25 June 2009, we are obligated to provide the following
information.
Your device contains an alkaline battery.
Depleted batteries should never be disposed of along with normal or household waste. Old batteries can
contain harmful substances that are hazardous to health and damaging to the environment. Please return
the batteries. Return is free of charge and required by law. Please only dispose of discharged batteries in
the designated containers.
Note: The battery used in this unit poses a re hazard and can cause chemical injuries if it is used impro-
perly. Neither the battery nor the battery cells should be opened or disassembled and they should not be
burned or exposed to temperatures exceeding 100°C.
If the battery needs to be disposed of, it can be removed as described under point (8.7). The disposal regu-
lations specied above are also applicable for old devices.
All batteries can be recycled. Precious raw materials such as iron, zinc and nickel can therefore be reused.
The symbol has the following meaning: Batteries and rechargeable batteries should not be disposed of
along with normal or household waste.
11. Cleaning, maintenance and storage
- Only use a suitable cloth or sponge to clean the unit.
Do not use any solvents or abrasive cleaning agents when cleaning.
- It is recommended that you have the functionality and accuracy of the unit inspected by the manufacturer
every two years.
- The battery should be removed from the unit during prolonged periods of storage.
- Do not allow storage temperatures to exceed or fall below the permitted range of –20 °C to +65 °C!
12. Warranty and liability
In accordance with our general terms and conditions ecom instruments GmbH provides a warranty for this
product for a period of two years regarding its function and material under the specied and permitted ope-
rating and maintenance conditions. This excludes all parts that are subject to wear and tear (e.g. batteries,
probes, calibration units, LEDs, etc.).
This warranty does not extend to products that are used improperly, have been modied, neglected, dama-
ged in accidents or exposed to abnormal operating conditions and inappropriate handling.
Any claims under the warranty can be invoked by returning the defective equipment to the manufacturer.
We reserve the right to repair, adjust or exchange the device.
The aforementioned warranty conditions are the sole and only right of the purchaser to compensation,
are exclusively valid and replace all other contract or legal warranty obligations. ecom instruments GmbH
assumes no liability for special, direct, indirect, accompanying or consequential damage as well as losses
including the loss of data, independent of whether it was caused by violation of the warranty obligation,
lawful or unlawful action, action in good faith or any other action.

10
If the limitation of a legal warranty as well as the exclusion or limitation of accompanying or consequential
damage is not permissible in certain countries, it may be that the aforementioned limitations and exclu-
sions are not valid for every purchaser. Should any such clause of these warranty conditions be declared
ineffective or unenforceable by a competent court, then the effectiveness or enforceability of any of the
other conditions of these warranty conditions shall remain unaffected by such a judgement.
13. EC Declaration of Conformity
Further details regarding the Declaration of Conformity can be found at:
http://www.ecom-ex.com in the product download area.

68
ecom instruments GmbH | Industriestraße 2 | 97959 Assamstadt | Germany
Tel.: +49 62 94 / 42 24-0 | Fax: +49 62 94 / 42 24-100 | E-Mail: sales@ecom-ex.com | www.ecom-ex.com
1/2
EG-Konformitätserklärung / EC-Declaration of Conformity
Déclaration de conformité CE
Wir, ecom instruments GmbH, erklären in alleiniger Verantwortung, dass unser Produkt auf welches sich diese Erklärung
bezieht, den nachfolgenden Daten, Richtlinien und Normen entspricht:
We, ecom instruments GmbH, declare under our sole responsibility, that the product to which this declaration relates, is
in accordance with the provision of the following standards, directives and norms:
Nous, ecom instruments GmbH, déclarons sous notre seule responsabilité, que le produit pour lequel cette déclaration se
rapporte, est conforme à la disposition des normes et des directives suivantes:
Name & Anschrift des Herstellers innerhalb der
europäischen Gemeinschaft
Name & address of the manufacturer in the EC
Nom et adresse du fabricant dans l‘Union Européenne
ecom instruments GmbH
Industriestraße 2
97959 Assamstadt
Germany
Beschreibung der Geräte
Description of the devices
Description des appareils
Eigensicheres Temperaturmessgerät
Intrinsically-Safe Non-Contact Temperature Meter
Thermomètre Ex à mesure sans contact
Typ/Type: Ex-MP4 a
Ex-Kennzeichnung
Ex-designation
Marquage Ex
II 2 G Ex ia op is IIC T4
EG-Baumusterprüfbescheinigung
EC-Type Examination Certicate
Déclaration de conformité CE
EPS 10 ATEX 1 242 X
Zertizierungsstelle
Notied body
Organisme notié
Bureau Veritas Consumer Product Services GmbH
Businesspark A96
86842 Türkheim
Germany
Kenn-Nummer / Identication number /
Numéro d‘identication: 2004
Name & Anschrift der überwachenden Stelle
a) Qualitätssicherungssystem ISO 9001
Name & address of auditing body
a) Quality Management System according to ISO 9001
Nom et adresse de l‘organisme certié
a) Système d‘assurance qualité selon ISO 9001
DEKRA Certication GmbH
Handwerkstraße 15
70565 Stuttgart
Germany

69
ecom instruments GmbH | Industriestraße 2 | 97959 Assamstadt | Germany
Tel.: +49 62 94 / 42 24-0 | Fax: +49 62 94 / 42 24-100 | E-Mail: sales@ecom-ex.com | www.ecom-ex.com
2/2
1120ME02A02
WEEE-Reg.-Nr.: DE 934 99306
Name & Anschrift der überwachenden Stelle
b) Qualitätssicherungssystem der Produktion gemäß
Richtlinie 94/9/EG (ATEX)
Name & address of auditing body
b) Quality Management System of production according to
directive 94/9/EC (ATEX)
Nom et adresse de l‘organisme certié
b) Système d‘assurance qualité pour la fabrication selon la
directive 94/9/CE (ATEX)
Physikalisch-Technische Bundesanstalt
Bundesallee 100
38116 Braunschweig
Germany
Kenn-Nummer / Identication number /
Numéro d‘identication: 0102
Angewandte harmonisierte Standards
Applied harmonized standards
Standards harmonisés appliqués
EN 60079-0:2012+A11:2013
EN 60079-11:2012
EN 60079-28:2007
Angewandte europäische Richtlinien
Applied European directives
Directives générales européennes appliquées
94/9/EG
ecom instruments GmbH
Assamstadt, 11. 09. 2014
Patrick Sander
Vice President Products
Unterzeichnet /
signed

70

71

72
Copyright© 2014 ecom instruments GmbH.
The contents of this document, either in part or in their entirety, may not be
reproduced, forwarded, distributed or saved in any way without ecom instruments
GmbH prior written consent.
The contents of this document are presented as they currently exist. ecom instruments
GmbH does not provide any express or tacit guarantee for the correctness or
completeness of the contents of this document, including, but not restricted to, the
tacit guarantee of market suitability or tness for a specic purpose unless applicable
laws or court decisions make liability mandatory. ecom instruments GmbH reserves
the right to make changes to this document or to withdraw it any time without prior
notice.
PA000378 / / 03 / 2015
WEEE-Reg.-Nr. DE 934 99306
1120AL01A07
Subject to change.
ecom instruments GmbH | Industriestraße 2 | 97959 Assamstadt | Germany
Phone +49 62 94 42 24 0 | Fax +49 62 94 42 24 100
sales@ecom-ex.com | www.ecom-ex.com
Table of contents
Other Ecom Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

FroliCat
FroliCat DART duo PTY45-14259 operating guide

Flomec
Flomec OM025 instruction manual

FlowLine
FlowLine DataLoop LI24 Series quick start

ABB
ABB FieldIT AquaMaster Quick reference guide

Gossen MetraWatt
Gossen MetraWatt SECULIFE ES TECH operating instructions

Tektronix
Tektronix TriMode P7520A instruction manual