FOR-A IR18 User manual

SPECIFICATIONS
Dimensions: 120mm(L) x 42mm(W) x 53.3mm(H)
Weight: 64 g
Battery: CR2032 3V lithium battery
Displayed Temperature Range: 89.6°F to 109.4°F (32°C to
43°C)
Display Resolution: 0.1°F/0.1°C
Accuracy: Meet the accuracy requirement specified in ASTM
E1965-98
±0.4°F (±0.2°C) for the range of 96.8°F to 102.2°F
(36.0°C to 39.0°C)
±0.5°F (±0.3°C ) for the range of < 96.8°F (36.0°C) or >
102.2°F (39.0°C)
Temperature Unit: °F
Operating Temperature Range: 50°F to 104°F (10°C to
40°C)
Operating Humidity: 95% RH or less
Storage Temperature Range: -4°F to 140°F (-20°C to
60°C)
Storage Humidity: 95% RH or less
Memory Capacity: 10 measurements
The specifications may be changed without prior notice.
TROUBLESHOOTING
All error messages below would be shown together with a red
backlight. Please follow “what to do” to resolve problems. If
the problem still exists, please call your local dealer for help.
What It Means What To Do
Appear when
environmental
temperature is below
or above system
operation range.
Battery is low.
Thermometer errors.
Appears when the
battery can’t provide
enough power for a
test.
Put the thermometer
under operating
temperature range of
50°F to 104°F (10°C
to 40°C).
Replace the battery with
a new one immediately.
If the problem persists,
please contact your
dealer.
Replace the old battery
with a new one
immediately.
Message
RECALLING THE MEMORY
1
The thermometer comes with one CR2032 3V lithium
battery. Replace it when “ ” appears. Please follow the
steps for battery replacement.
1. Remove the battery cover.
2. Place the new battery in the battery compartment and
press it in until the battery are secured.
3. Reattach the battery cover.
NOTE:
● Although the thermometer works when “ ” appears, we
still recommend that you change the battery to obtain an
accurate result.
● Remove the battery if stored for a long period of time.
● The battery should be kept out of children. If they are
swallowed, promptly see a doctor for help.
● Properly dispose of the battery.
REPLACING THE BATTERY
1 32
CR2032
IR18
EAR THERMOMETER
Operating Instructions
311-1118300-005 ver 1.0 2016/03
Distributed by ForaCare
ForaCare, Inc.
893 Patriot Dr., Suite D, Moorpark, CA 93021 USA
Products made in Taiwan
Toll Free: 1-888-307-8188
(7:00 am - 6:00 pm PST, Mon.-Fri.)
For assistance outside of these hours,
please contact your healthcare professional.
www.foracare.com
Read instructions before use.
Your thermometer stores the last 10 readings in the
memory.
1. Be sure the thermometer is OFF before recalling the
memory.
2. Press the On/Memory button to turn on the
thermometer.
3. Each time you press the On/Memory button, a result will
be displayed in the order of dates (latest result shown first),
together with “ ” and a number (from 1 to 10). When the
memory is full, the oldest result is deleted as the new one is
added. When the last record is displayed in the LCD, press
On/Memory button again to return to the first one.
4. Exit the memory. Press and hold On/Memory button
again, and then the LCD will show “ OFF ” to exit memory.
NOTE:
● The device will automatically turn off if left without
pressing any button within 30 seconds.
● When using the thermometer for the first time, the first
memory result will display “---”. It indicates that there are
no test results in memory.
persons with compromised immune systems, persons taking
antibiotics, steroids or antipyretics (aspirin, ibuprofen,
acetaminophen), or persons with certain chronic illnesses.
Please consult your physician when you feel ill even if you
do not have a fever.
*Chamberlain, J. M. Terndrup, T. E., New Light on
Thermometer Readings, Contemporary Pediatrics, March
1994.
● The probe is not waterproof. Please wipe it with a clean
and dry cotton swab.
● The body of the thermometer is not water-resistant.
Never put the thermometer under a running tap or
submerge it into water. Use a soft and dry cloth to clean
it. Do not use abrasive cleaners.
● Store the thermometer in a cool and dry location that is
free from dust and away from direct sunlight.
● Keep the probe clean, as build-up of earwax and grease
may affect the measurement.
CARE & CLEANING
Normal Temperature Range*
95.9°F to 99.5°F (35.5°C to 37.5°C)
94.5°F to 99.1°F (34.7°C to 37.3°C)
97.9°F to 100.4°F (36.6°C to 38.0°C)
96.4°F to 100.4°F (35.8°C to 38.0°C)
Body Site
Oral
Axillary (underarm)
Rectal
Ear
ABOUT NORMAL BODY TEMPERATURE & FEVER
Body temperature can vary from person to person. It also
varies on different parts of the body and time of day. The
following table shows the statistics of normal ranges from
different body parts. Please keep in mind that temperatures
measured from different parts, even at the same time,
should not be directly compared.
Fever indicates that the body temperature is higher than
normal. This symptom may be caused by infection,
overdressing or immunization. Some people may not
experience fever even when they are ill. These include, but
are not limited to, infants younger than 3 months old,
12 13

Temperature Reading
Ear Temperature Scanning Progress
Memory Mode
Record Numbers
Hi Temperature Indicator
Temperature Unit
Low Battery Indicator
1
2
3
4
5
6
7
INTRODUCTION
Thank you for choosing FORA IR18 ear thermometer. This
innovative medical device relies on advanced infrared (IR)
technology to measure temperature instantly.
HOW DOES IT WORK
The thermometer measures the infrared heat generated by
the eardrum and its surrounding tissue. The thermometer
then converts it into a temperature value shown on the LCD
screen.
IMPORTANT SAFETY INSTRUCTIONS
READ THIS BEFORE USE. KEEP THESE INSTRUCTIONS IN A
SAFE PLACE.
The following basic safety precautions should always be
taken.
1. Close supervision is necessary when the thermometer is
used by, on, or near children, handicapped persons or
invalids.
2. Use the thermometer only for the intended use described
in this manual.
3. Do not use the thermometer if it is not working properly,
or if it has suffered any damage.
4. Do not use accessories which are not supplied or
recommended by the manufacturer.
5. Proper maintenance is essential to the longevity of your
device. If you are concerned about your accuracy of
measurement, please contact the local customer service
or place of purchase for help.
CAUTIONS AND WARNINGS
● As with any thermometer, proper technique is crucial to
getting accurate temperature readings. Please read this
manual thoroughly and carefully before use.
● Always operate the thermometer in temperatures
between 50°F to 104°F (10°C to 40°C) and relative
humidity less than 95%.
● Always store the thermometer in a cool and dry place:
temperatures between -4°F to 140°F (-20°C to 60°C);
relative humidity less than 95%. Avoid direct sunlight.
● Avoid dropping the thermometer.
● Use only FORA IR18 probe covers for the infrared ear
thermometer.
● For proper hygiene, do not share or reuse probe covers.
Damaged probe covers may result in an error display.
● Do not touch the probe cover tip.
● Do not disassemble the thermometer.
● Basic safety precautions should always be observed,
especially when the thermometer is used on or near
children and disabled persons.
● This thermometer is not intended to be a substitution for
a consultation with your physician.
● Keep probe covers out of reach of children.
● Temperature measurements may differ from left to right
ear. Always measure by using the same ear.
RESTRICTIONS OF USE
This thermometer is clinically proven to produce accurate
temperature measurements. However, please be advised if
you have the following situations:
● The accuracy cannot be ensured for a person who has a
deformity in their ear, which prevents the thermometer
probe from properly being inserted into the ear canal.
● The accuracy cannot be ensured when blood or drainage
is found in the ear canal.
● If ear drops or medications have been placed in one ear,
take measurements from the other ear.
● For a person who wears an ear plug or hearing aid,
remove the device and wait for 15 minutes before taking
temperature.
NOTE: Never try to clean inside your ears. You may
accidentally damage the eardrum or its surrounding tissues.
Remove excess earwax only when you can reach it with a
clean cloth. Consult a physician if you suspect the presence
of excess earwax.
APPEARANCE AND KEY FUNCTIONS OF THE
THERMOMETER
Scan Button
Probe
LCD Screen
1
2
3
On/Memory Button
Battery Cover
Probe Cover
4
5
6
1
2
6
3
4
5
HINTS ON TAKING TEMPERATURES
1. As with other thermometers, you may observe slight
variations in consecutive measurements. It is
recommended that you take 3 temperature readings and
use the highest one for the following situations:
● Infants younger than 3 months old.
● Children younger than 3 years old and who have a
compromised immune system and the presence /
absence of fever is critical.
● When you are learning to use the thermometer.
2. Do not take a reading while the patient is moving and/or
talking. Wait for 15 minutes after any of the following
situations before taking a measurement:
● When the ear has been covered.
● After exercising, swimming, or taking a bath.
● When the patient has been exposed to extreme
temperature.
3. To take accurate readings, the ear must be free from
excessive earwax build-up.
USING THE DEVICE
Replacing the Probe Cover
1. Check if the probe cover is clean and new.
2. Secure the cover to the probe.
NOTE:
● The used probe cover should be removed after each
measurement to ensure accuracy.
● To ensure proper hygiene, attach a new probe cover each
time and do not touch its tip.
21
INTENDED USE
FORA IR18 ear thermometer is intended for the intermittent
measurement and monitoring of human body temperature
from the ear canal. The device is intended for home use by
people of all ages.
LCD SCREEN
32 4 5 6 7
1
RESULT INDICATORS
“Lo”
Result with green
backlight
Result with red
backlight
“Hi”
Temperature ≤ 89.5°F (31.9°C)
89.6°F (32°C) ≤ Temperature
< 100.3°F (37.9°C)
100.4°F (38°C) ≤ Temperature
< 109.4°F (43°C)
Temperature ≥ 109.5°F (43.1°C)
5
1. Press and release the On/Memory button to turn on the
thermometer. When ready, the thermometer displays the
last measurement.
2. Gently fit the probe with a clean probe cover into the ear
canal.
3. Press and release the Scan button. You will hear a beep
to indicate the measurement is complete and the
thermometer can be removed.
4. Read the result. “ ” is shown together with a
temperature value.
5. Discard the used probe cover into a trash can.
NOTE: Turn off the thermometer by pressing On/Memory
button twice. It will automatically turn off if left idle for 30
seconds.
MEASURING TEMPERATURE
2 3
11
14
Table of contents
Other FOR-A Thermometer manuals

FOR-A
FOR-A IR42 User manual

FOR-A
FOR-A TM10 User manual

FOR-A
FOR-A IR20 Series User manual

FOR-A
FOR-A IR20b User manual

FOR-A
FOR-A IR10 User manual

FOR-A
FOR-A IR21 User manual

FOR-A
FOR-A IR20 Series User manual

FOR-A
FOR-A MT Series User manual

FOR-A
FOR-A IR20 Series User manual

FOR-A
FOR-A IR20 Series User manual