Setting optional program parameters
}Open a program or a program template for editing. Select a step.
}Program loops: Define to which step the program shall return under GoTo. Define the
number of reiterations under Cycles. Maximum: 999
}Define a temperature increment (±0.1to20°C) under ΔT by which the temperature shall
increase/decrease per cycle.
}Define a time increment (1to240s) under Δt by which the holding time shall increase
per cycle.
}Set the heating and cooling rate precisely in steps of 0.1°C/s under ΔR.
}In order to optimize the annealing temperature, define the temperature gradient in de-
grees by which the block temperature shall increase or decrease from one column to the
next, see below.
}If prompted, confirm your entries. Tap Done. Tap Save to save the program.
Program the temperature gradient (gradient-enabled models)
}Open a program or a program template for editing. Select the annealing step.
}Tap the field for the block temperature. Tap Degrees on the keypad.
Or: In the graphical/spreadsheet view, first tap Edit and then Degrees.
}Tap Lin./Std. to choose between the Linear Gradient Tool and the standard gradient pro-
gramming feature.
Annealing Temp.:
Column
Grad. incr.:
}Linear Gradient Tool: Enter the annealing temperature under Annealing Temp.:.
This temperature is applied to the column which is indicated under Annealing Temp.:.
Enter a temperature increment under Grad. incr.:.
}Standard gradient programming: Enter the temperature for the first and the last column.
}Confirm your entries. Tap Done. Tap Save to save the program.
Copying and deleting programs
Option Icon Description
Copy all Copy all programs of the selected user
Copy Copy the selected program
Delete all Delete all programs of the selected user
Delete Delete the selected program
Starting, stopping and pausing programs
}Use the quick-start function on the login screen/Home screen, to start the PCR run.
}Or: Tap Programs on the Home screen. Select the user directory and the program on the
page Program overview.
User: Admin dd.mm.yyyy hh:mm
Home EditDelete
Program overview
Copy Start
User
users3
USB
Admin
John Doe
Programs
in directory3 Preview
Step
Step
GoTo
Program information
Name
Block Type:
Lid:
Program Runtime:
Lin-grad-dag
Gradient 96
99 °C Preheat Lid: On
01h 15m 22s
30x 2
}Use the arrow keys to navigate through the list, if the number of users/programs is
greater than what can be displayed.
Or: Use to display all users/programs.
}Use the preview function to validate the program and tap Edit to adjust it, if required.
}Insert samples into the thermal block. Use replicates of a sample to optimize the anneal-
ing temperature. Close the lid. Establish the contact pressure: Turn the wheel clockwise
until you hear a clicking sound.
}Tap Start to start the program.
}Tap Pause to stop the currently executed program. Tap Continue to resume the program.
Tap Skip to skip a step. Tap Stop program to abort a PCR run.