Short Manual Biometra TRIO
Analytik Jena GmbH
Konrad-Zuse-Strasse 1
07745 Jena ⋅ Germany
Phone +49 3641 77 70
www.analytik-jena.com
Short Manual Biometra TRIO
Version: A (07/2022)
34-3600-003-23
1/2
Safety instructions
Only use the quick reference guide in combination with the detailed operating manual.
WARNING!Risk of electric shock! Check whether the voltage selector switch (bottom) is
set to the correct mains voltage.
CAUTION!Risk of burns! Close the heated lid before starting the program. Use protective
equipment when handling hot samples.
NOTICE!Reduce the contact pressure for opening: Turn the wheel counterclockwise and
then press the button. After closing the lid, apply the contact pressure via the lid wheel.
Design
¡Wheel for exerting the contact pressure(1)
¡Lid sealing button(2)
¡Touch screen display with control software(3)
¡USB port(4)
Login and Home screen
After the initial self test, the software first opens the login screen and then the Home screen.
Option Icon Description
Login Logging in a user
Quick Start Block Display of the current block status. If the block is free:
Tap to open the quick-start function
DE/EN/CN /¡Change login screen language
¡Set the software language in the user management
Programs ¡Create or edit PCR programs
¡Select a program and start the PCR run
Protocol Wizard ¡Automatically create a PCR program
¡Calculate the annealing temperature and holding
time for the elongation step
Incubate Incubate the samples at constant temperature
Tools ¡Customize the software settings
¡Set up the user management
¡View documentation, self test
¡Create Service info file, carry out a backup
Free
Quick Start
Programs Incubate Tools
Logout
28.0°C
Block 1
User: John Doe
dd.mm.yyyy hh:ss
Protocol Wizard
Quick Start
28.0°C
Free
Quick Start
28.0°C
Free
Block 2 Block 3
Creating or editing PCR programs
}Tap Programs on the Home screen.
}Select a program on the page Program overview and edit it. Or: Tap New from tem-
plate. Select a program template and tap Open template to open it.
}Use or to select the graphical or the spreadsheet programming mode.
Name: Lid: Preheat Lid: On
Home Edit Insert Delete Done Back
Step
Go To
Cycles
∆T
∆t
∆R
30x
1 2 3 4 5 6Step Step Step Step Step
Go To
Temperature [°C]
Three-step
dd.mm.yyyy hh:mm
Program Runtime: 1h 15m 10s
User: John Doe
Name: Lid: Preheat Lid: On
Home Edit Insert Delete Done Back
Loops Step Block Temp. (°C)
Hold Time
(h:mm:ss) Go To Cycles ∆T(°C) ∆t(s) ∆R(°C/s)
Three-step
dd.mm.yyyy hh:mm
Program Runtime: 1h 15m 10s
User: John Doe
Graphical programming mode Spreadsheet programming mode
}Enter the program name under Name:. Define the temperature for the heated lid under
Lid:. Select On to activate the preheating function of the heated lid.
}Use Insert and Delete to add or delete steps. Maximum number of steps: 30
}Select a program step for editing. Change individual parameters in graphical/spreadsheet
view. Or: Tap Edit and change all parameters.
}Define the block temperature (3to99°C) and the holding time (max. ≤9h59min59s).
Use ∞ to program a pause.
}If required, define optional program parameters.
}Tap Done. Select a user directory under Copy/save programs.
}Save the program. Or: Enter a new name under Save as.