FOR-A IR42 User manual

Model No.: IR42
Dimension & Weight: 155.46 (L) x 40.14 (W) x 39.45 (H) mm,
61.8g (without battery)
Power Source: 2 x 1.5V AAA alkaline batteries
Battery Life: With new batteries, approx. 3,000 measurements.
Displayed Temperature range: 89.6°F to 109.4°F (32°C to 43°C)
Display Resolution: 0.1°F/0.1°C
Accuracy: Meet the accuracy requirement specified in ASTM
E1965-98
±0.4°F (±0.2°C) for the range of 95°F to 107.6°F (35.0°C to 42.0°C)
/ ±0.5°F (±0.3°C) for the range of <95°F (35.0°C) or >107.6°F
(42.0°C)
Reference to Standards: ASTM E1965-98; IEC 60601-1; IEC
60601-1-2 (EMC)
Temperature unit: °F (Default) or °C
Operating temperature range: 50°F to 104°F (10°C to 40°C)
Operating humidity: 85% RH or less
Storage / transportation temperature range: -4°F to 140°F (-20°C
to 60°C)
Storage / transportation humidity: 85% RH or less
Memory capacity: 30 measurements
Expected service life: 3 years
The specifications may be changed without prior notice.
IR42
Forehead Thermometer
Operating Instructions

1. Probe lens
2. Display screen
3. Power/Scan button
4. Protective sensor cap
5. Mode/Set button
6. Memory button
7. Battery Cover
Thank you for choosing FORA IR42 Forehead Thermometer.
Please read this instruction manual first so you can use this
thermometer safely and correctly. Please keep this instruction
manual for future reference. This innovative medical device uses
advanced infrared (IR) technology to measure temperature
instantly and accurately on the forehead/surface. FORA IR42
Forehead Thermometer delivers a body temperature reading
from the thermal radiation emitted from the forehead without
contact to the body.
FORA IR42 Forehead Thermometer is intended for the
intermittent measurement and monitoring of human body
temperature from the forehead. The device is intended for use of
all ages for home use by one with a good understanding of the
operation instruction, where the patient may be an operator.
INTRODUCTION
INTENDED USE
APPEARANCE AND KEY FUNCTIONS OF
THE THERMOMETER
4
1
2
3
5
6
7

1. Close supervision is necessary when the thermometer is used
by, on, or near children, handicapped persons or invalids.
2. Use the thermometer only for the intended use described in
this manual.
3. Do not use the thermometer if it is not working properly, or if it
has sustained any damage.
4. Keep the sensor end clean and free of debris. See
Maintenance section for instructions.
5. Do not use ethylene oxide gas, heat, autoclave, or any other
harsh methods to sterilize the device.
6. Put in place the protective sensor cap on the sensor end
when not in use.
7. Do not use the device shortly after exercise, bathing or
coming indoors.
8. If coming from an environment of warmer or cooler
temperature or after a period of exertion, allow the user and
the thermometer to acclimate to room temperature for 20
minutes prior to taking a measurement.
9. As the forehead temperature may be affected by sweat, oil
and the surrounding temperature, the reading shall be taken
as a reference only.
10. Do not use in presence of flammable anesthetic mixtures.
11. Do not use accessories which are not supplied or
recommended by the manufacturer.
12. Proper maintenance is essential to the longevity of your
device. If you are concerned about the accuracy of
measurement, please contact the local customer service or
place of purchase for help.
IMPORTANT SAFETY INSTRUCTIONS
READ THIS BEFORE USING AND KEEP THESE
INSTRUCTIONS IN A SAFE PLACE
As with any thermometer, proper technique is crucial to getting
accurate temperature readings. Please read this manual
thoroughly and carefully before use.
WARNING AND PRECAUTIONS

5
4
3
27
8
9
6
1
1. Adult body temperature
indicator
2. Child body temperature
indicator
3. Object surface temperature
indicator
4. Error message
5. Memory mode
6. Hold Symbol
7. Battery indicator
8. Temperature display
9. Temperature unit
LCD SCREEN
Always operate the thermometer in an operating temperature
range 50°F to 104°F (10°C to 40°C), and relative humidity less
than 85%.
Always store the thermometer in a cool and dry place:
temperatures between -4°F to 140°F (-20°C to 60°C); relative
humidity less than 85%. Avoid direct sunlight.
Avoid dropping the thermometer.
Basic safety precautions should always be observed,
especially when the thermometer is used on or near children
and disabled persons.
This thermometer is not intended to be a substitution for a
consultation with your physician.
1. Remove battery cover by pressing down at the arrow mark
and slide in the direction of the arrow as shown in the figures
below.
2. Install (2) AAA alkaline batteries and close the battery cover.
3. If thermometer will be stored without use, remove batteries.
When the thermometer is off, press MODE key for 2 seconds to
enter selection mode. Press MODE key again to switch between
°F and °C. Press POWER key.
USING THE DEVICE
Install Battery
Selecting the measuring unit

4321
1. Remove protective cap. The
forehead should be clear of hair
and perspiration.
2. Aim at the center forehead area
1.18 to 2.75" away from skin
surface (3 to 7cm). Be sure the
thermometer is perpendicular to
the skin surface.
3. Press and release the SCAN
button to take a measurement. A
double “beep” sound indicates a
reading has been taken and
displayed on the LCD screen.
Measuring Temperature for Adults
When a reading is greater
than 100.4°F (38°C), a
warning symbol will flash
with Red backlight.

4. To take another measurement, follow step 2 and 3.
5. The thermometer turns off automatically after 30 seconds.
Replace the sensory cap when finished.
1. Press the SCAN button to turn on
the thermometer.
2. Press and hold the MODE button for
1 second before releasing. An adult
facial profile should be flashing,
indicating that the preset mode is for
adults.
3. Press the MODE key to switch to
Child mode, indicated by a smiley face
icon.
Measuring Temperature for Children
WARNING AND PRECAUTIONS
If the reading is < 89.5°F (31.9°C), the display will show
“Lo”.
If the reading is ≧89.6°F (32.0°C) and ≦100.3°F (37.9 °C),
the display will show the reading with green backlight.
If the reading is ≧100.4 °F (38°C) and ≦109.4 °F (43°C),
the display will show the reading with red backlight.
If the reading is ≧109.5 °F (43.1°C), the display will show
“Hi”.
4. Aim the scanner at the center of the child’s forehead or temple
1.18 to 2.75" away from skin surface (3 to 7cm). Be sure the
thermometer is perpendicular to the skin surface. Press and
release the SCAN button to take a measurement.

FORA IR42 stores 30 most recent readings.
1. Press and release the SCAN button to turn on the
thermometer.
2. Press and hold the MEMORY button for 1 second to enter the
memory mode indicated by a flashing " " symbol. The most
recent reading stored will be displayed.
5. The thermometer automatically switches off when left idle for
30 seconds.
6. Replace the sensor cap when finished.
Red backlight with a warning symbol indicates a reading greater
than 99.8°F (37.6 °C).
Note:
●If the reading is ≦89.5°F (31.9°C), the display will show
“Lo”.
●If the reading is ≧89.6°F (32°C) and ≦99.8°F (37.6 °C),
the display will show the reading with green backlight.
●If the reading is ≧99.9°F (37.7 °C) and ≦109.4°F (43°C),
the display will show the reading with red backlight.
●If the reading is ≧109.5°F (43.1°C), the display will show
“Hi”.
●Parents should not rely only on temperature readings. If
you have concerns, please seek medical advice.
1. Press the SCAN button to turn on the thermometer.
2. Press and hold the MODE button for 1 second before
releasing. An adult facial profile should be flashing, indicating
that the preset mode is for adults.
3. Press the MODE key to switch to Surface mode, indicated by
a thermometer icon.
Measuring Surface Temperature

FORA IR42 stores 30 most recent readings.
1. Press and release the SCAN button to turn on the
thermometer.
2. Press and hold the MEMORY button for 1 second to enter the
memory mode indicated by a flashing " " symbol. The most
recent reading stored will be displayed.
4. Make sure the probe is flat and close to the object surface, not
at an angle. Perform an object measurement with a distance
within 1.9 in (5 cm). Press and hold the SCAN button as you
move the meter along the surface. The HOLD symbol will be
flashing.
5. Release the button and read the result. If the reading is ≧
32°F (0°C) and ≦212.2°F (100.1°C), the display will show
the reading with green backlight.
Note:
● If the reading is ≦32°F (0°C), the display will show “Lo”.
● If the reading is ≧212.2°F (100.1°C), the display will show
“Hi”.
3. Press and release the MEMORY button to cycle through
older readings.
4. When left idle for 30 seconds, the thermometer automatically
switches off.
RECALLING PAST READINGS

Note:
When the memory is full, the oldest result is deleted as the new
ones are added. While recalling previous readings, you may
take a measurement by pressing the SCAN button.
• FORA IR42 has no user serviceable internal parts except
battery replacement.
• Always replace Sensor Cap (or place in cradle) when not in
use.
• Store in a dry location free of dust and away from direct
sunlight.
• Use a soft dry cloth to clean the plastic casing or a cloth
dampened with a solution of water and mild detergent.
Occasionally, 70% isopropanol solution may be used. Never
submerge in liquid.
• The sensor window is recessed to assist in keeping it clean
and free of debris. Inspect the lens and remove any debris.
Smudges may be cleaned by gently wiping the window with a
small foam-tipped swab (non-linting) moistened with 70%
alcohol. Wait 10 minutes prior to taking temperatures.
Body temperature can vary from one individual/person to the
next. It also varies by location on the body and time of day. Fever
indicates that the body temperature is higher than normal. This
symptom may be caused by infection, overdressing or
immunization. Some people may not experience fever even when
they are ill. These include, but are not limited to, infants younger
than 3 months old, persons with compromised immune systems,
persons taking antibiotics, steroids or antipyretics (aspirin,
ibuprofen, acetaminophen), or persons with certain chronic
illnesses. Please consult your physician if you are concerned
about your body temperature readings.
MAINTENANCE
CLEANING AND DISINFECTION
ABOUT NORMAL BODY
TEMPERATURE & FEVER

Note:
When the memory is full, the oldest result is deleted as the new
ones are added. While recalling previous readings, you may
take a measurement by pressing the SCAN button.
Body temperature can vary from one individual/person to the
next. It also varies by location on the body and time of day. Fever
indicates that the body temperature is higher than normal. This
symptom may be caused by infection, overdressing or
immunization. Some people may not experience fever even when
they are ill. These include, but are not limited to, infants younger
than 3 months old, persons with compromised immune systems,
persons taking antibiotics, steroids or antipyretics (aspirin,
ibuprofen, acetaminophen), or persons with certain chronic
illnesses. Please consult your physician if you are concerned
about your body temperature readings.
Electromagne�c environment-guidance
(for home healthcare environment)
Emission test Compliance
Manufacturer’s declara�on-electromagne�c emissions
The IR42 is intended for use in the electromagnetic environment (for home healthcare)
specified below.
The customer or the user of the IR42 should assure that it is used in such an environment.
The IR42 uses RF energy only for its internal
function. Therefore,its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
The IR42 is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
/flicker emissions
IEC 61000-3-3
Group 1
Class B
Not applicable
Not applicable
Manufacturer’s declara�on-electromagne�c immunity
The IR42 is intended for use in the electromagnetic environment (for home healthcare)
specified below.
The customer or the user of the IR42 should assure that it is used in such an environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Immunity test IEC 60601
test level
Compliance level Electromagnetic environment-
guidance (for home healthcare
environment)
Floors should be wood,concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%
Mains power quality should be
that of a typical home
healthcare environment.
Mains power quality should be
that of a typical home
healthcare environment.
Surge IEC
61000-4-5
Voltage Dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50, 60 Hz)
magnetic field
IEC 61000-4-8
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power
supply lines
± 1kV for input/
output lines
+ 0.5kV, +1kV
line(s) to line(s)
+ 0.5kV, +1kV,
+ 2kV line(s) to
earth
30 A/m
50 Hz or 60 Hz
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT;
25/30 cycles
Voltage
interruptions:
0 % UT;
250/300 cycle
Voltage dips:
Not applicable
Not applicable
Not applicable
Voltage
interruptions:
Not applicable
Mains power quality should be
that of a typical home
healthcare environment.
If the user of the IR42 requires
continued operation during
power mains interruptions,
it is recommended that the
IR42 be powered from an
uninterruptible power supply
or a battery.
The IR42 power frequency
magnetic fields should be at
levels characteristic of a typical
location in a typical home
healthcare environment.
Not applicable
Not applicable
Not applicable
Not applicable
30 A/m
50 Hz and 60 Hz
Electrostatic
discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,
±8 kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,
±8 kV,±15 kV

Manufacturer’s declara�on-electromagne�c immunity
The IR42 is intended for use in the electromagnetic environment (for home healthcare)
specified below.
The customer or the user of the IR42 should assure that it is used in such and environment.
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Electromagnetic environment-
guidance (for home healthcare
environment)
Portable and mobile RF
communications
equipment should be used no
closer to any part of the IR42
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
d = 1,2
d = 1,2 80MHz to 800 MHz
d = 2,3 800MHz to 2,7 GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and
d is the recommended
separation distance in
metres (m).
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601
test level
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Not applicable
Not applicable
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Compliance level
Recommended separa�on distance between
portable and mobile RF communica�ons equipment and the IR42
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
Rated maximum
output power
of transmitter
W
80 MHz to 800 MHz 800 MHz to 2,7 GHz
The IR42 is intended for use in an electromagnetic environment (for home healthcare) in
which radiated RF disturbances are controlled.The customer or the user of the IR42 can help
prevent electromagnetic interference by maintaining a minimum distance between portable
and mobile RF communications equipment (transmitters) and the IR42 as recommended
below, according to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where p is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
d =1,2√Pd =1,2√Pd =2,3√P
0,01
0,1
1
10
100
N/A
N/A
N/A
N/A
N/A
0,12
0,38
1,2
3,8
12
0,23
0,73
2,3
7,3
23

Test
frequency
(MHz)
385
450
2 450
710
745
780
5 240
5 500
5 785
810
870
930
1 720
1 845
1 970
380 –390
430 – 470
704 – 787
800 – 960
1 700 –
1 990
2 400 –
2 570
5 100 –
5 800
TETRA 400
GMRS 460,
FRS 460
LTE
Band 13,
17
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE
Band 1, 3,
4, 25; UMTS
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
WLAN 802.11
a/n
Pulse
modulation b)
18 Hz
Pulse
modulation b)
217 Hz
Pulse
modulation b)
217 Hz
Pulse
modulation b)
217 Hz
Pulse
modulation b)
18 Hz
FM c)
±5 kHz
deviation
1 kHz sine
Pulse
modulation b)
217 Hz
1,8
2
2
0,2
0,3 27 27
0,3 28 28
0,3 28 28
20,3
0,3
28 28
2
0,30,2
28 28
0,3 9 9
9 9
Band a)
(MHz)
Service a) Modulation b) Maximum power
(W)
Distance
(m)
IMMUNITY TEST
LEVEL
(V/m)
Compliance LEVEL
(V/m)
(for home healthcare)
Manufacturer’s declara�on-electromagne�c immunity
Test specifica�ons for ENCLOSURE PORT IMMUNITY to RF wireless communica�ons equipment
The IR42 is intended for use in the electromagnetic environment (for home healthcare)
specified below.
The customer or the user of the IR42 should assure that it is used in such an environment.
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m.The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.

Model No.: IR42
Dimension & Weight: 155.46 (L) x 40.14 (W) x 39.45 (H) mm,
61.8g (without battery)
Power Source: 2 x 1.5V AAA alkaline batteries
Battery Life: With new batteries, approx. 3,000 measurements.
Displayed Temperature range: 89.6°F to 109.4°F (32°C to 43°C)
Display Resolution: 0.1°F/0.1°C
Accuracy: Meet the accuracy requirement specified in ASTM
E1965-98
±0.4°F (±0.2°C) for the range of 95°F to 107.6°F (35.0°C to 42.0°C)
/ ±0.5°F (±0.3°C) for the range of <95°F (35.0°C) or >107.6°F
(42.0°C)
Reference to Standards: ASTM E1965-98; IEC 60601-1; IEC
60601-1-2 (EMC)
Temperature unit: °F (Default) or °C
Operating temperature range: 50°F to 104°F (10°C to 40°C)
Operating humidity: 85% RH or less
Storage / transportation temperature range: -4°F to 140°F (-20°C
to 60°C)
Storage / transportation humidity: 85% RH or less
Memory capacity: 30 measurements
Expected service life: 3 years
The specifications may be changed without prior notice.
TROUBLESHOOTING
What it meansMessage What to do
Appear when
ambient temperature
is outside of the
operating
temperature range.
Only operate the
thermometer in
ambient
temperature
range of 50°F to
104°F (10°C to
40°C).
Measured temperature
below the measurement
range (below 89.6°F).
Review instructions and
repeat measurement.
Make sure distance
from forehead is not
greater than 2”, and
that the forehead is
clear.
Measured temperature
above the measurement
range (above 109.4°F).
Surface temperature
measured outside of the
measurement range
(32.0°F ~ 212.2°F).
Review the instructions
and re-start the
measurement
procedure.
Low or no power. Replace with new
batteries.
Review instructions and
repeat measurement.
Consult with a physician
if the problem persists.
Distributed by ForaCare, Inc.
893 Patriot Drive Suite D
Moorpark, CA 93021 USA
Products made in Taiwan
Toll Free: 1-888-307-8188
(8:30 am - 5:00 pm PST, Mon.-Fri.)
For assistance outside of these hours,
please contact your healthcare professional.
www.foracare.com
Read instructions before use.

Model No.: IR42
Dimension & Weight: 155.46 (L) x 40.14 (W) x 39.45 (H) mm,
61.8g (without battery)
Power Source: 2 x 1.5V AAA alkaline batteries
Battery Life: With new batteries, approx. 3,000 measurements.
Displayed Temperature range: 89.6°F to 109.4°F (32°C to 43°C)
Display Resolution: 0.1°F/0.1°C
Accuracy: Meet the accuracy requirement specified in ASTM
E1965-98
±0.4°F (±0.2°C) for the range of 95°F to 107.6°F (35.0°C to 42.0°C)
/ ±0.5°F (±0.3°C) for the range of <95°F (35.0°C) or >107.6°F
(42.0°C)
Reference to Standards: ASTM E1965-98; IEC 60601-1; IEC
60601-1-2 (EMC)
Temperature unit: °F (Default) or °C
Operating temperature range: 50°F to 104°F (10°C to 40°C)
Operating humidity: 85% RH or less
Storage / transportation temperature range: -4°F to 140°F (-20°C
to 60°C)
Storage / transportation humidity: 85% RH or less
Memory capacity: 30 measurements
Expected service life: 3 years
The specifications may be changed without prior notice.
SPECIFICATIONS
Review instructions and
repeat measurement.
Make sure distance
from forehead is not
greater than 2”, and
that the forehead is
clear.
Review instructions and
repeat measurement.
Consult with a physician
if the problem persists.
Distributed by ForaCare, Inc.
893 Patriot Drive Suite D
Moorpark, CA 93021 USA
Products made in Taiwan
Toll Free: 1-888-307-8188
(8:30 am - 5:00 pm PST, Mon.-Fri.)
For assistance outside of these hours,
please contact your healthcare professional.
www.foracare.com
Read instructions before use. Version 2.0 2020/05
311-1242100-019
Table of contents
Other FOR-A Thermometer manuals

FOR-A
FOR-A Diamond Nexus User manual

FOR-A
FOR-A IR19 User manual

FOR-A
FOR-A IR20 Series User manual

FOR-A
FOR-A Diamond Nexus Mini User manual

FOR-A
FOR-A IR20 Series User manual

FOR-A
FOR-A IR20 Series Reference guide

FOR-A
FOR-A IR20 Series User manual

FOR-A
FOR-A TM10 User manual

FOR-A
FOR-A IR20 Series User manual

FOR-A
FOR-A FocusTemp IR42a User manual
Popular Thermometer manuals by other brands

Patton
Patton 55CCE123 Manual instruction

Safety Vision
Safety Vision SV-IR-TEMP user guide

Sealey
Sealey Auto Service Line VS900.V3 instructions

Medisana
Medisana 76120 FTN instruction manual

Ames Instruments
Ames Instruments 63980 Owner's manual & safety instructions

Diagnosis
Diagnosis Dr Check FC500 Instructions for use